Search Filters

Search Results

Found 5 results

510(k) Data Aggregation

    K Number
    K251615

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-09-02

    (98 days)

    Product Code
    Regulation Number
    868.5630
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Aerogen® Pro-X1 controller System is a portable medical device system inclusive of the Aerogen Solo Nebulizer (Single-patient, Multiple use) which is intended to aerosolize prescribed medications for inhalation to paediatric (29 days and older) and adult patients during invasive ventilation, non-invasive respiratory support or to conscious and spontaneous breathing patients in the hospital environment.

    Device Description

    The Aerogen® Pro-X1 controller System is a portable medical device system inclusive of the Aerogen Solo Nebulizer (Single-patient, Multiple use) which is intended to aerosolize prescribed medications for inhalation to paediatric (29 days and older) and adult patients during invasive ventilation, non-invasive respiratory support or to conscious and spontaneous breathing patients in the hospital environment.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K232507

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-05-10

    (266 days)

    Product Code
    Regulation Number
    868.5630
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Aerogen® Solo Nebulizer System includes the Aerogen® Solo Nebulizer, which is intended to aerosolize physicianprescribed medications for inhalation to patients on and off ventilation or other positive pressure breathing assistance in the hospital environment.

    Aerogen® Solo Nebulizer is for single-patient use. The Aerogen Solo nebulizer is for pediatric (29 days or older) and adult patients.

    The Aerogen® USB Controller System includes the Aerogen® Solo Nebulizer, which is intended to aerosolize physicianprescribed medications for inhalation to patients on and off ventilation or other positive pressure breathing assistance in the hospital environment, and on vent only in the homecare environment.

    Aerogen® Solo Nebulizer is for single patient use. The Aerogen Solo nebulizer is for pediatric (29 days or older) and adult patients.

    Device Description

    The Aerogen® Solo Nebulizer System consists of the Aerogen® Solo Nebulizer, which has the proposed modification outlined in these 510(k) applications, and the Aerogen® Pro-X Controller (K133360 for both components and system) or the Aerogen Solo Nebulizer can also be used alternatively with the reference device - Aerogen® USB Controller (K143719).

    The Aerogen® Solo Nebulizer is intended to aerosolize physician-prescribed medications for inhalation which are approved for use with a general-purpose nebulizer. It is intended for use on and off mechanical ventilation or other positive pressure breathing assistance and is intended for both intermittent and continuous nebulization.

    The Aerogen® Solo Nebulizer is designed to operate in-line with standard ventilator circuits and mechanical ventilators. It operates without changing patient ventilator parameters and can be refilled without interrupting ventilation.

    This submission is for a modification of the attachment process of the Aperture Plate (AP) to the Oscillator washer, both are components of the core for the Aerogen® Solo Nebulizer. The current method of attachment of the Aperture Plate to the Oscillator washer is by means of Brazing, a metal fusion process, where a copper/gold filler washer is melted in the brazing process to form the bond. The piezo is then attached using a heat cured epoxy adhesive.

    AI/ML Overview

    The provided documentation is a 510(k) premarket notification for a medical device, specifically the Aerogen® Solo Nebulizer System. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving efficacy or performance against a set of predefined acceptance criteria for a novel device. As such, the document does not contain information on rigorous acceptance criteria and the detailed study design typically seen for a new device's proven performance.

    However, it does present "Comparative Particle Characterization Performance" data (Table 2) which can be interpreted as the performance data for the modified device in comparison to its predicate. The "Acceptance Criteria" in this context would be the demonstration that the modified device performs similarly to the predicate device, thereby maintaining substantial equivalence.

    Here's an attempt to extract the requested information based on the provided document, interpreting "acceptance criteria" as "similarity to predicate device performance" for the purpose of 510(k) clearance:

    1. A table of acceptance criteria and the reported device performance

    Since this is a 510(k) showing substantial equivalence, the "acceptance criteria" for the modified device are implicitly that its performance parameters should be within a similar range or deemed not clinically significant compared to the predicate device. The document explicitly states "Similar Cascade Impactor performance and NGI performance" in the comparison column, implying this is the "acceptance" based on the presented data.

    FeaturesAcceptance Criteria (Predicate Range @ 28 lpm)Reported Device Performance (Modified @ 28 lpm)Reported Device Performance (Modified @ 15 lpm - NGI)Comparison (Achieved Acceptance)
    Particle size (MMAD) (microns)Albuterol: 2.90 – 3.23 Ipratropium: 3.07 – 3.42 Budesonide: 3.45 – 3.79Albuterol: 2.80 – 3.05Albuterol: 4.39 – 4.53 Ipratropium: 3.76 – 4.02 Budesonide: 4.90 – 5.01Similar Cascade Impactor performance and NGI performance
    Geometric Std. Dev. (GSD)Albuterol: 2.09 – 2.35 Ipratropium: 1.80 – 1.93 Budesonide: 1.92 – 2.14Albuterol: 2.26 – 2.36Albuterol: 2.09 – 2.16 Ipratropium: 2.28 – 2.41 Budesonide: 2.06 – 2.14Similar Cascade Impactor performance and NGI performance
    Total Respirable Dose (0.0-5 microns)Albuterol: 67.66 – 73.50 Ipratropium: 71.78 – 76.69 Budesonide: 62.32 – 66.90Albuterol: 70.88 – 73.94Albuterol: 54.42 – 56.10 Ipratropium: 59.57 – 63.06 Budesonide: 49.32 – 50.91Similar Cascade Impactor performance and NGI performance
    Coarse Particle Dose (>4.7 micron)Albuterol: 27.00 – 31.11 Ipratropium: 23.62 – 28.21 Budesonide: 32.31 – 36.12Albuterol: 28.73 – 32.03Albuterol: 46.69 – 48.78 Ipratropium: 39.87 – 43.38 Budesonide: 53.51 – 54.99Similar Cascade Impactor performance and NGI performance
    Fine Particle Dose (<4.7 micron)Albuterol: 66.33 – 72.07 Ipratropium: 68.58 – 73.84 Budesonide: 59.36 – 64.17Albuterol: 67.97 – 71.27Albuterol: 51.23 – 53.31 Ipratropium: 56.62 – 60.13 Budesonide: 45.01 – 46.49Similar Cascade Impactor performance and NGI performance
    Ultra-fine Particle Dose (< 1.0 micron)Albuterol: 5.91 – 9.93 Ipratropium: 1.85 – 4.19 Budesonide: 2.36 – 4.51Albuterol: 10.13 – 12.02Albuterol: 2.86 – 3.70 Ipratropium: 8.24 – 12.89 Budesonide: 2.78 – 3.41Similar Cascade Impactor performance and NGI performance

    Note: For "Total Dose (ug)", the predicate values are "N/A" and the comparison states "Additional data provided," which doesn't directly fit the "acceptance criteria" comparison. The acceptance is implied by the provision of this new data for the modified device at 15 Lpm.

    2. Sample sized used for the test set and the data provenance

    The document does not specify the exact sample sizes (number of nebulizers tested) used for the particle characterization testing. It refers to 95% confidence intervals, which implies statistical analysis was performed on a sample. The data provenance is not explicitly mentioned (e.g., country of origin), but it is bench testing ("Non-clinical Performance Testing") rather than patient data, and it is prospective in the sense that the tests were conducted on the modified device for the purpose of this submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This document describes a 510(k) submission for a nebulizer, a physical medical device. The "ground truth" for its performance is established through objective engineering and laboratory measurements (e.g., cascade impactor, NGI) of physical parameters like particle size and dose. This is not a study involving human experts establishing ground truth for diagnostic accuracy (e.g., radiologists for medical images). Therefore, this question is not applicable in this context.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This is not a study involving human adjudication of results; it's a bench performance study.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a nebulizer, not an AI-powered diagnostic tool for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not a software algorithm; it's a physical device. The performance tests ("Aerosol Performance Inter- and Intra-sample variability") are analogous to standalone performance of the physical nebulizer.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this device's performance is established by physical measurements using standardized laboratory equipment (Cascade Impactor, NGI) and methods to characterize aerosol properties. This is a form of objective, empirical measurement rather than subjective expert consensus or clinical outcomes data.

    8. The sample size for the training set

    Not applicable. This device is a nebulizer, not an AI/machine learning model that requires a training set. The "training" for a physical device involves its design, engineering, and manufacturing processes, which are guided by established principles, standards, and previous device iterations (like the predicate).

    9. How the ground truth for the training set was established

    Not applicable. As explained above, there is no "training set" in the context of a physical device like a nebulizer in the way it's understood for AI/ML.

    Ask a Question

    Ask a specific question about this device

    K Number
    K133360

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-10-17

    (350 days)

    Product Code
    Regulation Number
    868.5630
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Aeroneb® Solo Nebulizer System is a portable medical device for single patient use that is intended to aerosolize physician-prescribed solutions for inhalation to patients on and off ventilation or other positive pressure breathing assistance.

    The Aeroneb® Solo Adapter is an accessory specific to the Aeroneb® Solo Nebulizer. It facilitates intermittent and continuous nebulization and optional supply of supplemental Oxygen to pediatric (29 days or older) and adult patients in hospital use environments via a mouthpiece or aerosol mask. If supplemental oxygen is used, for pediatric patients under 18 years of age, a maximum flow rate of 2 LPM should be used.

    Note: The mouthpiece should not be used for children under 5 years of age.

    Device Description

    The Aeroneb® Solo Adapter is designed to operate with the Aeroneb® Solo Nebulizer System, enabling efficient delivery of aerosol therapy to non-ventilated patients via an aerosol mask or mouthpiece. The device is composed of a valve controlled chamber with ports for connection of the Aeroneb® Solo Nebulizer via a mouthpiece or facemask. Air is drawn into the device and exhausted through distal and proximal valves respectively. The mouthpiece is interchangeable with a facemask (not supplied). The Aeroneb Solo Adapter is equipped with an oxygen port for the delivery during aerosol therapy of supplementary oxygen via the Oxygen Tubing supplied. When using oxygen, the flow rate should be set between 1-6 LPM for adult use, and at a maximum rate of 2 LPM for pediatric patients less than 18 years of age.

    Inclusion of a reference in the labeling (Instructions for Use); to the use of Nasal Cannula (not supplied with the device) as an alternate patient interface on/off ventilator when the Aeroneb® Solo Nebulizer is placed dry side of a humidifier.

    AI/ML Overview

    The provided text is a 510(k) Summary for the Aeroneb® Solo Nebulizer System / Aeroneb® Solo Adapter. This document details the device's characteristics and the non-clinical tests performed to demonstrate substantial equivalence to a predicate device, rather than a clinical study evaluating its effectiveness against specific acceptance criteria in a human population.

    Therefore, many of the requested sections (2, 3, 4, 5, 6, 7, 8, 9) regarding clinical study details (sample size, ground truth, expert opinions, MRMC studies, etc.) cannot be extracted as they are not present in this type of regulatory submission. The document focuses on demonstrating physical, chemical, and performance characteristics in a non-clinical setting.

    Here's the information that can be extracted or inferred:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally implied by demonstrating the device performs comparably to the predicate device and meets established standards for safety and performance (e.g., biocompatibility standards, aerosol characterization within expected ranges). The reported performance is based on non-clinical tests.

    Characteristic / TestAcceptance Criteria (Implied)Reported Device Performance
    BiocompatibilityMeet requirements of ISO 10993-1, ISO 10993-5, ISO 10993-10, ISO 10993-11 for externally communicating tissue contact with limited contact duration.Acceptable results for the Aeroneb® Solo Adapter in Cytotoxicity, ISO Systemic Toxicity, ISO Intracutaneous, and ISO Maximization Sensitization studies. Tests were GLP compliant and carried out at NAMSA, Northwood, Ohio, USA.
    CleaningNot applicable (disposable device); however, the IFU should provide direction for rinsing to maintain functional performance.The Aeroneb® Solo Adapter is a non-sterile disposable device for single patient use and does not require cleaning. The IFU directs users to rinse with sterile water to maintain the functional performance of the device, particularly the valves.
    Simulated Life Testing (Valve Operating Pressure)Pressure within the device should not exceed +1.5 cm H2O with 15 l/min exhalation flow or decrease below -1.5 cm H2O with 15 l/min inhalation flow after 20 treatments (90%/90% confidence/reliability).After 20 treatments (at a rate of four 3 ml doses per day over 5 days), the pressure within the device did not exceed +1.5 cm H2O when connected to a steady exhalation flow of 15 l/min or decrease below -1.5 cm H2O when connected to a steady inhalation flow of 15 l/min (based on 90%/90% confidence/reliability).
    Simulated Life Testing (Inlet/Exhaust Valve Flaps)Inlet or exhaust valve flaps should not dislodge after rinsing during intermittent therapy.The inlet or exhaust valve flaps did not dislodge after rinsing during intermittent therapy.
    USP 661 Leachables TestingAll extractables should be within acceptable limits for drug-contacting polymer-based materials as per USP <661>. Amounts of COPC should not be toxicologically significant.All drug-contacting plastic materials tested met USP <661> acceptable limits. Exhaustive Extraction and Simulated Use Extraction demonstrated that potential leachates were not toxicologically significant and would not elicit an adverse biological response. The entire Aeroneb® Solo Adapter was extracted at 50°C for 72 hours in 132mL of purified water, with a control.
    Aerosol Characterization (Particle Size)Implied comparability to predicate or accepted nebulizer performance. Specific medical relevance often dictates acceptable ranges.Albuterol Sulphate (2mg/ml): 2.90 - 3.23 µm Budesonide (0.5mg/ml): 3.07 - 3.42 µm Ipratropium (0.25mg/ml): 3.45 - 3.79 µm (indicated ranges correspond to 95% confidence intervals).
    Aerosol Characterization (Geometric Standard Deviation (GSD))Implied comparability.Albuterol Sulphate (2mg/ml): 2.09 - 2.35 Budesonide (0.5mg/ml): 1.80 - 1.93 Ipratropium (0.25mg/ml): 1.92 - 2.14 (95% confidence intervals).
    Aerosol Characterization (Emitted Dose)Implied comparability.Albuterol Sulphate (2mg/ml): 97.23 - 99.30% of fill Budesonide (0.5mg/ml): 97.61 - 98.64% of fill Ipratropium (0.25mg/ml): 94.12 - 97.84% of fill (95% confidence intervals).
    Aerosol Characterization (Respirable Dose (0.5 – 5.0 µm))Implied comparability.Albuterol Sulphate (2mg/ml): 67.66 - 73.50% of fill Budesonide (0.5mg/ml): 71.78 - 76.69% of fill Ipratropium (0.25mg/ml): 62.32 - 66.90% of fill (95% confidence intervals).
    Aerosol Characterization (Coarse Particle Dose (>4.7 µm))Implied comparability.Albuterol Sulphate (2mg/ml): 27.00 - 31.11% of fill Budesonide (0.5mg/ml): 23.62 - 28.21% of fill Ipratropium (0.25mg/ml): 32.31 - 36.12% of fill (95% confidence intervals).
    Aerosol Characterization (Fine Particle Dose (<4.7 µm))Implied comparability.Albuterol Sulphate (2mg/ml): 66.33 - 72.07% of fill Budesonide (0.5mg/ml): 68.58 - 73.84% of fill Ipratropium (0.25mg/ml): 59.36 - 64.17% of fill (95% confidence intervals).
    Aerosol Characterization (Ultra-fine Particle Dose (<1.0 µm))Implied comparability.Albuterol Sulphate (2mg/ml): 5.91 - 9.93% of fill Budesonide (0.5mg/ml): 1.85 - 4.19% of fill Ipratropium (0.25mg/ml): 2.36 - 4.51% of fill (95% confidence intervals).

    Regarding the study that proves the device meets acceptance criteria:

    The "study" refers to a series of non-clinical tests performed to demonstrate substantial equivalence of the Aeroneb® Solo Adapter. These tests include:

    • Biocompatibility Testing: According to ISO 10993-1, ISO 10993-5, ISO 10993-10, and ISO 10993-11.
    • Cleaning: Evaluation of instructions for use (IFU) and material characteristics (disposable).
    • Simulated Life Testing: Evaluating valve operating pressure and the integrity of valve flaps after repeated use.
    • USP 661 Leachables Testing: To determine the level of plastic leachables from drug-contacting materials.
    • Aerosol Characterization: Using an 8-stage cascade impactor to measure particle size distribution, emitted dose, and respirable dose for various common nebulized medications.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Test Set Sample Size: Not explicitly stated as "sample size" in the context of human subjects or cases. For individual non-clinical tests:
      • Biocompatibility: Animal models (mice, rabbits) were used, but the specific number isn't quantified in this summary.
      • Simulated Life Testing: "20 treatments" were performed for the pressure and valve integrity tests.
      • USP 661 Leachables: The "entire Aeroneb® Solo Adapter test Article" was extracted.
      • Aerosol Characterization: Not explicitly stated how many nebulizer units or runs were conducted for each drug, but results are given as 95% confidence intervals, implying statistical analysis of multiple measurements.
    • Data Provenance:
      • Biocompatibility testing was carried out at NAMSA, Northwood, Ohio, USA.
      • Other tests' locations are not specified, but the manufacturer is Aerogen Limited in Ireland.
    • Retrospective or Prospective: These are non-clinical, controlled laboratory tests, not retrospective or prospective studies on human or patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. Ground truth, in the clinical sense, is not established for these non-clinical performance and safety tests. The "ground truth" for these tests is defined by regulatory standards (e.g., ISO, USP) and established measurement protocols.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. This concept applies to human expert review in clinical studies, not to non-clinical laboratory testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. No MRMC study was performed as this is a non-clinical evaluation of a nebulizer accessory, not an AI-assisted diagnostic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. There is no algorithm in this medical device. The device itself is the "standalone" component being tested for its physical and performance characteristics.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • For biocompatibility: Adherence to ISO 10993 standards and lack of adverse biological responses in in-vitro and animal tests.
    • For physical/performance tests (e.g., simulated life, aerosol characterization): Quantitative measurements against pre-defined engineering and regulatory specifications and comparison to the predicate device's expected performance. The "ground truth" is derived from these objective measurements and standard compliance.

    8. The sample size for the training set

    • Not applicable. This is not an AI/machine learning device, so there is no "training set."

    9. How the ground truth for the training set was established

    • Not applicable. As there is no training set for an AI/ML model.
    Ask a Question

    Ask a specific question about this device

    K Number
    K120939

    Validate with FDA (Live)

    Device Name
    AERONEB PRO
    Manufacturer
    Date Cleared
    2012-04-26

    (28 days)

    Product Code
    Regulation Number
    868.5630
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Aeroneb® Professional Nebulizer System is a portable medical device for multiple patient use that is intended to aerosolize physician-prescribed solutions for inhulple patients on and off ventilation or other positive pressure breathing assistance. The Aeroneb® Professional Nebulizer System is suitable for use in adult and pediatric patients.

    Device Description

    Not Found

    AI/ML Overview

    This document is a 510(k) premarket notification decision letter from the FDA regarding the Aeroneb Professional Nebulizer System. It confirms that the device is substantially equivalent to legally marketed predicate devices.

    However, this document does not contain information about acceptance criteria, device performance testing, study design (sample size, data provenance, expert ground truth, adjudication, MRMC studies, standalone performance), or training set details typically associated with AI/ML-based medical devices or complex diagnostic systems that require such extensive validation studies.

    The Aeroneb Professional Nebulizer System is a medical device (nebulizer) that delivers aerosolized medications. Its substantial equivalence determination is based on manufacturing, performance specifications (likely particle size, delivery rate, durability), and safety related to its intended use, not on complex data-driven diagnostic or interpretative capabilities that would require the kind of clinical validation study described in the prompt.

    Therefore, I cannot provide the requested information from the provided text. The document does not describe the type of study that would involve:

    • A table of acceptance criteria and reported device performance related to a diagnostic or interpretative function.
    • Sample sizes for test sets or training sets in the context of machine learning.
    • Ground truth established by experts, adjudication methods, or MRMC studies.
    • Standalone algorithm performance.

    The submission is for a traditional medical device (nebulizer) where performance is typically assessed through engineering tests, biocompatibility, and perhaps limited clinical trials demonstrating drug delivery effectiveness or safety, but not statistical performance on diagnostic tasks that rely on "ground truth" derived from expert consensus or pathology, as seen in AI/ML validation.

    Ask a Question

    Ask a specific question about this device

    K Number
    K103635

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-05-13

    (151 days)

    Product Code
    Regulation Number
    868.5630
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Aeroneb Professional Nebulizer System is a portable medical device for multiple patient use that is intended to aerosolize physician-prescribed solutions for inhalation to patients on and off ventilation or other positive pressure breathing assistance. The Aeroneby Professional Nebulizer System is suitable for use in adult and pediatric patients.

    Device Description

    The Aeroneb Professional Nebulizer System is a portable medical device for multiple patient use that is intended to aerosolize physician-prescribed solutions for inhalation to patients on and off ventilation or other positive pressure breathing assistance.

    AI/ML Overview

    I apologize, but the provided text is a 510(k) clearance letter from the FDA for the Aeroneb Professional Nebulisation System. It does not contain information about the acceptance criteria, study details, or device performance metrics as requested in your prompt.

    Specifically, the document is a regulatory approval letter and does not include:

    • A table of acceptance criteria and reported device performance.
    • Sample sizes for test or training sets, data provenance, or details about ground truth establishment.
    • Information on expert adjudication, MRMC studies, or standalone algorithm performance.

    Therefore, I cannot fulfill your request using the provided input.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1