Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K252867

    Validate with FDA (Live)

    Date Cleared
    2025-10-09

    (30 days)

    Product Code
    Regulation Number
    862.3100
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VINScreen Urine Drug Test Cup is lateral flow immunoassays for rapid detection of multiple commonly abused drugs in human urine. The detectable drugs and their cutoff concentrations are listed below:

    AbbreviationAnalyteCalibratorCut-off (ng/mL)
    6AM6-Acetylmorphine6-Monoacetylmorphine10
    AMPAmphetamined-Amphetamine500 or 1000
    BARSecobarbitalSecobarbital300
    BUPBuprenorphineBuprenorphine10
    BZOOxazepamOxazepam300
    COCCocaineBenzoylecgonine150 or 300
    EDDPEDDP2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine300
    FENFentanylFentanyl1
    FENNorfentanylNorfentanyl5
    MDMAEcstasyd,l-Methylenedioxymethamphetamine500
    METMethamphetamined-Methamphetamine500 or 1000
    MORMorphineMorphine300
    MTDMethadoned/l-Methadone300
    OPIOpiatesMorphine2000
    OXYOxycodoneOxycodone100
    PCPPhencyclidinePhencyclidine25
    PPXPropoxyphenePropoxyphene300
    TCANortriptylineNortriptyline1000
    THCMarijuana11-nor-Δ9-THC-9 COOH20 or 50

    The single or multi-test cup can include any combination of the analytes listed above, with and without on-board adulteration tests. However, only one cut-off concentration can be included per analyte per device.

    This in vitro diagnostic device provides only a preliminary test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical test result. GC/MS or LC/MS is the preferred confirmatory method.

    The VINScreen Urine Drug Home Test Cup is lateral flow immunoassays for rapid detection of multiple commonly abused drugs in human urine. The detectable drugs and their cutoff concentrations are listed below:

    DrugCut-off (ng/mL)Abbreviation
    6-monoacetylmorphine106AM
    Amphetamine500 / 1000AMP
    Secobarbital300BAR
    Buprenorphine10BUP
    Oxazepam300BZO
    Cocaine150 / 300COC
    EDDP300EDDP
    Fentanyl1FEN
    Norfentanyl5FEN
    Ecstasy500MDMA
    Methamphetamine500 / 1000MET
    Methadone300MTD
    Opiates300 / 2000MOR/OPI
    Oxycodone100OXY
    Phencyclidine25PCP
    Propoxyphene300PPX
    Nortriptyline1000TCA
    Marijuana20 / 50THC

    The single or multi-test cup can include any combination of the analytes listed above, with and without on-board adulteration tests. However, only one cut-off concentration can be included per analyte per device.

    This device provides only a preliminary test result. A more specific alternate method must be used in order to obtain a confirmed analytical test result. GC/MS or LC/MS is the preferred confirmatory method.

    Device Description

    VINScreen Urine Drug Test Cup and VINScreen Urine Drug Home Test Cup are immunochromatographic assays that use a lateral flow system for the qualitative detection of single or multiple drugs in human urine.

    The devices are a cup format. Each test device is sealed with sachets of desiccant in an aluminum pouch. The device is in a ready-to-use format and no longer requires assembly before use.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K201494

    Validate with FDA (Live)

    Date Cleared
    2020-09-17

    (104 days)

    Regulation Number
    862.3650
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ATTEST Drug Screen Cup and the ATTEST Drug Screen Dip Card are rapid lateral flow immunoassays for the qualitative detection of 6-Acetylmorphine, d-Amphetamine, Benzoylecgonine, EDDP, d/1-Methadone, d-Methamphetamine, d/Methylenedioxymethamphetamine, Nortriptyline, Oxazepam, Oxycodone, Phencyclidine, d-Propoxyphene, Secobarbital and THC-COOH in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:

    AnalyteCalibratorCutoff (ng/mL)
    6-Acetylmorphine6-monoacetylmorphine10
    Amphetamined-Amphetamine500
    Amphetamined-Amphetamine1,000
    SecobarbitalSecobarbital300
    OxazepamOxazepam300
    BuprenorphineBuprenorphine10
    EDDP2-ethylidene-1,5-dimethyl-3-3- diphenylpyrrolidine300
    CocaineBenzoylecgonine150
    CocaineBenzoylecgonine300
    Ecstasyd,l-Methylenedioxymethamphetamine500
    Methamphetamined-Methamphetamine500
    Methamphetamined-Methamphetamine1,000
    Marijuana11-nor-Δ9-THC-9-COOH20
    Marijuana11-nor-Δ9-THC-9-COOH50
    Methadoned/l-Methadone300
    OpiatesMorphine300
    OpiatesMorphine2,000
    OxycodoneOxycodone100
    PhencyclidinePhencyclidine25
    PropoxyphenePropoxyphene300
    NortriptylineNortriptyline1,000

    The single or multi-test panels can consist of the above listed analytes in any combination, up to a maximum of 16 analytes, with and without on-board adulteration/specimen validity tests (SVT) in the cup format. The drug screen tests are intended for prescription use only.

    The tests provide only a preliminary result. To obtained a confirmed analytical result, a more specific alternative chemical method should be used. Gas Chromatography / Mass Spectrometry (GC/MS), Liquid Chromatography / Mass Spectrometry (LC/MS) and their tandem mass-spectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results.

    Device Description

    For prescription use, the devices consist of:

    • a. 10 or 25 test cups or dip cards with or without adulteration/specimen validity tests
    • b. Package Insert
    • c. Procedure Card
    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the ATTEST Drug Screen Cup and Dip Card, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly state "acceptance criteria" as a separate, quantitative target for each metric before presenting the results. Instead, it presents the performance data and implies that these results demonstrate the device's suitability. The primary performance metric presented is agreement with gold-standard methods (GC/MS, LC/MS or equivalent), and for most categories, the reported agreement is 100% or very close to it.

    For the purpose of this analysis, I will infer the "acceptance criteria" from the reported performance, assuming that the FDA's acceptance is based on these high agreement rates, especially at or further from the cutoff concentrations.

    CharacteristicAcceptance Criteria (Inferred from data)Reported Device Performance (Agreement with Gold Standard) - ATTEST Cup & Dip Card
    Precision/Reproducibility>99% correlation at +/-50% of each assay cutoff across multiple sites.>99% correlation (across Advin Biotech and 3 external sites).
    Analytical SpecificityNo or minimal cross-reactivity for structurally similar compounds; no positive interference from chemically dissimilar compounds, endogenous agents, or pH/specific gravity variations.Detailed tables provided showing specific cross-reactivity percentages (e.g., d-Amphetamine 100%, l-Amphetamine 1%), and non-interference for numerous substances, pH (4-9), and specific gravity (1.003-1.030).
    Method Comparison/AccuracyHigh agreement with gold-standard methods (GC/MS, LC/MS) across drug-free, near-cutoff, and beyond-cutoff concentrations.Generally 93.2% - 100% agreement for negative and positive results, with specific discordant cases identified and quantified relative to the cutoff.
    Read Time StabilityStable results expected for a reasonable reading window (e.g., up to 75 minutes).Test results stable for up to 75 minutes.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: The method comparison/accuracy study (which serves as the "test set" for performance evaluation) uses the following sample sizes per drug test/cutoff combination:

      • Drug-free: 40 samples
      • -50% C/O to <-25% C/O: Varies (e.g., 4 for 6-AM/10, 22 for THC/20)
      • -25% C/O to C/O: Varies (e.g., 1 for 6-AM/10, 6 for THC/20)
      • C/O to +25% C/O: Varies (e.g., 1 for 6-AM/10, 2 for THC/20)
      • >+25% C/O to +50% C/O: Varies (e.g., 4 for 6-AM/10, 5 for THC/20)
      • >+50% C/O: Varies (e.g., 35 for 6-AM/10, 46 for THC/20)
      • Total samples per drug test/cutoff category appears to be around 80-100 samples.
    • Data Provenance: The document states that the studies used "clinical urine specimens previously quantitated for the target drugs of abuse by gold-standard methods (GC/MS, LC/MS or equivalent)." There is no explicit mention of the country of origin for these clinical specimens or if they were retrospective or prospective. Given the context of a 510(k) submission, these are typically retrospective studies using archived, de-identified clinical samples.

    3. Number of Experts Used to Establish the Ground Truth and Qualifications

    • The document does not mention the use of human experts to establish the ground truth for the test set.
    • Ground truth for the method comparison/accuracy study was established by "gold-standard methods (GC/MS, LC/MS or equivalent)", which are analytical laboratory techniques, not expert human interpretation.

    4. Adjudication Method for the Test Set

    • Since the ground truth was established by instrumental analytical methods (GC/MS, LC/MS), there was no human adjudication (e.g., 2+1, 3+1) involved in determining the "true" presence or absence of drugs for the test set. The results from the gold-standard methods served as the definitive determination.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No MRMC comparative effectiveness study was done. The ATTEST Drug Screen Cup and Dip Card are qualitative lateral flow immunoassays and function as standalone diagnostic devices. Their performance is evaluated against chemical reference methods, not against human readers. Therefore, the concept of "effect size of how much human readers improve with AI vs without AI assistance" is not applicable.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance)

    • Yes, a standalone performance evaluation was done. The ATTEST Drug Screen Cup and Dip Card are designed to be used as standalone qualitative tests. The performance data presented (precision, analytical specificity, and method comparison/accuracy tables) directly reflect the device's diagnostic capability without human in-the-loop interaction for result interpretation, beyond simply reading a positive or negative line visually as designed by a qualitative immunoassay. The results are based on the device's output (presence or absence of control/test lines).

    7. Type of Ground Truth Used

    • The ground truth used was analytical gold-standard methods: Gas Chromatography / Mass Spectrometry (GC/MS), Liquid Chromatography / Mass Spectrometry (LC/MS), and their tandem mass-spectrometer versions. The document explicitly states these are the "preferred confirmatory methods."

    8. Sample Size for the Training Set

    • The document does not specify the sample size used for the training set. As this device is a rapid lateral flow immunoassay (a chemical test, not an AI/machine learning algorithm), the concept of a "training set" in the context of predictive modeling is not directly applicable in the same way. The device's components and antibodies are developed and optimized through laboratory work and validation, rather than "training" on a large dataset of results in the way an AI algorithm would be.

    9. How the Ground Truth for the Training Set Was Established

    • Given that this is a chemical immunoassay, the concept of a "training set" for ground truth establishment, as it pertains to AI/machine learning, is not directly relevant. The "ground truth" during the development and optimization of such assays would involve:

      • Known concentrations of target analytes and potential cross-reactants: Used to optimize antibody binding, cutoff concentrations, and minimize false positives/negatives.
      • Control samples: Including negative urine samples (drug-free) and positive urine samples spiked with known drug concentrations.
      • Reference materials: Calibrators and controls with established values, often verified by GC/MS or LC/MS.

      The "precision/reproducibility" and "analytical specificity" studies described in the document reflect parts of the characterization and validation process that would inform the final design and performance claims of the device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K182123

    Validate with FDA (Live)

    Date Cleared
    2019-03-29

    (235 days)

    Product Code
    Regulation Number
    862.3650
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ATTEST Drug Screen Cup and the ATTEST Drug Screen Dip Card are rapid lateral flow immunoassays for the qualitative detection of 6-Acetylmorphine, d-Ampletamine, Benzoylecgonine, EDDP, d/1-Methadone, d-Methamphetamine, d/Methylenedioxymethamphetamine, Nortriptyline, Oxazepam, Oxycodone, Phencyclidine, d-Propoxyphene, Secobarbital and THC-COOH in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:

    AnalyteCalibratorCutoff (ng/mL)
    6-Acetylmorphine6-monoacetylmorphine10
    Amphetamined-Amphetamine500
    Amphetamined-Amphetamine1,000
    BarbituratesSecobarbital/Pentobarbital300
    BenzodiazepinesOxazepam300
    BuprenorphineBuprenorphine10
    EDDP2-ethylidene-1,5-dimethyl-3-3- diphenylpyrrolidine300
    CocaineBenzoylecgonine150
    CocaineBenzoylecgonine300
    Ecstasyd,l-Methylenedioxymethamphetamine500
    Methamphetamined-Methamphetamine500
    Methamphetamined-Methamphetamine1,000
    Marijuana11-nor-Δ9-THC-9-COOH50
    Methadoned/l-Methadone300
    OpiatesMorphine300
    OpiatesMorphine2,000
    OxycodoneOxycodone100
    PhencyclidinePhencyclidine25
    PropoxyphenePropoxyphene300
    Tricyclic AntidepressantsNortriptyline1,000

    The single or multi-test panels can consist of the above listed analytes in any combination, up to a maximum of 16 analytes, with and without on-board adulteration/specimen validity tests (SVT) in the cup format. The drug screen tests are intended for prescription use only.

    The tests provide only a preliminary result. A more specific alternative chemical method should be used in order to obtain a confirmed presumptive positive result if the donor doesn't admit use or anytime required by testing procedures. Gas Chromatography / Mass Spectrometry (GC/MS), Liquid Chromatography / Mass Spectrometry (LC/MS) and their tandem mass-spectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results.

    Device Description

    For professional use, the devices consist of:

    • a. 10 or 25 test cups or dip cards with or without adulteration/specimen validity tests
    • b. Package insert
    • c. Procedure Card
    AI/ML Overview

    The provided text describes the performance characteristics of the ATTEST Drug Screen Cup and ATTEST Drug Screen Dip Card, which are rapid lateral flow immunoassays for the qualitative detection of various drugs in human urine.

    Here's an analysis of the acceptance criteria and the study proving the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for qualitative drug screening devices are typically based on agreement with a gold-standard confirmatory method (like GC/MS, LC/MS) at and around the cut-off concentrations. The concept of "agreement" translates to sensitivity and specificity for qualitative tests. While explicit numeric acceptance criteria (e.g., "must achieve >95% positive agreement") aren't directly stated as "acceptance criteria" in a separate table, they are implied by the performance statistics presented in the "Method Comparison/Accuracy Studies" section.

    The reported device performance is presented in the two large tables under "Method Comparison/Accuracy Studies" (Pages 14-16), showing the agreement of the ATTEST devices (Cup and Dip Card) with GC/MS or LC/MS results.

    Let's interpret the tables to deduce the implicit acceptance criteria and the reported performance. The "Agreement" column in these tables represents the percentage of agreement between the device's qualitative result (Negative or Positive) and the quantitative result from the gold-standard method, particularly around the cutoff and at significant concentrations above and below it.

    For each drug assay (e.g., 6-AM/10, AMP/500, etc.), the data shows the number of samples that fell into different concentration ranges relative to the cutoff (e.g., "Drug-free", "-50% C/O to <-25% C/O", "C/O to +25% C/O", ">+50% C/O"). The "Agreement" percentage is then provided for both Negative and Positive results.

    Implicit Acceptance Criteria (Deduced from Reported Performance):

    Drug Test/Cutoff (ng/mL)Acceptance Criteria (Implicit)Reported Device Performance (Agreement %) - ATTEST Cup & Dip Card
    GeneralHigh agreement (ideally 100% or very close to it) for drug-free samples and samples with concentrations significantly above the cutoff. Very high agreement is also expected for samples significantly below the cutoff that should test negative, and samples significantly above the cutoff that should test positive.Achieved for all assays as detailed below.
    Specific to Each AssayNegative Agreement: Expected to be high for drug-free samples and samples definitively below the cutoff. Positive Agreement: Expected to be high for samples definitively above the cutoff.
    6-AM/10Expected high agreement for Negatives (samples < 10 ng/mL) and Positives (samples >= 10 ng/mL).Neg: 100%, Pos: 100%
    AMP/500Expected high agreement for Negatives (samples < 500 ng/mL) and Positives (samples >= 500 ng/mL).Neg: 97.7%, Pos: 100%
    AMP/1000Expected high agreement for Negatives (samples < 1000 ng/mL) and Positives (samples >= 1000 ng/mL).Neg: 100%, Pos: 100%
    BAR/300Expected high agreement for Negatives (samples < 300 ng/mL) and Positives (samples >= 300 ng/mL).Neg: 95.2%, Pos: 100%
    BUP/10Expected high agreement for Negatives (samples < 10 ng/mL) and Positives (samples >= 10 ng/mL).Neg: 95.5%, Pos: 100%
    BZO/300Expected high agreement for Negatives (samples < 300 ng/mL) and Positives (samples >= 300 ng/mL).Neg: 93.2%, Pos: 100% (This is the lowest negative agreement)
    COC/150Expected high agreement for Negatives (samples < 150 ng/mL) and Positives (samples >= 150 ng/mL).Neg: 97.70%, Pos: 100%
    COC/300Expected high agreement for Negatives (samples < 300 ng/mL) and Positives (samples >= 300 ng/mL).Neg: 100%, Pos: 100%
    EDDP/300Expected high agreement for Negatives (samples < 300 ng/mL) and Positives (samples >= 300 ng/mL).Neg: 93.2%, Pos: 100% (Similar to BZO/300)
    MDMA/500Expected high agreement for Negatives (samples < 500 ng/mL) and Positives (samples >= 500 ng/mL).Neg: 95.5%, Pos: 100%
    MET/500Expected high agreement for Negatives (samples < 500 ng/mL) and Positives (samples >= 500 ng/mL).Neg: 93.2%, Pos: 100% (Similar to BZO/300 & EDDP/300)
    MET/1000Expected high agreement for Negatives (samples < 1000 ng/mL) and Positives (samples >= 1000 ng/mL).Neg: 100%, Pos: 100%
    MTD/300Expected high agreement for Negatives (samples < 300 ng/mL) and Positives (samples >= 300 ng/mL).Neg: 95.5%, Pos: 100%
    OPI/300Expected high agreement for Negatives (samples < 300 ng/mL) and Positives (samples >= 300 ng/mL).Neg: 97.72%, Pos: 100%
    OPI/2000Expected high agreement for Negatives (samples < 2000 ng/mL) and Positives (samples >= 2000 ng/mL).Neg: 93.2%, Pos: 100% (Similar to BZO/300, EDDP/300, MET/500)
    OXY/100Expected high agreement for Negatives (samples < 100 ng/mL) and Positives (samples >= 100 ng/mL).Neg: 93.2%, Pos: 100% (Similar to OPI/2000)
    PCP/25Expected high agreement for Negatives (samples < 25 ng/mL) and Positives (samples >= 25 ng/mL).Neg: 97.7%, Pos: 100%
    PPX/300Expected high agreement for Negatives (samples < 300 ng/mL) and Positives (samples >= 300 ng/mL).Neg: 95.3%, Pos: 100%
    TCA/1000Expected high agreement for Negatives (samples < 1000 ng/mL) and Positives (samples >= 1000 ng/mL).Neg: 95.5%, Pos: 100%
    THC/50Expected high agreement for Negatives (samples < 50 ng/mL) and Positives (samples >= 50 ng/mL).Neg: 97.7%, Pos: 100%

    Note: The performance percentages across the two formats (Cup and Dip Card) are presented identically in the provided combined tables for all assays, suggesting the performance data is pooled or very similar for both. The discordant results tables (pages 15-17) provide a more granular view for specific instances where the device result differed from the GC/MS or LC/MS result, primarily for samples near the cutoff. For example, for AMP/500, a 97.7% negative agreement means 2.3% of negative samples were discordant (likely false negative or near-cutoff negative that read positive by device). For Amp/500, 3 "Neg" results were in the "-50% C/O to <-25% C/O" range (i.e. between 250 and 375 ng/ml). This discrepancy is precisely what explains the 97.7% agreement and is not necessarily a "failure" given the nature of a qualitative test and its proximity to the cutoff.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for the Test Set:

      • For the main Method Comparison/Accuracy study (Table on pages 14-15), the sample size for each assay varies slightly but can be calculated from the row sums.
        • For "Negative" results: typically 40 drug-free samples + varying numbers of samples below cutoff (e.g., for 6-AM/10 Neg, 40+4+1 = 45 samples).
        • For "Positive" results: varying numbers of samples above and around cutoff (e.g., for 6-AM/10 Pos, 1+4+35 = 40 samples).
        • Total sample size per assay appears to be around 80-90 samples (e.g., 6-AM/10 has 45 Neg + 40 Pos = 85 samples). This includes drug-free samples and samples spiked at various concentrations relative to the cutoff (e.g., -50% C/O to <-25% C/O, -25% C/O to C/O, C/O to +25% C/O, >+25% C/O to +50% C/O, >+50% C/O).
      • The "Summary of Discordant Results" tables (pages 15-17) list specific instances of samples where the device's qualitative result did not match the quantitative GC/MS/LC/MS result (e.g., "Positive" device result for a sample with a drug concentration slightly below the cutoff, or "Negative" for a sample slightly above). These are qualitative examples, not the full dataset.
    • Data Provenance: The studies used clinical urine specimens "previously quantitated for the target drugs of abuse by gold-standard methods (GC/MS, LC/MS or equivalent)." The text does not specify the country of origin of the data or whether the samples were retrospective or prospective collections. However, given the nature of a 510(k) submission for a device to be marketed in the US, it is generally assumed that the clinical studies are relevant to the US population.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This type of device (lateral flow immunoassay for drug screening) does not rely on human experts for establishing ground truth for the test set. The ground truth is established by analytical methods (Gas Chromatography / Mass Spectrometry (GC/MS), Liquid Chromatography / Mass Spectrometry (LC/MS) and their tandem mass-spectrometer versions), which are considered the preferred confirmatory methods for drug testing. Therefore, there were no human "experts" in the sense of radiologists interpreting images to establish medical ground truth.

    4. Adjudication Method for the Test Set

    Not applicable. Since the ground truth is established by objective analytical methods (GC/MS, LC/MS), there's no need for an adjudication process involving human readers or experts. The assay results are compared directly against the quantitative chemical analysis.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, an MRMC comparative effectiveness study was not done. The device is a rapid diagnostic test for drug detection, not an AI-powered image analysis tool for human interpretation. The study is a comparison of the device's performance against a gold-standard chemical analysis method, not a study of human readers. Therefore, there is no effect size related to human readers improving with AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, the study described is essentially a standalone (algorithm/device-only) performance evaluation. The device's qualitative results are directly compared to the quantitative results from GC/MS/LC/MS, without human interpretation as an intermediate step. The "intended professional users" mentioned in Section M.a.iv (Detection Limit/sensitivity) are those who read the test result (presence/absence of a line), not those who interpret complex data as an AI model would.

    7. The Type of Ground Truth Used

    The ground truth used for the test set was analytical confirmation by gold-standard chemical methods: Gas Chromatography / Mass Spectrometry (GC/MS), Liquid Chromatography / Mass Spectrometry (LC/MS), or equivalent. This is explicitly stated in "Method Comparison/Accuracy Studies" (Page 14).

    8. The Sample Size for the Training Set

    The document describes a 510(k) submission for a diagnostic device. Unlike AI/ML models, these traditional diagnostic devices do not typically have a "training set" in the sense of a data set used to train an algorithm. The device's formulation and antibodies are developed through a research and development process, and its performance is then validated (tested) using various samples.

    The document mentions "The precision, reproducibility and sensitivity of the ATTEST Drug Screen Cup and the ATTEST Drug Screen Dip Card were evaluated at Advin Biotech and at three (3) external testing sites. Data obtained at all testing sites across the intended use populations indicate >99% correlation at +/-50% of each assay cutoff." (Page 8-9). This suggests internal development and testing, but not explicitly a "training set" as understood in machine learning. The stability and specificity studies also involve testing with numerous samples.

    9. How the Ground Truth for the Training Set was Established

    As there is no "training set" for this type of device in the AI sense, this question is not directly applicable. The device's underlying chemistry and design are based on established immunochromatographic principles. The process involves:

    • Standardization: The assays are harmonized with DHHS/SAMHSA screening cutoffs (Page 8-9).
    • Analytical Specificity: This is established by testing the device with contrived solutions of chemically related/structurally similar compounds and other potential interferents (Pages 9-13). This involves verifying that the device reacts appropriately to the target analyte and does not cross-react significantly with other substances.
    • Precision/Reproducibility: Evaluated through repeated testing at various concentrations and across different sites (Page 8).

    These evaluations contribute to the overall confidence in the device's performance and design, akin to "establishing ground truth" for its analytical properties, but it's fundamentally different from training an AI model on a labeled dataset.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1