Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K200057

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-06-04

    (143 days)

    Product Code
    Regulation Number
    886.4300
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The bioli™ IOL Delivery System is a single-use, sterile device intended to insert an approved single-piece foldable acrylic intraocular lens (IOL) into the human eye through an incision.

    The bioli™ IOL Delivery System is for implantation of qualified Lenstec Softec 1 IOL and IOL models validated for use with this device as indicated in the IOL approved labeling.

    Device Description

    The bioli™ IOL Delivery System is an intraocular lens delivery system used for progressive folding and delivering intraocular lenses into the eye for replacement of the human crystalline lens. The bioli™ IOL Delivery System is intended for use during cataract surgery.

    The bioli™ IOL Delivery System is composed of two parts:

    • (1) a cartridge that holds the intraocular lens in position in preparation for loading and folding, and
    • (2) a single-use injector that houses a silicone-tipped plunger used to advance the lens into the capsular bag

    The cartridge incorporates a smooth, internal geometry that progressively folds the lens for delivery through conventional nozzle geometry. A coating is applied to the cartridge including the inner surfaces to increase lubricity.

    The cartridge is loaded by positioning the intraocular lens into the cartridge, folding the trailing haptic onto the anterior side of the optic, and finally pushing the optic edge of the lens to position it as far into the cartridge as the forceps will permit. After the cartridge is attached onto the injector, the system is now ready for the advancement of the plunger to deliver the lens.

    AI/ML Overview

    The document provided is a 510(k) summary for the bioli™ IOL Delivery System. It describes the device, its intended use, technological characteristics, and performance data to demonstrate substantial equivalence to a legally marketed predicate device.

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly provide a formal table of acceptance criteria alongside reported device performance for all tests. However, it states that various tests "met the defined acceptance criteria and requirements," or "showed passing results," or "demonstrated that the device can successfully deliver" or "was able to."

    Here's a summary of the performance claims, with acceptance criteria implied by the "passing" or "successful delivery" statements:

    TestAcceptance Criteria (Implied)Reported Device Performance
    BiocompatibilityAcceptable biocompatibility per ISO 10993 parts 5, 10, and 11"The biocompatibility testing was found acceptable." This included cytotoxicity, sensitization, ocular irritation, acute systemic toxicity, and material-mediated pyrogenicity.
    IOL Delivery Performance (Mechanical)Successful delivery of IOLs without affecting functionality, damage, or out-of-spec dimensions."All lenses were delivered through the bioli™ IOL Delivery System according to the loading and delivery instructions." "After delivery, lenses were observed for possible damages or scratches using a 10x microscope and showed passing results, and were within dimensional specifications. The bioli™ IOL Delivery System also showed no damage after lens delivery." "The resulting data from simulated surgical manipulation... showed that the device can successfully deliver Lenstec's hydrophilic Softec 1 IQLs of low to high diopters and IOL models validated for use with this device... without affecting the functionality of the lens."
    IOL Optical and Dimensional PropertiesMaintain optical properties, dimensional properties, and surface/bulk homogeneity before and after delivery per ISO 11979-3:2012.IOLs were "evaluated for optical properties, dimensional properties and overall surface and bulk homoqeneity before and after being surgically manipulated... IOL optical properties and overall surface and bulk homogeneity inspection were conducted in accordance with ISO 11979-3:2012... showed passing results, and were within dimensional specifications."
    Lens Opening TimeUnspecified, but evaluated to ensure proper IOL function.IOLs were "evaluated for... lens opening time after folding." (Results implied as acceptable by overall successful delivery statement)
    Cartridge PerformanceMaintain overall surface and bulk homogeneity."The bioli™ IOL Delivery System was also evaluated for its cartridge performance, such as overall cartridge surface and bulk homogeneity." (Results implied as acceptable by overall successful delivery statement)
    Shelf-Life (Package & Performance Stability)5-year shelf-life, meeting acceptance criteria for visual inspection, package seal, and package integrity per ASTM F1980-2007 (2011 reapproved), ASTM F88-09, ASTM F1929-12, ISO 11737-2:2009, USP 35 and ASTM F1140/F1140M-2013."Package stability and performance stability test results met the defined acceptance criteria and requirements for a shelf-life claim of 5 years. Visual inspection, package seal and package integrity of the sterile barrier system of bioli™ IOL Delivery System met the requirements in accordance with ASTM F1980-2007(2011 reapproved). The post aging tests were carried out in accordance with ASTM F88-09, ASTM F1929-12, ISO 11737-2:2009, USP 35 and ASTM F1140/F1140M-2013."
    Sterilization ValidationSterilization using EtO with residual EO within safe levels post-aeration period per ISO 11135:2014."Results from sterilization validation testing conducted in accordance with ISO 11135:2014 demonstrated that bioli™ IOL Delivery System is able to be sterilized using EtO sterilization parameters with residual EO found to be within safe levels post-aeration period."
    Transport StabilityResist physical damage from drop, vibration, and short extreme temperature impacts per ISO 2233:2001, ASTM D4169-16, ASTM F88/F88M-15, ASTM F2096-11, and ASTM F1929-15."The results demonstrated that bioli™ IOL Delivery System was able to resist the physical damage caused by drop, vibration and short extreme temperature impacts during transportation."

    2. Sample sizes used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: The document does not specify exact sample sizes (number of IOLs or devices) used for the non-clinical tests. It generally refers to "All lenses" being delivered and evaluated.
    • Data Provenance: The document does not specify the country of origin of the data, nor whether the non-clinical tests were conducted retrospectively or prospectively. The tests are described as "simulated surgical manipulation" and "performed," suggesting prospective testing in a laboratory setting.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This is a medical device submission, not an AI/algorithm-based diagnostic device. Therefore, the concept of "ground truth" established by experts in the context of diagnostic interpretation is not directly applicable. The "ground truth" for this device's performance relies on objective measurements and established engineering and biocompatibility standards. For instance, IOL damage/scratches were assessed using a "10x microscope," and optical/dimensional properties were assessed against ISO standards. The expertise would be in the technicians and engineers conducting these standardized tests.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable, as this is not an AI diagnostic device requiring expert adjudication of interpretations. The tests involve objective measurements against predefined criteria and standards.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. The bioli™ IOL Delivery System is a physical medical device (an intraocular lens injector), not an AI algorithm. Therefore, no MRMC study or assessment of human reader improvement with AI assistance was conducted or is relevant.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI algorithm. The performance described is of the physical device in laboratory simulations.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the bioli™ IOL Delivery System's performance is based on:

    • Objective measurements against established industry standards: Such as ISO 11979-3:2012 for IOL properties, ISO 10993 series for biocompatibility, ASTM standards for packaging and transport.
    • Visual inspection under magnification: For damage or scratches to IOLs and the device.
    • Functional tests: Ensuring successful delivery of IOLs "without affecting the functionality of the lens."

    8. The sample size for the training set

    Not applicable. This is a physical medical device, not a machine learning model. There is no "training set" in this context.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K182965

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-03-29

    (155 days)

    Product Code
    Regulation Number
    886.4300
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BL-Cart™ IOL Delivery Cartridge is a single-use, sterile device intended to insert an approved single-piece foldable acrylic intraocular lens (IOL) into the human eye through a surgical procedure. The cartridge is intended to be used in conjunction with Alcon Monarch® III IOL Delivery System injector. The BL-Cart™ IOL Delivery Cartidge is only for the insertion of Lenstec Softec I IOL and IOL models validated for use with this device as indicated in the approved IOL labeling.

    Device Description

    BL-Cart™ IOL Delivery Cartridge (Model: Type D) is an intraocular lens (IOL) delivery cartridge solution capable of supplying the ophthalmologist surgeon with a cartridge pre-coated with lubricious coating for use during a typical cataract surgery. The BL-Cart™ cartridge is used to deliver an approved IOL lens. The BL-Cart™ IOL Delivery Cartridge and referenced predicate device, Monarch® III D cartridge, utilize the same lens folding system design in which IOLs are folded and/or delivered from an injector system and the resulting optical and physical properties of the IOL remain unchanged as a result of folding/delivery. The BL-Cart™ IOL Delivery Cartridge comes with one pre-coated cartridge, packaged in a Tyvek pouch, placed in a 10 unit box as intended for marketing, and sterilized by ethylene oxide (EtO). The BL-Cart™ IOL Delivery Cartridge and the referenced legally marketed predicate device share the same principal method to implant an intraocular lens into the eye.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification for the BL-Cart™ IOL Delivery Cartridge. This device is classified as a Class I medical device (Intraocular Lens Guide). Class I devices generally have lower regulatory burden and often rely on substantial equivalence to a predicate device rather than extensive clinical studies.

    Here's an analysis of the acceptance criteria and supporting study information:

    Acceptance Criteria and Reported Device Performance

    Given the nature of a Class I device and the 510(k) submission, explicit "acceptance criteria" for performance metrics like sensitivity or specificity are not typically defined or required in the same way as for higher-risk devices or software. Instead, acceptance is based on demonstrating substantial equivalence to a legally marketed predicate device. This involves showing that the new device is as safe and effective as the predicate, often through shared technological characteristics, intended use, and performance testing that confirms it meets established standards or does not raise new safety/effectiveness concerns.

    For this device, the primary performance goals are related to its ability to deliver an IOL without damage to the IOL or the cartridge, and to maintain the IOL's properties.

    Acceptance Criterion (Implicit)Reported Device Performance (as demonstrated by non-clinical tests)
    BiocompatibilityPassed cytotoxicity, sensitization, ocular irritation, acute systemic toxicity, and material-mediated pyrogenicity tests in accordance with ISO 10993-5, 10, and 11.
    IOL Delivery Success (without affecting IOL functionality)All lenses were successfully delivered through the BL-Cart™ IOL Delivery Cartridge according to loading and delivery instructions.
    IOL Integrity Post-Delivery (optical, dimensional, surface)IOLs evaluated for optical properties, dimensional properties, and overall surface/bulk homogeneity before and after manipulation using the BL-Cart™ showed passing results and were within dimensional specifications. Lenses observed for possible damages or scratches using a microscope showed passing results.
    Cartridge Integrity Post-Delivery (surface, bulk homogeneity)BL-Cart™ showed no damage after lens delivery.
    IOL Opening TimeEvaluated this property after folding. (Specific results not given, but implied to be acceptable as part of overall positive conclusion).
    Sterilization (SAL 1x10-6)Achieved with Ethylene Oxide (EtO) sterilization.
    Shelf-LifeTesting was performed (specific results not detailed, but implied to be acceptable).

    Study Information

    1. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: The document does not explicitly state a numerical "sample size" for the non-clinical tests in terms of how many IOLs were tested or how many cartridges were used. It states, "All lenses were delivered through the BL-Cart™ IOL Delivery Cartridge," and that testing involved "Lenstec's hydrophilic Softec 1 IOLs of low to high diopters and IOL models validated for use with this device." This implies a sufficient number of tests to cover the range of intended use.
    • Data Provenance: The data is from non-clinical (bench) testing performed by the manufacturer, AST Products, Inc. There is no mention of country of origin for the data beyond being conducted by the submitting company in the USA. The data is prospective in the sense that it was generated specifically for this submission to demonstrate substantial equivalence.

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. This device is an IOL delivery cartridge, not a diagnostic or AI-driven device requiring expert interpretation or ground truth establishment in a clinical sense. The "ground truth" for its performance is determined by established engineering and material science standards (e.g., ISO 11979-3 for IOL properties, ISO 10993 for biocompatibility) and direct physical observation/measurement (e.g., micro-inspection for damage, dimensional checks). No human experts are described as establishing "ground truth" for the test set observations.

    3. Adjudication method for the test set:

    • Not applicable. As there are no human-interpreted "cases" or "ground truth" to adjudicate, no adjudication method is mentioned or required. Performance is determined by objective measurements and adherence to standards.

    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. An MRMC study is relevant for diagnostic AI devices where multiple readers interpret cases with and without AI assistance. This device is a manual surgical tool (cartridge) and does not involve AI or human readers for diagnostic purposes.

    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This device is purely mechanical/physical and does not involve any algorithms or AI components.

    6. The type of ground truth used:

    • The "ground truth" for this device's performance is based on International Standards (e.g., ISO 11979-3 for IOL mechanical properties, ISO 10993 for biocompatibility), engineering specifications (dimensional, optical properties of IOLs post-delivery), and physical observation (absence of damage to IOL or cartridge). The performance is assessed against these objective benchmarks rather than a clinical outcome or expert consensus on clinical findings.

    7. The sample size for the training set:

    • Not applicable. As this device does not involve AI or machine learning algorithms, there is no "training set."

    8. How the ground truth for the training set was established:

    • Not applicable. No training set exists for this type of device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K172228

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-12-15

    (143 days)

    Product Code
    Regulation Number
    886.4300
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The pioli™ IOL Delivery System is a single-use, sterile device intended to insert a single-piece foldable intraccular lens (IOL) into the human eye through a surgical procedure. The system provides a tubular pathway for lens implantation through an incision. The pioli™ IOL Delivery System is only for the Lenstec Softee I IOL and IOL models validated for use with this device as indicated in the IOL approved labeling.

    Device Description

    The pioli™ IOL Delivery System is a single-use, sterile device intended to insert a single-piece foldable intraocular lens (IOL) into the human eye through a surgical procedure. The system provides a tubular pathway for lens implantation through an incision. The pioli™ IOL Delivery System is only for the insertion of the Lenstec Softec I IOL and IOL models validated for use with this device as indicated on the IOL approved labeling.

    The set consists of a single-use, sterile svringe injector with a silicone tip and a single-use, sterile LubriMATRIX™-treated cartridge.

    The tip of the plunger, housed in the injector, is covered by a silicone cushion that provides a good contact to the lens to ensure a smooth delivery. The plunger is advanced by applying a direct forward motion.

    Note: LubriMATRIX™ is a proprietary treatment composed of a hydrophilic polymer and a lubricious polymer engrafted onto the surface of the cartridge via a chemical polymerization process. Refer to MAF-1963 for further information.

    AI/ML Overview

    The provided document is a 510(k) summary for the pioli™ IOL Delivery System. It describes the device, its intended use, and its equivalence to a predicate device. The document details performance data from non-clinical tests to demonstrate that the device meets its acceptance criteria.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    IOLs show no damage or scratches after delivery.All delivered lenses showed no damages or scratches.
    IOLs are within dimensional specifications after delivery.All delivered lenses were within dimensional specifications.
    Cartridges show no damage after lens delivery.All cartridges showed no damages after lens delivery.
    IOL optical properties, sagitta, and overall surface/bulk homogeneity are maintained after surgical manipulation with the pioli™ IOL Delivery System. (Evaluated per ISO 11979-2 and ISO 11979-3).The resulting data from simulated surgical manipulation showed that the pioli™ IOL Delivery System (Model: PIOLI-D) can successfully deliver Lenstec's hydrophilic Softec 1 IOLs of low to high diopters and IOL models validated for use with this device... without affecting the functionality of the lens.
    The pioli™ IOL Delivery System can successfully deliver Lenstec's hydrophilic Softec 1 IOLs of low to high diopters and other validated IOL models without affecting lens functionality.The resulting data... showed that pioli™ IOL Delivery System (Model: PIOLI-D) can successfully deliver Lenstec's hydrophilic Softec 1 IOLs of low to high diopters and IOL models validated for use with this device as indicated on the IOL approved labeling without affecting the functionality of the lens.
    Packaging validation: packaging seals meet specified parameters (e.g., appearance, peeling strength, adhesion).The resulting data from packaging validation showed that the most suitable parameter to seal the packaging of pioli™ IOL Delivery System was under the condition of 129°C, 3.6 seconds, with all blisters passing the appearance inspection, peeling strength and adhesion tests.
    Biocompatibility of components in direct/indirect patient contact is established. (Implied, as based on similarity to predicate device).The components of pioli™ IOL Delivery System that have direct and indirect contact with the patient are manufactured of the same materials and same manufacturing processes as the predicate device, lioli™ IOL Delivery System. Therefore, the biocompatibility testing on the predicate device is applicable to the subject device.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the exact sample size for the IOLs and cartridges tested in the non-clinical performance study. It mentions "All lenses were delivered..." and "All delivered lenses showed no damages..." and "all cartridges showed no damages...", which implies that a set of lenses and cartridges were tested, but the specific number is not provided.

    • Sample Size: Not explicitly stated (referred to as "All lenses" and "all cartridges").
    • Data Provenance: The study was a "simulated surgical manipulation" non-clinical test conducted by the device manufacturer (AST Products, Inc.). The country of origin of the data is not specified, but the manufacturer is based in Billerica, MA, USA. It is a prospective non-clinical study.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • The document does not mention the use of experts to establish ground truth for the non-clinical performance tests. The evaluation of IOL damage/scratches, dimensional specifications, and cartridge damage appears to be based on objective inspection and adherence to ISO standards, rather than expert interpretation of a 'ground truth' per se.

    4. Adjudication Method for the Test Set

    • Not applicable/Not mentioned. The non-clinical tests described involve objective measurements and observations against pre-defined criteria (e.g., ISO standards, presence of damage, dimensional specifications). It does not involve subjective assessment requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, a TRMC comparative effectiveness study was not done. The device is an IOL delivery system, not an AI-assisted diagnostic or interpretive device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Not applicable. The pioli™ IOL Delivery System is a physical medical device (an IOL delivery system), not an algorithm or AI system.

    7. The Type of Ground Truth Used

    • The "ground truth" for the non-clinical performance study was based on objective measurements and observations per established standards and specifications. This includes:
      • ISO standards: ISO 11979-2 (Optical properties and test methods) and ISO 11979-3 (Mechanical properties and test methods) for IOL evaluation.
      • Visual inspection: For damages, scratches, and homogeneity of IOLs and cartridges.
      • Dimensional specifications: For IOLs after delivery.
      • Functional assessment: The ability to successfully deliver IOLs without affecting their functionality.
      • Packaging tests: Appearance inspection, peeling strength, and adhesion tests.

    8. The Sample Size for the Training Set

    • Not applicable. This is a physical medical device, not a machine learning or AI algorithm that requires a training set. The "training" for such a device involves engineering design, material selection, and manufacturing processes validation.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable, as there is no "training set" in the context of an IOL delivery system as described.
    Ask a Question

    Ask a specific question about this device

    K Number
    K142056

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-04-09

    (254 days)

    Product Code
    Regulation Number
    886.4300
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The lioli™ IOL Delivery System is a single-use, sterile device intended to insert a one-piece foldable intraocular lens (IOL) into the human eye through a surgical procedure. The system provides a tubular pathway for lens implantation through an incision. Only for the insertion of IOL models validated for use with this device as indicated in the IOL approved labeling.

    Device Description

    The lioli™ IOL Delivery System is a single-use, sterile device intended to insert a one-piece foldable intraocular lens (IOL) into the human eye through a surgical procedure. The system provides a tubular pathway for lens implantation through an incision. Only for the insertion of IOL models validated for use with this device as indicated in the IOL approved labeling.

    The set consists of a single- use, sterile syringe injector with a silicone tip and a single-use, sterile LubriMATRIX™-treated cartridge.

    The tip of the plunger, housed in the injector, is covered by a silicone cushion that provides a good contact to the lens to ensure a smooth delivery. The plunger is advanced by applying a direct forward motion.

    AI/ML Overview

    The provided document describes the lioli™ IOL Delivery System and its substantial equivalence to a predicate device, but it does not contain detailed acceptance criteria or a comprehensive study report that directly proves the device meets specific performance metric acceptance criteria in a structured format.

    The document discusses performance data in a general sense, focusing on non-clinical tests to demonstrate the device's functionality and safety by showing that the delivered IOLs were undamaged and within specifications.

    Here's an attempt to extract and synthesize the requested information based on the provided text, noting where specific details are absent:


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Implied/Deduced)Reported Device Performance
    IOL Integrity Post-Delivery:
    No damages or scratchesAll delivered lenses showed no damages or scratches.
    Within dimensional specificationsAll delivered lenses were within dimensional specifications.
    Functionality not affectedLens functionality was not affected.
    Cartridge Integrity Post-Delivery:
    No damagesAll cartridges showed no damages after lens delivery.
    Biocompatibility:
    Safe for human contact (all contact materials)Performance testing for all contact materials has been tested for biocompatibility and was found to be safe.
    Sterilization:
    SAL level 1x10^-6 achieved (ETO)Achieved with ETO sterilization.
    Shelf-Life:
    Maintained throughout shelf lifeTesting was performed; results not explicitly detailed but stated as part of the conclusion for SE.

    Explanation of "Implied/Deduced" Criteria: The document describes the results of the tests ("all delivered lenses showed no damages"), from which we can infer the unstated acceptance criteria (e.g., "no damages or scratches on IOLs post-delivery").

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not explicitly stated. The document mentions "All lenses were delivered..." and "Each lioli™ IOL Delivery System were also evaluated...", which implies a sample was used but the exact number for IOLs or delivery systems is not provided.
    • Data Provenance: Retrospective, as these are pre-market notification tests conducted by the manufacturer for regulatory submission. The country of origin of the data is not specified, but the manufacturer is AST Products, Inc., located in Billerica, MA, USA.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    • Experts: Not applicable. The "ground truth" for the performance tests (e.g., integrity checks, dimensional specifications, optical properties) was established through objective measurements and standardized test methods (ISO standards), not through expert consensus or interpretation of subjective data.
    • Qualifications: Not applicable for establishing ground truth in this context.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable. The tests performed are objective measurements and observations (e.g., microscope inspection for damage, dimensional checks), rather than subjective assessments requiring adjudication among experts.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What War the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    • MRMC Study: No, an MRMC study was not done.
    • Effect Size: Not applicable. This device is an IOL delivery system, not an AI-powered diagnostic or interpretive tool that would involve human readers or AI assistance in interpretation.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Standalone Performance: Not applicable. This device is a mechanical delivery system for IOLs, not an algorithm. Its "performance" is its ability to successfully deliver an IOL without damage, which was assessed in non-clinical tests.

    7. The Type of Ground Truth Used

    • Ground Truth: Objective measurements against predefined specifications and visual inspection. For example:
      • IOL optical properties: Measured in accordance with ISO 11979-2.
      • IOL mechanical properties (including sagitta): Measured in accordance with ISO 11979-3.
      • Damages/scratches: Visually inspected using a microscope.
      • Biocompatibility: Tested against known safety standards for materials.
      • Dimensional specifications: Compared against manufacturing specifications.

    8. The Sample Size for the Training Set

    • Sample Size: Not applicable. This device is a mechanical medical device, not a machine learning model, and therefore did not require a "training set."

    9. How the Ground Truth for the Training Set Was Established

    • Ground Truth Establishment: Not applicable, as there was no training set.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1