Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K142336

    Validate with FDA (Live)

    Date Cleared
    2014-09-17

    (27 days)

    Product Code
    Regulation Number
    862.1345
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GLUCOCARD 01 Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips, or palm. Testing is done outside the body (In Vitro diagnostic use). It is indicated for use at home by persons with diabetes as an aid to monitor the effectiveness of diabetes control. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates. It is intended for single patient use and should not be shared with other individuals.

    The GLUCOCARD 01 SENSOR PLUS Blood Glucose Test Strips are intended to be used with the GLUCOCARD 01 Blood Glucose Meter for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips, or palm.

    For over the counter use only

    The ReliOn Confirm Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips, or palm. Testing is done outside the body (In Vitro diagnostic use). It is indicated for use at home by persons with diabetes as an aid to monitor the effectiveness of diabetes control. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates. It is intended for single patient use and should not be shared with other individuals.

    The ReliOn Confirm Plus Blood Glucose Test Strips are intended to be used with the ReliOn Confirm Blood Glucose Meter for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips, or palm.

    For over the counter use only

    Device Description

    The GLUCOCARD 01 Blood Glucose Monitoring System and ReliOn Confirm Blood Glucose Monitoring System consist of a meter, test strips, and control solution for use as an aid to monitor the effectiveness of diabetes control.

    AI/ML Overview

    The provided FDA 510(k) summary (K142336) describes the GLUCOCARD 01 Blood Glucose Monitoring System and ReliOn Confirm Blood Glucose Monitoring System. This submission focuses on modifications to certain exterior materials used in the manufacture of the meter. Therefore, the primary studies mentioned are related to demonstrating the continued safety and performance of the device despite these material changes, rather than a clinical effectiveness study of the glucose monitoring itself.

    Based on the provided document, here's an analysis of the acceptance criteria and supporting studies related to the modifications:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria/TestReported Device Performance
    Material Modifications SafetyViral elimination effectivenessStudies were conducted on the surfaces of the GLUCOCARD 01 and ReliOn Confirm Meters. (Specific results not detailed in this summary, but the conclusion states substantial equivalence.)
    Cleaning and Disinfection DurabilityWithstand multiple cleaning and disinfection cyclesTesting was performed to demonstrate that the GLUCOCARD 01 and ReliOn Confirm meters can withstand multiple cleaning and disinfection cycles. (Specific results not detailed in this summary, but the conclusion states substantial equivalence.)
    Physical DurabilityRobustness of meter case design (e.g., drop, button durability)Drop testing and button durability studies demonstrate that the material changes do not affect the robustness of the meter case design. (Specific results not detailed in this summary, but the conclusion states substantial equivalence.)

    2. Sample Size and Data Provenance for Test Set

    The document does not specify sample sizes for the viral elimination, cleaning/disinfection, or drop/button durability test sets. It also does not explicitly state the provenance (e.g., country of origin) of the data for these engineering/material tests. These are typically simulated or laboratory tests rather than patient data.

    3. Number of Experts to Establish Ground Truth and Qualifications

    This information is not provided in the document. For engineering/material tests, "ground truth" would typically be established by established test methodologies and internal validation, rather than expert consensus in the clinical sense.

    4. Adjudication Method

    This information is not provided as it is not relevant for the type of engineering/material testing described.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was mentioned. This type of study is more relevant for AI/radiology devices where human interpretation is a key component. The device described is a glucose meter, and the submitted changes are material-related.

    6. Standalone (Algorithm Only) Performance Study

    No standalone performance study of an algorithm was conducted or described in this summary. The device's primary function is a measurement system; the "performance" here refers to the physical and safety aspects after material changes.

    7. Type of Ground Truth Used

    For the safety and durability tests:

    • Viral elimination effectiveness: The ground truth would be established by predefined microbiological assay standards and protocols to determine the reduction of viral load on surfaces.
    • Cleaning and disinfection durability: The ground truth would be resistance to degradation or damage after exposure to specified cleaning agents and cycles, measured against physical and functional integrity criteria.
    • Physical durability (drop/button): The ground truth would be the device's ability to remain functional and intact after specified mechanical stresses, measured against established engineering and performance specifications.

    8. Sample Size for the Training Set

    No training set is mentioned in the context of these material modification studies. This typically refers to machine learning models, which are not described here.

    9. How Ground Truth for Training Set Was Established

    Not applicable, as no training set was identified.


    Summary of the K142336 Submission Context:

    It's important to note that this 510(k) submission (K142336) is for modifications to an already cleared predicate device (K124021). The "Reason for Submission" explicitly states "Modifications to certain exterior materials used in the manufacture of the meter." This means the core functionality and clinical accuracy of the glucose measurement system were likely established in the predicate submission. The studies presented here are specifically to demonstrate that the new materials do not adversely affect the device's safety (e.g., viral transmission risk, ability to be disinfected) or durability, thereby maintaining substantial equivalence to the predicate.

    Ask a Question

    Ask a specific question about this device

    K Number
    K142035

    Validate with FDA (Live)

    Date Cleared
    2014-09-05

    (39 days)

    Product Code
    Regulation Number
    862.1345
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GLUCOCARD 01-mini Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips, or palm. Testing is done outside the body (In Vitro diagnostic use). It is indicated for use at home by persons with diabetes as an aid to monitor the effectiveness of diabetes control. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates. It is intended for single patient use and should not be shared with other individuals.

    The GLUCOCARD 01 SENSOR PLUS Blood Glucose Test Strips are intended to be used with the GLUCOCARD 01-mini Blood Glucose Meter for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips, or palm.

    The ReliOn Micro Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips, or palm. Testing is done outside the body (In Vitro diagnostic use). It is indicated for use at home by persons with diabetes as an aid to monitor the effectiveness of diabetes control. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates. It is intended for single patient use and should not be shared with other individuals.

    The ReliOn Micro Plus Blood Glucose Test Strips are intended to be used with the ReliOn Micro Blood Glucose Meter for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips, or palm.

    Device Description

    The GLUCOCARD 01-mini Blood Glucose Monitoring System and ReliOn Micro Blood Glucose Monitoring System consist of a meter, test strips, and control solution for use as an aid to monitor the effectiveness of diabetes control.

    AI/ML Overview

    This document describes the 510(k) premarket notification for the GLUCOCARD 01-mini Blood Glucose Monitoring System and the ReliOn Micro Blood Glucose Monitoring System, which are intended for quantitative measurement of glucose in fresh capillary whole blood samples.

    Here's the breakdown of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document does not explicitly state the quantitative acceptance criteria for system accuracy in a table format, nor does it provide detailed numerical results of the device's performance against such criteria. It generally states that a "clinical study was done with persons with diabetes to evaluate system accuracy."

    However, glucose monitoring systems like this are typically evaluated against standards like ISO 15197 for accuracy. A common acceptance criterion for blood glucose meters, as per ISO 15197, often includes:

    • 95% of results falling within ±15 mg/dL of the reference method for glucose concentrations < 100 mg/dL.
    • 95% of results falling within ±15% of the reference method for glucose concentrations ≥ 100 mg/dL.

    Without the full study report, specific reported performance values against these criteria are not available in this document.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document mentions "A clinical study was done with persons with diabetes to evaluate system accuracy and to assess ease of use." However, it does not specify the sample size for this clinical study.
    • Data Provenance: The document states that the clinical study was conducted with "persons with diabetes," implying real-world data collection. It doesn't specify the country of origin, but the submitting company is ARKRAY Factory, Inc. from Japan, with a US contact. It also doesn't explicitly state if the study was retrospective or prospective, but clinical studies for device accuracy are typically prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not provide information on the number of experts or their qualifications used to establish the ground truth for the test set. For blood glucose monitoring systems, the ground truth is typically established using a laboratory reference method (e.g., an automated analyzer calibrated to a recognized standard) rather than expert consensus on interpretation.

    4. Adjudication Method for the Test Set

    As the ground truth is established by a laboratory reference instrument, there isn't typically an "adjudication method" in the sense of expert review or consensus for the test set results directly. The device's readings are compared against the reference method.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is typically relevant for medical imaging or diagnostic interpretation tasks where human readers evaluate cases, often with and without AI assistance, to measure the impact of the AI on reader performance. For a blood glucose monitoring system, the device directly outputs a quantitative measurement, and human "readers" are not involved in interpreting the results in a diagnostic imaging sense.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    Yes, the system accuracy evaluation is inherently a standalone performance study. The GLUCOCARD 01-mini and ReliOn Micro Blood Glucose Monitoring Systems are intended to provide a quantitative glucose measurement directly. The clinical study mentioned evaluates the system's ability to accurately measure glucose without human interpretation influencing the measurement itself. The "human-in-the-loop" aspect for a blood glucose meter primarily involves the user performing the test correctly and interpreting their own results, not in aiding the device's measurement.

    7. Type of Ground Truth Used

    The ground truth for blood glucose monitoring systems is typically established using a laboratory reference method that is highly accurate and traceable to an international standard. While not explicitly stated in this summary, it is standard practice for glucose meter validation to compare the device's readings against a clinical laboratory analyzer.

    8. Sample Size for the Training Set

    The document does not provide any information about a "training set." This device is a blood glucose meter, not an AI/machine learning algorithm that requires a training set in the typical sense for image recognition or predictive modeling. The device's performance is determined by its electrochemical measurement technology, calibration, and manufacturing consistently.

    9. How the Ground Truth for the Training Set Was Established

    Since there is no "training set" in the context of an AI/machine learning model for this medical device, there is no ground truth established for a training set. The device itself is calibrated during manufacturing to known glucose concentrations using reference methods.

    Ask a Question

    Ask a specific question about this device

    K Number
    K124021

    Validate with FDA (Live)

    Date Cleared
    2014-06-30

    (550 days)

    Product Code
    Regulation Number
    862.1345
    Age Range
    12 - 120
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GLUCOCARD 01 Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips, or palm. Testing is done outside the body (In Vitro diagnostic use). It is indicated for use at home (over the counter [OTC]) by persons with diabetes as an aid to monitor the effectiveness of diabetes control. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates. It is intended for single patient use and should not be shared with other individuals.

    The ReliOn Confirm Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips, or palm. Testing is done outside the body (In Vitro diagnostic use). It is indicated for use at home (over the counter [OTC]) by persons with diabetes as an aid to monitor the effectiveness of diabetes control. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates. It is intended for single patient use and should not be shared with other individuals.

    Device Description

    The GLUCOCARD 01 Blood Glucose Monitoring System and ReliOn Confirm Blood Glucose Monitoring System consist of a meter, test strips, and control solution for use as an aid to monitor the effectiveness of diabetes control.

    AI/ML Overview

    The GLUCOCARD 01 Blood Glucose Monitoring System and ReliOn Confirm Blood Glucose Monitoring System did not include AI, therefore much of this information is not applicable.

    1. A table of acceptance criteria and the reported device performance

    No explicit acceptance criteria (e.g., specific percentages for accuracy within certain glucose ranges) or detailed performance statistics were provided in the document beyond a general statement of positive clinical study results.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document mentions "A clinical study was conducted with persons with diabetes to evaluate system accuracy and to assess ease of use." However, the sample size for this clinical study and the data provenance (country of origin, retrospective or prospective nature) are not specified.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    Not applicable, as this device measures glucose and does not typically involve expert interpretation as a ground truth method.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable, as this device measures glucose and does not involve human readers interpreting cases or AI assistance in that context.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    The device is a standalone blood glucose monitoring system. Its "performance" refers to the accuracy of its glucose measurements. The clinical study evaluated "system accuracy," which inherently means the performance of the device itself.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For blood glucose monitoring systems, the ground truth for accuracy studies is typically established by comparing the device's readings to a laboratory reference method (e.g., a central lab analyzer) known for its high accuracy in measuring blood glucose. The document does not explicitly state the specific ground truth method used, but it's universally implied for such devices.

    8. The sample size for the training set

    Not applicable, as this device does not utilize an AI algorithm that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1