Found 2 results

510(k) Data Aggregation

    K Number
    K210248

    Validate with FDA (Live)

    Date Cleared
    2021-07-30

    (182 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Lap Bag with fabric pouch system is applied in minimally invasive abdominal procedures for the safe and convenient removal of the tissue, organ or specimen during surgery.

    Device Description

    The Lap Bag with fabric pouch is applied in minimally invasive abdominal procedures for the safe and convenient removal of tissue, organ or specimens during surgery.

    The device is a TPU and nylon bonded fabric pouch that minimizes spillage and intraoperative contamination by isolating and containing tissue specimens. It is designed for introduction and use through 11/12 and 15mm cannula.

    The product is a single-use, sterile, and biocompatible laparoscopic accessory device. It is packaged by the Tyvek® 1073B sterile pouch and its shelf life is 3 years.

    AI/ML Overview

    The provided document is a 510(k) Summary for a medical device called "Lap Bag with Fabric Pouch". It describes the device, its intended use, and substantial equivalence to a predicate device, supported by performance data. However, it does not include specific quantitative acceptance criteria or a detailed study directly proving the device meets those criteria in the way typically found in clinical trials or software validation studies.

    This document focuses on demonstrating substantial equivalence to a legally marketed predicate device (Laparoscopic Tissue Retrieval Bag, Model 24003-MF, K123728) through various engineering and scientific tests, rather than clinical performance metrics with specific acceptance thresholds.

    Therefore, many of the requested details about acceptance criteria, sample sizes for test/training sets, expert involvement, and comparative effectiveness studies are not present in this type of regulatory submission for this particular device.

    Here's an attempt to answer the questions based on the provided text, while acknowledging the limitations:

    1. A table of acceptance criteria and the reported device performance

    The document does not present a single table of "acceptance criteria" alongside specific "reported device performance" in a quantitative manner as one might expect for diagnostic accuracy or clinical outcomes. Instead, it lists various tests performed and states whether the device "Pass"ed them. The implicit acceptance criteria are typically compliance with the referenced standards or internal protocols, and the reported performance is simply passing these tests.

    Test NameApplicable Standards/ProtocolImplicit Acceptance CriteriaReported Device Performance
    Biocompatibility Testing
    Cytotoxicity TestISO 10993-5:2009No significant cytotoxic effectsPass
    Sensitization TestISO 10993-10:2010No significant sensitizationPass
    Intracutaneous Reactivity TestISO 10993-10:2010No significant irritation/reactivityPass
    Acute Systemic Toxicity TestISO 10993-11:2017No significant systemic toxic effectsPass
    Material-Mediated Pyrogenicity TestUSP <151>Non-pyrogenicPass
    Sterilization Validation & Shelf Life Study
    Ethylene oxide Sterilization Validation StudyISO 11135:2014, ISO 11737-1:2018, ISO 11737-2:2019Achieve Sterility Assurance Level (SAL) of 10-6Meets requirements
    Package Integrity Test (Dye penetration Test)ASTM F1929-15:2015Maintain sterile barrier integrityPass
    Seal Peel Strength TestASTM F88/F88M-15:2015Maintain seal strengthPass
    Product Sterility TestISO 11737-2:2019SterilePass
    Product Performance Test
    Loading capacity testN/A (followed by the internal testing protocol)Meet specified internal capacity requirements (e.g., hold a certain weight/volume without tearing)Pass
    Leak-resistant testN/A (followed by the internal testing protocol)Maintain leak resistance (e.g., no fluid leakage under specified conditions)Pass

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not explicitly state sample sizes for individual tests. The tests are bench tests and laboratory validations, not human clinical trials. Data provenance in terms of country of origin or retrospective/prospective is not applicable or stated for these types of engineering and biocompatibility tests. The manufacturer, Advanced Medical Design Co., Ltd., is based in Taiwan.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable to the types of tests reported. The "ground truth" for these tests are the objective results of standard laboratory procedures (e.g., measuring cytotoxicity, sterile barrier integrity, or leak resistance) against predefined ISO or ASTM standards or internal protocols. No human experts are described as establishing "ground truth" in the context of these evaluations.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This is not a study involving human interpretation or subjective assessment that would require an adjudication method.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This document describes a physical medical device (a laparoscopic retrieval bag), not an AI-powered diagnostic or assistive tool. Therefore, MRMC studies and "human readers improve with AI" metrics are irrelevant.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This is not an algorithm or AI device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the various tests are based on:

    • Biocompatibility: Standardized biological assays and in vivo tests as per ISO 10993 series and USP <151> standards, which have objective endpoints.
    • Sterilization Validation: Microbiological testing and physical indicators to verify a Sterility Assurance Level (SAL) of 10-6, according to ISO 11135 and ISO 11737.
    • Shelf Life Study: Physical testing of packaging integrity (dye penetration, seal peel strength) and product sterility after accelerated or real-time aging, per ASTM F1929, ASTM F88/F88M, and ISO 11737.
    • Product Performance Tests (Loading Capacity, Leak-resistant): Verification against internal testing protocols, implying objective measurement against predefined specifications.

    8. The sample size for the training set

    Not applicable. This device does not involve machine learning or a training set.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K181887

    Validate with FDA (Live)

    Date Cleared
    2019-01-22

    (193 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    AMD Anti-Fog Solution is a sterile and single-use laparoscopic accessory device for use prior to and during endoscopic and laparoscopic procedures. The product is intended to prevent condensation of the distal lenses of the endoscopic/ laparoscopic instruments during the surgery. The product is intended for prescription use only.

    Device Description

    The AMD Anti-Fog Solution is a single-use, sterile, and biocompatible laparoscopic accessory device. The product is packaged by the Tyvek" Pouch which includes one clear plastic screw top dispenser bottle with the volume of 6 mL and one polyurethane (PU) foam pad with x-ray detectable radiopaque ribbon. The solution is a clear/colorless, odorless, and aqueous solution which comprises of surfactant (≤2%), isopropyl alcohol (≤2%), and water (≥96%). The polyurethane (PU) foam pad is an adhesivebacked sponge with radiopaque ribbon that can be used to introduce the solution into the surface of the lenses.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document primarily focuses on demonstrating substantial equivalence to a predicate device, rather than defining explicit performance acceptance criteria with numerical targets for the AMD Anti-Fog Solution itself (beyond biocompatibility, sterility, and shelf life). However, the "Performance" row in Table 5.1 and the "Product Performance Test" section provide the core performance claim.

    Acceptance Criteria (Implied)Reported Device Performance
    Biocompatibility (no adverse biological reactions)All biocompatibility tests (Cytotoxicity, Sensitization, Intracutaneous Reactivity, Acute Systemic Toxicity, Material-Mediated Pyrogenicity) passed requirements.
    Sterility (Sterility Assurance Level < 10⁻⁶)Routine sterilization parameter at ≥25 kGy achieved SAL < 10⁻⁶.
    Shelf-life (maintenance of sterility and function for 5 years)All aging tests (Package Integrity, Seal Peel Strength, Product Sterility, Fog Resistance) passed for a proposed 5-year shelf life.
    Performance (Preventing condensation of distal lenses)"AMD Anti-Fog Solution is as effective as the predicate device."

    2. Sample Size for the Test Set and Data Provenance:

    • Biocompatibility Testing: The specific number of samples (e.g., number of animals for in vivo tests, number of cell cultures for in vitro tests) is not explicitly stated in the provided document for each test. The tests were performed according to ISO standards, which would typically specify sample sizes.
    • Sterilization Validation: The VDmax 25 method was used, implying a standard number of samples derived from product bioburden testing to establish the sterilization dose. The document states bioburden was < 1000 CFU/sample, but the number of samples tested for bioburden is not specified.
    • Product Aging Validation (Shelf Life Study): The number of samples for each test (Package Integrity, Seal Peel Strength, Product Sterility, Product Stability/Fog Resistance) is not explicitly stated. These tests were performed for a 5-year shelf life validation.
    • Product Performance Test (Fog Resistance Test): The sample size for this test is not explicitly stated.
    • Data Provenance: The studies were conducted by the manufacturer, Advanced Medical Design Co., Ltd., based in Taiwan. The location where the tests were performed (e.g., specific labs) is not detailed. The data is retrospective in the sense that it was completed prior to the 510(k) submission.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    • This information is not applicable in the context of this submission. The device is a physical product (anti-fog solution), and its performance (biocompatibility, sterility, fog resistance, shelf life) is evaluated through objective laboratory tests against established standards, not through human interpretation or expert-driven ground truth. Therefore, no "experts" were required to establish ground truth in the sense of clinical assessment or image interpretation.

    4. Adjudication Method for the Test Set:

    • Not applicable. As the performance is based on objective laboratory tests against specified standards, there is no need for an adjudication method as would be used for subjective assessments or disputes among human readers/interpreters.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done:

    • No. This type of study is typically used for diagnostic or screening devices where human interpretation plays a role. The AMD Anti-Fog Solution is a laparoscopic accessory intended to prevent fogging, not a diagnostic tool requiring human reader interpretation. Therefore, an MRMC study and analysis of AI vs. human reader improvement is not relevant.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) study was done:

    • Not applicable. The device is a physical solution, not an algorithm or AI system.

    7. The Type of Ground Truth Used:

    • Objective Laboratory Measurements and Standard Criteria:
      • Biocompatibility: Ground truth was established by recognized international standards (ISO 10993 series) which define acceptable biological responses.
      • Sterility: Ground truth was established by microbiological testing methods (ISO 11137-2, ISO 11737-1, ISO 11737-2) to achieve a specified sterility assurance level (SAL).
      • Shelf Life (Packaging Integrity, Seal Strength): Ground truth was established by recognized ASTM and ISO standards for packaging materials.
      • Shelf Life (Product Stability/Fog Resistance): Ground truth was established by an "internal testing protocol." The specifics of this protocol are not detailed, but it would involve a measurable criterion for fog resistance.
      • Overall Performance (Fog Resistance): The claim is based on equivalence to the predicate device, implying that its fog resistance is at least as good as the predicate's, which itself would have been demonstrated through specific test methods.

    8. The Sample Size for the Training Set:

    • Not applicable. This device is not an AI/ML algorithm or software that requires a training set.

    9. How the Ground Truth for the Training Set Was Established:

    • Not applicable. See point 8.