Search Results
Found 39 results
510(k) Data Aggregation
(184 days)
The Libre Rio Continuous Glucose Monitoring System is an over-the-counter (OTC) integrated continuous glucose monitoring (iCGM) device indicated for non-insulin using persons age 18 and older. The System detects trends and tracks patterns and aids in the detection of euglycemia, and hypoglycemia. The System is also intended to autonomously communicate with digitally connected devices.
The Libre Rio Continuous Glucose Monitoring System (herein referred to as the 'System') is an integrated continuous glucose monitoring system (iCGM) that provides continuous glucose measurements every minute to facilitate calculation of glucose values accompanied by trend information (glucose arrows) and historical glucose information (glucose graph). The System is intended for over-the-counter use in a home setting. The System consists of the following components: a Sensor which transmits via Bluetooth Low Energy (BLE), and a mobile application Libre Rio App that is downloaded to a compatible smartphone running iOS and Android operating system.
Here are the details regarding the acceptance criteria and the study that proves the device meets them, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The provided document does not explicitly present a table of acceptance criteria with corresponding performance metrics for the Libre Rio Continuous Glucose Monitoring System. It primarily focuses on the device's substantial equivalence to a predicate device and summarizes various performance testing categories. The "Reported Device Performance" for each category usually states that acceptance criteria were met, without detailing the specific metrics or thresholds.
However, based on the type of testing mentioned and common practices for CGM devices, we can infer some general areas where acceptance criteria would have been established and reportedly met:
| Area of Performance Testing | Implied Acceptance Criteria (General) | Reported Device Performance (Summary) |
|---|---|---|
| Software Verification & Validation | Adherence to IEC 62304 and FDA guidance for software functions, no critical bugs, software performs as intended. | "Results of executed protocols met the acceptance criteria and therefore support that the System software is acceptable for its intended use." |
| Cybersecurity | Adherence to FDA guidance for cybersecurity in medical devices, identification and mitigation of threat/vulnerability risks, protection of confidentiality, integrity, and availability of data. | "Appropriate risk mitigation controls have been implemented and tested." |
| Interoperability | Compliance with FDA guidance for interoperable medical devices, successful communication with digitally connected devices. | Approach "developed in alignment with FDA guidance," implying successful implementation. |
| Human Factors | Adherence to ANSI/AAMI/IEC 62366, IEC 60601-1-6, and FDA guidance for human factors, demonstrating usability and safety for intended users. | "The analysis and the study performed demonstrated that the changes implemented for the subject device meet the usability requirement for its intended use." |
| Bench Testing (CT, MRI, X-ray compatibility) | Device maintains functionality and accuracy under specific CT, MRI, and X-ray conditions; no adverse effects. | "The test results showed all functionality testing acceptance criteria was met." |
| Biocompatibility, Sterility, Shelf Life, Packaging, Electrical Safety, EMC, Mechanical Design, Clinical Performance | Each established for the predicate device, implying they met relevant standards and criteria. | "The Libre Rio Sensor is identical to the predicate FreeStyle Libre 2 Sensor and no design changes were introduced to allow compatibility to the Libre Rio App. Therefore, the following supportive performance characteristics established for the predicate device (K222447) is applicable to the subject device and is not impacted." These were implicitly met by the predicate and thus deemed met for the subject device. |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the sample sizes for the test sets used in the listed performance tests. It mentions "studies" and "testing" but does not quantify the number of participants or data points.
The data provenance is not explicitly mentioned for the reported tests. However, the study is for the Libre Rio Continuous Glucose Monitoring System, which appears to be a new device (or an updated version) from Abbott Diabetes Care, Inc., located in Alameda, CA, USA. This suggests the data would likely be generated in the USA, and primarily from prospective testing conducted specifically for this submission, especially for areas like human factors, bench testing for new contraindications, and software validation. The clinical performance data is stated to be derived from the predicate device (K222447), which would have its own provenance details.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
The document does not provide information on the number of experts used or their qualifications for establishing ground truth for any of the mentioned test sets. It broadly refers to "studies" and "testing" but does not detail the methodology of ground truth establishment for specific components like software, cybersecurity, or human factors.
4. Adjudication Method for the Test Set
The document does not specify any adjudication methods (e.g., 2+1, 3+1, none) used for establishing ground truth in the performance testing.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
No, the document does not mention a Multi-Reader Multi-Case (MRMC) comparative effectiveness study. The device is a continuous glucose monitoring system, which typically involves direct measurement and user interaction, not interpretation by multiple human readers of clinical cases. The "Human Factors" testing mentioned focuses on usability and user interface, not on comparative effectiveness with human interpretation of readings.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done
Yes, implicit in the description of the device and testing, especially for "Software Verification and Validation" and the "Sensor glucose algorithm," is that standalone algorithm performance testing was conducted. The device's "Sensor glucose algorithm" is stated to be the "ADC Glucose Algorithm established for the predicate device," which would have undergone rigorous standalone validation. The "Libre Rio App" is designed to autonomously receive and display glucose data, suggesting the algorithm operates independently of immediate human intervention for value generation. The bench testing of the sensor's compatibility with CT, MRI, and X-ray would also involve objective measurements without human interpretation in the loop to assess the sensor's function.
7. The Type of Ground Truth Used
The document does not explicitly state the type of ground truth used for each specific test. However, based on the nature of the device and the tests:
- Clinical Performance (inherited from predicate): For CGM devices, the gold standard for ground truth is typically blood glucose measurements obtained from a laboratory reference method (e.g., YSI analyzer) at various glucose levels.
- Software Verification and Validation: Ground truth would be based on functional specifications and expected outputs for given inputs.
- Cybersecurity: Ground truth involves adherence to security protocols and identified risk mitigations.
- Human Factors: Ground truth is established through user task completion rates, error rates, and subjective feedback against predefined usability goals and safety requirements.
- Bench Testing (CT, MRI, X-ray): Ground truth would be based on physical and electrical performance standards and expected device behavior under specific environmental conditions, potentially using calibrated instruments to verify sensor output accuracy.
8. The Sample Size for the Training Set
The document does not provide information about the sample size for the training set. Given that the sensor glucose algorithm is "ADC Glucose Algorithm established for the predicate device," the training data would be associated with the development of that original algorithm, not necessarily new training for the Libre Rio specifically (unless modifications were made, which is not indicated for the algorithm itself).
9. How the Ground Truth for the Training Set Was Established
The document does not specify how the ground truth for the training set (for the inherited "ADC Glucose Algorithm") was established. For glucose monitoring algorithms, ground truth for training data is typically established through paired comparisons with laboratory reference methods (e.g., YSI blood glucose measurements) across a diverse range of glucose values, patient populations, and physiological conditions.
Ask a specific question about this device
(172 days)
Libre 2 Sensor users:
The FreeStyle Libre 2 Flash Glucose Monitoring System is a continuous glucose monitoring (CGM) device with real time alarms capability indicated for the management of diabetes in persons age 4 and older. It is intended to replace blood glucose testing for diabetes treatment decisions, unless otherwise indicated.
The System also detects trends and tracks patterns and aids in the detection of episodes of hyperglycemia and hypoglycemia, facilitating both acute and long-term therapy adjustments. Interpretation of the System readings should be based on the glucose trends and several sequential readings over time.
The System is also intended to autonomously communicate with digitally connected devices. The System can be used alone or in conjunction with these digitally connected devices where the user manually controls actions for therapy decisions.
Libre 2 Plus Sensor users:
The FreeStyle Libre 2 Flash Glucose Monitoring System is a continuous glucose monitoring (CGM) device with real time alarms capability indicated for the management of diabetes in persons age 2 and older. It is intended to replace blood glucose testing for diabetes treatment decisions, unless otherwise indicated.
The System also detects trends and tracks patterns and aids in the detection of episodes of hyperglycemia and hypoglycemia, facilitating both acute and long-term therapy adjustments. Interpretation of the System readings should be based on the glucose trends and several sequential readings over time.
The System is also intended to autonomously communicate with digitally connected devices, including automated insulin dosing (AID) systems. The System can be used alone or in conjunction with these digitally connected devices for the purpose of managing diabetes.
Libre 3 Sensor users:
The FreeStyle Libre 3 Continuous Glucose Monitoring System is a real time continuous glucose monitoring (CGM) device with alarms capability indicated for the management of diabetes in persons age 4 and older. It is intended to replace blood glucose testing for diabetes treatment decisions, unless otherwise indicated.
The System also detects trends and tracks patterns and aids in the detection of episodes of hyperglycemia and hypoglycemia, facilitating both acute and long-term therapy adjustments. Interpretation of the System readings should be based on the glucose trends and several sequential readings over time.
The System is also intended to autonomously communicate with digitally connected devices. The System can be used alone or in conjunction with these digitally connected devices where the user manually controls actions for therapy decisions.
Libre 3 Plus Sensor users:
The FreeStyle Libre 3 Continuous Glucose Monitoring System is a real time continuous glucose monitoring (CGM) device with alarms capability indicated for the management of diabetes in persons age 2 and older. It is intended to replace blood glucose testing for diabetes treatment decisions, unless otherwise indicated.
The System also detects trends and tracks patterns and aids in the detection of episodes of hyperglycemia and hypoglycemia, facilitating both acute and long-term therapy adjustments. Interpretation of the System readings should be based on the glucose trends and several sequential readings over time.
The System is also intended to autonomously communicate with digitally connected devices, including automated insulin dosing (AID) systems. The System can be used alone or in conjunction with these digitally connected devices for the purpose of managing diabetes.
The FreeStyle Libre 3 Continuous Glucose Monitoring System (hereinafter also referred to as 'FSL3 System') and FreeStyle Libre 2 Flash Glucose Monitoring System (hereinafter also referred to as the 'FSL2 System') are integrated continuous glucose monitoring (iCGM) systems that provide continuous glucose measurements every minute to provide glucose levels, trends, and real-time alarms capability to aid in the management of diabetes. The FSL2 and FSL3 Systems also provide configurable alarms designed to warn the user of Low Glucose, High Glucose or Signal Loss. The user may make treatment decisions based in part on the sensor glucose results provided by both Systems. The FSL2 and FSL3 Systems require a prescription and are intended for home use.
The subject FSL3 System consists of a sensor (the FreeStyle Libre 3 Sensor (FSL3 Sensor) or FreeStyle Libre 3 Plus Sensor (FSL3 Plus Sensor)) and a primary display device (the FreeStyle Libre 3 Reader (FSL3 Reader) or the FreeStyle Libre App (FSL App) [iOS and Android] downloaded to a compatible phone). The FSL3 Reader and the FSL App do not interact with each other.
The subject FSL2 System consists of a sensor (the FreeStyle Libre 2 Sensor (FSL2 Sensor) or FreeStyle Libre 2 Plus Sensor (FSL2 Plus Sensor)) and a primary display device (the FreeStyle Libre 2 Reader (FSL2 Reader) or the FSL App (iOS and Android) downloaded to a compatible phone). The FSL2 Reader and FSL App do not interact with each other.
Both the FSL2 and FSL3 Systems are compatible with the Libre Data Sharing API cleared in K223537. The display device of the connected FSL2 or FSL3 Systems, which directly receives the data from the sensor, continues to serve as a primary display device for the glucose data and alarms.
This document is a 510(k) summary from Abbott Diabetes Care Inc. regarding their FreeStyle Libre 3 Continuous Glucose Monitoring System and FreeStyle Libre 2 Flash Glucose Monitoring System. It outlines the device details, indications for use, comparison to predicate devices, and a summary of performance testing.
Acceptance Criteria and Study Proving Device Meets Criteria
The provided text focuses on the substantial equivalence of the new FreeStyle Libre 2 Plus and 3 Plus sensors and updated app features to their predicate devices, rather than establishing new acceptance criteria for the entire system's glucose monitoring accuracy. The document states that clinical performance and human factors were "established in the predicate device (K223435) and are not affected by the introduction of the FSL App in this 510(k)." This implies that the glucose accuracy and user interface performance metrics (e.g., MARD, % in Zones A+B, usability outcomes) for the core glucose monitoring function were previously accepted for the predicate devices and are assumed to hold true for the current submission due to the nature of the changes being primarily related to sensor life, age ranges, and app features.
Therefore, the acceptance criteria and performance data discussed below are extrapolated from the types of performance studies conducted to demonstrate substantial equivalence for the modifications presented in this 510(k), particularly regarding safety and compatibility, and from the implied reliance on the predicate device's established clinical performance.
Please note: This response does not contain specific quantitative acceptance criteria or precise performance metrics for glucose accuracy (e.g., MARD values) as these details are explicitly referred to as having been established in the predicate device (K223435) and are not re-evaluated in this submission. The focus of this 510(k) is on the changes from the predicate.
1. Table of Acceptance Criteria and Reported Device Performance
Given the nature of this 510(k) (modifications to existing systems), the presented "acceptance criteria" are derived from the tests conducted to demonstrate that the changes do not adversely affect safety and effectiveness.
| Category | Specific Test/Feature | Acceptance Criteria (Implied/Stated) | Reported Device Performance/Conclusion |
|---|---|---|---|
| Software | Software Verification and Validation | Compliant with IEC 62304 and FDA Guidance "Content of Premarket Submissions for Device Software Functions" (June 14, 2023) and "Multiple Function Device Products: Policy and Considerations" (July 29, 2020). | Results of executed protocols met the acceptance criteria and therefore support that the system software is acceptable for its intended use. |
| Cybersecurity | Cybersecurity Risk Management | Compliant with FDA guidance "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions'' (September 27, 2023). Appropriate risk mitigation for identified threats and vulnerabilities. | Risk assessment performed, appropriate risk mitigation controls implemented and tested. |
| Bench Testing | Removal of CT/MRI Contraindications, X-ray Caution | Functionality testing acceptance criteria met for sensor performance after exposure to CT, MRI (MR conditional), and X-ray. | The test results showed all functionality testing acceptance criteria was met. (Previous mechanical, electrical, and functional testing established in the predicate device are not affected). |
| App Feature: Auto-Display (FSL2) | Automatic display of glucose results without user-initiated scan. | Functionality: Glucose data automatically displays via BLE without user scan, while NFC scan option remains. | Functionality confirmed. (Users may still perform NFC scan as an option, consistent with predicate FSL2 App behavior.) |
| App Feature: Alarms Escalation | Incremental volume increase for alarms (High, Low, Urgent Low, Signal Loss). | Volume increases incrementally over 30 seconds until maximum volume, equal to predicate device alarm volume. | Functionality confirmed: Alarms increase volume over 30 seconds to the maximum level of predicate device alarms. |
| Core Glucose Monitoring Performance | Clinical Performance (e.g., Accuracy) | Established in predicate device K223435. (No new evaluation specified in this 510(k) summary for these aspects). | Established in the predicate device (K223437) and not affected by the introduction of the FSL App in this 510(k). (This implies that the existing performance data from K223435 is deemed sufficient for these modified devices). |
| Human Factors | Usability and User Interface (new features) | Established in predicate device K223435. (No new evaluation specified in this 510(k) summary for these aspects beyond software V&V). | Established in the predicate device (K223437) and not affected by the introduction of the FSL App in this 510(k). (The app navigation change to "Bottom bar navigation menu" from "Side panel navigation", and the "App Stopped Alarm" for Android were likely part of software V&V and user acceptance testing, implied by overall software V&V acceptance). |
2. Sample Size Used for the Test Set and Data Provenance
The document does not provide specific sample sizes for the detailed technical tests (Software V&V, Cybersecurity, Bench Testing for CT/MRI/X-ray). These types of tests typically involve controlled laboratory environments rather than human subject test sets in the traditional sense for clinical performance.
For the core clinical performance (glucose accuracy, etc.), the document explicitly states these characteristics were "established in the predicate device (K223435) and are not affected." Therefore, any sample size for these fundamental performance aspects would refer to those used in the studies for the predicate device, which are not detailed in this 510(k) summary.
Data Provenance: The document does not specify the country of origin for any testing data. The nature of the tests (Software V&V, Cybersecurity, Bench Testing) suggests they are likely conducted in a controlled environment, potentially in-house or by a certified testing facility. The "retrospective or prospective" nature is not specified, but these are typically prospective tests conducted as part of the development and verification process.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This type of information (number of experts, qualifications, and ground truth establishment) is typically relevant for clinical studies involving expert adjudication (e.g., for diagnostic imaging devices). Given that the 510(k) summary indicates clinical performance was "established in the predicate device" and that the current submission primarily addresses technical and software improvements/modifications, there's no mention of a new clinical study involving multiple experts to establish ground truth for this specific submission.
For the bench testing related to CT/MRI/X-ray compatibility, ground truth would be established by physical measurements and functional checks against engineering specifications, not by human experts.
4. Adjudication Method for the Test Set
Adjudication methods (e.g., 2+1, 3+1) are typically used in clinical studies, particularly for diagnostic accuracy where multiple human readers interpret data. As explained above, this 510(k) focuses on technical and software changes, and refers back to the predicate device for clinical performance. Therefore, no information on an adjudication method for a test set is provided or expected in this context.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done
No, an MRMC comparative effectiveness study was not done, and it is not typically relevant for a CGM device. MRMC studies are primarily conducted for diagnostic imaging devices to assess how AI assistance impacts human reader performance (e.g., radiologists interpreting scans). CGM devices provide quantitative glucose measurements, and their performance is evaluated directly against a reference method (e.g., laboratory glucose measurements), not through human reader interpretation of complex images or cases.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done
Yes, for the core glucose algorithm, a standalone performance evaluation is inherent to a CGM device. The document explicitly states:
- Glucose Algorithm: "ADC Glucose Algorithm"
- Location of Glucose Algorithm: "FSL2 System: Receiver", "FSL3 System: Sensor"
This implies that the algorithm processes the sensor's raw electrical signals to calculate a glucose value and trend independently. The clinical performance of this algorithm (e.g., MARD against a reference method) would have been part of the studies for the predicate device (K223435). This 510(k) does not present new data for this, but rather relies on the predicate's established performance, stating that ADC (Abbott Diabetes Care) Glucose Algorithm is the "Same" for the subject device.
7. The Type of Ground Truth Used
For Continuous Glucose Monitoring (CGM) devices, the ground truth for establishing accuracy is typically obtained from laboratory reference methods for blood glucose concentration, such as:
- YSI analysis (Yellow Springs Instrument): This is a widely accepted laboratory reference method for measuring glucose in blood or plasma samples.
While not explicitly stated in this 510(k) summary for the ground truth method, it is the standard for CGM accuracy studies which would have been performed for the predicate device to which this submission refers for clinical performance.
8. The Sample Size for the Training Set
The document does not provide a sample size for the training set of the glucose algorithm. This information would typically be detailed in the original predicate device's submission which established the clinical performance. The current 510(k) refers to the "ADC Glucose Algorithm" as being the same as the predicate, indicating no retraining or significant algorithm changes that would necessitate reporting new training data.
9. How the Ground Truth for the Training Set was Established
Similar to the sample size, the specifics of how the ground truth was established for the training set would have been detailed in the original submission for the predicate device. For CGM algorithms, this involves comparing the sensor's raw electrical signals to paired blood glucose values obtained from a highly accurate laboratory reference method (like YSI), collected over time from a cohort of study participants under various glycemic conditions. This establishes the true glucose concentration against which the algorithm learns to convert sensor signals into accurate glucose readings.
Ask a specific question about this device
(150 days)
The FreeStyle Libre 2 Flash Glucose Monitoring System is a continuous glucose monitoring (CGM) device with real time alarms capability indicated for the management of diabetes in persons age 2 and older. It is intended to replace blood glucose testing for diabetes treatment decisions, unless otherwise indicated.
The System also detects trends and tracks patterns and aids in the detection of episodes of hyperglycemia and hypoglycemia, facilitating both acute and long-term therapy adjustments. Interpretation of the System readings should be based on the glucose trends and several sequential readings over time.
The System is also intended to autonomously communicate with digitally connected devices, including automated insulin dosing (AID) systems. The System can be used alone or in conjunction with these digitally connected devices for the purpose of managing diabetes.
The FreeStyle Libre 3 Continuous Glucose Monitoring System is a real time continuous glucose monitoring (CGM) device with alarms capability indicated for the management of diabetes in persons age 2 and older. It is intended to replace blood glucose testing for diabetes treatment decisions, unless otherwise indicated.
The System also detects trends and tracks patterns and aids in the detection of episodes of hyperglycemia and hypoglycemia, facilitating both acute and long-term therapy adjustments. Interpretation of the System readings should be based on the glucose trends and several sequential readings over time.
The System is also intended to autonomously communicate with digitally connected devices, including automated insulin dosing (AID) systems. The System can be used alone or in conjunction with these digitally connected devices for the purpose of managing diabetes.
The FreeStyle Libre 3 Continuous Glucose Monitoring System (hereinafter also referred to as 'FSL3 System') and FreeStyle Libre 2 Flash Glucose Monitoring System (hereinafter also referred to as the 'FSL2 System') are integrated continuous glucose monitoring (iCGM) systems that provide continuous glucose measurements every minute to provide glucose levels, trends, and real-time alarms capability to aid in the management of diabetes. The FSL2 and FSL3 Systems also provide configurable alarms designed to warn the user of Low Glucose, High Glucose or Signal Loss. The user may make treatment decisions based in part on the Sensor glucose results provided by both Systems. The FSL2 and FSL3 Systems require a prescription and are intended for home use.
The FSL3 System consists of the FreeStyle Libre 3 Sensor and either the FreeStyle Libre 3 Reader or the FreeStyle Libre 3 App (iOS and Android) downloaded to a compatible smartphone as a primary display device. The FSL3 Reader and FSL3 App do not interact with each other.
The FSL2 System consists of the FreeStyle Libre 2 Sensor and either the FreeStyle Libre 2 Reader or the FreeStyle Libre 2 App (iOS and Android) downloaded to a compatible smartphone as a primary display device. The FSL2 Reader and FSL2 App do not interact with each other.
Both the FSL2 and FSL3 Systems are compatible with the Libre Data Sharing API cleared in K223537. The display device of the connected FSL2 or FSL3 Systems, which directly receives the data from the Sensor, continues to serve as a primary display device for the glucose data and alarms.
1. A table of acceptance criteria and the reported device performance:
The document primarily focuses on demonstrating substantial equivalence to a predicate device rather than providing a detailed table of clinical acceptance criteria and specific numerical performance. However, it does state that "System accuracy was demonstrated to meet the iCGM special controls requirements per 21 CFR 862.1355." This regulation for Integrated Continuous Glucose Monitoring Systems outlines performance standards.
Based on the document, the following aspects were evaluated:
| Acceptance Criteria Category | Reported Device Performance |
|---|---|
| Accuracy (Clinical) | Met iCGM special controls requirements per 21 CFR 862.1355. |
| Sterility | Sterility Assurance Level (SAL) of 10^-6 achieved with minimum sterilization dose of 25 kGy, established by VDmax25 method and ISO 11137-1/11137-2. |
| Shelf-Life, Packaging Integrity, Shipping | Seal integrity, user accessibility, and device functionality met acceptance criteria after worst-case testing. |
| Electrical Safety | Compliant with IEC 60601-1: 2005(r)2012, IEC 60601-1-6:2010+A1:2013, and IEC 60601-1-11:2015. |
| Electromagnetic Compatibility (EMC) | Compliant with IEC 60601-1-2 and IEC CISPR 11. Wireless coexistence testing confirmed functionality within acceptable limits in presence of common radiating electronic devices (AAMI TIR69, ANSI C63.27, FDA Guidance). Compliant with FCC Regulations Part 15.225 and Part 15.247, and FAA Advisory Circular RTCA DO-160. |
| Mechanical Engineering | Met acceptance criteria for mechanical, electrical, and functional testing at system level and for individual Sensor Applicator components. |
| Biocompatibility | Demonstrated biocompatibility per ISO10993-1 and FDA Guidance "Use of International Standard ISO 10993-1." |
| Software Verification and Validation | Met acceptance criteria for executed protocols per IEC 62304 and FDA Guidance "Content of Premarket Submissions for Software Contained in Medical Devices." |
| Cybersecurity | Cybersecurity risk management documentation analyzed confidentiality, integrity, and availability of data, information, and software. Appropriate risk mitigation controls implemented and tested per October 2014 FDA Guidance. |
| Human Factors | Risk analysis demonstrated design changes met usability requirements per ANSI/AAMI/IEC 62366, IEC 60601-1-6, and FDA Guidance "Applying Human Factors and Usability Engineering to Medical Devices." |
| Interoperability | Incorporated an approach for interoperability in alignment with FDA Guidance "Design Considerations and Pre-market Submission Recommendations for Interoperable Medical Devices." |
2. Sample size used for the test set and the data provenance:
The document does not explicitly state the sample size for clinical test sets. For the "Clinical Performance" section, it only mentions that "System accuracy was demonstrated to meet the iCGM special controls requirements per 21 CFR 862.1355." This often implies a clinical trial or study, but specific numbers are not provided in this 510(k) summary. The provenance of the data (country of origin, retrospective/prospective) is also not detailed in this summary.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
This information is not provided in the document. For glucose monitoring systems, "ground truth" is typically established by laboratory reference methods (e.g., YSI glucose analyzer) rather than expert human interpretation.
4. Adjudication method for the test set:
This information is not provided in the document.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
This is not applicable to the FreeStyle Libre Continuous Glucose Monitoring System. This device is a standalone diagnostic tool that provides glucose readings directly to the user (or digitally connected devices), not a system designed to assist human readers in interpreting medical images or data. Therefore, an MRMC study comparing human reader improvement with/without AI assistance would not be conducted for this type of device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
Yes, the device is inherently designed to operate in a standalone manner (algorithm only) to provide glucose readings. The "Clinical Performance" section implicitly refers to this standalone performance when stating "System accuracy was demonstrated to meet the iCGM special controls requirements per 21 CFR 862.1355." The FSL3 System's glucose algorithm resides in the Sensor, and for FSL2, it resides in the Receiver (App or Reader), generating glucose values without human real-time intervention for each reading.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
For Continuous Glucose Monitoring (CGM) systems, the ground truth for accuracy studies is typically established by laboratory reference methods, such as those using a YSI glucose analyzer, which provides highly accurate and precise blood glucose measurements. While not explicitly stated in this summary, this is standard practice for CGM device validation.
8. The sample size for the training set:
The document does not provide information about the training set size. This 510(k) summary focuses on the substantial equivalence and performance of the final device, not the specifics of its development or algorithm training.
9. How the ground truth for the training set was established:
This information is not provided in the document. Similar to the test set, it would typically involve comparison to highly accurate laboratory reference methods.
Ask a specific question about this device
(200 days)
The FreeStyle Libre 2 Flash Glucose Monitoring System is a continuous glucose monitoring (CGM) device with real time alarms capability indicated for the management of diabetes in persons age 2 and older. It is intended to replace blood glucose testing for diabetes treatment decisions, unless otherwise indicated.
The System also detects trends and tracks patterns and aids in the detection of episodes of hyperglycemia and hypoglycemia, facilitating both acute and long-term therapy adjustments. Interpretation of the System readings should be based on the glucose trends and several sequential readings over time.
The System is also intended to autonomously communicate with digitally connected devices, including automated insulin dosing (AID) systems. The System can be used alone or in conjunction with these digitally connected devices for the purpose of managing diabetes.
The FreeStyle Libre 3 Continuous Glucose Monitoring System is a real time continuous glucose monitoring (CGM) device with alarms capability indicated for the management of diabetes in persons age 2 and older. It is intended to replace blood glucose testing for diabetes treatment decisions, unless otherwise indicated.
The System also detects trends and tracks patterns and aids in the detection of episodes of hyperglycemia and hypoglycemia, facilitating both acute and long-term therapy adjustments. Interpretation of the System readings should be based on the glucose trends and several sequential readings over time.
The System is also intended to autonomously communicate with digitally connected devices, including automated insulin dosing (AID) systems. The System can be used alone or in conjunction with these digitally connected devices for the purpose of managing diabetes.
The FreeStyle Libre 2 Flash Glucose Monitoring System (with modified Sensor tail), hereinafter also referred to as 'modified FSL2 System' and FreeStyle Libre 3 Continuous Glucose Monitoring System (with modified Sensor tail), hereinafter also referred to as 'modified FSL3 System' are integrated continuous glucose monitoring (iCGM) systems that provide continuous glucose measurements every minute to provide glucose levels, trends, and real-time alarms capability to aid in the management of diabetes. The Systems also provide configurable alarms designed to warn the user of Low Glucose or Signal Loss. The user may make treatment decisions based in part on the Sensor glucose results provided by the Systems require a prescription and are intended for home use.
The FreeStyle Libre 2 System consists of the FreeStyle Libre 2 Sensor and either the FreeStyle Libre 2 Reader or the FreeStyle Libre 2 App downloaded to a compatible smartphone as a primary display device. The FreeStyle Libre 3 System consists of the FreeStyle Libre 3 Sensor and the FreeStyle Libre 3 App downloaded to a compatible smartphone as a primary display device.
Both the FreeStyle Libre 2 and FreeStyle Libre 3 Systems are compatible with the Libre Data Sharing API cleared in K223537. The display device of the connected FreeStyle Libre 2 or FreeStyle Libre 3 Systems, which directly receives the data from the Sensor, continues to serve as a primary display device for the glucose data and alarms.
The Abbott FreeStyle Libre 2 Flash Glucose Monitoring System and FreeStyle Libre 3 Continuous Glucose Monitoring System sought clearance for modified sensor tails to reduce vitamin C interference, extend sensor wear to 15 days, and enable compatibility with automated insulin dosing (AID) systems.
Here's an analysis of the acceptance criteria and supporting studies:
1. Table of Acceptance Criteria & Reported Device Performance:
The document doesn't explicitly list a table of "acceptance criteria" with specific numeric targets for device performance. Instead, it states that the devices meet the iCGM special controls for clinical performance set forth in 21 CFR 862.1355. This regulation broadly outlines requirements for accuracy, reliability, and safety of integrated Continuous Glucose Monitoring systems.
However, based on the provided text, we can infer some key performance objectives and the reported outcome:
| Acceptance Criteria (Inferred from Changes & Regulations) | Reported Device Performance |
|---|---|
| Accuracy across measuring range (15-day wear) | The clinical study demonstrated that the subject devices meet the iCGM special controls for clinical performance. Specific accuracy metrics (e.g., MARD) are not provided in this summary but would be part of the full submission. |
| Reduced Vitamin C Interference | An additional clinical study evaluated the effects of high doses of vitamin C, and the results showed the devices met iCGM special controls. Labeling was updated to reflect Vitamin C interference at doses >1000 mg/day (improved from >500 mg/day for the predicate). |
| Extended Sensor Wear Duration (from 14 to 15 days) | The subject devices (modified FreeStyle Libre 2 and 3 Sensors) are designed for a 15-day wear duration, and performance was evaluated throughout this duration in the clinical study. |
| Compatibility with Automated Insulin Dosing (AID) systems | Interoperability testing and an interoperability communication plan were provided to ensure secure and reliable data transfer with digitally connected devices, including AID systems. |
| Software Functionality and Safety | Software regression testing was conducted in accordance with established specifications and IEC 62304. Results met acceptance criteria. |
| Non-clinical Performance | Non-clinical performance testing was conducted as per internal design control process requirements, and results met acceptance criteria. |
| Known Performance of Unaffected Aspects | Sterilization, biocompatibility, shelf life stability, packaging integrity/shipping integrity, mechanical engineering, electrical safety and electromagnetic compatibility, and human factors were established in predicate/reference devices and are not affected. Cybersecurity for FSL3 also relies on existing data. |
2. Sample Size Used for the Test Set and Data Provenance:
- Sample Size: The clinical study evaluated performance in adult (18 years and older) and pediatric (2 to 17 years) participants with diabetes. Specific numbers of participants within these age groups are not provided in this summary.
- Data Provenance: The document does not specify the country of origin. It indicates the studies were clinical studies, implying a prospective nature.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:
The document mentions that the clinical study evaluated performance with respect to reference venous plasma sample YSI measurements.
- Ground Truth Providers: YSI measurements are obtained from laboratory instruments, which are considered the gold standard for glucose measurement. The "experts" involved would be trained lab technicians and medical professionals responsible for collecting and processing blood samples and operating the YSI instrumentation.
- Qualifications: Not explicitly stated, but they would be certified laboratory personnel and healthcare providers responsible for patient care and sample collection within a clinical study setting.
4. Adjudication Method for the Test Set:
Not applicable. The ground truth was established by YSI laboratory measurements, not subjective expert review or adjudication.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
A Human-in-the-loop or MRMC study is not applicable for this device type (Continuous Glucose Monitoring System). The device provides continuous glucose values, and its performance is assessed against a laboratory reference, not human interpretation of images or other data that would involve multiple readers.
6. Standalone Performance:
Yes, the studies described are primarily focused on the standalone performance of the algorithm (as embedded in the device) in measuring glucose levels against a reference method (YSI venous plasma samples). The "clinical study" directly assesses how well the device's sensor and algorithm perform without direct human intervention in the glucose reading process itself, other than the placement and use of the device by participants.
7. Type of Ground Truth Used:
The ground truth used was venous plasma sample YSI measurements. This is a highly accurate and accepted laboratory reference method for glucose concentration.
8. Sample Size for the Training Set:
The document does not specify the sample size for the training set used for the device's glucose algorithm. It mentions that the "ADC Glucose Algorithm established for the predicate device" is used, implying that the core algorithm was developed and trained previously, and this submission focuses on modifications and re-validation.
9. How the Ground Truth for the Training Set Was Established:
The document does not specify how the ground truth for the training set of the ADC Glucose Algorithm was established. However, given that the clinical performance evaluation uses venous plasma YSI measurements as ground truth, it is highly probable that similar high-accuracy reference methods would have been employed during the development and training of the original algorithm.
Ask a specific question about this device
(90 days)
The FreeStyle Libre 2 Flash Glucose Monitoring System is a continuous glucose monitoring (CGM) device with real time alarms capability indicated for the management of diabetes in persons age 4 and older. It is intended to replace blood glucose testing for diabetes treatment decisions, unless otherwise indicated.
The System also detects trends and tracks patterns and aids in the detection of episodes of hyperglycemia and hypoglycemia, facilitating both acute and long-term therapy adjustments. Interpretation of the System readings should be based on the glucose trends and several sequential readings over time.
The System is also intended to autonomously communicate with digitally connected devices. The System can be used alone or in conjunction with these digitally connected devices where the user manually controls actions for therapy decisions.
The FreeStyle Libre 3 Continuous Glucose Monitoring System is a real time continuous glucose monitoring (CGM) device with alarms capability indicated for the management of diabetes age 4 and older. It is intended to replace blood glucose testing for diabetes treatment decisions, unless otherwise indicated.
The System also detects trends and tracks patterns and aids in the detection of episodes of hyperglycemia and hypoglycemia, facilitating both acute and long-term therapy adjustments. Interpretation of the System readings should be based on the glucose trends and several sequential readings over time.
The System is also intended to autonomously communicate with digitally connected devices. The System can be used alone or in conjunction with these digitally connected devices where the user manually controls actions for therapy decisions.
The FreeStyle Libre 2 and FreeStyle Libre 3 are integrated continuous glucose monitoring (iCGM) Systems designed to be used alone or in conjunction with digitally connected devices. The FreeStyle Libre 2 System consists of a Sensor and either a Reader or the FreeStyle Libre 2 App downloaded to a compatible smartphone as a primary display device. The FreeStyle Libre 3 System consists of a Sensor and the FreeStyle Libre 3 App downloaded to a compatible smartphone as a primary display device. Both Systems can communicate glucose data and other information wirelessly and securely to and from these digitally connected devices as described below:
- Wireless communication from the FreeStyle Libre 2 Sensor or FreeStyle Libre 3 Sensor directly to an interoperable receiver device, which connects with the Sensor using the near field communication (NFC) and Bluetooth Low Energy wireless interfaces provided by the Sensor
- The FreeStyle Libre 2 App or FreeStyle Libre 3 App communicates through the cloud to another software device, such as LibreView.
Compared to the respective predicate devices, the proposed subject devices include an additional software component, the Libre Data Sharing API. The Libre Data Sharing API is a cloud-based application programming interface (API) that enables communication of glucose data including alarms through the cloud from the FreeStyle Libre 2 System or FreeStyle Libre 3 System to authorized client software on digitally connected devices. The data transmitted by the API to authorized client software can be used for specific and permitted use cases, including nonmedical device applications, medical device data analysis, CGM secondary display alarm, active patient monitoring, and treatment decisions. Use of the Libre Data Sharing API and the CGM information it transmits is limited by the indications for use of the iCGM systems with which it is used.
The Libre Data Sharing API does not have any command or control over the client software, nor does it allow for the client software to have any command or control over the FreeStyle Libre 2 or FreeStyle Libre 3 Systems. Additionally, glucose data and alarms from the connected iCGM system are not modified or manipulated by the Libre Data Sharing API through its transmission to the authorized client software.
The display device of the connected FreeStyle Libre 2 or FreeStyle Libre 3 Systems, which directly receives the data from the Sensor, continues to serve as a primary display device for the glucose data and alarms. The current components of the FreeStyle Libre 2 and FreeStyle Libre 3 Systems (sensor/applicator and primary display devices) have not been modified as a result of the added the Libre Data Sharing API.
This document is a 510(k) Summary for the FreeStyle Libre 2 Flash Glucose Monitoring System and FreeStyle Libre 3 Continuous Glucose Monitoring System. It describes the devices and argues for their substantial equivalence to previously cleared predicate devices.
Here's an analysis of the acceptance criteria and study information provided in the document:
1. Acceptance Criteria and Reported Device Performance:
The document doesn't present a table of numerical acceptance criteria with corresponding performance statistics. Instead, it argues for substantial equivalence based on the technological characteristics and intended use being the same as the predicate devices, with the addition of the Libre Data Sharing API.
The key acceptance criteria, implicitly, are that the devices continue to perform according to specifications and meet their technological and performance criteria, as demonstrated by verification and validation, aligning with the predicate devices' cleared performance.
| Acceptance Criteria Category (Implicit) | Reported Device Performance |
|---|---|
| Technological Characteristics | Identical to predicate devices (amperometric measurement, glucose oxidase chemical reaction). |
| Intended Use / Indications for Use | Same as predicate devices. The Libre Data Sharing API enables communication of iCGM data for specific, permitted use cases, but does not change the core indications. |
| Primary Display Device Function | Unchanged. Continues to act as the primary display and issue glucose alarms. |
| Safety and Effectiveness | No impact from the Libre Data Sharing API. The API has no command/control over client software or the iCGM systems, and glucose data/alarms are not modified during transmission. |
| Adherence to Special Controls | Conforms to iCGM special controls per 21 CFR 862.1355. |
2. Sample Size and Data Provenance:
The document states: "The proposed subject devices with the Libre Data Sharing API were verified and validated according to ADC's internal design control process and in accordance with the applicable special controls for integrated continuous glucose monitoring systems. The testing demonstrated that the subject devices conform to the iCGM special controls per 21 CFR 862.1355 and that they performed according to specifications and met their technological and performance criteria."
- Sample Size: The document does not specify the sample sizes used for the test set or the training set. It refers broadly to "performance testing" and "data provided in this pre-market notification."
- Data Provenance: The document does not explicitly state the country of origin or whether the studies were retrospective or prospective. Given it's a 510(k) submission for a medical device, it's highly likely that the underlying performance data comes from clinical studies, often multi-site and prospective, but this document does not detail them. The focus of this 510(k) is specifically on the added software component (Libre Data Sharing API) and its impact on substantial equivalence, rather than re-proving the core glucose monitoring accuracy which would have been established in previous 510(k) clearances.
3. Number of Experts and Qualifications for Ground Truth:
The document concerns the addition of a data sharing API to existing glucose monitoring systems. It does not mention the use of experts to establish ground truth in the context of a reader study for the new component. The primary ground truth for glucose monitoring devices is typically derived from lab reference methods (e.g., YSI analyzer for blood glucose), rather than expert adjudication of images or clinical assessments.
4. Adjudication Method for the Test Set:
Not applicable in the context of this submission. The submission is about an API for device data sharing, not a diagnostic imaging or clinical assessment device requiring expert adjudication.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
No, an MRMC study was not conducted or mentioned. This type of study is relevant for AI-assisted diagnostic tools that impact human reader performance, typically in radiology or pathology. This submission is for a continuous glucose monitoring system and an API to share its data.
6. Standalone (Algorithm Only) Performance:
The document states: "The Libre Data Sharing API does not have any command or control over the client software, nor does it allow for the client software to have any command or control over the FreeStyle Libre 2 and FreeStyle Libre 3 Systems. Additionally, glucose data and alarms from the FreeStyle Libre 2 and FreeStyle Libre 3 Systems are not modified or manipulated by the Libre Data Sharing API through its secured transmission to the authorized client software."
This implies that the core algorithm performance (i.e., glucose measurement accuracy) of the FreeStyle Libre 2 and 3 systems, which operates independently of the Libre Data Sharing API, has already been established in previous clearances (K210943, K213996, K212132). This submission focuses on the safety and effectiveness of the data sharing mechanism, not the standalone performance of the glucose measurement algorithms themselves. The API itself is primarily a data transmission layer, not a
diagnostic algorithm.
7. Type of Ground Truth Used:
For the underlying glucose monitoring systems (FreeStyle Libre 2 and 3), the ground truth for performance studies is typically laboratory reference measurements (e.g., plasma glucose values obtained from a YSI glucose analyzer) against which the accuracy of the interstitial glucose readings are compared. The document does not specifically detail how the ground truth was established for the "performance testing" related to the API, as the API's function is data transmission, not direct measurement. The ground truth for the API's performance would likely focus on the integrity, security, and timeliness of data transfer.
8. Sample Size for the Training Set:
The document does not specify the sample size for any training set. This submission is for a device modification (adding an API), not a de novo AI/ML algorithm that typically undergoes distinct training and test phases. The core glucose algorithms were already developed and cleared.
9. How the Ground Truth for the Training Set was Established:
Not applicable, as this submission pertains to an API for data sharing and refers to "verification and validation" for "performance testing" of the modified device, rather than the development and training of a new algorithm requiring a specific ground truth establishment for a training set. The underlying glucose measurement algorithms would have had their own ground truth established during their prior development and clearance, likely using reference lab methods.
Ask a specific question about this device
(157 days)
The FreeStyle Libre 3 Continuous Glucose Monitoring System is a real time continuous glucose monitoring (CGM) device with alarms capability indicated for the management of diabetes in persons age 4 and older. It is intended to replace blood glucose testing for diabetes treatment decisions, unless otherwise indicated.
The System also detects trends and tracks patterns and aids in the detection of episodes of hyperglycemia and hypoglycemia, facilitating both acute and long-term therapy adjustments. Interpretation of the System readings should be based on the glucose trends and several sequential readings over time.
The System is also intended to autonomously communicate with digitally connected devices. The System can be used alone or in conjunction with these digitally connected devices where the user manually controls actions for therapy decisions.
The FreeStyle Libre 3 Continuous Glucose Monitoring System (herein referred to as the 'FreeStyle Libre 3 System' or 'System') is an integrated continuous glucose monitoring system (iCGM) that provides real time continuous glucose measurements every minute to provide glucose levels, trends, and alarms. The System requires a prescription and is intended for home use. The System consists of the following components: a Sensor which transmits via Bluetooth Low Energy (BLE), and a mobile application, FreeStyle Libre 3 App, downloaded to a compatible smartphone running on Android operating system. The FreeStyle Libre 3 System provides the user with real-time glucose measurements (glucose values) accompanied by trend information (glucose arrows) and historical glucose information (glucose graph). The user may make treatment decisions based in part on the Sensor glucose results provided by the System. The System also provides fixed and configurable alarms designed to warn the user of Low Glucose, High Glucose, or Signal Loss.
The document details the FreeStyle Libre 3 Continuous Glucose Monitoring System and its performance testing to support substantial equivalence to its predicate device, the FreeStyle Libre 2 Flash Glucose Monitoring System.
Here's an analysis of the acceptance criteria and study information provided:
1. Table of Acceptance Criteria and Reported Device Performance
The document states that a bridging clinical study was conducted, and the "combined System accuracy of the FreeStyle Libre 3 System and FreeStyle Libre 2 System met the iCGM special controls requirements per 21 CFR 862.1355." However, specific numerical acceptance criteria (e.g., MARD percentage, accuracy at different glucose ranges) and the exact reported performance of the FreeStyle Libre 3 System are not explicitly listed in a detailed table within the provided text segments. Instead, a general statement of meeting the iCGM special controls is given.
To fill this, we must infer that the implicit acceptance criteria are those defined by the iCGM special controls (21 CFR 862.1355) for integrated continuous glucose monitoring systems. These typically include accuracy metrics like MARD (Mean Absolute Relative Difference) against a reference method, as well as performance across different glucose ranges (hypoglycemic, euglycemic, hyperglycemic). Without the actual study report, we cannot provide specific numbers for the FreeStyle Libre 3's performance or exact acceptance criteria.
General Interpretation of iCGM Special Controls (Implicit Acceptance Criteria):
iCGM special controls generally require demonstration of:
- Accuracy: Sufficient accuracy across the glucose range (e.g., MARD values below a certain threshold when compared to a laboratory reference).
- Safety: Acceptable safety profile with minimal adverse events.
- Performance in various conditions: Performance evaluated across different patient populations, wear sites, and duration of use.
- Alarm performance: Accuracy and reliability of hypoglycemia and hyperglycemia alarms.
- Interoperability: Functionality with digitally connected devices.
2. Sample Size Used for the Test Set and Data Provenance
The document states: "ADC conducted a bridging clinical study to demonstrate comparability of the performance of the FreeStyle Libre 3 System to the predicate FreeStyle Libre 2 System, cleared under K210943."
- Sample Size for Test Set: Not explicitly stated in the provided text.
- Data Provenance: Not explicitly stated in the provided text (e.g., country of origin, retrospective or prospective). It is implied to be prospective as it's a clinical study for FDA clearance.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This information is not provided in the text. For CGM devices, the "ground truth" (reference glucose values) is typically established using a highly accurate laboratory reference method (e.g., YSI glucose analyzer) on blood samples, not expert consensus interpreting images or clinical cases. Therefore, the concept of "experts establishing ground truth" in the way it applies to image interpretation is not directly applicable here. The ground truth would be the laboratory-measured glucose values.
4. Adjudication Method for the Test Set
The concept of an "adjudication method" (like 2+1 or 3+1) is primarily relevant for studies where multiple human readers interpret data (e.g., radiology images) and a consensus or tie-breaking mechanism is needed for the ground truth. For a CGM device, the ground truth is established by a quantitative, objective measurement (laboratory blood glucose). Therefore, an adjudication method in this context is not applicable/not mentioned as it's not a reader study.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done
No, an MRMC comparative effectiveness study was not done. The FreeStyle Libre 3 is a standalone device that measures glucose; it does not involve human readers interpreting its output in a diagnostic context that would warrant an MRMC study. The "comparative effectiveness" mentioned refers to its performance against a predicate device and standard laboratory methods, not a comparison of human reader performance with and without AI assistance.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done
Yes, a standalone study was done. The entire premise of the submission is to demonstrate the performance of the FreeStyle Libre 3 System as a standalone integrated continuous glucose monitoring device. The bridging clinical study, designed to show "comparability of the performance of the FreeStyle Libre 3 System to the predicate FreeStyle Libre 2 System," inherently evaluates the algorithm's performance without direct human intervention in the glucose measurement and calculation process. The FreeStyle Libre 3 delivers "real-time continuous glucose measurements every minute" and its "System accuracy" was evaluated.
7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)
The ground truth for a CGM device is typically a highly accurate laboratory reference method (e.g., YSI glucose analyzer) performed on venous blood samples taken concurrently with the CGM readings. While not explicitly stated, this is the standard for CGM accuracy studies.
8. The Sample Size for the Training Set
The document mentions that the "Sensor Glucose Algorithm established for the predicate device" is the same for the FreeStyle Libre 3. It also states that the FreeStyle Libre 3 contains a "software algorithm for conversion of the raw glucose measurements from the Sensor to calculate glucose results." This implies the algorithm was developed and trained prior to this submission, likely using data from previous FreeStyle Libre sensor generations.
The sample size for the training set is not provided in this document.
9. How the Ground Truth for the Training Set was Established
Given that the "Sensor Glucose Algorithm" is the same as the predicate device, it can be inferred that the ground truth for training this algorithm was established similarly to the predicate device's clearance. This would typically involve pairing sensor readings with laboratory reference blood glucose measurements (e.g., YSI) taken from human subjects during previous clinical studies. The specific details of how this ground truth was established for the training set are not provided in the given text.
Ask a specific question about this device
(322 days)
The FreeStyle Libre 3 Continuous Glucose Monitoring System is a real time continuous glucose monitoring (CGM) device with alarms capability indicated for the management of diabetes in persons age 4 and older. It is intended to replace blood glucose testing for diabetes treatment decisions, unless otherwise indicated.
The System also detects trends and tracks patterns and aids in the detection of episodes of hyperglycemia and hypoglycemia, facilitating both acute and long-term therapy adjustments. Interpretation of the System readings should be based on the glucose trends and several sequential readings over time.
The System is also intended to autonomously communicate with digitally connected devices. The System can be used alone or in conjunction with these digitally connected devices where the user manually controls actions for therapy decisions.
The FreeStyle Libre 3 Continuous Glucose Monitoring System (herein referred to as the 'FreeStyle Libre 3 System' or 'System') is an integrated continuous glucose monitoring system (iCGM) that provides real time continuous glucose measurements every minute to provide glucose levels, trends, and alarms. The System requires a prescription and is intended for home use. The System consists of the following components: a Sensor which transmits via Bluetooth Low Energy (BLE), and a mobile application, FreeStyle Libre 3 App, downloaded to a compatible smartphone running iOS operating system. The FreeStyle Libre 3 System provides the user with real-time glucose measurements (glucose values) accompanied by trend information (glucose arrows) and historical glucose information (glucose graph). The user may make treatment decisions based in part on the Sensor glucose results provided by the System. The System also provides fixed and configurable alarms designed to warn the user of Low Glucose, High Glucose, or Signal Loss.
Here's a breakdown of the acceptance criteria and study information for the FreeStyle Libre 3 Continuous Glucose Monitoring System, based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
The document doesn't provide a specific table of numerical acceptance criteria for clinical performance (like MARD values with defined thresholds). Instead, it states that the device "met the iCGM special controls requirements per 21 CFR 862.1355." However, it does list several performance characteristics that were "confirmed to support substantial equivalence," implying that specific acceptance criteria were met for each. Without the actual criteria for each, here's a general table based on the information provided:
| Performance Characteristic | Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|---|
| Sterility | Sterility Assurance Level (SAL) of 10^-6 | Achieved SAL of 10^-6 with minimum sterilization dose of 25 kGy (established by VDmax25 method described in ISO 11137-2). |
| Shelf-Life & Packaging | Seal integrity, user accessibility, and device functionality met | Attributes related to seal integrity, user accessibility, & device functionality (including sterile barrier system integrity) met acceptance criteria after subjecting units to worst-case sealing, sterilization, shipping, storage, handling, and transit challenges. |
| Electrical Safety | Compliance to IEC 60601-1, IEC 60601-1-6, IEC 60601-1-11 | Demonstrated compliance to IEC 60601-1: 2005(r)2012, IEC 60601-1-6:2010+A1:2013, and IEC 60601-1-11:2015. |
| Electromagnetic | Withstand EMI and emissions (IEC 60601-1-2, CISPR 11) | Verified system withstands EMI and emissions in compliance with IEC 60601-1-2 and IEC CISPR 11. Wireless coexistence testing confirmed functionality within acceptable limits in presence of common radiating electronic devices (FDA Guidance, AAMI TIR69, ANSI C63.27). Complied with FCC Part 15.225, Part 15.247, and FAA Advisory Circular RTCA DO-160. |
| Mechanical Engineering | Mechanical, electrical, and functional testing met | Test results showed mechanical, electrical, and functional testing all met acceptance criteria at system and component level. |
| Biocompatibility | Compliance with ISO 10993-1 and FDA Guidance | Evaluation and testing performed in accordance with ISO10993-1 and FDA Guidance. |
| Software V&V | Met established specifications and IEC 62304 | Results of executed protocols met acceptance criteria, supporting the System's embedded software is acceptable for intended use (in accordance with IEC 62304 and FDA Guidance). |
| Cybersecurity | Analysis of confidentiality, integrity, availability, risk mgmt. | Provided cybersecurity risk management documentation (analysis of confidentiality, integrity, availability for data, info, & software per FDA Draft Guidance). Risk assessment performed for identified threats/vulnerabilities, and appropriate risk mitigation controls implemented/tested. |
| Clinical performance | Met iCGM special controls requirements per 21 CFR 862.1355 | A bridging clinical study demonstrated comparability of performance to the predicate FreeStyle Libre 2 System, and the combined System accuracy of the FreeStyle Libre 3 and FreeStyle Libre 2 System met the iCGM special controls. |
| Human Factors | Met usability requirements for intended use (ANSI/AAMI/IEC 62366) | Risk analysis of design and user interface (in accordance with ANSI/AAMI/IEC 62366, IEC 60601-1-6, and FDA Guidance) demonstrated design changes met usability requirements. |
| Interoperability | Alignment with FDA Guidance "Design Considerations..." | Incorporated an approach for interoperability developed in alignment with FDA Guidance. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: The document mentions "a bridging clinical study" but does not explicitly state the number of subjects or data points used in this study.
- Data Provenance: Not explicitly stated (e.g., country of origin, retrospective or prospective). The phrasing "bridging clinical study" generally implies a prospective study designed to bridge results from a previous study or device.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This information is not provided in the document. For glucose monitoring systems, ground truth is typically established using a highly accurate reference method (e.g., YSI glucose analyzer) rather than expert consensus on images.
4. Adjudication Method for the Test Set
This information is not provided in the document.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No MRMC study was mentioned. The FreeStyle Libre 3 is a Continuous Glucose Monitoring (CGM) system, which provides direct glucose readings to the user, not an AI system that assists human "readers" in interpreting diagnostic images or data where MRMC studies are typically performed. The device is intended to replace blood glucose testing for treatment decisions, not to aid human interpretation of complex medical cases.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
- Yes, a standalone performance assessment was done. The document states, "The subject device calculates glucose information identically to the predicate device, and the combined System accuracy of the FreeStyle Libre 3 System and FreeStyle Libre 2 System met the iCGM special controls requirements per 21 CFR 862.1355." This refers to the accuracy of the algorithm itself in determining glucose values from the sensor data. The "clinical performance" testing directly assesses the device's accuracy without human intervention influencing the glucose measurement.
7. The Type of Ground Truth Used
- Reference Glucose Methods: While not explicitly named, for CGM devices, the standard ground truth for clinical accuracy studies is typically venous blood samples analyzed by a laboratory reference method, such as a YSI glucose analyzer. The document states accuracy was demonstrated, which implies comparison to such a reference.
8. The Sample Size for the Training Set
- Not provided. The document describes a "bridging clinical study" for performance validation but does not detail the size of any training sets used for algorithm development. It does state that the "Sensor Glucose Algorithm established for the predicate device" is the same, implying the algorithm's core might have been trained previously for the FreeStyle Libre 2.
9. How the Ground Truth for the Training Set was Established
- Not provided. Given that the subject device uses the "ADC Glucose Algorithm established for the predicate device," the ground truth establishment for any original training would have likely involved comparisons to laboratory reference glucose measurements in clinical studies. However, details of such a training process or ground truth for a training set are not in this document.
Ask a specific question about this device
(237 days)
The FreeStyle Libre 2 Flash Glucose Monitoring System is a continuous glucose monitoring (CGM) device with real time alarms capability indicated for the management of diabetes in persons age 4 and older. It is intended to replace blood glucose testing for diabetes treatment decisions, unless otherwise indicated.
The System also detects trends and tracks patterns and aids in the detection of episodes of hyperglycemia and hypoglycemia, facilitating both acute and long-term therapy adjustments. Interpretation of the System readings should be based on the glucose trends and several sequential readings over time.
The System is also intended to autonomously communicate with digitally connected devices. The System can be used alone or in conjunction with these digitally connected devices where the user manually controls actions for therapy decisions.
The System can be used with the FreeStyle Libre 2 Sensor (14 day) or the FreeStyle Libre 2 MediRx Sensor (10 day).
The FreeStyle Libre 2 Flash Glucose Monitoring System with the FreeStyle Libre 2 App -Android (herein referred to as the 'FreeStyle Libre 2 System' or 'System') is an integrated continuous glucose monitoring system (iCGM) that provides continuous glucose measurements every minute to provide glucose levels, trends, and real time alarms capability to aid in the management of diabetes. The System requires a prescription and is intended for home use. The System consists of the following components: a Sensor which transmits yia Bluetooth Low Energy (BLE), a BLE enabled display device (Reader), and an Android or iOS mobile app (FreeStyle Libre 2 App) downloaded to a compatible smartphone. Scanning of the Sensor via Reader or App provides the user with real-time glucose measurements (glucose values) accompanied by trend information (glucose arrows) and historical glucose information (glucose graph). The user may make treatment decisions based in part on the Sensor glucose results provided by the System. The System also provides configurable alarms designed to warn the user of Low Glucose, High Glucose or Signal Loss.
This 510(k) summary describes the FreeStyle Libre 2 Flash Glucose Monitoring System (with FreeStyle Libre 2 App), focusing on the addition of an Android application to the existing system. The document states that the Android app utilizes the identical algorithm and implements the same wireless interfaces with the sensor as the previously cleared Reader and iOS app. Therefore, no additional clinical data was used in this 510(k) to support a determination of substantial equivalence for clinical performance. The original clinical performance data was established in predicate devices K193371 and K211102.
Since this 510(k) relies on the clinical performance data established in prior submissions (K193371 and K211102) and explicitly states that no new clinical performance data was generated for this specific submission (K210943) regarding the Android App, the detailed acceptance criteria and study information for clinical accuracy are not directly present within this provided text for K210943. The document focuses on non-clinical performance testing for the Android app.
However, based on the information provided regarding the non-clinical performance of the Android App within this specific 510(k) (K210943):
Table of Acceptance Criteria and Reported Device Performance (Non-Clinical)
| Acceptance Criteria Category | Reported Device Performance |
|---|---|
| Software Verification and Validation | Results of executed protocols met the acceptance criteria, supporting that the Android App software is acceptable for its intended use. (Conducted in accordance with IEC 62304 and FDA Guidance "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices," May 11, 2005.) |
| Human Factors | Results demonstrated that the System met usability requirements. (Conducted in accordance with ANSI/AAMI/IEC 62366, IEC 60601-1-6, and FDA Guidance "Applying Human Factors and Usability Engineering to Medical Devices," dated February 3, 2016.) |
| Wireless Coexistence | Test results showed the System could tolerate interference generated by tested RF interfering devices and still meet the target performance criteria. (Consistent with AAMI TIR69 and ANSI C63.27, including in-band interference sources and other expected wireless interference sources from the intended use environment.) |
| Cybersecurity | Appropriate risk mitigation controls have been implemented and tested. (Risk management documentation includes analysis of confidentiality, integrity, and availability for data, information, and software, with risk assessment and mitigation for identified threats and vulnerabilities.) |
| Electrical Safety and Electromagnetic Compatibility (EMC) | Results of the evaluation demonstrated that the System complies with electrical safety and EMC requirements. (Per IEC 60601-1:2005(r)2012 and IEC 60601-1-2:2014, respectively.) |
Due to the nature of this 510(k) being primarily for the addition of the Android app, the following information regarding clinical performance criteria, sample sizes, expert involvement, and ground truth would have been established in the predicate device studies (K193371 and K211102) and are not detailed in this document (K210943):
- Sample size used for the test set and the data provenance: Not specified in K210943, as clinical data was referenced from predicate devices. For iCGM, clinical studies typically involve a significant number of participants (e.g., hundreds) across varied demographics to assess accuracy across the glycemic range. Data provenance would likely be from multi-center prospective clinical trials conducted in various regions.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not specified in K210943. In iCGM clinical trials, ground truth is usually established by highly precise laboratory reference methods (e.g., YSI glucose analyzer) rather than expert consensus on device readings.
- Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable for iCGM clinical accuracy studies where the "ground truth" is typically a laboratory reference method.
- If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable for iCGM clinical accuracy studies as described. The device provides direct glucose values, not interpretations requiring human reader assistance in the same way as imaging diagnostics. It replaces blood glucose testing for treatment decisions.
- If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Yes, the core accuracy of the algorithm is evaluated in a standalone manner against laboratory reference values in the predicate device studies. The FreeStyle Libre 2 System itself, including the sensor and its algorithm, performs autonomously to produce glucose readings.
- The type of ground truth used (expert consensus, pathology, outcomes data, etc): For iCGM devices, the gold standard ground truth for accuracy studies is typically laboratory reference methods, such as those performed by a YSI glucose analyzer, which measures glucose concentration in blood plasma.
- The sample size for the training set: Not specified in K210943. Training data for the glucose algorithm would have been part of the original development and validation of the FreeStyle Libre 2 system (K193371 and K211102), and would involve a large, diverse dataset of sensor readings and corresponding reference glucose values.
- How the ground truth for the training set was established: Not specified in K210943. Similar to the test set, the ground truth for training would have been established using highly accurate laboratory reference methods (e.g., YSI analyzer) from blood samples collected concurrently with sensor readings.
Ask a specific question about this device
(120 days)
The FreeStyle Libre 2 Flash Glucose Monitoring System is a continuous glucose monitoring (CGM) device with real time alarms capability indicated for the management of diabetes in persons age 4 and older. It is intended to replace blood glucose testing for diabetes treatment decisions, unless otherwise indicated.
The System also detects trends and tracks patterns and aids in the detection of episodes of hyperglycemia and hypoglycemia, facilitating both acute and long-term therapy adjustments. Interpretation of the System readings should be based on the glucose trends and several sequential readings over time.
The System is also intended to autonomously communicate with digitally connected devices. The System can be used alone or in conjunction with these digitally connected devices where the user manually controls actions for therapy decisions.
The System can be used with the FreeStyle Libre 2 Sensor (14 day) or the FreeStyle Libre 2 MediRx Sensor (10 day).
The FreeStyle Libre 2 Flash Glucose Monitoring System is an integrated continuous glucose monitoring (iCGM) system that provides continuous glucose measurements every minute to provide glucose levels, trends, and real-time alarms capability to aid in the management of diabetes. The FreeStyle Libre 2 System consists of two primary components: a Sensor that transmits via Bluetooth Low Energy (BLE), and a BLE enabled display device (Reader). User initiated RFID scanning of the Sensor via Reader provides the user with real-time glucose measurements (glucose values) accompanied by trend information (glucose arrows) and historical glucose information (glucose graph). Users may use the Sensor glucose results and information provided by the System in making treatment decisions. The System also provides configurable alarms designed to warn the user of Low Glucose, High Glucose or Signal Loss. The system is intended for single-patient use at home and requires a prescription.
The Sensor is single use, disposable, and powered by a silver oxide battery. The Sensor is provided as two secondary components, Sensor Applicator and Sensor Pack (electron beam sterilized device) which are used to assemble and apply the Sensor to the back of the user's arm. During Sensor application, the sensor tail is inserted about 5.5 millimeters below the surface of the skin through the guidance of a needle. The needle is retracted back into the applicator after insertion, and the Sensor remains attached to the skin with a medical grade adhesive. The Sensor continuously measures glucose concentration in interstitial fluid and has an 8-hour memory capacity. The Sensor is factory calibrated, does not require fingerstick calibration, and can be worn for up to 14 days.
The Reader is a small handheld device that is powered by a lithium-ion rechargeable battery and uses RFID communication to start new Sensors and to scan Sensors to display and record data and uses BLE communication to issue alarms that notify the user to scan his/her sensor when glucose values pass a high or low glucose threshold. The Reader also has a built-in strip port with blood glucose functionality (that is intended to work with the FreeStyle Precision Neo Blood Glucose test strips, cleared under K171941), and a user interface that includes event logging features.
The proposed subject device is a modified FreeStyle Libre 2 Flash Glucose Monitoring System that adds compatibility with the FreeStyle Libre 2 MediRx Sensor, which can be worn for up to 10 days. The FreeStyle Libre 2 MediRx Sensor design is unchanged from that of the predicate FreeStyle Libre 2 Sensor, which remains compatible with the modified System. In addition, the Sensor glucose algorithm and Reader design of the modified System remain unchanged from those of the predicate.
The alternate 10-day wear duration of the FreeStyle Libre 2 MediRx Sensor is achieved by changing a Sensor configuration parameter at manufacturing, which is detected by the predicate Reader to automatically determine the wear duration and accordingly adjust the user interface display of remaining Sensor wear time and ensure the Sensor cannot report data beyond the configured wear duration. In addition, each Sensor type has an end of life parameter, which determines when the Sensor will automatically shut down. This functionality is already built into the Sensor and Reader and was validated as part of previously conducted software validation under K193371.
Other than the differences related to wear duration, the FreeStyle Libre 2 MediRx Sensor is identical to the predicate Sensor, and the predicate Reader functions as intended with either the predicate FreeStyle Libre 2 Sensor (14 day) or FreeStyle Libre 2 MediRx Sensor (10 day).
The provided text is related to a 510(k) premarket notification for the FreeStyle Libre 2 Flash Glucose Monitoring System, specifically for a modification to include compatibility with a 10-day wear sensor (FreeStyle Libre 2 MediRx Sensor).
The document does not describe an AI/ML-based device where the performance is presented in terms of AI metrics (e.g., accuracy, precision, recall, AUC, etc.) or a multi-reader multi-case (MRMC) study. Instead, it concerns a medical device for continuous glucose monitoring (CGM). The "algorithm" mentioned (Sensor Glucose Algorithm) refers to the internal processing of sensor signals to derive glucose values, not an AI/ML model in the context of clinical decision support or image analysis.
Therefore, many of the requested criteria (e.g., sample size for test/training set in AI context, number of experts for ground truth, adjudication method, MRMC study, standalone performance for AI, type of ground truth for AI) are not applicable to this device's submission.
However, I can extract information related to the device's clinical performance evaluation based on the provided text, which supports its substantial equivalence.
Here's a summary of the relevant information:
1. Acceptance Criteria and Device Performance:
The document states: "Clinical data from the adult and pediatric iCGM clinical studies that supported clearance of the predicate device were re-analyzed to show that use of the subject device with the FreeStyle Libre 2 MediRx Sensor for a 10-day wear duration meets the iCGM special controls for clinical performance set forth in 21 CFR 862.1355."
This indicates that the acceptance criteria are based on the iCGM special controls outlined in 21 CFR 862.1355. The document does not provide a specific table of numerical acceptance criteria or reported device performance metrics (e.g., MARD values, clinical accuracy zones) for the FreeStyle Libre 2 MediRx Sensor in this specific 510(k) submission. It relies on the re-analysis of data from the predicate device's clearance.
| Acceptance Criteria (General) | Reported Device Performance (as stated) |
|---|---|
| Meets iCGM special controls for clinical performance set forth in 21 CFR 862.1355 for 10-day wear duration. | "Clinical data from the adult and pediatric iCGM clinical studies that supported clearance of the predicate device were re-analyzed to show that use of the subject device with the FreeStyle Libre 2 MediRx Sensor for a 10-day wear duration meets the iCGM special controls..." |
2. Sample Size and Data Provenance for the Test Set:
- Sample Size Used for Test Set: Not explicitly stated for this particular 510(k). The clinical performance is based on "re-analyzed" data from "clinical studies that supported clearance of the predicate device (K193371)." To find specific sample sizes, one would need to refer to the original K193371 submission.
- Data Provenance: Not explicitly stated regarding country of origin. The data is based on "clinical studies that supported clearance of the predicate device." It's retrospective in the sense that existing data was re-analyzed for the new sensor configuration.
3. Number of Experts and Qualifications for Ground Truth:
- This question is not applicable in the context of this device. The "ground truth" for a glucose monitoring system is typically a high-accuracy reference method for blood glucose (e.g., YSI analyzer in a controlled lab setting), not expert consensus from radiologists or similar.
4. Adjudication Method for the Test Set:
- Not applicable for this type of device. Adjudication methods like '2+1' or '3+1' are common in image analysis studies where human readers provide interpretations and discrepancies are resolved. This is a continuous glucose monitoring device where performance is measured against reference glucose values.
5. MRMC Comparative Effectiveness Study:
- No, an MRMC comparative effectiveness study was not done as described for AI assistance. This device is a direct-to-patient glucose monitoring system, not an AI-assisted diagnostic tool that human readers would use to improve their performance.
6. Standalone Performance:
- Yes, in essence, standalone performance was done. The device itself provides glucose readings without human interpretation or intervention in the measurement process. The "re-analysis" of clinical data to meet iCGM special controls essentially evaluates this standalone performance of the FreeStyle Libre 2 MediRx Sensor. The output of the device (glucose value) is directly compared against a reference method.
7. Type of Ground Truth Used:
- The ground truth used for glucose monitoring devices is typically central laboratory reference glucose measurements (e.g., from a YSI glucose analyzer) taken from blood samples during clinical studies. The document states a re-analysis of "clinical studies," implying the use of such a reference method from the predicate device's trials.
8. Sample Size for the Training Set:
- Not explicitly stated/applicable in the context of AI/ML training. The "Sensor Glucose Algorithm" mentioned is likely a deterministic or model-based algorithm, not a trainable deep learning model in the sense of a "training set" for AI. If the algorithm involves parameters that were "trained" or optimized, the document does not specify the sample size used for this internal process. The primary evaluation here is of the modified sensor with an unchanged algorithm.
9. How the Ground Truth for the Training Set Was Established:
- Not explicitly stated/applicable in the context of AI/ML training. As above, the ground truth for any underlying algorithm development would refer to the reference glucose measurements used to build or validate that algorithm, but this is not typically referred to as a "training set" in the AI sense for this type of device. The document explicitly states: "the Sensor glucose algorithm... of the modified System remain unchanged from those of the predicate."
Ask a specific question about this device
(396 days)
The FreeStyle Libre 2 Flash Glucose Monitoring System is a continuous glucose monitoring (CGM) device with real time alarms capability indicated for the management of diabetes in persons age 4 and older. It is intended to replace blood glucose testing for diabetes treatment decisions, unless otherwise indicated.
The System also detects trends and tracks patterns and aids in the detection of episodes of hyperglycemia and hypoglycemia, facilitating both acute and long-term therapy adjustments. Interpretation of the System readings should be based on the glucose trends and several sequential readings over time.
The System is also intended to autonomously communicate with digitally connected devices. The System can be used alone or in conjunction with these digitally connected devices where the user manually controls actions for therapy decisions.
The FreeStyle Libre 2 Flash Glucose Monitoring System with the FreeStyle Libre 2 App (herein referred to as the 'FreeStyle Libre 2 System' or 'System') is an integrated continuous glucose monitoring system (iCGM) that provides continuous glucose measurements every minute to provide glucose levels, trends and alerts. The System requires a prescription and is intended for home use. The System consists of the following components: a sensor which transmits via Bluetooth Low Energy (BLE), a BLE enabled display device (Reader), and an iOS mobile app (FreeStyle Libre 2 App) downloaded to a compatible smartphone. Scanning of the sensor via Reader or App provides the user with real-time glucose measurements (glucose values) accompanied by trend information (glucose arrows) and historical glucose information (glucose graph). The user may make treatment decisions based in part on the sensor glucose results provided by the System. The System also provides configurable alarms designed to warn the user of Low Glucose, High Glucose or Signal Loss.
The provided text describes the FreeStyle Libre 2 Flash Glucose Monitoring System (with FreeStyle Libre 2 App) and its substantial equivalence to a predicate device. However, the document primarily focuses on explaining the device, its intended use, comparison with a predicate, and the general types of performance testing conducted to support substantial equivalence for the app component.
Crucially, the acceptance criteria and detailed study results proving the device meets those criteria (specific performance metrics, sample sizes for clinical trials, ground truth establishment, expert adjudication, MRMC studies, etc.) are NOT explicitly present in this excerpt. The document states "no additional clinical data beyond that provided in K193371 was used in this 510(k) to support a determination of substantial equivalence" for clinical performance, indicating reliance on prior submissions for the predicate device. Therefore, a comprehensive answer for the requested points regarding the study that proves the device meets the acceptance criteria cannot be fully extracted from this document alone.
However, based on the information provided, I can infer and answer parts of your request.
Here's an attempt to answer your questions based on the provided text, and where information is missing, I will clearly state that:
Device: FreeStyle Libre 2 Flash Glucose Monitoring System (with FreeStyle Libre 2 App)
Predicate Device: FreeStyle Libre 2 Flash Glucose Monitoring System (K193371)
A Table of Acceptance Criteria and Reported Device Performance:
The document does not provide a specific table of acceptance criteria with corresponding performance metrics. It generally states that "Results of executed protocols met the acceptance criteria" for various tests, but does not quantify these criteria or the results.
| Acceptance Criteria (Inferred from general statements) | Reported Device Performance (General Statement in document) |
|---|---|
| Software Verification and Validation: Meets established specifications and IEC 62304 guidance. | "Results of executed protocols for the met the acceptance criteria and therefore supports that the System's embedded software is acceptable for its intended use." |
| Human Factors: Meets usability requirements. | "Results demonstrated that the System met usability requirements." |
| Wireless Coexistence: Tolerates interference from various RF interfering devices and meets target performance criteria. | "Test results showed the System could tolerate interference generated by these RF interfering devices and still meet the target performance criteria." |
| Cybersecurity: Risk mitigation controls are implemented and tested. | "Appropriate risk mitigation controls have been implemented and tested." |
| Electrical Safety and EMC: Complies with IEC 60601-1:2005(r)2012 and IEC 60601-1-2:2014. | "demonstrated that the System complies with electrical safety and EMC requirements." |
| Clinical Performance (Accuracy) | "The App utilizes the identical algorithm and implements the same wireless interfaces with the Sensor as used by the Reader. As the App calculates glucose information identically to the Reader, no additional clinical data beyond that provided in K193371 was used in this 510(k) to support a determination of substantial equivalence." (This implies that the clinical performance established for the predicate device is considered applicable and meets criteria, but specific numbers are not here). |
1. Sample sized used for the test set and the data provenance:
- Sample Size for Test Set: Not specified in this document. The document states that no additional clinical data was used for this 510(k) beyond what was presented in K193371 (the predicate device's submission). The performance testing described (Software V&V, Human Factors, Wireless Coexistence, Cybersecurity, Electrical Safety/EMC) are generally system-level or component-level tests, not direct patient clinical trials with specific "test sets" in the sense of clinical data.
- Data Provenance: Not specified for the clinical data that supports the predicate device (K193371). For the testing described in this document (Software, Human Factors, etc.), it's implied to be internal testing by Abbott Diabetes Care Inc. and conducted in laboratory environments. The document does not specify if it's retrospective or prospective.
2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- The document does not describe specific "ground truth" establishment for a clinical test set in this submission, as it defers to a previous 510(k) (K193371) for clinical performance data. Therefore, the number and qualifications of experts for establishing clinical ground truth are not provided here.
- For Human Factors testing, "usability requirements" were met, implying expert evaluation or user testing, but the specifics of who established the "ground truth" or expert qualifications are not detailed.
3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not applicable/Not specified. Since clinical data specifics are not provided in this submission (deferring to K193371), there's no mention of an adjudication method commonly used in clinical trials for ground truth.
4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This device is a Continuous Glucose Monitoring (CGM) system, not an AI-assisted diagnostic imaging device that involves human "readers" interpreting cases. Therefore, an MRMC study is not relevant to this device type.
5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- The document implies that standalone performance of the algorithm was assessed for the core glucose measurement. It states, "The App utilizes the identical algorithm and implements the same wireless interfaces with the Sensor as used by the Reader. As the App calculates glucose information identically to the Reader..." This suggests that the algorithm's performance in converting raw sensor measurements to glucose values was established and relied upon from the predicate device's data (K193371). No specific standalone performance metrics are provided in this document, but the fact that the algorithm is "identical" means its standalone performance is assumed to be the same as the predicate.
6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- For the clinical performance of a CGM, the ground truth typically involves lab-based reference glucose measurements (e.g., YSI analyzer or similar highly accurate methods) from blood samples taken concurrently with sensor readings. However, none of these specifics for ground truth are provided in this document, as it refers back to K193371 for clinical data.
7. The sample size for the training set:
- Not specified. The document does not discuss the training of the glucose algorithm. It merely confirms that the app uses the "identical algorithm" as the predicate device.
8. How the ground truth for the training set was established:
- Not specified. As the document does not detail the training set, it also does not elaborate on how its ground truth was established.
In summary, this 510(k) summary focuses on demonstrating substantial equivalence for the addition of an app to an already cleared CGM system. It relies heavily on the clinical performance and established ground truth from the predicate device's prior submission (K193371) for the core glucose measurement accuracy. The testing described in this particular document pertains mostly to software validation, human factors, wireless communication, cybersecurity, and electrical safety/EMC, which are specific to the new app component and its interaction with the existing system, rather than new clinical accuracy studies.
Ask a specific question about this device
Page 1 of 4