510(k) Data Aggregation

    K Number
    K231384

    Validate with FDA (Live)

    Date Cleared
    2023-09-22

    (133 days)

    Product Code
    Regulation Number
    892.2080
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Annalise Enterprise is a device designed to be used in the medical care environment to aid in triage and prioritization of studies with features suggestive of the following finding: mass effect . The device analyzes studies using an artificial intelligence algorithm to identify the finding. It makes study-level output available to an order and imaging management system for worklist prioritization or triage. The device is not intended to direct attention to specific portions of an image and only provides notification for the suspected finding. Its results are not intended: to be used on a standalone basis for clinical decision making ● ● to rule out specific findings, or otherwise preclude clinical assessment of CTB studies

    Device Description

    Annalise Enterprise CTB Triage Trauma is a software workflow tool which uses an artificial intelligence (AI) algorithm to identify suspected findings on non-contrast brain CT studies in the medical care environment. The findings identified by the device include mass effect. Radiological findings are identified by the device using an AI algorithm - a convolutional neural network trained using deep-learning techniques. Images used to train the algorithm were sourced from datasets that included a range of equipment manufacturers and models. The performance of the device's AI algorithm was validated in a standalone performance evaluation, in which the case-level output from the device was compared with a reference standard ('ground truth'). This was determined by two ground truthers, with a third truther used in the event of disagreement. All truthers were US board-certified neuroradiologists. The device interfaces with image and order management systems (such as PACS/RIS) to obtain noncontrast brain CT studies for processing by the AI algorithm. Following processing, if any of the radiological findings of interest are identified in a non-contrast brain CT study, the device provides a notification to the image and order management system for prioritization of that study in the worklist. This enables users to review the studies containing features suggestive of these radiological findings earlier than in the standard clinical workflow. It is important to note that the device will never decrease a study's existing priority in the worklist. This ensures that worklist items will never have their priorities downgraded based on AI results. The device workflow is performed parallel to and in conjunction with the standard clinical workflow for interpretation of non-contrast brain CTs. The device is intended to aid in prioritization and triage of radiological medical images only.

    AI/ML Overview

    Here is a summary of the acceptance criteria and study proving device performance, based on the provided text:

    Device Name: Annalise Enterprise CTB Triage Trauma
    Manufacturer: Annalise-AI Pty Ltd.


    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" for sensitivity and specificity in a separate table. Instead, it provides the results of the standalone performance study in a table, implying these values meet the unstated criteria for demonstrating safety and effectiveness. The comparison section further states that these results are "substantially equivalent to those of the predicate device."

    FindingSlice Thickness RangeOperating PointSensitivity % (Se) (95% CI)Specificity % (Sp) (95% CI)
    Mass Effect≤1.5mm0.16019597.0 (95.3, 98.4)88.7 (83.5, 94.0)
    ≤1.5mm0.22148496.6 (94.9, 98.2)89.5 (84.2, 94.0)
    Mass Effect>1.5mm & ≤5.0mm0.12094496.8 (95.3, 98.1)89.3 (84.5, 93.5)
    0.16019595.3 (93.6, 97.0)92.9 (88.7, 96.4)

    Additional Performance Metric (Triage Effectiveness):

    • Triage Turn-around Time: 81.6 seconds (95% CI: 80.3 - 82.9)

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size:
      • For slice thickness ≤1.5mm: 626 cases (493 positive, 133 negative for mass effect)
      • For slice thickness >1.5mm & ≤5.0mm: 762 cases (594 positive, 168 negative for mass effect)
    • Data Provenance: Retrospective, anonymized study. Collected consecutively from five US hospital network sites.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • Number of Experts: At least two neuroradiologists for initial review, with a third neuroradiologist for adjudication in case of disagreement.
    • Qualifications: All experts were US board-certified neuroradiologists, ABR-certified, and protocol-trained.

    4. Adjudication Method for the Test Set

    • Method: Consensus determined by two ground truthers. If there was a disagreement between the initial two ground truthers, a third ground truther was used to reach consensus (2+1 method).

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • An MRMC study was not explicitly described as being performed for AI-assisted human reader performance improvement.
    • The performance assessment focused on standalone performance of the AI algorithm and a triage effectiveness (turn-around time) study, which was an internal bench study. The document states the AI "enables users to review the studies containing features suggestive of these radiological findings earlier than in the standard clinical workflow," implying an effect on human workflow, but doesn't quantify reader improvement in an MRMC setting.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Yes, a standalone performance evaluation was done. The sections under "Performance Testing" describe comparing the device's case-level output with a reference standard (ground truth). The results in the table above (Sensitivity, Specificity) are from this standalone performance evaluation.

    7. The Type of Ground Truth Used

    • Type: Expert consensus (from multiple board-certified neuroradiologists).

    8. The Sample Size for the Training Set

    • The document states, "The test dataset used during the standalone performance evaluation was newly acquired and independent from the training dataset used in model development."
    • The sample size of the training set is not explicitly provided in the given text. It only mentions that "Images used to train the algorithm were sourced from datasets that included a range of equipment manufacturers and models."

    9. How the Ground Truth for the Training Set Was Established

    • The document does not explicitly describe how the ground truth for the training set was established. It only details the ground truth establishment for the test set used in the standalone performance evaluation.