510(k) Data Aggregation

    K Number
    K243092

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-03-19

    (170 days)

    Product Code
    Regulation Number
    868.2375
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    AcuPebble Ox200 is a wearable device intended for use in the recording, analysis, displaying, exporting, and storage of biophysical parameters to aid in the evaluation of adult patients with, or with suspected, obstructive sleep apnea (OSA). The device is primarily intended for home setting use (although can also be used in healthcare settings) under the direction of a Healthcare Professional (HCP).

    Device Description

    AcuPebble Ox200 is prescribed by a Health Care Professional for the patient to use in the home as a 'home sleep apnea test' (HSAT). AcuPebble Ox200 comprises: 1- A small microphones/accelerometer sensor, identical to the one in the previously cleared devices AcuPebble SA100 (K210480) and AcuPebble Ox100 (K222950) that is worn on the front of the neck around the suprasternal notch area (the "AcuPebble SA100 sensor"), and attaches with a single use biocompatible adhesive tape (same one as for "AcuPebble SA100" and "AcuPebble Ox100)); 2- Another PPG/accelerometer sensor ("AcuPebble_oximetry sensor") worn either on the finger or on the forehead; 2- A mobile device app (the "AcuPebble Ox200 app") that guides the patient through the steps of the test, collects the data from the sensors, and uploads them in the cloud ; 3- A cloud-based software that analyses the collected signals as in the previously cleared devices; 4- A web app user interface for healthcare professionals where they can set up a study and see the results, including different physiological channel traces and OSA diagnostic parameters, as in the previously cleared devices.

    The differences between AcuPebble Ox200 and the previously cleared device (K222950) are:

    • A new sensor for sensing the PPG signal and extraction of the SpO2 and pulse rate channel from. This new sensor can be placed either on the finger or the forehead.
    • The addition of new parameters extracted from the acoustic neck sensor including position, respiratory efforts and respiratory rate channels.

    The OSA diagnostic indexes are still extracted from the signals measured with the identical sensor as in the predicate (neck sensor only), and with the same algorithms, remain unchanged with respect to the predicate device (K222950) and hence have not been separately tested in this device.

    AI/ML Overview

    Here's a summary of the acceptance criteria and the studies that prove the device meets those criteria, based on the provided text:

    AcuPebble Ox200 Acceptance Criteria and Performance Study Summary

    1. Table of Acceptance Criteria and Reported Device Performance

    ParameterAcceptance Criteria (Target or Reference Performance)Reported Device Performance (AcuPebble Ox200)
    OSA Diagnostic IndexesAHI (3% desaturation criteria): Diagnostic Sensitivity > 81.3% (from reference device)92.73%
    AHI (4% desaturation criteria): Diagnostic Sensitivity > 81.3% (from reference device)95.92%
    AHI (3% desaturation criteria): Diagnostic Specificity > 82.1% (from reference device)96.84%
    AHI (4% desaturation criteria): Diagnostic Specificity > 82.1% (from reference device)97.03%
    Pulse Rate (Acoustics)50-120bpm, RMS error < 3.62bpm (from predicate device)50-120bpm, 3.62bpm validation accuracy (from acoustics)
    Pulse Rate (PPG)50-120bpm, RMS error < 2 bpm (from predicate device)Arms (validation accuracy) < 1.23bpm (for both forehead and finger, average RMSE of 1.23 bpm across varying skin tones), worst case RMSE 1.66 bpm (using functional simulator)
    Respiratory Rate4-35 bpm, accuracy ±2.7 bpm (from reference device Airmod K2427984)Overall RMSD 2.52 bpm (4-30bpm, lab study with capnography); RMSD 2.46 bpm (during natural sleep, compared to Embletta MPR chest/abdomen bands). Claimed range: 4-30 bpm, resolution 1 bpm, accuracy ±3 bpm.
    Respiratory EffortNot inferior to reference device (Embletta MPR RIP bands) for differentiating no effort vs. effortAgreement of 96.8% with manual scoring by two clinicians when distinguishing apnoea with no effort vs. with effort. Outperformed reference RIP bands of Embletta MPR (K122516) with 96.3% accuracy vs. 75.9% accuracy when using Embla RemLogic (K162140) automatic scoring as reference.
    Body Position DetectionNot inferior to reference device (implied standard from reference device for accelerometer-based position sensing)Lab study: 93.93% overall accuracy (supine, prone, left, right, upright). Natural sleep study: 79.00% accuracy overall; 83.59% accuracy in detecting supine vs. non-supine.
    SpO2 Accuracy (Arms)Comparable to predicate (implied standard, likely ISO 80601-2-61:2017 standards for accuracy with varying skin tones for pulse oximeters)1.86 (forehead), 2.25 (finger) in the range of 70% to 100%. Lightest skin tone (Fitzpatrick 1, Monk 1): 1.63 (forehead), 1.44 (finger). Darkest skin tone (Fitzpatrick 6, Monk 10): 0.66 (forehead), 1.36 (finger). (All values for SpO2 accuracy (Arms) are within typical acceptance for pulse oximetry).

    2. Sample Sizes and Data Provenance

    • SpO2 and Pulse Rate: 12 subjects with varying skin tone. The study was performed by an independent testing lab. (Country of origin not specified, but the device manufacturer is UK-based.) Data appears to be prospective clinical validation.
    • Respiratory Rate:
      • Study 1: 31 healthy subjects. Conducted under laboratory conditions. (Country of origin not specified, but the device manufacturer is UK-based.) Data appears to be prospective clinical validation.
      • Study 2: 150 patients referred for OSA diagnosis. Conducted during natural sleep. Royal Free London Hospital NHS Foundation Trust (UK). Data appears to be prospective clinical validation, with a trial registration number (NCT03544086).
    • Respiratory Effort: 150 patients referred for OSA diagnosis. Conducted during natural sleep. Royal Free London Hospital NHS Foundation Trust (UK). Data appears to be prospective clinical validation, with a trial registration number (NCT03544086).
    • Body Position Detection:
      • Study 1: 17 volunteers. Conducted under laboratory conditions. (Country of origin not specified, but the device manufacturer is UK-based.) Data appears to be prospective clinical validation.
      • Study 2: 60 patients referred for OSA diagnosis. Conducted during natural sleep. Virgen Macarena Hospital, Spain. Data appears to be prospective clinical validation, with a trial registration number (NCT04028011).

    3. Number of Experts and Qualifications for Ground Truth (Test Set)

    • Respiratory Effort: "manual scoring by two clinicians". Qualifications of the clinicians are not explicitly stated (e.g., specific specialties or years of experience), but they are described as "clinicians." For the automatic scoring comparison, an FDA-cleared device (Embla RemLogic, K162140) served as a reference.
    • Other parameters (SpO2, Pulse Rate, Respiratory Rate, Body Position): The ground truth relied on established reference devices and standards (e.g., functional simulator, capnography, RIP bands, ISO 80601-2-61:2017), rather than expert consensus on individual cases for the test set. For OSA diagnostic indexes, the performance is inherited from the predicate device, so the ground truth setup for that specific parameter is not detailed for AcuPebble Ox200 in this document.

    4. Adjudication Method (Test Set)

    • Respiratory Effort: Manual scoring was done by "two clinicians," and the agreement between them or a further adjudication process (e.g., 2+1) is not specified beyond "the agreement in distinguishing." However, a primary comparison for superiority was made against an FDA-cleared device's automatic scoring (Embla RemLogic).
    • For other parameters, adjudication is not applicable as ground truth was established by reference devices/standards.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study is explicitly mentioned as having been performed to assess how much human readers improve with or without AI assistance for this device. The evaluations focus on the standalone performance of the device's outputs.

    6. Standalone Performance (Algorithm Only)

    Yes, standalone performance was done for several new parameters:

    • SpO2 and Pulse Rate Output: Validated as per ISO 80601-2-61:2017 using an independent testing lab.
    • Respiratory Rate: Validated in two studies against capnography and respiratory inductance plethysmography (RIP) bands.
    • Respiratory Effort Detection: Validated against RIP bands and compared to automatic scoring of an FDA-cleared device.
    • Body Position Detection: Validated in laboratory conditions and during natural sleep.
    • The OSA diagnostic indexes (AHI) are explicitly stated to be "extracted from the signals measured with the identical sensor as in the predicate (neck sensor only), and with the same algorithms, remain unchanged with respect to the predicate device (K222950) and hence have not been separately tested in this device." Their performance is inherited from the predicate.

    7. Type of Ground Truth Used (Test Set)

    The ground truth varied by parameter:

    • SpO2 and Pulse Rate: Reference pulse oximetry and functional simulator, following ISO 80601-2-61:2017 standards for accuracy assessment.
    • Respiratory Rate: Capnography (K150272) in a lab setting, and chest and abdomen respiratory effort bands of Embletta MPR Sleep Data Recording System (K122516) during natural sleep.
    • Respiratory Effort: Manual scoring by two clinicians based on RIP bands (Respiratory Inductance Plethysmography) and automatic scoring from the FDA-cleared device Embla RemLogic (K162140).
    • Body Position Detection: Presumed to be observational ground truth established by study personnel in lab, and potentially manually annotated based on video/observation in natural sleep studies, or potentially using other reference accelerometers. The document does not specify the exact method for ground truth for body posture.

    8. Sample Size for the Training Set

    The document does not explicitly state the sample size used for the training set for any of the algorithms. It focuses on the validation studies.

    9. How the Ground Truth for the Training Set Was Established

    As the training set information is not provided, the method for establishing its ground truth is also not detailed in this document. The document primarily describes the validation of the device's performance against established clinical standards and reference devices.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222950

    Validate with FDA (Live)

    Device Name
    AcuPebble OX100
    Manufacturer
    Date Cleared
    2023-02-03

    (129 days)

    Product Code
    Regulation Number
    868.2375
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    AcuPebble Ox100 is a wearable device intended for use in the recording, analysis, displaying, exporting, and storage of biophysical parameters to aid in the evaluation of adult patients with, or with suspected, obstructive sleep apnea (OSA). The device is primarily intended for home settings use (although can also be used in healthcare settings) under the direction of a Healthcare Professional (HCP).

    Device Description

    AcuPebble Ox100 is prescribed by a Health Care Professional for the patient to use in the home as a 'home sleep apnea test' (HSAT). AcuPebble Ox100 comprises: 1- A small physiological sensing device, identical to the previously cleared device AcuPebble SA100 (K210480) that is worn on the front of the neck around the suprasternal notch area (the "AcuPebble SA100 sensor"), and attaches with a single use biocompatible adhesive tape (same one as for "AcuPebble SA100" too); 2- Another physiological wireless sensing device ("AcuPebble finger oximetry sensor") worn on the finger; 3- A mobile device app (the "AcuPebble Ox100 app") that guides the patient through the steps of the test, collects the data from the sensor (continuously in the case of the "AcuPebble SA100 sensor", and at the end of the test in the case of the "AcuPebble finger oximetry sensor"), and uploads them in the cloud ; 4- A cloud-based software that analyses the signals from the "AcuPebble SA100 sensor" (already cleared as part of K210480); 5- A web app user interface for healthcare professionals where they can set up a study and see the results, including different physiological channel traces and OSA diagnostic parameters. Outputs of the device relevant to its intended use are further described in the comparison table below.

    AI/ML Overview

    The provided text describes the AcuPebble Ox100 device and its substantial equivalence to a predicate device (AcuPebble SA100, K210480) based on various performance data and testing. However, it does not contain a detailed study specifically proving that the device meets defined acceptance criteria for its effectiveness in aiding the evaluation of Obstructive Sleep Apnea (OSA).

    The information provided primarily focuses on:

    • Comparison to Predicate and Reference Devices: Highlighting similarities in indications for use, regulation, target population, environment of use, and certain performance metrics previously claimed for the predicate or a reference device (NightOwl, K191031).
    • Compliance with Recognized Consensus Standards: For biocompatibility, electrical safety (EMC), home healthcare safety, risk management, and software development.
    • High-level performance metrics: AHI, ODI, Diagnostic LR+, LR-, Sensitivity, Specificity, and Pulse rate/SpO2 performance, which appear to be inherited or similar to the predicate/reference device claims rather than outcomes of a new primary study for AcuPebble Ox100's specific combination of sensors.

    Information NOT explicitly available in the provided text:

    • A dedicated table of acceptance criteria for AcuPebble Ox100 with its own reported performance to demonstrate compliance.
    • Detailed information about a specific clinical study conducted for AcuPebble Ox100 to prove its diagnostic performance. The diagnostic metrics listed (Sensitivity, Specificity, etc.) are presented as "Same" compared to the predicate, implying these are existing claims rather than results from a new study specific to Ox100 to justify a de novo clearance.
    • Sample sizes for a test set specifically for demonstrating the performance of AcuPebble Ox100 in diagnosing OSA.
    • Data provenance (country of origin, retrospective/prospective) for a study validating AcuPebble Ox100's diagnostic capabilities.
    • Number of experts, their qualifications, or adjudication methods for establishing ground truth of a test set for AcuPebble Ox100's diagnostic performance.
    • Whether a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done for AcuPebble Ox100, or any effect size of AI assistance.
    • The type of ground truth used for a study validating AcuPebble Ox100's diagnostic capabilities.
    • Sample size and ground truth establishment for the training set of AcuPebble Ox100's diagnostic algorithm.

    Based on the provided text, here's what can be extracted, and what is missing or implied:

    1. A table of acceptance criteria and the reported device performance:

    The document doesn't explicitly present a standalone table of "acceptance criteria" for the AcuPebble Ox100 followed by "reported device performance" from a new study of this specific device. Instead, it compares the AcuPebble Ox100's characteristics and performance to its predicate (AcuPebble SA100) and a reference device (NightOwl) in a tabular format. The performance metrics listed are presented as "Same" to the predicate, implying that these are the established performance levels from the predicate device that the AcuPebble Ox100 is deemed to meet by being substantially equivalent.

    Here's a reconstruction of the relevant performance metrics mentioned, noting that these are largely presented as being "Same" as the predicate:

    Performance MetricAcceptance Criteria (Implied / From Predicate)Reported Device Performance (Likely Inherited from Predicate)
    Diagnostic LR+Not explicitly stated as acceptance criteria, but listed as 29.36 /32.29 for predicate29.36 /32.29 (Stated as "Same")
    Diagnostic LR-Not explicitly stated as acceptance criteria, but listed as 0.08 /0.04 for predicate0.08 /0.04 (CI- 0.03 to 0.19/0.01 to 0.16) (Stated as "Same")
    Diagnostic SensitivityNot explicitly stated as acceptance criteria, but listed as 92.73% /95.92% for predicate92.73% /95.92% (CI-82.41% to 97.98% /86.02% to 99.50%) (Stated as "Same")
    Diagnostic SpecificityNot explicitly stated as acceptance criteria, but listed as 96.84% /97.03% for predicate96.84% /97.03% (Stated as "Same")
    Pulse rate (from heart sounds)Not explicitly stated as acceptance criteria, but listed as 50-120bpm for predicate50-120bpm (3.62bpm validation accuracy) (Stated as "Same")
    Pulse rate (from PPG)±2bpm or ±2%Meets requirement of ±2bpm or ±2%.
    SpO2 performance (80%-100% range)Compliant with ISO 80601-2-61 (2% root-mean-square difference)2% root-mean-square difference
    SpO2 performance (70%-79% range)Compliant with ISO 80601-2-61 (3% root-mean-square difference)3% root-mean-square difference
    SpO2 performance (70-100% SaO2)Compliant with ISO 80601-2-61 (≤4% root-mean-square difference) (from reference device)(Implied to meet this based on specific ranges above)
    BiocompatibilityISO 10993 requirements (sensitization, irritation, cytotoxicity, USP class VI)Complies (passing results for finger sensor, neck sensor same as cleared predicate)
    General Safety and EMCIEC 60601-1:2005:MOD, IEC 60601-1-2:2014Complies
    Home Healthcare Environment SafetyIEC 60601-1-11:2015Complies
    Risk ManagementISO 14971:2019Complies
    Software Verification and ValidationIEC 62304:2006/A1:2016, FDA Cybersecurity GuidanceComplies (software level of concern: "moderate")

    2. Sample size used for the test set and the data provenance:

    • Not specified for a new effectiveness study of AcuPebble Ox100. The performance metrics (Diagnostic LR+, LR-, Sensitivity, Specificity) are stated to be the "Same" as the predicate device (AcuPebble SA100). Therefore, any sample size and data provenance would refer to the previous clearance of K210480. This information is not provided in the document for K210480.
    • For the SpO2 performance, compliance with ISO 80601-2-61 is stated. Such compliance typically requires specific testing against a reference oximeter in a clinical setting with a certain number of subjects across defined SpO2 ranges, but the specific sample size, country of origin, or retrospective/prospective nature of this verification for AcuPebble Ox100 is not detailed.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not specified for a new effectiveness study of AcuPebble Ox100. This information would be relevant for the diagnostic performance metrics, which are stated to be "Same" as the predicate. Details for the predicate (K210480) are not in this document.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not specified for a new effectiveness study of AcuPebble Ox100. This would relate to the diagnostic performance metrics, which are stated to be "Same" as the predicate. Details for the predicate (K210480) are not in this document.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No such study is mentioned or described for AcuPebble Ox100. The device is intended to aid in evaluation, but no comparative effectiveness study with human readers (with vs. without AI assistance) is detailed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • The diagnostic performance metrics listed (Sensitivity, Specificity, LR+, LR-) usually reflect standalone algorithm performance against a gold standard. However, these are listed as "Same" to the predicate device (AcuPebble SA100). The document does not provide details of a new standalone study for AcuPebble Ox100 itself to derive these numbers. The "cloud-based software that analyses the signals" and "web app user interface for healthcare professionals" suggest an algorithmic analysis component.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Not specified for a new effectiveness study of AcuPebble Ox100. For diagnostic assessments of OSA, polysomnography (PSG) scored by experts according to AASM guidelines is the typical gold standard/ground truth. However, the document does not confirm this for a specific study validating AcuPebble Ox100's performance.

    8. The sample size for the training set:

    • Not specified. The document refers to "A cloud-based software that analyses the signals" but does not provide details on its training set or how it was developed for AcuPebble Ox100.

    9. How the ground truth for the training set was established:

    • Not specified.
    Ask a Question

    Ask a specific question about this device

    K Number
    K210480

    Validate with FDA (Live)

    Device Name
    AcuPebble SA
    Manufacturer
    Date Cleared
    2021-07-06

    (137 days)

    Product Code
    Regulation Number
    868.2375
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    AcuPebble SA100 is indicated to sense, record, and interpret a patient's physiological signals (including respiratory pattern) during sleep for the purpose of prescreening patients for obstructive sleep apnea (OSA) syndrome. The designed for use in home-screening of adults with suspected possible sleep breathing disorders (although it can also be used in clinic). Results are used to assist the healthcare professional's in the patient's evaluation.

    The system is not intended as a substitute for full polysomnography when additional parameters such as sleep stages, limb movements, or EEG activity are required.

    Device Description

    AcuPebble SA100 comprises a sensor that is worn on the front of the neck around the suprasternal notch area (the "AcuPebble SA100 sensor"), a mobile device app that guides the patient through the steps of the test, collects the data from the sensor, and uploads them in the "AcuPebble SA100 app"), and a cloud-based analysis software (the "AcuPebble SA100 algorithms". The "AcuPebble SA100 Sensor" consists of a small biocompatible enclosure with very low power sensors and electronics inside. It is attached to the front of the neck (somewhere between the Adam's apple and the suprasternal notch area) by single-use biocompatible adhesive tape. The sensor measures physiological body sounds. The data sensed is streamed via a Bluetooth link to the "AcuPebble SA100 app" on a smart device (mobile phone/tablet). The data can be uploaded into the cloud where it is analyzed, any time after the sleep study test is finished at the tap of a button in the app. AcuPebble SA100 signal processing algorithms produce a number of OSA diagnostic traces and parameters. These can be accessed by the Health Care Professional through a healthcare professional portal, which can be accessed from any computer or smart device with an Internet connection.

    AcuPebble SA100 is not intended to be used with patients with pacemakers or other implantable devices, or with patients with known or suspected arrhythmias. This device has not been validated in patients with congestive heart failure or patients with neuromuscular disorders. Patients with significant cardiopulmonary or neurological disorders need to be excluded from using the device.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:

    Device Name: AcuPebble SA100
    Intended Use: To sense, record, and interpret a patient's physiological signals (including respiratory pattern) during sleep for the purpose of prescreening patients for obstructive sleep apnea (OSA) syndrome. Designed for home-screening of adults with suspected possible sleep breathing disorders to assist healthcare professionals in patient evaluation. Not a substitute for full polysomnography.


    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state pre-defined "acceptance criteria" with numerical thresholds for diagnostic accuracy (PPV, NPV, LR+, LR-). Instead, it presents the results of the clinical validation study for several diagnostic indices (AHI3, AHI4, ODI3, ODI4) and heart rate accuracy. The implicit acceptance is that these values demonstrate substantial equivalence to the predicate device.

    Implicit Acceptance Criteria (Performance Benchmarks from Predicate/Context): Not explicitly stated with numerical targets in the provided text. The submission focuses on demonstrating substantial equivalence to the predicate (Drowzle, K173974) and reference devices (NightOwl, K191031; Sleep Strip II, K112822). The reported values in the table are the result of meeting the substantial equivalence standard, rather than being against pre-set numerical acceptance criteria.

    Reported Device Performance:

    Diagnostic IndexPositive Predictive Value (PPV)Negative Predictive Value (NPV)Positive Likelihood Ratio (LR+)Negative Likelihood Ratio (LR-)
    AHI3 (for airflow reductions and 3% desaturations)94.44% (CI 84.78%- 98.11%)95.83% (CI 89.94%- 98.34%)29.36 (CI 9.62 to 89.64)0.08 (CI 0.03 to 0.19)
    AHI4 (for 4% desaturations)94.00% (CI 83.69%- 97.95%)98.00% (CI 92.65% to 99.48%)32.29 (CI 10.58 to 98.59)0.04 (CI 0.01 to 0.16)
    ODI3 (considering 3% desaturations)93.42% (CI 85.87%- 97.07%)90.54% (CI 82.48%- 95.11%)13.11 (CI 5.61 to 30.62)0.10 (CI 0.05 to 0.20)
    ODI4 (considering 4% desaturations)85.71% (CI 75.50%- 92.11%)98.94% (CI 93.03%- 99.85%)12.37 (CI 6.35-24.08)0.02 (CI 0.00-0.15)

    Heart Rate Accuracy:

    Bias (µ) ≤ ±1 bpmBias (µ) ≤ ±3 bpmBias (µ) ≤ ±5 bpmStandard Deviation (σ) ≤ 2 bpmStandard Deviation (σ) ≤ 5 bpmStandard Deviation (σ) ≤ 8 bpm
    62.0%94.0%98.7%28.7%93.3%98.7%

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: 150 consecutive evaluation patients.
    • Data Provenance: Not specified in terms of country of origin. The study was a "Clinical Investigation," conducted in accordance with ISO 14155:2011 and 21 CFR 812.28(a)(1) (GCP for clinical investigations). This implies a prospective study design for data collection, as it refers to "consecutive evaluation patients."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • The document does not specify the number or qualifications of experts used to establish the ground truth. It states that the ground truth for diagnostic accuracy was based on "the AASM for airflow reductions and desaturations" and "established thresholds, for OSA (i.e. 5, 15 and 30)". For heart rate, the ground truth was the "PPG signal of the polygraphy system." This implies that the ground truth was derived from a gold-standard PSG (Polysomnography) system, rather than expert interpretation of the device's output alone.

    4. Adjudication Method for the Test Set

    • The document does not mention any specific adjudication method (e.g., 2+1, 3+1, none) for the test set. Ground truth appears to be derived from the gold standard PSG, rather than expert review of the device's output.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    • No, an MRMC comparative effectiveness study involving human readers with and without AI assistance was not described. The clinical validation focuses on the device's accuracy compared to a polysomnography system, not on assisting human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    • Yes, the clinical validation study appears to be a standalone performance evaluation. The device (AcuPebble SA100) generates diagnostic indices (AHI, ODI) and heart rate output, and these are compared directly against a polygraphy system, which serves as the ground truth. The role of a human healthcare professional is described as using the results to "assist" in patient evaluation, implying the device provides direct outputs.

    7. The Type of Ground Truth Used

    • Type of Ground Truth: The ground truth for OSA diagnostic indices (AHI, ODI) was established using definitions from the "AASM" (American Academy of Sleep Medicine) for airflow reductions and desaturation criteria (3% and 4%). This indicates comparison against polysomnography (PSG), which is the gold standard for sleep apnea diagnosis. For heart rate, the ground truth was "the heart rate obtained from the PPG signal of the polygraphy system," further confirming PSG as the reference.

    8. The Sample Size for the Training Set

    • The document does not specify the sample size used for the training set. The clinical validation section only refers to the "150 consecutive evaluation patients" used for the accuracy validation clinical study (i.e., the test set).

    9. How the Ground Truth for the Training Set Was Established

    • The document does not specify how the ground truth for the training set was established, as the training set details are not provided.