510(k) Data Aggregation
(63 days)
The Aevumed FENIX™ Suture Anchors are intended to be used for suture or tissue fixation in the foot, ankle, knee, hand, wrist, elbow, shoulder, and hip. Specific indications are listed below:
Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clabicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction
Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Metatarsal Ligament Repair, Hallux Valgus Reconstruction, Digital Tendon Transfers, Mid-foot Reconstruction
Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis
Hand/Wrist: Scapholunate Ligament Reconstruction, Carpal Ligament Reconstruction, Repair/Reconstruction of Collateral Ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP, and MCP joints for all Digits, Digital Tendon Transfers
Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction
Hip: Capsular repair, Acetabular Labral Repair
The Aevumed FENIX™ Sutures Anchor is a suture anchor manufactured from HS Fiber™ UHMWPE/PET material and are preloaded on a disposable inserter assembly intended for fixation of soft tissue to bone. The Aevumed FENIX™ Suture Anchors are available in diameter sizes: 2.6mm and 1.8mm. They are offered sterile and are for single use only.
N/A
Ask a specific question about this device
(26 days)
The Aevumed PROTEKT™ Suture Anchors are intended to be used for suture or tissue fixation in the foot, ankle, knee, hand, wrist, elbow, shoulder, and hip. Specific indications are listed below:
Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clabicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction
Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Metatarsal Ligament Repair, Hallux Valgus Reconstruction, Digital Tendon Transfers, Mid-foot Reconstruction
Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis
Hand/Wrist: Scapholunate Ligament Reconstruction, Carpal Ligament Reconstruction, Repair/Reconstruction of Collateral Ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP, and MCP joints for all Digits, Digital Tendon Transfers
Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction
Hip: Capsular repair, Acetabular Labral Repair
The Aevumed PROTEKT™ Sutures Anchor with HS Fiber™ suture is a suture anchor manufactured from polyetheretherketone (PEEK) material and are preloaded on a disposable inserter assembly intended for fixation of soft tissue to bone. The Aevumed PROTEKT™ Suture Anchors is available in diameter size: 6.5 mm. It is offered sterile and is for single use only.
N/A
Ask a specific question about this device
(29 days)
The Aevumed FASE™ Suture Anchors are intended to be used for suture or tissue fixation in the foot, ankle, knee, hand, wrist, elbow, shoulder, and hip. Specific indications are listed below:
Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clabicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction
Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Metatarsal Ligament Repair, Hallux Valgus Reconstruction, Digital Tendon Transfers, Mid-foot Reconstruction
Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis
Hand/Wrist: Scapholunate Ligament Reconstruction, Carpal Ligament Reconstruction, Repair/Reconstruction of Collateral Ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP, and MCP joints for all Digits, Digital Tendon Transfers
Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction
Hip: Capsular repair, Acetabular Labral Repair
The Aevumed FASE™ Sutures Anchor with HS Fiber™ suture is a suture anchor manufactured from polyetheretherketone (PEEK) material and are preloaded on a disposable inserter assembly intended for fixation of soft tissue to bone. The Aevumed FASE™ Suture Anchors are available in diameter sizes: 4.75mm and 5.5mm. They are offered sterile and are for single use only.
N/A
Ask a specific question about this device
(317 days)
The Aevumed PHANTOM™ Suture Anchors are intended to be used for suture or tissue fixation in the foot, ankle, knee, hand, wrist, elbow, shoulder, and hip. Specific indications are listed below:
Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clabicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction
Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Metatarsal Ligament Repair, Hallux Valgus Reconstruction, Digital Tendon Transfers, Mid-foot Reconstruction
Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis
Hand/Wrist: Scapholunate Ligament Reconstruction, Carpal Ligament Reconstruction, Repair/Reconstruction of Collateral Ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP, and MCP joints for all Digits, Digital Tendon Transfers
Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction
Hip: Capsular repair, Acetabular Labral Repair
The Aevumed PHANTOM™ Sutures Anchor with HS Fiber™ suture is a suture anchor manufactured from polyetheretherketone (PEEK) material and are preloaded on a disposable inserter assembly intended for fixation of soft tissue to bone. The Aevumed PHANTOM™ Suture Anchors are available in diameter sizes: 3.5mm and 4.5mm. They are offered sterile and are for single use only.
The document provided is an FDA 510(k) clearance letter for the Aevumed PHANTOM Suture Anchors. This type of device (a bone fixation fastener) is a physical implant and not an AI/ML-driven diagnostic or assistive technology. Therefore, the questions related to AI/ML device performance (such as acceptance criteria for AI performance, sample size for test sets, data provenance, expert adjudication, MRMC studies, standalone performance, ground truth establishment for training and test sets, and training set size) are not applicable to this submission.
The 510(k) summary focuses on demonstrating substantial equivalence to a predicate device (Aevumed RAPID Suture Anchors, K223878) based on non-clinical performance and material characteristics, rather than diagnostic accuracy or AI effectiveness.
Here's an analysis based on the information provided, tailored to a medical device clearance:
1. Acceptance Criteria and Reported Device Performance
| Acceptance Criteria (General for Suture Anchors) | Reported Device Performance (Aevumed PHANTOM Suture Anchor) |
|---|---|
| Mechanical Performance: | Maximum Pullout Strength: Demonstrated significantly higher pullout strength compared to the predicate RAPID™ Anchor (K223878). |
| - Adequate pullout strength | |
| - Cyclic tensile strength | Tested. No specific values provided, but implied to be acceptable and comparable or superior to predicate. |
| - Static tensile strength | Tested. No specific values provided, but implied to be acceptable and comparable or superior to predicate. |
| Material: | Manufactured from Polyetheretherketone (PEEK). Identical to predicate. |
| - Biocompatibility | Deemed identical to the predicate device (Aevumed RAPID™ Suture Anchor, K223878). |
| Sterilization: | Identical to the predicate device. |
| Packaging: | Identical to the predicate device. |
| Pyrogenicity/Endotoxin Monitoring: | Identical to the predicate device. |
| Shelf Life: | Identical to the predicate device. |
| Intended Use/Indications: | Identical to predicate for suture or tissue fixation in various anatomical locations (shoulder, foot/ankle, knee, hand/wrist, elbow, hip) |
| Operational Principle: | Identical to predicate. |
| Geometric Design: | Similar to predicate. Minor differences do not raise new questions of safety and effectiveness. Available in 3.5mm and 4.5mm diameters. |
Study Proving Device Meets Acceptance Criteria:
The study conducted was a non-clinical mechanical testing study. This involved direct comparative testing of the Aevumed PHANTOM™ Suture Anchor against its primary predicate device, the Aevumed RAPID™ Suture Anchor (K223878).
- Tests performed: Insertion testing, cyclic tensile testing, and static tensile testing.
- Key finding: The Aevumed PHANTOM™ Suture Anchor demonstrated "significantly higher pullout strength" in comparison to the predicate RAPID™ Anchor.
- Conclusion drawn: The data supports the relative safety of the device and demonstrates that it performs as intended and comparably (or superiorly in key mechanical aspects) to the predicate device.
2. Sample Size Used for the Test Set and Data Provenance:
- Sample Size: The document does not specify the exact number of devices (suture anchors) used in the non-clinical mechanical tests. These tests typically involve a statistically significant number of samples to ensure repeatability and reliability of the results.
- Data Provenance: The data is generated from laboratory-based non-clinical mechanical testing. This means the "data provenance" isn't in terms of patient population (e.g., country of origin, retrospective/prospective) but rather controlled experimental conditions.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:
- Not Applicable. For a physical medical device like a suture anchor, "ground truth" for non-clinical mechanical testing is established by engineering standards and validated testing methodologies, not by expert interpretation of patient data or images. The "ground truth" would be the measured physical properties (e.g., pullout strength).
4. Adjudication Method for the Test Set:
- Not Applicable. As mentioned above, this device does not involve human interpretation of data for its primary evaluation for 510(k) clearance. Mechanical tests have quantitative outputs, not subjective interpretations requiring adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done:
- No. An MRMC effectiveness study is relevant for diagnostic imaging or AI-assisted diagnostic devices where human readers interpret cases. This is not applicable to a physical implantable device like a suture anchor.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done:
- Not Applicable. This is not an algorithm-based device.
7. The Type of Ground Truth Used:
- The "ground truth" for demonstrating substantial equivalence for this device is based on objective engineering and mechanical performance data (e.g., measured pullout strength, tensile strength, and material properties) obtained from non-clinical laboratory testing, alongside adherence to established standards for biocompatibility and sterilization.
8. The Sample Size for the Training Set:
- Not Applicable. This device is not an AI/ML model, so there is no "training set."
9. How the Ground Truth for the Training Set Was Established:
- Not Applicable. There is no training set for this type of device.
Ask a specific question about this device
(80 days)
The Aevumed RAPID™ Suture Anchors are intended to be used for suture or tissue fixation in the foot, ankle, knee, hand, wrist, elbow, shoulder, and hip. Specific indications are listed below:
Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clabicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction
Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Metatarsal Ligament Repair, Hallux Valgus Reconstruction, Digital Tendon Transfers, Mid-foot Reconstruction
Knee: Medial Collateral Ligament Repair, Lateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis
Hand/Wrist: Scapholunate Ligament Reconstruction, Carpal Ligament Reconstruction, Repair/Reconstruction of Collateral Ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP, and MCP joints for all Digits, Digital Tendon Transfers
Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction
Hip: Capsular repair, Acetabular Labral Repair
The Aevumed RAPIDTM Suture Anchor with HS FiberTM suture is a suture anchor manufactured from polyetheretherketone (PEEK) material preloaded on a disposable inserter assembly intended for fixation of soft tissue to bone. The Aevumed RAPIDTM Suture Anchors are available in diameter sizes: 2.4mm and 3.0mm. They are offered sterile and are for single use only
The provided text is a 510(k) premarket notification for a medical device (Aevumed RAPID™ Suture Anchors). It details the device's characteristics, indications for use, and a comparison to predicate devices, but does not describe a study involving AI or software, nor does it provide acceptance criteria and performance data in the context of an AI/software device evaluation.
Therefore, I cannot extract the information required in your request (acceptance criteria, study details, sample sizes, expert qualifications, etc.) because the document describes a hardware medical device (suture anchors) and its mechanical testing, not a diagnostic or AI-driven system.
The "Non-clinical Tests" section mentions mechanical testing for pullout strength and states:
- Study: "Both the Aevumed RAPID™ Suture Anchor and predicate SutureTak™ Anchor (K140855) were mechanically tested for maximum pullout strength."
- Result: "The Aevumed RAPID™ Suture Anchor demonstrated significantly higher pullout strength in comparison to the predicate SutureTak™ Anchor (K140855)."
This is the only "performance" information related to a "study" in the document, and it pertains to the physical device's mechanical properties, not an algorithmic performance.
Ask a specific question about this device
(116 days)
Ask a specific question about this device
(99 days)
The Aevumed PHANTOM™ suture anchors are intended to be used for suture and/or soft tissue fixation to bone in the shoulder, foot/ankle, knee, hand/wrist, and elbow in, but not limited to, the following procedures:
· Shoulder: Rotator Cuff Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction.
· Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Midfoot Reconstruction, Metatarsal Ligament Repair.
· Knee: Anterior Cruciate Ligament Repair, Medial Collateral Ligament Repair, Lateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair and Illiotibial Band Tenodesis.
• Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar or Radial Collateral Ligament Reconstruction.
· Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction.
The Aevumed PHANTOM ™ Anchor with HS Fiber ™ suture is a threaded suture anchor manufactured from polyetheretherketone (PEEK) material preloaded on a disposable driver assembly intended for fixation of soft tissue to bone. The Aevumed PHANTOM ™ Anchor with HS Fiber ™ are pre-loaded with or without needles. The Aevumed PHANTOM ™ Suture Anchor are available in diameter sizes: 5.5mm and 6.5mm. They are offered sterile and are for single use only.
The provided text is a 510(k) Premarket Notification for a medical device (AEVUMED PHANTOM Suture Anchors). This document describes the device, its intended use, and a comparison to a predicate device to demonstrate substantial equivalence.
However, the provided text DOES NOT contain information about acceptance criteria and a study that proves the device meets those criteria in the context of AI/ML-based medical devices. The document is for a traditional mechanical medical device (suture anchors) and the "NON-CLINICAL TESTS" section only mentions mechanical testing (Torsional Strength, Driving Torque, Pullout Strength) as the basis for substantial equivalence.
Therefore, I cannot extract the information required by your request, as the request pertains to the evaluation of an AI/ML device, and the provided document describes a non-AI/ML device.
Here's what I can tell you from the document regarding the "study" that proves the device meets the acceptance criteria, even though it's not an AI/ML study:
- Study Type: Non-clinical mechanical testing.
- Tests Conducted: Torsional Strength, Driving Torque, and Pullout Strength.
- Purpose: To demonstrate substantial equivalence to the predicate device (Arthrex Corkscrew FT) by showing comparable mechanical performance.
- Ground Truth: The mechanical properties of the predicate device serve as the de facto "ground truth" for comparison.
- Acceptance Criteria: While specific numerical acceptance criteria are not explicitly listed in this summary, the implication is that the Aevumed PHANTOM™ Suture Anchors must perform comparably to the predicate device in these mechanical tests to establish substantial equivalence.
To directly answer your numbered points based on the provided text, I must state that this information is not available for an AI/ML device study:
- A table of acceptance criteria and the reported device performance: Not applicable for an AI/ML study. The document mentions mechanical tests (Torsional Strength, Driving Torque, Pullout Strength), but doesn't provide a table of acceptance criteria or specific numerical performance results for these non-clinical tests. It only states they were tested.
- Sample sized used for the test set and the data provenance: Not applicable for an AI/ML study. For the mechanical tests, the sample sizes and provenance of the test articles are not specified in this summary.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable for an AI/ML study. Ground truth for mechanical testing is based on physical measurements, not expert consensus.
- Adjudication method for the test set: Not applicable for an AI/ML study.
- If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is a mechanical device, not an AI/ML device.
- If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is a mechanical device.
- The type of ground truth used: For this mechanical device, the "ground truth" is likely defined by the established mechanical standards and performance of the predicate device.
- The sample size for the training set: Not applicable. There is no AI/ML training set.
- How the ground truth for the training set was established: Not applicable. There is no AI/ML training set.