510(k) Data Aggregation

    K Number
    K161239

    Validate with FDA (Live)

    Device Name
    MTA2.2 MATERIAL
    Manufacturer
    Date Cleared
    2016-09-01

    (122 days)

    Product Code
    Regulation Number
    872.3820
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MTA2.2 MATERIAL is indicated for dental procedures that contact pulp and periradicular tissues, as well as obturation and sealing of root canals.

    Device Description

    The MTA2.2 MATERIAL is designed and developed for dental clinicians to use in contact with vital pulp tissue and periradicular tissue, including sealing and obturation of root canals. This material can be used for dental procedures contacting pulpal or periradicular tissue such as: Pulp capping, Cavity lining, Base material in a cavity, Pulpotomies, Root-end filling, Apexification, Perforation repair, Root resorption, and Obturation (pulpectomy) in primary or secondary teeth.

    The dentist will mix the powder and water-based gel of MTA2.2 and place the mixed MTA2.2 MATERIAL into the space created by the procedure. The procedure may be part of caries treatment, root canal treatment, or periapical surgery. These procedures are performed on primary and secondary dentition and are color stable over time. MTA2.2 MATERIAL is tinted yellow, pink and gray.

    AI/ML Overview

    This document describes the premarket notification (510(k) submission) for the MTA2.2 MATERIAL, a root canal filling resin. However, it does not present a study with acceptance criteria and reported device performance in the way typically expected for a medical imaging or diagnostic device involving statistical metrics like sensitivity, specificity, or reader studies.

    Instead, this submission focuses on demonstrating substantial equivalence to a predicate device (MTA2.1 MATERIAL, K140955) based on material properties, design, indications for use, and conformance to established dental standards.

    Therefore, many of the requested sections about acceptance criteria, study design, sample sizes for test/training sets, expert ground truth, adjudication, and MRMC studies are not applicable or cannot be extracted from this document, as they pertain to clinical performance evaluations not present here.

    Here's a breakdown of what can be extracted based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The "acceptance criteria" here are conformance to recognized consensus standards for dental materials. The "reported device performance" refers to the MTA2.2 MATERIAL meeting these standards.

    Acceptance Criteria (Standard Conformance)Reported Device Performance (MTA2.2 Material)
    ISO 6876: Endodontic sealing materials
    FlowConforms to standard
    Film thicknessConforms to standard
    SolubilityConforms to standard
    Setting timeConforms to standard
    Working timeConforms to standard
    RadiopacityConforms to standard
    ADA 57: Endodontic Sealing Material(Same as above, as 57 relates to ISO 6876)
    RadiopacityConforms to standard
    SolubilityConforms to standard
    Dimensional stabilityConforms to standard
    Film thicknessConforms to standard
    FlowConforms to standard
    Working timeConforms to standard
    Setting timeConforms to standard
    ISO 9917
    Compressive strengthConforms to standard
    Leaching of arsenic and leadConforms to standard
    Biocompatibility
    CytotoxicityEvaluated, no adverse findings reported
    Washout testsConfirmed stability
    Discoloration testsConfirmed color stability

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not applicable for a "test set" in the context of diagnostic performance. The document describes bench testing of material properties. The specific sample sizes for each bench test (e.g., number of specimens for compressive strength) are not provided.
    • Data Provenance: Not specified, but likely from laboratory bench testing conducted by the manufacturer, Avalon Biomed Inc.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. Ground truth in this context refers to established material properties measured according to standardized laboratory protocols, not expert consensus on clinical findings.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. No expert adjudication process is described for batch testing of material properties.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. This is a material safety and effectiveness submission, not an AI or diagnostic imaging device. No MRMC study was conducted or is relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • No. This is not an algorithmic device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • The "ground truth" for the performance data in this document is derived from measurements of physical and chemical properties according to established and recognized consensus standards (ISO 6876, ADA 57, ISO 9917).

    8. The sample size for the training set

    • Not applicable. This is not a machine learning or AI device that requires a training set.

    9. How the ground truth for the training set was established

    • Not applicable. See point 8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K140955

    Validate with FDA (Live)

    Device Name
    MTA2.1 MATERIAL
    Manufacturer
    Date Cleared
    2014-06-11

    (58 days)

    Product Code
    Regulation Number
    872.3820
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MTA2.1 MATERIAL is indicated for dental procedures that contact pulp and periradicular tissues, as well as obturation and sealing of root canals.

    Device Description

    The MTA2.1 MATERIAL is designed and developed for dental clinicians to use in contact with vital pulp tissue and periradicular tissue, including sealing and obturation of root canals. This material can be used for dental procedures contacting pulpal or periradicular tissue such as: Pulp capping, Cavity lining, Base material in a cavity, Pulpotomies, Root-end filling, Apexification, Perforation repair, Root resorption, and Obturation (pulpectomy) in primary or secondary teeth. The dentist will mix the powder and water-based gel and place the mixed MTA2.1 MATERIAL into the space created by the procedure. The procedure may be part of caries treatment, root canal treatment, or periapical surgery. These procedures are performed on primary and secondary dentition and are color stable over time. MTA2.1 MATERIAL is tinted yellow, pink and gray.

    AI/ML Overview

    The provided document describes the Avalon Biomed Inc.'s MTA2.1 MATERIAL and its substantial equivalence determination to a predicate device (MTA2 ROOT AND PULP MATERIAL, K122892). This submission is for a material, not an AI/ML powered medical device, therefore, many aspects of the requested information (e.g., sample size for test set, number of experts, adjudication methods, MRMC studies, standalone performance, training set details) are not applicable.

    However, I can extract the acceptance criteria and the studies performed to demonstrate the device meets those criteria, focusing on the available non-clinical performance data.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the MTA2.1 MATERIAL are based on established dental material standards, specifically ADA 57 and ISO 9917. The document states that the MTA2.1 MATERIAL meets these standards.

    Acceptance Criterion (Standard)Reported Device Performance
    Radiopacity (ADA 57)Meets standard
    Solubility (ADA 57)Meets standard
    Dimensional Stability (ADA 57)Meets standard
    Film Thickness (ADA 57)Meets standard
    Flow (ADA 57)Meets standard
    Working Time (ADA 57)Meets standard
    Setting Time (ADA 57)Meets standard
    Compressive Strength (ISO 9917)Conforms to standard
    Leaching of Arsenic (ISO 9917)Conforms to standard
    Leaching of Lead (ISO 9917)Conforms to standard
    Washout StabilityConfirmed stability
    Discoloration StabilityConfirmed color stability
    Biocompatibility (Cytotoxicity)Evaluated, passed

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not specified. The document refers to "bench testing" and "design validation" for these non-clinical tests, implying laboratory experiments. The specific number of samples for each test (e.g., radiopacity, solubility) is not detailed.
    • Data Provenance: The tests were conducted as part of the device's "design validation." This implies laboratory testing performed by or for the manufacturer (Avalon Biomed Inc.). The country of origin of the data is not explicitly stated beyond the company being in Bradenton, FL, USA, and the submission being to the US FDA. The studies are prospective in the sense that they were conducted specifically for this regulatory submission to evaluate the material's properties against predefined standards.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    Not applicable. The "ground truth" for material properties like radiopacity, solubility, or compressive strength is established through standardized laboratory measurement methods as defined by ADA 57 and ISO 9917, rather than expert consensus.

    4. Adjudication Method for the Test Set

    Not applicable for material property testing.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. This type of study is typically relevant for interpretative diagnostic devices where human readers evaluate cases with and without AI assistance. This submission is for a dental material, not an AI system.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    No, this is not applicable as the device is a dental material, not an algorithm or AI system.

    7. The Type of Ground Truth Used

    The "ground truth" for the material's performance relies on standardized measurement methods specified by ADA 57 and ISO 9917. These standards define the procedures and expected ranges for various physical and chemical properties of dental root canal filling and pulp materials. For example:

    • Radiopacity: Measured according to the method described in ADA 57.
    • Solubility, Dimensional Stability, Film Thickness, Flow, Working Time, Setting Time: Measured according to the methods described in ADA 57.
    • Compressive Strength, Leaching of Arsenic and Lead: Measured according to the methods described in ISO 9917.
    • Biocompatibility: Evaluated through cytotoxicity tests.

    8. The Sample Size for the Training Set

    Not applicable. This is a dental material, not an AI/ML model that requires a training set.

    9. How the Ground Truth for the Training Set was Established

    Not applicable. (See #8)

    Ask a Question

    Ask a specific question about this device

    K Number
    K122892

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-12-11

    (82 days)

    Product Code
    Regulation Number
    872.3820
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MTA2 ROOT AND PULP MATERIALS are intended for use in dental procedures that contact pulp and periradicular tissues, as well as obturation and sealing of root canals.

    Device Description

    The MTA2 ROOT AND PULP MATERIALS are designed and developed for dentists to use in dental procedures that contact with vital pulp tissue and periradicular tissue, as well as obturation and sealing of root canals. The MTA2 ROOT AND PULP MATERIALS can be used for dental procedures such as: pulp capping, cavity lining, base material in a cavity, pulpotomies, root-end filling, apexification, perforation repair, root resorption, and sealing/obturation. The dentist mixes the powder with the provided gel and places the mixed MTA2 material into the space created by the procedure. The procedure may be part of caries treatment, root canal treatment, or periapical surgery. The MTA2 ROOT AND PULP MATERIALS are available in white or gray.

    AI/ML Overview

    The provided text describes the 510(k) summary for a dental device, the MTA2 ROOT AND PULP MATERIALS. This device is a cement primarily used for root canal procedures and interaction with pulp and periradicular tissues.

    Here's an analysis of the acceptance criteria and supporting study details based on the provided input:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Standard / Test)Reported Device Performance (MTA2 ROOT AND PULP MATERIALS)
    ADA 57 (Radiopacity)Met the ADA 57 standard
    ADA 57 (Solubility)Met the ADA 57 standard
    ADA 57 (Dimensional Stability)Met the ADA 57 standard
    ADA 57 (Film Thickness)Met the ADA 57 standard
    ADA 57 (Flow)Met the ADA 57 standard
    ADA 57 (Working Time)Performed to measure, compared to predicates (results not explicitly stated but implied meeting similar performance)
    ADA 57 (Setting Time)Performed to measure, compared to predicates (results not explicitly stated but implied meeting similar performance)
    ISO 9917 (Compressive Strength)Tests performed to show conformance to ISO 9917
    ISO 9917 (Leaching of Arsenic)Tests performed to show conformance to ISO 9917
    ISO 9917 (Leaching of Lead)Tests performed to show conformance to ISO 9917
    Biocompatibility (Cytotoxicity)Evaluated (with the gel)
    Biocompatibility (Implantation Tests)Evaluated (with the gel)
    Compositional Analysis (X-ray diffraction)Performed
    Compositional Analysis (X-ray fluorescence)Performed
    WashoutPerformed
    Particle Size AnalysisPerformed
    Hydroxyapatite FormationPerformed (induced precipitation of hydroxyapatite crystals in synthetic body fluid)
    Substantial Equivalence to Predicate K011009 (White MTA Material)Demonstrated through bench testing and similarities in inorganic powder composition, hydration mechanism, and intended use for pulp/periapical tissue contact.
    Substantial Equivalence to Predicate K080203 (MTA root canal sealer)Demonstrated through bench testing and similarities in inorganic powder composition, hydration mechanism, and intended use for sealing/obturation of root canals, including presence of a gel.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document primarily describes non-clinical performance (bench testing). There is no information provided about a "test set" in the context of human subjects or a clinical study. The data provenance is laboratory testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. This device's evaluation was based on non-clinical, bench testing against recognized standards (ADA 57, ISO 9917). There was no "test set" requiring expert ground truth in the traditional sense of diagnostic accuracy studies. The standards themselves serve as the 'ground truth' for performance.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. Without a clinical test set or expert evaluation component, there is no adjudication method described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI/CAD device. It is a dental material.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This is not an AI/CAD device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this device's performance was established by adherence to recognized international and national standards for dental materials: ADA 57, ISO 6876, and ISO 9917. These standards define the acceptable physical, chemical, and biological properties for such materials.

    8. The sample size for the training set

    Not applicable. This is not a machine learning or AI device that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. This is not a machine learning or AI device.