    K Number
    K213046

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-06-27

    (643 days)

    Product Code
    Regulation Number
    878.5000
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ace Connex™ Pre-Sutured Fascia is intended for use as a component in soft tissue surgical procedures where constructs including those with allograft tissues are used for reconstruction, replacement, or augmentation of the hip labrum.

    Device Description

    The Labrum Replacement and Augmentation Device, marketed as AceConnex™ Pre-Sutured Fascia subject device is a pre-assembled surgical construct comprised of allograft tissue and non-absorbable, synthetic suture. The allograft fascia lata tissue is terminally cleaned and disinfected using a proprietary process. The fascia lata tissue is pre-sutured with Force Fiber® ultra-high molecular weight polyethylene (UHMWPE) non-absorbable surgical suture cleared under K063778, and the device is terminally sterilized by low dose electron beam irradiation.

    The device has a smooth side for articulating against the rotating cartilage surface and a sutured side which is secured against the bone and comes in three sizes to accommodate the needs of surgeons performing segmental (40-60mm and 60-100mm) and full circumferential (100-140mm) augmentations or reconstructions. Each device contains two adjustable sections where the device may be cut to provide adjustment options so that the device may be trimmed to fit the needs of patient and surgeon.

    AI/ML Overview

    This document does not contain information about the acceptance criteria and study proving the device meets the acceptance criteria in the manner requested. The provided text is a 510(k) summary for the AceConnex™ Pre-Sutured Fascia, which focuses on demonstrating substantial equivalence to a predicate device rather than providing a detailed study report with specific acceptance criteria and performance data in a tabular format.

    Here's why the requested information cannot be extracted from this document:

    1. Table of Acceptance Criteria and Reported Device Performance: While the document states that "the subject device passed all product performance tests" and lists some characteristics (e.g., "suture pullout strength greater than the estimated force on the labrum in a healthy hip during jogging"), it does not provide a formal table with quantitative acceptance criteria and corresponding quantitative device performance results.
    2. Sample Size, Data Provenance, Expert Information, Adjudication Method, MRMC Study, Standalone Study, Ground Truth Type, Training Set Size, Training Set Ground Truth: None of these details, which are typical for clinical performance studies, are present in this 510(k) summary. This type of submission relies on demonstrating substantial equivalence through comparison to an already cleared predicate device, often supported by bench testing (product performance testing) rather than extensive clinical studies with human subjects and expert evaluations.

    Based on the provided text, the following information can be extracted:

    • Device Name: AceConnex™ Pre-Sutured Fascia
    • Device Type: Pre-assembled surgical construct comprised of allograft tissue and non-absorbable, synthetic suture.
    • Intended Use: As a component in soft tissue surgical procedures where constructs including those with allograft tissues are used for reconstruction, replacement, or augmentation of the hip labrum.

    Regarding the "study that proves the device meets the acceptance criteria," the document mentions "Product Performance Testing," which included:

    • Visual, physical, mechanical, and clinical properties assessment.
    • Outcomes reported (without specific quantitative acceptance criteria or detailed study methodology):
      • Device integrity maintained during surgical preparation and application.
      • Device length was adjustable.
      • Device conformed to the acetabular rim.
      • Device was securable with knotted and knotless suture anchors.
      • Suture pullout strength greater than the estimated force on the labrum in a healthy hip during jogging.
      • Retains necessary tensile strength after two years frozen storage.
      • Similar physical properties of surgeon fascia allografts.
    • Bacterial Endotoxin Testing: Resulted in less than 20 EU/Device, meeting bacterial endotoxin testing requirements (standards listed: USP chapter <85>, ANSI/AAMI ST72:2011, FDA Guidance for Industry: Pyrogen and Endotoxins Testing).

    In summary, the provided document does not contain the detailed clinical study information requested. It focuses on engineering and bench testing results to support substantial equivalence.

    Ask a Question

    Ask a specific question about this device

    K Number
    K170957

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-03-20

    (354 days)

    Product Code
    Regulation Number
    878.5000
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Pre-sutured allograft tendons are intended for use as a construct in anterior cruciate ligament and posterior cruciate ligament reconstruction.

    Device Description

    This device consists of a combination of tendons that have been cleaned and disinfected using a proprietary process. The tendons are pre-sutured and terminally sterilized by low dose electron beam irradiation. The device may include anterior tibialis, posterior tibialis, peroneus longus, semitendinosus and/or gracilis tendons and is pre-sutured with Force Fiber UHWMPE nonabsorbable surgical suture. Through a contractual agreement with Teleflex Medical, AlloSource has the license rights to manufacture pre-sutured tendons using their UHMWPE non-absorbable surgical suture cleared 15-Sept-2009 under pre-market clearance K092533. All products are provided sterile and for single patient use.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device called "ReConnex™ Pre-Sutured Tendon". It details the device's characteristics, intended use, and the justification for its substantial equivalence to a predicate device.

    However, the provided document does not contain any information about an AI/ML device, nor does it discuss acceptance criteria and study results in the context of an AI/ML algorithm's performance.

    The document primarily focuses on:

    • The regulatory classification of the device (Class II, surgical suture).
    • The components of the device (pre-sutured allograft tendons).
    • Its intended use (ACL/PCL reconstruction).
    • Comparison to a predicate device (Arthrex Suture Grafting Kit) and a reference device (Force Fiber UHWMPE Non-absorbable Surgical Suture).
    • Biocompatibility requirements (donor eligibility, infectious disease screening).
    • Performance testing related to physical properties of the tendon/suture construct (visual characteristics, tensile strength, suture pull-out, knot pull strength).
    • A cadaver feasibility study demonstrating biomechanical equivalence to a surgeon-sutured construct and feasibility of implantation.
    • Bacterial endotoxin testing.

    Therefore, I cannot fulfill the request to describe the acceptance criteria and the study that proves the device meets the acceptance criteria for an AI/ML device based on the provided text. The prompt's requirements (e.g., sample size for test set, data provenance, number of experts for ground truth, MRMC study, AI assistance effect size, standalone algorithm performance, training set details) are all relevant to the validation of an AI/ML system, which is not what this document is describing.

    Ask a Question

    Ask a specific question about this device

    K Number
    K103036

    Validate with FDA (Live)

    Device Name
    ALLOFUSE PLUS
    Manufacturer
    Date Cleared
    2011-01-10

    (89 days)

    Product Code
    Regulation Number
    888.3045
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For orthopedic use, AlloFuse Plus Paste and Putty are intended for use as an autograft extender (i.e. extremities, posterolateral spine and pelvis) and as a bone void filler (i.e. extremities and pelvis) for bony voids or gaps that are not intrinsic to the stability of the bony structure. The AlloFuse Plus products are indicated to be packed gently into bony defects of the skeletal system. These defects may be surgically created or from the result of traumatic injury to the bone.

    Device Description

    AlloFuse Plus is derived from selected donated human bone tissue that has been processed into particles. The bone particles are subsequently demineralized using a hydrochloric acid process. The demineralized bone matrix (DBM) is combined with a reverse phase carrier, cancellous chips from the same donor and formulated into a paste or putty-like consistency.

    The carrier is a solution of polvethylene oxide polypropylene oxide block copolymer dissolved in water exhibiting reverse phase characteristics (i.e. an increase in viscosity as temperature increases).

    AI/ML Overview

    This is a 510(k) premarket notification for the AlloFuse Plus Paste and Putty, which are resorbable calcium salt bone void filler devices. The submission focuses on demonstrating substantial equivalence to previously cleared predicate devices, rather than establishing de novo safety and effectiveness through extensive clinical trials.

    The provided document does not contain a traditional table of acceptance criteria and reported device performance in the manner one would expect for a diagnostic or AI-driven medical device. Instead, the acceptance criteria are implicitly tied to demonstrating substantial equivalence and meeting established manufacturing and biological safety standards for tissue-based products.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific CriteriaReported Device Performance / Evidence
    Technological EquivalenceDevice design, materials of construction, and function are the same as previously cleared predicate devices (K070751, K050642).The proposed device is stated to be "the same device in design, materials of construction and function as the previously cleared devices of 510(k) Notification K070751 cleared 15-Oct-2007 and K050642 cleared 05-Dec-2005." This is supported by a contractual agreement with IsoTis Orthobiologics for an exclusive license to manufacture and market.
    Biological Safety - Viral InactivationProcessing methods for Demineralized Bone Matrix (DBM) and cancellous chips, as well as final product sterilization (electron beam), demonstrate significant viral inactivation potential for a wide range of human viruses."The methods for processing the DBM and cancellous chips contained in AlloFuse Plus were evaluated for their viral inactivation potential as well as the electron beam sterilization process for final product. A select panel of viruses representing various virus types, sizes, shapes, and genomes was evaluated. Both the DBM processing method (demineralization), the proprietary cleaning process for the cancellous bone and the sterilization process were determined to provide significant viral inactivation potential for a wide range of potential human viruses."
    Biological Safety - Osteoninductive PotentialThe product exhibits osteoinductive potential."AlloFuse Plus Paste and Putty have been shown to have osteoinductive potential in athymic rats. Every lot of final product is tested via an in vivo assay to ensure osteoinductive potential of the final product." (Note: The document explicitly states this should not be interpreted to predict clinical performance in humans).
    Biological Safety - Endotoxin/LALThe product meets endotoxin/LAL requirements."Product safety and effectiveness is adequately supported by the substantial equivalence information and test data including osteoinductive potential, viral inactivation and endotoxin/LAL provided in this Premarket Notification." (This is a summary statement, specific data is not provided in the excerpt).
    OsteoconductivityThe device is osteoconductive."The proposed and predicate devices are osteoconductive..." (This is a declarative statement of characteristic, not a test result).

    2. Sample Size Used for the Test Set and Data Provenance

    • Viral Inactivation Validation: "A select panel of viruses representing various virus types, sizes, shapes, and genomes was evaluated." The exact number of viruses and the specific experimental setup (e.g., how many replicates, sample sizes per condition) are not specified in this document. This is a laboratory-based study, not a human clinical test.
    • Osteoinductive Potential: The test set involves "athymic rats." The exact number of rats used in the initial validation is not specified. For ongoing lot testing, it states "Every lot of final product is tested via an in vivo assay."
    • Data Provenance: The studies are laboratory and animal studies conducted for regulatory submission purposes. The geographic origin of the lab or specific animal facility is not mentioned. These are prospective, controlled laboratory/animal studies designed to evaluate specific biological characteristics.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    There is no mention of experts being used to establish "ground truth" in the context of human clinical data or image interpretation for this device. The evaluations described (viral inactivation, osteoinductivity, endotoxin/LAL) are laboratory or animal-based analyses typically performed by trained scientists and technicians specializing in those respective fields (e.g., virologists, histologists, toxicologists). These are not studies that require expert adjudication of clinical outcomes or images.

    4. Adjudication Method for the Test Set

    Not applicable. The studies described are laboratory and animal experiments with objective readouts (e.g., viral titer reduction, bone formation presence, endotoxin levels), not studies requiring human rater adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    Not applicable. This is not an AI-driven or diagnostic device. No MRMC study was conducted or mentioned.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This is not an algorithm or AI device.

    7. The Type of Ground Truth Used

    • Viral Inactivation: The "ground truth" would be established by standard virological assays (e.g., cell culture infectivity assays) to quantify viral reduction logs.
    • Osteoinductive Potential: The "ground truth" is typically established by histological examination of tissue sections from the athymic rat implants to confirm the presence of new bone formation (ectopic bone).
    • Endotoxin/LAL: The "ground truth" is measured quantitatively using Limulus Amebocyte Lysate (LAL) assay, against established limits.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/ML device that requires a training set. The "training" in this context refers to the development and optimization of the manufacturing process (demineralization, cleaning, formulation) that enables the described performance.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" in the AI/ML sense. The process parameters (e.g., demineralization time, temperatures, concentrations) for manufacturing DBM and the final product are established through extensive research and development work, guided by scientific principles and iterative testing to achieve desired product characteristics (e.g., demineralization, retention of growth factors, sterility, rheology). This involves rigorous process validation, where various parameters are tested and optimized to consistently produce a product meeting specifications.

    Ask a Question

    Ask a specific question about this device

    K Number
    K071849

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2008-12-04

    (518 days)

    Product Code
    Regulation Number
    888.3045
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    AlloFuse is indicated for orthopedic applications as filler for gaps or voids that are not intrinsic to the stability of the bony structure. AlloFuse® is indicated to be packed gently into bony gaps in the skeletal system as a bone graft extender (extremities, spine, and pelvis) and as bony void filler of the extremities and pelvis. These defects may be surgically created or from the result of traumatic injury to the bone.

    Device Description

    AlloFuse is derived from selected donated human bone tissue that has been processed into particles. The bone particles are subsequently demineralized using a hydrochloric acid process. The demineralized bone matrix (DBM) is combined with an inert reverse phase carrier and formulated into a gel or putty-like consistency.

    The carrier is a solution of polyethylene oxide polypropylene oxide block copolymer dissolved in water exhibiting reverse phase characteristics (i.e. an increase in viscosity as temperature increases).

    AI/ML Overview

    Here's an analysis of the provided text, outlining the acceptance criteria and the studies mentioned for the AlloFuse Gel and Putty device:

    Executive Summary:

    The provided document describes the 510(k) submission for AlloFuse Gel and Putty, which claims substantial equivalence to the predicate device, DynaGraft® II. The primary "acceptance criteria" is demonstrating substantial equivalence due to the proposed device being the same in design, materials, and function as the predicate device. This is supported by:

    1. Viral Inactivation Validation: Testing of the DBM processing method to ensure viral inactivation.
    2. Osteoinductive Potential: An in vivo assay in athymic rats confirms osteoinductive potential for every lot.
    3. Product Performance Testing: Reference to studies on the predicate device (DynaGraft® II) in rabbit and sheep models (radiographic and histological) and human clinical studies for spinal fusions.

    1. Table of Acceptance Criteria and Reported Device Performance

    It's important to note that the document does not explicitly present a table of numerical "acceptance criteria" and direct performance metrics for the AlloFuse device itself in the way one might see for a diagnostic AI model. Instead, the acceptance is based on demonstrating substantial equivalence to a predicate device. The performance data presented primarily pertains to the predicate device or the processing methods.

    Acceptance Criterion (Implicit)Reported Device Performance
    Premarket Requirement: Demonstrated Substantial Equivalence to Predicate Device (DynaGraft® II)The proposed device is stated to be "the same device in design, materials of construction and function" as the previously cleared DynaGraft® II. AlloSource also holds an exclusive license to manufacture the predicate device. This is the primary claim for substantial equivalence.
    Material Safety (Viral Inactivation)The DBM processing method (for AlloFuse) was evaluated for its viral inactivation potential, demonstrating "significant viral inactivation potential for a wide range of potential human viruses" using a panel of viruses.
    Biological Activity (Osteoinductive Potential)AlloFuse has shown "osteoinductive potential in athymic rats." "Every lot of final product is tested via an in vivo assay to ensure osteoinductive potential."
    Clinical Performance / EfficacyNot directly tested on AlloFuse in human trials for this submission. Instead, the submission relies on: - Product Performance Testing (Predicate): DBM in RPM carrier (DynaGraft® II formulation) evaluated in rabbit and sheep models by radiographic and histological methods. - Clinical Studies (Predicate): Clinical studies using DynaGraft® II DBM Putty and Gel for spinal fusions "demonstrating acceptable outcomes."
    Sterility & Donor Tissue QualityAlloFuse is provided sterile and for single patient use. Donor bone meets AATB requirements.

    2. Sample Size Used for the Test Set and Data Provenance

    Due to the nature of this 510(k) submission, direct "test sets" for clinical performance of the new device (AlloFuse) are not reported. The submission relies on data from the predicate device and specific lab tests for the new device:

    • Viral Inactivation Validation:
      • Sample Size: Not specified for the "select panel of viruses," but assumed to be multiple viruses representing various types.
      • Provenance: Lab-based (in-vitro) testing of the DBM processing method.
    • Osteoinductive Potential:
      • Sample Size: Assumed to be adequate for "an in vivo assay to ensure osteoinductive potential of the final product," performed for every lot of final product.
      • Provenance: Athymic rat model (animal study), conducted by the manufacturer.
    • Product Performance Testing (Predicate Device - DynaGraft® II):
      • Sample Size: Not specified for the rabbit and sheep models.
      • Provenance: Animal models (rabbit, sheep), nature (retrospective/prospective) not specified but typically prospective for such studies.
    • Clinical Studies (Predicate Device - DynaGraft® II):
      • Sample Size: Not specified.
      • Provenance: Human clinical studies, typically prospective. Country of origin not specified, but likely US-based given the FDA submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Given this is a 510(k) for a device claiming substantial equivalence rather than a new drug or complex diagnostic, explicit "ground truth" established by a panel of human experts for a specific test set (like an AI study) is not detailed.

    • Viral Inactivation: Ground truth would be based on established virology protocols and detection methods, likely performed by qualified microbiologists/virologists in a laboratory setting.
    • Osteoinductive Potential: Ground truth is established by histological and potentially radiographic evaluation of new bone formation in the athymic rat model, assessed by experts in pathology and histology.
    • Product Performance Testing (Predicate - Animal Models): Ground truth was established by radiographic and histological methods, interpreted by relevant veterinary or medical experts (e.g., veterinary radiologists, pathologists).
    • Clinical Studies (Predicate - Human): Ground truth (acceptable outcomes in spinal fusions) would have been established by clinicians (e.g., orthopedic surgeons, radiologists) based on follow-up examinations, imaging, and patient outcomes.

    4. Adjudication Method for the Test Set

    No explicit adjudication method (e.g., 2+1, 3+1) is mentioned, as this is not an AI diagnostic study relying on human reader consensus. Outcomes from animal and human studies would be assessed via standard clinical or laboratory practices, which often involve multiple reviewers but not necessarily a formal adjudication process as seen in AI studies.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study was done. This 510(k) is for a bone void filler device, not an AI diagnostic tool. Therefore, there's no discussion of human readers improving with or without AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This submission is for a medical device (bone void filler), not an algorithm or AI system.

    7. The Type of Ground Truth Used

    • Viral Inactivation Validation: Established lab-based measurements of viral reduction.
    • Osteoinductive Potential: Histological evidence of new bone formation in an in vivo athymic rat model.
    • Product Performance Testing (Predicate): Radiographic and histological findings in animal models.
    • Clinical Studies (Predicate): Clinical outcomes (e.g., successful spinal fusion) and patient follow-up data.

    8. The Sample Size for the Training Set

    Not applicable. There is no "training set" in the context of an AI algorithm for this device. The development of the manufacturing processes and material formulation would be based on general scientific knowledge and prior product development, not a labeled dataset for training an algorithm.

    9. How the Ground Truth for the Training Set was Established

    Not applicable. As there is no AI training set, there is no ground truth established for it.