Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K250212

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-04-15

    (81 days)

    Product Code
    Regulation Number
    884.4050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LapBox Power Tissue Containment System is intended for use as a multiple instrument port and tissue containment system during minimally invasive gynecologic laparoscopic surgery to enable the isolation and containment of tissue considered benign, resected during single-port or multi-site laparoscopic surgery during power morcellation and removal. The LapBox is compatible with electromechanical laparoscopic power morcellators that are between 15 mm and 18 mm in shaft outer diameter and 135 mm and 180 mm in shaft working length and which have an external component that allows for the proper orientation of the laparoscope to perform a contained morcellation. When used in women with fibroids, the LapBox Power Tissue Containment System is for women who are pre-menopausal and under age 50.

    Device Description

    The LapBox Power Tissue Containment System is a single use sterile device. It is comprised of a double wall inflatable polyurethane chamber which is mounted on an insertion shaft and is provided with two port sizes. Once the shaft is inserted to the abdominal cavity, the chamber is deployed and the organ to be morcellated is placed within the chamber. The chamber is then inflated using an external handpump and the sleeve of the chamber is exteriorized. The selected port is then placed over the sleeve in the incision site and the organ can be morcellated using a power morcellator. Once morcellation is complete, the port is removed, and the chamber is deflated and removed from the patient.

    AI/ML Overview

    The provided FDA 510(k) clearance letter describes a medical device, the LapBox Power Tissue Containment System, which falls under the regulation of medical devices, not AI/ML-based software. Therefore, the information typically found in a study proving an AI device meets acceptance criteria (such as sample sizes for training/test sets, expert adjudication, MRMC studies, or specific ground truth methodologies for AI models) is not applicable here.

    However, I can extract the acceptance criteria and performance data for the LapBox Power Tissue Containment System based on the non-clinical performance testing described in the document.


    Overview of Device and Study Context

    The LapBox Power Tissue Containment System is a medical device intended for use during minimally invasive gynecologic laparoscopic surgery to contain and isolate tissue during power morcellation. It is a physical device, not an AI/ML software. The provided document details the non-clinical testing performed to demonstrate its safety and effectiveness for 510(k) clearance, asserting substantial equivalence to a predicate device (PneumoLiner).

    1. Table of Acceptance Criteria and Reported Device Performance

    Since this is a physical medical device and not an AI/ML model, the "acceptance criteria" and "device performance" are focused on engineering, biological, and usability metrics rather than AI-specific performance metrics like sensitivity, specificity, or AUC. The document outlines various tests conducted and confirms whether the device met the prespecified acceptance criteria for each test. The actual quantitative "performance" data (e.g., specific burst pressures, detailed force measurements) are generally not fully presented in a 510(k) summary but rather summarized as meeting criteria.

    Acceptance Criterion (Category / Test)Reported Device Performance (Summary from Document)
    Sterilization & Shelf-Life
    Sterility Assurance Level (SAL)Validated in accordance with ISO 11135 to a SAL of $10^{-6}$.
    Shelf-Life (18 months)Validation performed following accelerated aging and simulated transit conditioning. Testing evaluated packaging integrity and device functionality, which met acceptance criteria.
    Biocompatibility
    CytotoxicityDevice was non cytotoxic.
    SensitizationDevice was non-sensitizer.
    Intracutaneous IrritationDevice was non-irritant.
    Acute Systemic ToxicityDevice was not systemically toxic.
    PyrogenicityDevice was non-pyrogenic.
    Non-Clinical Performance Testing (Bench)
    Bacterial Immersion Test (Impermeability)Test samples did not show any evidence of growth of Brevundimonas Diminuta after simulated use, disinfection, sterilization, and immersion. (Positive and negative controls performed as intended).
    Dimensional Verification (Initial)A total of 30 non-aged units tested for proximal handle outer diameter, inner tube inner diameter, outer tube inner diameter, and inflation chamber inner diameter. All samples met the prespecified acceptance criteria. (Supplemental to K221365 tests).
    Pressure Relief Valve TestingA total of 30 aged and 30 non-aged valves tested to ensure relief above set target. All samples met the prespecified acceptance criteria.
    Design and Performance Validation Test30 aged and 27 non-aged units tested by 2 trained surgeons in a simulator with bovine tongue. All samples passed a dye leak test. (Design changes validated in supplemental tests).
    Puncturing Force Comparative TestResults obtained with LapBox supported non-inferiority when compared to the predicate (72 predicate material samples; 72 aged and 72 non-aged LapBox material samples tested).
    Air Tube Bond Connection & Handpump Assembly (Bond Strength)30 aged units tested. All samples met the prespecified acceptance criteria. (Supplemental to K221365 tests).
    Burst Pressure Evaluation (Modified Air Tube)10 non-aged units with modified new air tube tested. All samples met the prespecified acceptance criteria. (Supplemental to K221365 tests).
    Inflation Pressure Range (Shape Maintenance)Customized jig developed. One LapBox unit used. Demonstrated that the chamber will maintain its shape at a pressure below that relieved by the pressure relief valve.
    Comparison Force Test (Outward Force)Customized jig developed. One predicate and one LapBox utilized. Results demonstrate that a uniform distribution of force by LapBox does not pose a greater risk of compression on adjacent organs when compared to the predicate.
    Corrosion Resistance (K221365)30 units of 2 different metal components, double sterilized and 18 months accelerated aged. No corrosion was noted on any samples.
    Burst Pressure Evaluation (K221365)30 aged and 30 non-aged units tested for burst pressure of the chamber. All tested units met the prespecified acceptance criteria.
    Bond Strength Tests (K221365)30 non-aged units and 30 aged units tested for tensile strength of various bonds. All samples met the prespecified acceptance criteria.
    Dimensional Verification (K221365)36 non-aged units tested for shaft effective length, inner sleeve height, chamber double wall height, chamber outer diameter, and chamber opening. All samples met the prespecified acceptance criteria.
    Viral Penetration (ASTM F 1671) (K221365)30 material samples with 12 months equivalent accelerated aging. All samples met the prespecified acceptance criteria.
    Maximum Organ Size Testing (K221365)2 aged devices utilized. (Specific findings not detailed, but considered in the clearance).
    Training Validation and Usability Studies
    Initial Usability Study (31 physicians)31 physicians (varying levels of experience) performed 40 procedures in a simulator with bovine tongue. No device samples had observed leakage of dye. Minor device issues noted and addressed via manufacturing/labeling changes.
    Supplementary Usability Study (6 physicians)6 physicians performed 33 procedures in a simulator. (Units used in bacterial immersion testing). No evidence of any device leak.
    Training Validation (Porcine Model)16 physicians (varying experience) completed 32 morcellation procedures. Device samples subjected to dye leak test. No samples leaked. Minor issues with IFU adherence noted, leading to revision of training and IFU.
    Follow-up Usability Study (15 physicians)15 physicians performed 30 power morcellation procedures in a simulator. Demonstrated that revisions to training and IFU were effective, ensuring safe and effective use.
    Additional Animal Studies
    Performance Characterization (In-Vivo)Part 1: One animal, 3 devices. Part 2: One animal, 5 devices. (Purpose: define performance characteristics; specific quantitative results not detailed in summary, but support clearance).

    2. Sample Sizes Used for the Test Set and the Data Provenance

    As this is a physical device, the "test set" refers to the samples of the device itself and the human users/animals involved in performance testing. There isn't "data provenance" in the sense of patient data for AI.

    • Bacterial Immersion Test: 32 device samples. Data provenance: Controlled laboratory test.
    • Dimensional Verification Tests (Supplemental): 30 non-aged units. Data provenance: Controlled laboratory test.
    • Pressure Relief Valve Testing: 30 aged and 30 non-aged units (total 60). Data provenance: Controlled laboratory test.
    • Design and Performance Validation Test: 30 aged and 27 non-aged units (total 57). Data provenance: Simulated environment using bovine tongue.
    • Puncturing Force Comparative Test: 72 predicate material samples, 72 aged LapBox material samples, and 72 non-aged LapBox material samples (total 216 material samples). Data provenance: Controlled laboratory test.
    • Bond Strength Tests (Supplemental): 30 aged units. Data provenance: Controlled laboratory test.
    • Burst Pressure Evaluation (Modified Air Tube): 10 non-aged units. Data provenance: Controlled laboratory test.
    • Inflation Pressure Range: One LapBox unit. Data provenance: Controlled laboratory test.
    • Comparison Force Test: One predicate device and one LapBox device. Data provenance: Controlled laboratory test.
    • Corrosion Resistance (from K221365): 30 units of 2 different metal components (total 60). Data provenance: Controlled laboratory test.
    • Burst Pressure Evaluation (from K221365): 30 aged and 30 non-aged units (total 60). Data provenance: Controlled laboratory test.
    • Bond Strength Tests (from K221365): 30 non-aged units and 30 aged units (total 60). Data provenance: Controlled laboratory test.
    • Dimensional Verification (from K221365): 36 non-aged units. Data provenance: Controlled laboratory test.
    • Viral Penetration ASTM Method F 1671 (from K221365): 30 material samples. Data provenance: Controlled laboratory test.
    • Maximum Organ Size Testing (from K221365): 2 aged devices. Data provenance: Controlled laboratory test.
    • Initial Usability Study: 31 physicians, performing a total of 40 procedures. Data provenance: Simulator with bovine tongue.
    • Supplementary Usability Study: 6 physicians, performing a total of 33 procedures. Data provenance: Simulator.
    • Training Validation: 16 physicians, completing 32 procedures. Data provenance: Porcine model with bovine tongue.
    • Follow-up Usability Study: 15 physicians, performing 30 procedures. Data provenance: Simulator.
    • Performance Characterization Animal In-Vivo Study: Part 1: 1 animal, 3 devices. Part 2: 1 animal, 5 devices. Data provenance: Porcine model (in-vivo).

    Data provenance details (e.g., country of origin, retrospective/prospective) are not directly applicable as these are laboratory, simulator, or animal studies of a physical device, not patient data for an AI model.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    For a physical medical device, "ground truth" often relates to objective measurements (e.g., integrity, dimensions) or expert assessment of functionality and usability.

    • Design and Performance Validation Test: 2 trained surgeons. Qualifications not explicitly detailed beyond "trained surgeons."
    • Usability Studies (Initial, Supplementary, Training Validation, Follow-up): Physicians of "varying levels of experience" were involved. The exact number of "experts" (or the most senior/experienced) who formally established performance metrics or approved the "ground truth" (i.e., whether a procedure was successful or a leak occurred) is not specified beyond their participation as test users. The methods (e.g., dye leak test) were objective.

    4. Adjudication Method for the Test Set

    Adjudication methods (like 2+1, 3+1) are typically used for subjectively labeling data in AI ground truth establishment. For this physical device, the "adjudication" is inherent in the test protocols and measurements.

    • Dye leak tests: This is an objective visual assessment.
    • Dimensional, pressure, strength tests: These rely on objective measurement against prespecified criteria, not human consensus on subjective labels.
    • Usability Studies: While surgeons performed the procedures, the outcomes like "observed leakage of dye" or "device leak" are objective observations rather than subjective interpretations requiring adjudication. "Minor device issues noted" would have been observed directly by the testers or study coordinators.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. MRMC studies are specific to evaluating human reader performance (e.g., radiologists interpreting images) with and without AI assistance on complex medical cases. This device is a physical surgical tool, not an AI software tool for interpretation.

    6. If a Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Was Done

    This question is not applicable. The device is a physical product, not an algorithm. Its performance is always in the context of human use (e.g., surgeons performing procedures with it). The "non-clinical performance testing (bench)" and "animal studies" could be considered "standalone" in that they test the device's inherent properties and function, but not in the AI sense of algorithm-only performance.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance is established through a combination of:

    • Objective Measurement against Engineering Specifications: For tests like dimensional verification, burst pressure, puncturing force, bond strength, pressure relief valve performance, corrosion resistance.
    • Controlled Biological/Physical Outcomes: For bacterial immersion (absence of growth) and viral penetration (meeting criteria).
    • Expert Observational Assessment of Functionality: In simulated and in-vivo environments, such as dye leak tests after morcellation, maintenance of pneumoperitoneum, adequate space/visualization, and observation of proper deployment and removal without spillage. Usability studies assessed successful completion of tasks and adherence to IFU.

    There is no "pathology" or "outcomes data" in the sense of clinical patient outcomes or histopathology for establishing ground truth for individual performance evaluations outlined.

    8. The Sample Size for the Training Set

    This question is not applicable for a physical device. There is no "training set" in the context of machine learning. The studies were validation tests.

    9. How the Ground Truth for the Training Set Was Established

    This question is not applicable, as there is no training set for an AI model.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221365

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-08-25

    (105 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LapBox Tissue Containment Removal System is indicated to contain and isolate tissue during, or prior to, surgical removal and/or extracorporeal manual morcellation.

    Device Description

    The LapBox Tissue Containment Removal System is a single use sterile device. It is comprised of a double wall inflatable polyurethane chamber which is mounted on an insertion shaft and is provided with two port sizes. Once the shaft is inserted to the abdominal cavity, the chamber is deployed and the organ to be morcellated is placed within the chamber. The chamber is then inflated using an external handpump and the sleeve of the chamber is exteriorized. The selected port is then placed over the sleeve in the incision site and the organ can be manually morcellated. Once morcellation is complete, the port is removed, and the chamber is deflated and removed from the patient.

    AI/ML Overview

    The FDA 510(k) summary for the LapBox Tissue Containment Removal System does not contain the level of detail typically found in reports of diagnostic device performance, such as sensitivity, specificity, or accuracy. This is a surgical device designed for tissue containment and removal, rather than a diagnostic AI-powered device. Therefore, many of the requested fields are not applicable or cannot be extracted from the provided text.

    Here's an attempt to answer the questions based on the information available in the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document primarily focuses on demonstrating substantial equivalence to a predicate device through various performance tests, rather than providing specific numerical acceptance criteria and performance data for a diagnostic algorithm. The acceptance criteria described are often a "pass" or "same as predicate" for various engineering and biocompatibility tests.

    Acceptance Criteria CategorySpecific Test/EvaluationReported Device Performance (LapBox)Comparison to Predicate (K142427)
    Regulatory & ClassificationRegulation Number876.1500Same
    Product CodeGCJSame
    Indications for UseTo contain and isolate tissue during, or prior to, surgical removal and/or extracorporeal manual morcellation. Contraindicated for laparoscopic power morcellation during gynecologic procedures and with powered cutting devices.Same
    BiocompatibilityCytotoxicitySuccessfully passedSame (predicate also tested)
    SensitizationSuccessfully passedSame (predicate also tested)
    IrritationSuccessfully passedSame (predicate also tested)
    Acute Systemic ToxicitySuccessfully passedNot explicitly stated for predicate in table, but overall "Same" implied for biocompatibility
    PyrogenicitySuccessfully passedNot explicitly stated for predicate in table, but overall "Same" implied for biocompatibility
    Infrared SpectroscopySuccessfully passedNot explicitly stated for predicate in table, but overall "Same" implied for biocompatibility
    Material/Design PropertiesChamber/bag & Port/Guard MaterialsTPU (Polyurethane) film, Nylon reinforced fabric and polyurethane portSimilar (Predicate: TPU and coiled HDPE port)
    Single useYesSame
    SterilityEtOSame
    Puncture Force (chamber/Bag)Tested against predicate, yielded superior resultsSuperior results for LapBox
    Puncture Force (Port/Guard)Tested against predicate, yielded superior resultsSuperior results for LapBox
    Viral Penetration per ASTM F 1671PassSame
    Functionality (Non-Clinical Performance Testing)Performance Characterization In-Vivo StudyPerformed and successfully passedNot explicitly in comparison table, but overall supports substantial equivalence.
    Corrosion ResistancePerformed and successfully passedNot explicitly in comparison table, but overall supports substantial equivalence.
    Pressure Relief Valve TestingPerformed and successfully passedNot explicitly in comparison table, but overall supports substantial equivalence.
    Burst Pressure EvaluationPerformed and successfully passedNot explicitly in comparison table, but overall supports substantial equivalence.
    Bond Strength TestPerformed and successfully passedNot explicitly in comparison table, but overall supports substantial equivalence.
    Dimensional Verification TestPerformed and successfully passedNot explicitly in comparison table, but overall supports substantial equivalence.
    Closure Integrity Test: Bubble TestPerformed and successfully passedNot explicitly in comparison table, but overall supports substantial equivalence.
    Design and Performance Validation TestPerformed and successfully passedNot explicitly in comparison table, but overall supports substantial equivalence.
    Clinical Simulation StudyPerformed and successfully passedNot explicitly in comparison table, but overall supports substantial equivalence.
    Training Validation StudyPerformed and successfully passedNot explicitly in comparison table, but overall supports substantial equivalence.
    Maximum organ size testingPerformed and successfully passedNot explicitly in comparison table, but overall supports substantial equivalence.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document lists "Performance Characterization In-Vivo Study" and "Clinical Simulation Study" as non-clinical performance testing, but does not specify the sample size, type of data (e.g., animal, cadaveric, human), or provenance for these studies. These are likely bench or animal studies rather than human clinical trials based on the context of a 510(k) for this type of device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not provided in the document. Given the nature of the device (surgical tissue containment), "ground truth" in the diagnostic sense is not directly applicable. For performance testing, experts might have been involved in assessing the functionality or usability, but details are absent.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not provided and is generally not applicable to the type of device and testing described in this 510(k) summary. Adjudication methods are typically relevant for human clinical reads in diagnostic studies.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC study was conducted or is applicable as this is not an AI-powered diagnostic device designed to assist human readers.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable, as this is a physical surgical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For a surgical device, ground truth usually relates to the successful and safe performance of its intended function. The document mentions an "In-Vivo Study" and "Clinical Simulation Study," which would generate data on performance. However, the specific type of "ground truth" or endpoints (e.g., successful tissue containment, absence of leakage, ease of use) and how it was established are not detailed for these studies. For biocompatibility, the ground truth is typically laboratory assay results.

    8. The sample size for the training set

    Not applicable, as this is a physical surgical device, not an AI model requiring a training set.

    9. How the ground truth for the training set was established

    Not applicable, as this is a physical surgical device, not an AI model.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1