510(k) Data Aggregation

    K Number
    K212073

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-07-14

    (377 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Alma Diode Tabletop Laser is intended for use in dermatologic and general surgical procedures.
    Indications for Use
    The Alma Diode Tabletop Laser includes three possible diode laser modules depending on the customer order.
    Diode Laser Modules:
    The indications for use for the 810 nm Alma Diode Tabletop Laser include:
    -The Alma 810 rm diode tabletop laser is indicated for endoluminal or endovenous laser surgery for saphenous incompetent veins.
    The indications for use for the 980 nm Alma Diode Tabletop Laser include:
    -The Alma 980 nm diode tabletop laser is indicated for use in endovenous occlusion of the greater saphenous vein in patients with superficial vein reflux. The Alma 980 mm diode tabletop laser assisted lipolysis.
    The indications for use for the 1470 nm Alma Diode Tabletop Laser include:
    -The Alma 1470 nm diode tabletop laser is indicated for use in endovenous occlusion of the greater saphenous vein in patients with superficial vein reflux. The Alma 1470 nm diode tabletop laser is further indicated for laser assisted lipolysis.

    Device Description

    The Diode Tabletop System is a small, tabletop laser console that can be sold with one of three wavelengths - 1470nm, 980mm or 810mm. It is comprised of 4 major components. The main console unit contains the laser moduler, cooling and user interface. The Footswitch is used to activate the laser. The diode laser modules are contained in the main console and emit laser energy in the required wavelength. All of the accessories connect via a fiber to the SMA connection port on the laser console. A single use, radial emitting fiber will be provided sterile by an OEM manufacturer. The fiber is sterilized using ETO. It is for prescription use only.

    AI/ML Overview

    The provided text pertains to a 510(k) premarket notification for a medical device called the "Alma Diode Tabletop Laser." It describes the device, its intended use, comparison to a predicate device, and the non-clinical testing performed.

    However, the document does not contain information related to acceptance criteria, device performance against those criteria, sample sizes for test or training sets, data provenance, expert involvement for ground truth, adjudication methods, multi-reader multi-case studies, or standalone algorithm performance, as would be expected for a product relying on AI/ML or complex diagnostic capabilities.

    The core of this 510(k) submission is for a laser surgical instrument, and the significant change described is the addition of a temperature sensing thermistor to a cannula used during laser lipolysis. The testing for this modification focused on bench testing to verify the performance of the temperature sensor and its accuracy in an ex vivo porcine model, rather than complex AI/ML performance metrics.

    Therefore, many of the requested elements of your prompt are not applicable or cannot be extracted from this specific document.

    Here's what can be extracted based on the provided text, and where gaps exist:


    Acceptance Criteria and Study Details for the Alma Diode Tabletop Laser (K212073)

    Summary of Device and Key Change:
    The Alma Diode Tabletop Laser is a laser surgical instrument used in dermatologic and general surgical procedures, specifically for saphenous incompetent veins and laser-assisted lipolysis. The K212073 submission describes a modification to a previously cleared device (K160952), primarily the addition of a temperature-sensing thermistor to a cannula used during laser lipolysis procedures. This thermistor is designed to detect tissue temperature in near real-time and alert the user if a preset temperature is exceeded.

    Study Type:
    The study described is non-clinical bench testing to verify the performance of the modified device, particularly the added temperature sensing functionality.

    Missing Information:

    • This document does not describe the testing of an AI/ML component. Therefore, acceptance criteria regarding AI diagnostic performance (e.g., sensitivity, specificity, AUC) are not present.
    • No human reader studies (MRMC) were conducted as the device's function does not involve diagnostic interpretation by human readers.
    • No standalone algorithm performance metrics are provided as it's not an AI-driven diagnostic algorithm.
    • Ground truth establishment methods (expert consensus, pathology, outcome data) are not relevant in the context of this device's functional testing.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not present a formal table of quantitative acceptance criteria and corresponding reported performance values for properties like sensitivity, specificity, or similar metrics typically associated with AI/ML products. Instead, the performance testing described relates to the functionality of the new temperature sensor.

    Acceptance Criteria (Inferred from testing)Reported Device Performance
    Demonstrate time required for temperature sensor to register tissue temperature and notify user.Performed (details not explicitly quantified in this summary)
    Demonstrate accuracy of temperature detection in an ex vivo porcine model.Performed (details not explicitly quantified in this summary)
    Demonstrate relationship between tissue temperature at thermistor and temperature at a specified distance from the thermistor (in ex vivo porcine model).Performed (details not explicitly quantified in this summary)
    Meet general safety and performance standards for medical electrical equipment and laser products.Passed (compliant with IEC 60601-1, -1-2, -2-22; IEC 60825-1; ISO 17665-1; ISO 10993)
    Software Validation & Verification TestPassed
    Biocompatibility evaluationPassed

    2. Sample Size Used for the Test Set and Data Provenance

    The document mentions "ex vivo porcine model" for performance testing of the temperature sensor.

    • Sample Size: Not specified (e.g., number of porcine samples or measurements).
    • Data Provenance: Ex vivo porcine model (experimental, not human clinical retrospective or prospective data). The country of origin for the data is not specified.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. The ground truth for the performance of a temperature sensor is typically established by calibrated reference instruments, not expert human readers or consensus, especially in an ex vivo setting.


    4. Adjudication Method for the Test Set

    Not applicable. Adjudication methods (e.g., 2+1, 3+1) are used for resolving disagreements among human readers in diagnostic studies, which is not relevant here.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. An MRMC study was not conducted. This device is a surgical laser, and the testing described is not related to human diagnostic performance or AI assistance in interpretation.


    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    No. This is not an AI/ML algorithm or diagnostic device tested for standalone performance. The testing involved the functional performance of a physical component (temperature sensor) within a laser system.


    7. The Type of Ground Truth Used

    The ground truth for the temperature sensor's performance would be derived from calibrated temperature measurement devices in the ex vivo porcine model, comparing the thermistor's readings against these reference measurements. This is a type of "bench testing" or "experimental setup" ground truth, not expert consensus, pathology, or outcomes data.


    8. The Sample Size for the Training Set

    Not applicable. This device is not an AI/ML product that undergoes a training phase with a dedicated dataset.


    9. How the Ground Truth for the Training Set was Established

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K203441

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-07-02

    (221 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Alma Hybrid Laser System, Delivery Devices, Applicators and Accessories are intended for use in surgical applications requiring the ablation, vaporization, incision, and coagulation of soft tissue in medical specialties including: aesthetic surgery (dermatology and plastic surgery).

    HyLight-CO2

    The Alma Hybrid CO2 non-fractional applicator, with wavelength of 10600 nm is cleared for use for the particular indications as follows:

    Dermatology & Plastic Surgery

    The ablation, vaporization, excision, and coagulation of soft tissue in dermatology and plastic surgery in the performance of:

    • laser skin resurfacing
    • laser derm-abrasion
    • laser burn debridement.

    Laser skin resurfacing (ablation and/or vaporization) of soft tissue for the reduction, removal, and/or treatment of: warts, acne scars, nevi epidermal, syringoma. Vaporization/coagulation of warts.

    Pixel

    The Alma Hybrid Pixel CO2 fractional applicator, with wavelength of 10600 nm is indicated for: The ablation, vaporization, and coagulation of soft tissue in dermatology and plastic surgery in the performance of skin resurfacing.

    ProScan

    The Alma Hybrid ProScan CO2 fractional applicator, with wavelength of 10600 nm is indicated for:

    · Laser skin resurfacing (ablation and/or vaporization) of soft tissue.

    The Alma Hybrid ProScan 1570nm fractional applicator, with wavelength of 1570 nm, is indicated for:

    · Use in dermatological procedures requiring fractional skin resurfacing and coagulation of soft tissue

    The Alma Hybrid ProScan CO2 &1570nm fractional applicator, with wavelengths of 10600 nm & indicated for laser skin resurfacing (ablation and/or vaporization) of soft tissue.

    Device Description

    The Alma Hybrid Laser System consists of:

    • Laser system console (containing the optical bench assembly and laser, the ● microcontroller control electronics and system software, the high voltage power supply, the laser cooling system, the compressed air-purge system, and the service panel)
    • LCD control panel with touch-screen technology: the LCD display provides information . on the status and settings of the The Alma Hybrid Laser System with touch-screen technology provided to input commands into the system.
    • Two wavelengths – CO2 and 1570 nm
    • . Articulated arm
    • Footswitch
    • Delivery devices (Pixel applicators, Hylight applicators, Proscan applicator) ●
    AI/ML Overview

    The provided text is a 510(k) Pre-Market Notification for the Alma Hybrid Laser System. Within this document, the main objective is to demonstrate substantial equivalence to predicate devices, rather than establishing new efficacy or safety criteria for the device itself. Therefore, it does not contain a typical "acceptance criteria" table for device performance against specific metrics, nor does it detail a clinical study designed to prove the device meets such criteria in the way one might expect for a novel device.

    However, based on the document, we can infer some "acceptance criteria" through the comparisons made with predicate devices and the performance testing conducted. The study described focuses on histology testing on porcine animals to justify technical differences in some applicators.

    Here's an attempt to extract and frame the information requested, with the understanding that the document's purpose is a 510(k) submission:

    1. Table of Acceptance Criteria and Reported Device Performance

    Since explicit acceptance criteria for efficacy in humans are not provided (as it's a substantial equivalence claim), the "acceptance criteria" are implied by demonstrating similar technical characteristics to the predicate devices and confirming tissue effects in an animal model meet expected physiological responses.

    Acceptance Criteria (Implied)Reported Device Performance (Alma Hybrid Laser System)
    Technical Equivalence to Predicate Devices:
    - Wavelengths- HyLight CO2 & Pixel CO2: 10600 nm (Same as predicate)
    - ProScan 1570nm: 1570 nm (Close to predicate's 1565 nm)
    - Output Power/Energy- HyLight CO2: 30W, 70W (Same as predicate)
    - Pixel CO2: 5mJ/Pixel - 150 mJ/Pixel (Same as predicate)
    - ProScan CO2: Up to 1000 mJ (Predicate) vs 120 (30W model)/240 mJ (70W model) (The document states "Output energy" vs "Up to 1000 mJ" for predicate. It also later clarifies "max energy output is within the range of the predicate device" for Pixel CO2, and for ProScan CO2, differences in spot size, beam density and pulse duration are noted, with histology supporting these differences.)
    - ProScan 1570nm: 24mJ-144 mJ/pixel (Predicate: 10 mJ per beam – 70 mJ per beam. Hybrid system has different range, but is considered substantially equivalent.)
    - Spot Size, Repetition Rate, Pulse Duration, Fluence, Beam DensityDifferences noted for some applicators, but documented as being within acceptable ranges or justified by performance testing (e.g., histology). For HyLight-CO2, spot size, pulse duration, and output energy differ, but fluence is within predicate range. For Pixel CO2, pulse duration differs but max energy output is within predicate range. For ProScan, differences in spot size, beam density, and pulse duration are noted.
    Safety and Performance Standards Compliance:- Complied with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, IEC 60825-1, ISO 10993-1, ISO 17665-1, IEC 62304, ISO 14971.
    Expected Histological Tissue Response (for ProScan Applicator):- Histology results complied with FDA requirements at 0, 3, 7, and 14 days.
    - Re-epithelialization observed three days after radiation in all specimens.
    - No adverse events or unexpected complications in the swines.
    Device Functionality:- Functioned as intended and results observed were as expected.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set (Histology Study): 3 Domestic female (Mixed Landrace & Large White) crossbred swine.
    • Data Provenance: The study was implicitly prospective for the animal testing, conducted by Alma Lasers. The document does not specify the country of origin of the data beyond "Alma also performed histology testing on porcine animals."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    The document does not specify the number of experts or their qualifications for evaluating the histology results. It states, "histology results complied with the FDA requirements." This implies interpretation by qualified individuals, but no details are given.

    4. Adjudication Method for the Test Set

    The document does not describe an adjudication method for the animal histology results. It simply states that the "histology results complied with the FDA requirements" and "Re-epithelialization was observed three days after radiation in all specimens." This suggests a direct evaluation against predefined criteria rather than a consensus process among multiple independent reviewers.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. The document explicitly states: "Based on the similarities in the safety and effectiveness profiles of the subject and the predicate, no human clinical studies were deemed needed to support this submission." The study described is an animal histology study, not a human reader study.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable. The device is a laser system, not an AI or algorithm-based diagnostic tool. The performance described relates to the physical and biological effects of the laser, not a computational algorithm's performance.

    7. The Type of Ground Truth Used

    For the animal study, the ground truth was histology from punch biopsies taken from the radiated points at 0, 3, 7, and 14 days, examined for re-epithelialization and other tissue effects.

    8. The Sample Size for the Training Set

    This question is not applicable in the context of this 510(k) submission. The device is a physical laser system, not an AI or machine learning algorithm that requires a "training set" of data.

    9. How the Ground Truth for the Training Set Was Established

    This question is not applicable for the same reasons as point 8.