510(k) Data Aggregation

    K Number
    K242689

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-01-13

    (129 days)

    Product Code
    Regulation Number
    878.4040
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AAMI Level 4 Protective Gown is intended to protect health care personnel and patients from the transfer of microorganisms, body fluids and particulate material. Not intended for use in the operating room. The AAMI Level 4 Protective Gown meets the barrier protection requirements of AAMI Level 4 per ANSI/AAMI PB70:2012 - Liquid Barrier Performance and Classification of Protective Apparel and Drapes Intended for Use in Health Care Facilities; but has an open back which is non-protective.

    The AAMI Level 4 Protective Gown is a single use, disposable medical device and is provided non-sterile.

    Device Description

    The AAMI Level 4 Protective Gown is a single use, disposable medical device and is provided non-sterile. It has an open back which is non-protective.

    AI/ML Overview

    This document is a 510(k) clearance letter for an AAMI Level 4 Protective Gown. It is not a study report for a diagnostic device or an AI/software-as-a-medical-device (SaMD) product. Therefore, the requested information about acceptance criteria, study details, sample sizes, ground truth establishment, expert qualifications, adjudication methods, and comparative effectiveness studies is not applicable to this document.

    The document states that the device "meets the barrier protection requirements of AAMI Level 4 per ANSI/AAMI PB70:2012 - Liquid Barrier Performance and Classification of Protective Apparel and Drapes Intended for Use in Health Care Facilities". This indicates the performance standard the device is intended to meet, but it doesn't provide the detailed acceptance criteria or a study write-up in the format requested for a software device.

    In summary, the provided text does not contain the information necessary to answer the specific questions about acceptance criteria, study design, and performance metrics as it pertains to a diagnostic or AI device. It is a regulatory clearance for a physical medical device (a protective gown).

    Ask a Question

    Ask a specific question about this device

    K Number
    K211538

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-11-17

    (183 days)

    Product Code
    Regulation Number
    878.4040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AZAC Protect X Level 3, Single Use, Non-Sterile, Protective Gown is intended to protect health care personnel and patients from the transfer of microorganisms, body fluids and particulate material. Not intended for use in the operating room.

    The AZAC Protect X Level 3, Single Use, Non-Sterile, Protective Gown gown meets the barrier protection requirements of AAMI Level 3 per ANSI/AAMI PB70:2012 – Liquid Barrier Performance and Classification of Protective Apparel and Drapes Intended for Use in Health Care Facilities.

    The AZAC Protect X Level 3, Single Use, Non-Sterile, Protective Gown is a single use, disposable medical device and is provided non-sterile.

    Device Description

    The AZAC Protect X Level 3, Single Use, Non-Sterile, Protective Gown is used in moderate risk situations such as venous blood draws, inserting IVs, and emergency room use. The gown is not sterile; is not an isolation gown and is not indicated for use in the Operating Room. The AZAC Protect X Level 3, Single Use, Non-Sterile, Protective Gown is made of one-piece chlorinated polyethylene elastomer non-woven tri-laminate. The Level 3 gown meets the ANSI/ AAMI PB70 Level 3 standards. The gown is a single use, disposable device.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device: an AZAC Protect X Level 3, Single Use, Non-Sterile, Protective Gown. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than proving clinical effectiveness through new clinical trials as would be required for a premarket approval (PMA) application for a novel, high-risk device.

    Therefore, the information typically requested about AI/algorithm performance studies (e.g., sample sizes for test and training sets, expert ground truth establishment, MRMC studies, effect sizes of AI assistance, standalone performance) is not applicable to this document. This document describes the testing of a physical medical device (a protective gown), not an AI/algorithm-based diagnostic or assistive system.

    The acceptance criteria and performance data provided relate to the physical properties and barrier performance of the gown, not to an algorithm's accuracy or clinical utility.

    Here's the information extracted from the document that is relevant to the acceptance criteria and the non-clinical study proving the device meets them:

    1. A table of acceptance criteria and the reported device performance:

    Test Method (Acceptance Criteria)PurposeCriteria (Acceptance Criteria)Result (Reported Device Performance)
    AAMI PB70:2012Overall barrier performanceMeet requirements for classification as AAMI PB70 Level 3 gownPass
    Hydrostatic Pressure Test (AATCC 127:2018)Resistance to static liquidNLT 50.0 cm H2OPass
    Impact Penetration (AATCC 42-2017)Resistance to liquid streamNo penetrationPass
    ASTM D 3776 - Mass/unit areaBasis weight of material26 g/m²Pass
    ASTM D5034 Breaking Strength - Grab TestDurability: Grab testLength – 27 N, Width – 31.5 NPass
    ASTM D5587 Tear Strength - Trapezoid TestDurability: TrapezoidLength – 10.5 N, Width 14.0 NPass
    16 CFR 1610 - FlammabilityFlammabilityClass 1Pass
    ISO 10993-5 CytotoxicityBiocompatibility: Cytotoxicity potentialUnder the conditions of the study, non-cytotoxicPass
    10993-10 Sensitization/IrritationBiocompatibility: Sensitization and irritationUnder the conditions of the study, not a sensitizer; Under the conditions of the study, non-irritatingPass

    2. Sample sized used for the test set and the data provenance:

    • Sample Size: The document does not specify the exact sample sizes (number of gowns or material samples) used for each non-clinical test. It states that "Results confirm that the design inputs and performance specifications are met."
    • Data Provenance: The tests were conducted internally by AZAC Group or by labs on their behalf, based on "internal requirements and applicable standards." The country of origin and whether the data is retrospective or prospective are not detailed. These are standard laboratory tests for physical properties.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. For physical and material property testing, "ground truth" is established by adhering to published ASTM, AATCC, AAMI, and ISO standard test methods and their specified procedures, rather than by human expert consensus or interpretation of complex data. The results are objective measurements or categorical determinations (e.g., pass/fail).

    4. Adjudication method for the test set:

    • Not applicable. Adjudication methods like 2+1 or 3+1 are typically used for expert review of medical images or clinical cases to establish ground truth in diagnostic studies. For these non-clinical material tests, the results are determined by the specified test protocols and criteria.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This document is for a physical protective gown, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This document is for a physical protective gown, not an algorithm.

    7. The type of ground truth used:

    • For the non-clinical performance data, the "ground truth" is defined by the objective pass/fail criteria established by recognized industry standards (e.g., AAMI PB70:2012, AATCC 127:2018, ASTM D 3776, ISO 10993-5). These standards specify the test methodology and the performance thresholds required for a "Pass" result.

    8. The sample size for the training set:

    • Not applicable. There is no "training set" as this is not an AI/machine learning device.

    9. How the ground truth for the training set was established:

    • Not applicable. There is no "training set."