Search Results
Found 1 results
510(k) Data Aggregation
(136 days)
The VPOD™ Intrathoracic Pressure Regulator is indicated for the temporary decrease in intrathoracic pressure to increase blood circulation, as prescribed by a physician, licensed practitioner, or qualified technician. Recommended duration of use is up to four hours.
The proposed VPOD Intrathoracic Pressure Regulator (ITPR) is also intended to be used with patients who would benefit from increased venous return back to the heart for treatment of a number of different clinical conditions associated with severe hypotension but generates its own vacuum via the VPOD Motor and connects with a positive pressure source such as a separate ventilator or manual resuscitator bag.
The provided text is a 510(k) summary for the VPOD™ Intrathoracic Pressure Regulator. It describes the device, its indications for use, and its comparison to a predicate device (CirQlator™). The document outlines various tests conducted to demonstrate substantial equivalence but does not explicitly provide a table of acceptance criteria with reported device performance in the format requested.
However, based on the non-clinical testing and comparative performance sections, we can infer some of the performance aspects and the studies that support them.
Here's an attempt to structure the information as requested, though some details like specific numerical acceptance criteria are not explicitly stated in the document.
Acceptance Criteria and Study Details for VPOD™ Intrathoracic Pressure Regulator
The provided document, a 510(k) summary for the VPOD™ Intrathoracic Pressure Regulator (K172388), describes various tests and comparisons to establish substantial equivalence with its predicate device, the CirQlator™ (K070490). While explicit numerical acceptance criteria are not tabulated, the document details performance characteristics that were compared and found to be equivalent or superior.
1. Table of Acceptance Criteria and Reported Device Performance:
| Feature/Metric | Acceptance Criteria (Inferred from Predicate Equivalence) | Reported Device Performance (VPOD™ ITPR) |
|---|---|---|
| Biocompatibility | Non-cytotoxic, non-irritating, non-mutagenic, margin of safety > 1 per ISO 10993. | PASS: Materials were found to be non-cytotoxic, non-irritating, non-mutagenic, and with a margin of safety > 1 for Externally Communicating, Tissue, Limited durations (<24 hours) per ISO 10993-1. Specific tests included: Cytotoxicity (ISO 10993-5), Sensitization (ISO 10993-10), Irritation (ISO 10993-10), Acute Systemic Toxicity (ISO 10993-11), Gas emission VOC with risk based assessment, Particulate Matter (PM2.5). |
| Electrical Safety, EMC | Compliance with AAMI ANSI ES 60601-1, IEC 60601-1-2. | PASS: Device tested for compliance. |
| Alarms | Compliance with IEC 60601-1-8. | PASS: Device tested for compliance. |
| Ventilator Compatibility | Compliance with applicable requirements of ISO 80601-2-12. | PASS: Device tested for compliance. |
| Vacuum Accuracy | Capable of achieving negative intrathoracic pressure up to -9 mmHg (-12 cmH2O). | PASS: Achieved negative expiratory airway pressure of approximately -9 mmHg (-12 cmH2O). Continuously variable from -2 to -12 cmH2O. Monitored by microprocessor and pressure sensors. Any out of range performance (greater than -9 mmHg) causes Motor to shut down. |
| Mean Arterial Pressure (MAP) Improvement | Comparable improvement to predicate CirQlator in compromised circulation model. | PASS: Both VPOD and CirQlator improved mean arterial pressure by 17 mmHg (after bleed and recovery) at minute 30 of device use in an animal study. |
| Maximum Airway Pressure | No significant difference compared to predicate during device use. | PASS: No significant difference between the positive airway pressures of the CirQlator and the VPOD during device use. |
| Maximum Negative Airway Pressure (Expiratory) | No significant difference compared to predicate. Capable of -9 mmHg (-12 cmH2O). | PASS: No significant differences between the negative airway pressures achieved with either device, demonstrating substantial equivalence. Both devices achieved a negative expiratory airway pressure of approximately -9 mmHg (-12 cmH2O). |
| Perturbation of Ventilator Monitors (Exhaled Volume) | No perturbation of exhaled volume or other ventilator monitors. | PASS: VPOD does not perturb exhaled volume or other ventilator monitors. (In contrast, the predicate CirQlator directs exhaled gas to wall suction, activating ventilator alarms requiring silencing/disabling). |
| Adjustable Vacuum Level | (Predicate fixed at -12 cmH2O) | PASS: VPOD allows user to adjust expiratory airway pressure in increments of -1 cmH2O from -2 to -12 cmH2O, offering more clinical latitude. |
| Cleaning, Shelf-life, Aging Effects | Ensured validation and verification of VPOD Motor and effects of aging. | PASS: Validation and verification of VPOD Motor effects of aging were performed. |
| Other Bench Tests | Accuracy of airway pressure measurement, flow rates, simulated altitude, resistance to flow, compatibility with ventilator (pressure/flow), mechanical drop, dead space, blower leakage, induced temperature rise/change. | PASS: These bench performance tests were conducted. (Specific results not detailed beyond "Device performance testing included"). |
| Battery Safety | Compliance with UL 2054. | PASS: Device tested for compliance. |
| Usability | Demonstrated usability. | PASS: Usability testing was conducted. (Specific results not detailed). |
2. Sample Size Used for the Test Set and Data Provenance:
-
Animal Study: The document explicitly mentions an "Animal Study" was conducted comparing the subject device to the predicate.
- Sample Size: Not explicitly stated.
- Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). It is a "comparative performance" study.
-
Bench Performance: No explicit sample sizes are provided for the various bench tests listed (e.g., Accuracy – airway pressure, flow rates, vacuum accuracy, etc.). These tests are presumed to be conducted in a laboratory setting.
- Sample Size: Not explicitly stated.
- Data Provenance: Laboratory testing.
-
Biocompatibility: Tests were performed on materials.
- Sample Size: Not explicitly stated for each specific test, but representative material samples would have been used.
- Data Provenance: Laboratory testing.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts:
This information is not provided in the document. The studies described are primarily technical performance tests (bench and animal), not clinical studies requiring expert ground truth for interpretation of outcomes from patient-derived data (e.g., image interpretation).
4. Adjudication Method for the Test Set:
This information is not applicable as the described tests are technical performance evaluations (bench and animal studies), not studies involving human interpretation or adjudication of diagnostic findings.
5. If a Multi-reader Multi-case (MRMC) Comparative Effectiveness Study was Done:
No, an MRMC comparative effectiveness study was not conducted. The studies described are an animal study comparing device performance and various bench tests.
- Effect size of how much human readers improve with AI vs without AI assistance: Not applicable, as this is neither an AI device nor a human reader study.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study was Done:
This question is not directly applicable as the VPOD™ Intrathoracic Pressure Regulator is a medical device, not an AI algorithm. Its standalone performance was evaluated through bench testing and, presumably, the animal study where its intrinsic function was assessed without human intervention in the mechanism of action. The device itself is designed to be "standalone" in its vacuum generation, while still being used with assisted ventilation by a clinician.
7. The Type of Ground Truth Used:
- Biocompatibility: Laboratory standard tests for cytotoxicity, irritation, sensitization, etc., against established ISO standards.
- Bench Performance: Engineering specifications, physical measurements, and compliance with relevant IEC/AAMI standards.
- Animal Study: Physiological parameters (e.g., mean arterial pressure, airway pressure measurements) directly measured in an animal model of compromised circulation, compared against predicate device performance.
8. The Sample Size for the Training Set:
This information is not applicable. The VPOD™ is a hardware medical device with integrated microprocessor control, not a machine learning model that requires a "training set" of data in the typical sense.
9. How the Ground Truth for the Training Set Was Established:
This information is not applicable for the same reason as #8. The device's control system and operational parameters would be based on engineering design, physiological principles, and iterative testing rather than a data training set.
Ask a specific question about this device
Page 1 of 1