510(k) Data Aggregation

    K Number
    K222442

    Validate with FDA (Live)

    Date Cleared
    2022-09-28

    (47 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DISPOSABLE VINYL EXAMINATION GLOVE (Pink, Black, Blue) is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

    Device Description

    The subject device is powder free vinyl examination glove. The subject device's color is pink, black and blue. It can be available in four specifications: S, M, L and XL. The subject device is non-sterile.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification for Disposable Vinyl Examination Gloves. It focuses on demonstrating substantial equivalence to a predicate device based on bench testing (non-clinical testing) rather than clinical studies involving human patients or complex AI algorithms.

    Therefore, many of the requested points related to AI, MRMC studies, expert ground truth adjudication, and clinical trial specifics are not applicable to this submission.

    Here's a breakdown of the requested information based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    Test MethodPurposeAcceptance CriteriaReported Device Performance
    ASTM D5250Physical Dimensions TestLength (mm): S/M/L/XL: ≥230Width (mm): S: 85±5; M: 95±5; L: 105±5; XL: 115±5Thickness (mm): Finger: ≥0.08, Palm: ≥0.08Length (mm): >230 / PassWidth (mm): S: 83-87 / Pass; M: 92-97 / Pass; L: 104-106 / Pass; XL: 112-117 / PassThickness (mm): Finger: 0.13-0.14 / Pass; Palm: 0.09-0.11 / Pass
    ASTM D5151Watertightness Test for Detection of HolesMeet the requirements of ASTM D5151 AQL 2.50/125 / Pass (This indicates 0 holes found in 125 samples tested)
    ASTM D6124Powder ContentMeet the requirements of ASTM D6124 < 2.0mg0.17~0.26mg / Pass
    ASTM D412Physical properties (Before Aging)Tensile StrengthUltimate ElongationBefore Aging:Tensile Strength: ≥11MPaUltimate Elongation: ≥300%Before Aging:Tensile Strength: 1222MPa / PassUltimate Elongation: 331589% / Pass
    ASTM D412Physical properties (After Aging)Tensile StrengthUltimate ElongationAfter Aging:Tensile Strength: ≥11MPaUltimate Elongation: ≥300%After Aging:Tensile Strength: 1119MPa / PassUltimate Elongation: 325544% / Pass
    ISO 10993-5CytotoxicityIn Vitro Cytotoxicity: Not cytotoxicUnder conditions of the study, device extract is not cytotoxic. / Pass
    ISO 10993-10IrritationNon-irritatingUnder the conditions of the study, not an irritant. / Pass
    ISO 10993-10SensitizationNon-sensitizingUnder conditions of the study, not a sensitizer. / Pass

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: The document explicitly mentions "0/125 / Pass" for the ASTM D5151 Watertightness Test, indicating a sample size of 125 for this test. For other "Pass" results, the exact sample sizes are not explicitly stated, but the tests are typically performed on a statistically relevant number of samples to demonstrate compliance with the ASTM/ISO standards.
    • Data Provenance: The tests are non-clinical (bench testing) performed by the manufacturer. The company is Anhui Bytech Medical Supplies Co., Ltd., located in Anhui Province, China. The data is from the manufacturer's internal testing as part of their premarket notification. The document indicates these are "non-clinical tests."

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    • Not Applicable. This is a medical device 510(k) submission for examination gloves, which are low-risk devices. The "ground truth" for proving the device meets acceptance criteria comes from standardized laboratory bench testing (e.g., measuring physical properties, mechanical strength, and biocompatibility), not from expert interpretation of medical images or clinical outcomes. Therefore, no medical experts were involved in establishing "ground truth" for the performance tests.

    4. Adjudication Method for the Test Set

    • Not Applicable. As the "ground truth" is established via objective physical and chemical measurements (bench testing) according to international standards (ASTM, ISO), there is no need for expert adjudication. The results are quantitative and directly compared to pre-defined acceptance criteria.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, and Effect Size of Human Readers Improvement with AI vs. Without AI Assistance

    • Not Applicable. This device is an examination glove, not an AI-powered diagnostic or assistive tool. MRMC studies are relevant for evaluating the impact of AI on human readers' performance in tasks like image interpretation, which is not pertinent to this device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    • Not Applicable. There is no AI algorithm involved with this device; it is a physical product (gloves).

    7. The Type of Ground Truth Used

    • The "ground truth" for this device's performance is based on objective, quantitative measurements from standardized laboratory tests according to recognized international consensus standards (ASTM and ISO). Examples include:
      • Direct measurement of dimensions (length, width, thickness).
      • Quantification of tensile strength and ultimate elongation.
      • Detection of holes using a water integrity test.
      • Measurement of residual powder content.
      • Biological performance testing (cytotoxicity, irritation, sensitization) in vitro, which involves observing cellular responses under controlled lab conditions.

    8. The Sample Size for the Training Set

    • Not Applicable. This submission is for a physical medical device (gloves), not an AI/machine learning model. Therefore, there is no "training set" in the context of machine learning. The manufacturing process itself involves quality control and process validation, which might involve sample testing, but this is distinct from "training data" for an algorithm.

    9. How the Ground Truth for the Training Set was Established

    • Not Applicable. As there is no training set for an AI model, this question is not relevant to the provided document.
    Ask a Question

    Ask a specific question about this device

    K Number
    K201504

    Validate with FDA (Live)

    Date Cleared
    2021-05-10

    (339 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Single-use medical poly (vinyl chloride) examination glove is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

    Device Description

    The proposed device is Powder Free Vinyl Patient Examination Gloves which is made of poly vinyl chloride. The proposed device is clear, non-colored. The design of proposed device is addressing the standards as ASTM D6124, ASTM D5151, and ASTM D5250. The proposed device is non-sterile.

    AI/ML Overview

    This document is a 510(k) Pre-Market Notification for a medical device: "Single-use medical poly (vinyl chloride) examination glove." It's primarily a comparison to a predicate device to establish substantial equivalence, not a study proving device meeting acceptance criteria in the context of advanced AI/medical imaging devices. Therefore, many of the requested categories are not applicable.

    Here's an analysis based on the provided text, focusing on the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally based on meeting the specifications of various ASTM and ISO standards for medical examination gloves. The reported device performance indicates that the proposed device meets these criteria.

    Test ItemAcceptance Criteria (Pass Criteria)Reported Device Performance (Verdict)
    Dimension (mm)ASTM D5250Pass
    Overall LengthMin 230>230/ Pass
    WidthXS: 76±6; S: 85±5; M: 95±5; L: 105±5; XL: 115±5XS: 79-81/ Pass; S: 85-87 /Pass; M: 95-97/ Pass; L: 105-107/ Pass; XL: 115-117/ Pass
    Finger thickness0.050.09-0.12/ Pass
    Palm thickness0.080.08-0.09/ Pass
    Freedom from HolesMeet the requirements of ASTM D5151 AQL 1.50/125, 1/ 125, 4/315, 5/500, 3/200 leaks / Pass (Implies AQL 1.5 met)
    Powder ContentMeet the requirements of ASTM D6124 < 2.0mg0.3 – 0.42/ Pass
    Physical PropertiesASTM D412Pass
    Tensile Strength (MPa)Meet the requirements of ASTM D412 Min 1114-19/ Pass
    Ultimate Elongation (%)Meet the requirements of ASTM D412 Min 300%390 – 430/ Pass
    Biocompatibility
    IrritationNon-irritating (ISO 10993-10)Under the conditions of the study, not an irritant/ Pass
    SensitizationNon-sensitizing (ISO 10993-10)Under conditions of the study, not a sensitizer./ Pass
    CytotoxicityNon-cytotoxic (ISO 10993-5)Under conditions of the study, did not show potential toxicity to L-929 cells./ Pass

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: The document provides sample sizes for the "Freedom from Holes" test (e.g., 0/125, 1/125, 4/315, 5/500, 3/200 leaks). These numbers likely refer to the number of failed units out of the total tested for specific batches or sample groups to demonstrate compliance with AQL (Acceptable Quality Level) 1.5. For other tests like Dimensions, Powder Content, Physical Properties, and Biocompatibility, specific sample numbers are not explicitly stated, but are implied to be sufficient for compliance with the stated ASTM and ISO standards.
    • Data Provenance: The document does not specify the country of origin of the data. Given the applicant is Anhui Bytech Medical Supplies Co., Ltd. in China and the submission correspondent is Shanghai SUNGO Management Consulting Co., Ltd. also in China, it's highly probable the testing was conducted in China. The data is based on non-clinical performance testing, effectively a prospective quality control assessment of manufactured gloves against established standards.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

    • This section is not applicable to this type of device submission. The "ground truth" for these tests are objective measurements against established technical standards (ASTM, ISO), not expert consensus on medical interpretation.

    4. Adjudication Method for the Test Set

    • This section is not applicable. The tests performed are objective measurements against defined criteria in standards, not subjective assessments requiring adjudication.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

    • No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic imaging devices where human readers interpret results with and without AI assistance. This document describes a physical medical device (examination glove).

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is not an AI algorithm. The device is a physical glove. Performance is measured directly from the physical characteristics of the glove.

    7. The Type of Ground Truth Used

    • The "ground truth" for the non-clinical performance and biocompatibility tests are the specifications and acceptable limits defined within the referenced international standards (ASTM D5250, ASTM D5151, ASTM D6124, ASTM D412, ISO 10993-10, ISO 10993-5). This is an objective, standardized basis for evaluation.

    8. The Sample Size for the Training Set

    • This concept is not applicable here. There is no "training set" in the context of an AI/machine learning model. The device's manufacturing process is designed to produce gloves meeting specifications, not "trained" on data.

    9. How the Ground Truth for the Training Set was Established

    • Not applicable, as there is no training set for this device.