Found 7 results

510(k) Data Aggregation

    K Number
    K120209

    Validate with FDA (Live)

    Date Cleared
    2012-04-12

    (79 days)

    Product Code
    Regulation Number
    880.5725
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This device is an accessory for a syringe or infusion pump (not manufactured by AIV, Inc.). The healthcare worker connects the device to the corresponding pump. The patient uses the switch on the cable to signal the pump to deliver medication consistent with the parameters entered into the pump by the healthcare worker.

    Connects to designated syringe or infusion pump model used in hospital, ambulatory and home care environments.

    Device Description

    AIV's Patient Pendant Bolus Cables are a replacement for similar cables manufactured by the Original Equipment Manufacturers (OEM) for their respective syringe and infusion pumps. AIV does not manufacture the pumps. These Patient Pendant Bolus Cables are an accessory for the pump and are also available as a replacement part. These cables include a switch and are plugged into a mating connector on the syringe or infusion pump to allow the patient to request medication within the parameters entered into the syringe pump by the healthcare provider and the designed limits on the pumps.

    The AIV Patient Pendant Bolus Cables use the same type of construction and have the same technological characteristics as the predicate device – the accessory supplied with the original pump. These are passive devices that contain no electronic components. They are powered by the mating equipment using low voltages to detect changes in the state of the push button. Shielding and insulation of these devices is substantially equivalent to the predicate device.

    AIV Patient Pendant Bolus Cables use Biocompatible PVC for the cables, cable jackets and over molded connectors with an integral strain relief.

    AI/ML Overview

    The provided document describes the predicate devices as "Patient Controlled Analgesia (PCA) Button Cables," and focuses on demonstrating that the AIV Patient Pendant Bolus Cables are substantially equivalent to these existing, legally marketed devices. It is explicitly stated that this device "is not sterile, has no software component, has no electronic components and is not a kit." Therefore, the details requested about clinical studies, MRMC studies, training/test sets, and ground truth are not applicable to this type of device and submission.

    Here's the breakdown based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Performance Features)Reported Device Performance
    Verification Testing: Construction and operation consistent with technical specifications.Consistent with technical specifications.
    Storage and Cleaning Tests: Operation remains consistent with technical specifications after repeated cycles.Operation remains consistent with technical specifications after repeated cleaning cycles and exposure to storage environmental limits.
    Performance Testing:A switch activation test pattern and pump test program were used, showing that under identical activation patterns and pump programs, the number of dose requests is equivalent to the predicate devices, and the fluid dispensed is consistent with the accuracy of the target pump (which implies the cable accurately transmits the patient's request).
    Safety: Double insulation of conductors.Same as predicate devices.
    Operating Principle: Patient presses button to signal medication request to pump.Same as predicate devices. The device functions as intended: registering a patient request when the button is pressed and not registering one when it is not activated.
    Operating Voltage: 3.3 VDC to 5 VDC.Same as predicate devices.
    Biocompatibility: Use of Biocompatible PVC.Biocompatible PVC used for cables, jackets, and over-molded connectors with integral strain relief. (Referenced Section F for biocompatibility test reports, which were not provided in the excerpt but are stated as completed).
    Intended Use & Indications for Use: As an accessory/replacement part for syringe/infusion pumps.Same as predicate devices; connects to designated syringe or infusion pump models in various healthcare environments. The patient uses the switch to signal the pump to deliver medication consistent with entered parameters and pump limitations.
    Technological Characteristics: Same construction and passive nature.Uses the same type of construction and has the same technological characteristics as the predicate devices. These are passive devices with no electronic components, powered by mating equipment using low voltages to detect push-button state changes. Shielding and insulation are substantially equivalent to the predicate device.
    Device Failure Consequences: Same as OEM devices.Due to extreme similarities, the consequences of a modified device and a device failure are the same for both AIV and OEM devices. (A risk management file was developed and used throughout the development process, referenced Section G, not provided here).
    Clinical Testing: Not performed.The non-clinical testing demonstrated the AIV Patient Pendant Bolus Cables are as safe, as effective, and perform as well as the legally marketed accessories for syringe and infusion pumps. No clinical testing was performed as the device is an accessory with no direct patient interaction other than the button press, and its function relies on the primary pump's accuracy, not its own.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size: The document does not specify a distinct "test set" sample size in terms of patient numbers or data records. The testing described focuses on the physical devices themselves. The tests involved "each of the AIV Patient Pendant Bolus Cables models" with specific test patterns and programs. This implies testing each manufactured cable model rather than a statistical sample size of patient data.
    • Data Provenance: The tests were conducted by AIV at their facility in Harmans, MD, USA. The testing is non-clinical, involving device verification and performance testing, not patient data. Therefore, the concept of "data provenance" related to patient data (e.g., country of origin, retrospective/prospective) is not applicable here.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • This information is not applicable. The device is a simple mechanical accessory (a cable with a button). The "ground truth" for its operation is whether the button press is correctly registered by the connected pump and whether it withstands cleaning/storage. This would be verified through engineering tests rather than expert interpretation of complex data.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • This information is not applicable. The device validation involves objective pass/fail criteria for physical and functional performance, not subjective review requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, an MRMC comparative effectiveness study was not done. This type of study is relevant for AI or diagnostic devices where human interpretation is involved. The AIV Patient Pendant Bolus Cable is a non-AI, non-diagnostic accessory.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • This information is not applicable. The device is a cable with a button; it does not contain an algorithm or software. Its function is to facilitate human-in-the-loop interaction (patient presses button) with the main infusion pump.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • The "ground truth" for this device's performance is its direct functional operation according to engineering specifications and its equivalence to predicate devices. This includes:
      • Functional Correspondence: Verification that when the button is pressed, the signal is sent, and when not pressed, no signal is sent.
      • Durability and Environmental Stability: Performance remains consistent after cleaning and storage cycles.
      • Material Biocompatibility: Conformance to biocompatibility standards (referenced in Section F).
      • Physical and Electrical Equivalence: Matching construction, operating voltage, and safety features of predicate devices.

    8. The sample size for the training set

    • This information is not applicable. There is no AI component or machine learning model that requires a training set.

    9. How the ground truth for the training set was established

    • This information is not applicable, as there is no training set for this device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K070232

    Validate with FDA (Live)

    Date Cleared
    2007-03-27

    (62 days)

    Product Code
    Regulation Number
    870.2900
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    These devices are intended to be used with various electrocardiograph recorders/monitors for both diagnostic and monitoring purposes. They are solely intended to be used between the electrode in contact with the patient's skin and the recording/monitoring device. This cabling facilitates the conduction of signals between the patient and the monitoring device. AIV cables are limited by the Indications for Use of the connected recording/monitoring device.

    Device Description

    AlV's Electrocardioaraph (ECG) cables (also sometimes referred to as trunk or leadwires) are replacements for similar cables manufactured by the Original Equipment Manufacturers (OEM) and other third party after market manufacturers for their respective monitors. The Trunk Cables connect the OEM patient monitor to the patient leadwires connect the trunk cable to the skin mounted ECG electrodes.

    These cables consist of connectors on each cable end and a shielded bulk cable. The cables are used to transfer the signals from the electrodes to the patient monitor. The Trunk Cables have limited skin contact with the patient, while the Leadwires attached on the patient's chest have more continuous skin contact.

    The AIV cables use the same type of construction and have the same technological characteristics as the predicate devices. They use a medical grade PVC cable jacket with medical grade PVC overmolded connectors with an integral strain relief.

    AI/ML Overview

    The provided document describes the 510(k) summary for AIV's ECG Cables, Trunk and Patient Lead Wires, seeking substantial equivalence to existing devices. As such, the study described here is NOT an AI/ML device study. The acceptance criteria and "device performance" in this context refer to engineering specifications and compliance with recognized standards for medical devices, rather than an AI model's predictive accuracy.

    Here's the information extracted and formatted according to your request, with clarifications due to the nature of the device:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Standard / Key Characteristic)Reported Device Performance (Compliance / Equivalence)
    Connector Retention Force (ANSI/AAMI EC53A-1998/(R)2001)Declared conformity to standard; equivalent to predicates.
    Electrical Performance and Safety (ANSI/AAMI EC53-1995/(R)2001)Declared conformity to standard; equivalent to predicates.
    Design/Appearance (Cables with various connectors)Same as predicate devices.
    Type of Construction (Flexible, shielded, multi-conductor)Same as predicate devices.
    Connector Design (Keyed, snap and grabber)Same as predicate devices.
    Cable Length (Various specified lengths)Same as predicate devices.
    Wire Colors (Snappers and grabbers color coded)Same as predicate devices.
    Wire Material (Braided shield, tin/copper with elastomer jacket)Same as predicate devices.
    Sterility (Used non-sterile)Same as predicate devices.
    Intended Use (Conduct impulse signals from electrode to patient monitor)Same as predicate devices.
    Patient Usage (Reusable)Same as predicate devices.
    Anatomical Sites (Attached to electrodes at specified locations on chest wall)Same as predicate devices.
    Technological Characteristics (Construction, materials)Same as predicate devices.

    2. Sample size used for the test set and the data provenance

    • Sample Size: Not explicitly stated as a "test set" in the context of an AI/ML study. The evaluation was based on bench testing of the AIV cables. The sample size for these physical tests is not specified in the provided summary.
    • Data Provenance: Not applicable in the sense of data origin for an AI model. This is a physical device and the "data" comes from bench testing of the manufactured cables.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not applicable. This is not an AI/ML device requiring expert ground truth for image interpretation or similar. The "ground truth" for these cables is compliance with engineering standards (ANSI/AAMI EC53 and EC53A) and functional equivalence to legally marketed predicate devices, assessed through bench testing. The evaluation was done by the manufacturer (AIV) and reviewed by the FDA.

    4. Adjudication method for the test set

    • Not applicable. There was no "test set" requiring adjudication in the AI/ML sense. Compliance was determined by meeting consensus standards and demonstrating equivalence to predicate devices through bench testing.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. This is not an AI/ML device. Therefore, no MRMC study with human readers or AI assistance was performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is not an algorithm. The "standalone performance" for this device would be its ability to physically and electrically perform its intended function (signal conduction) when bench tested, which was stated to be "as intended."

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • The "ground truth" for this device is based on compliance with recognized consensus standards (ANSI/AAMI EC53 and EC53A) for electrical and mechanical performance, and demonstrated functional and technological equivalence to legally marketed predicate devices.

    8. The sample size for the training set

    • Not applicable. This is not an AI/ML device, so there is no training set.

    9. How the ground truth for the training set was established

    • Not applicable. There is no training set for this device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K070342

    Validate with FDA (Live)

    Date Cleared
    2007-03-19

    (41 days)

    Product Code
    Regulation Number
    880.2910
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The devices are intended to be used in a clinical setting, physicians' offices or other alternate care settings. The devices are reusable and are intended for the quantitative detection of body temperature, which can be obtained orally, under the arm (axially) or rectally by trained medical personnel.

    Device Description

    AIV's Clinical Electronic Thermometers (also referred to as Temperature Probes) are replacement thermometers for temperature monitoring/measuring systems. The device is reusable. The devices are electronic thermometers using a thermistor temperature sensor. The devices are placed inside a protective sheath (not manufactured or supplied by AIV). Then the thermometer is used on the patient by trained medical personnel. The sensor is displayed on a monitor/measuring system (not manufactured or supplied by AIV). The probe consists of a thermistor located in a metal tipped plastic shaft is attached to a plastic body that allows the protective sheath (not supplied component that actually touches the patient) to be gripped and ejected off the probe. A shielded multi-conductor cable is used to conduct thermistor signals to the connector for the monitor/measuring system. Some a heating resistor that allows a more rapid temperature measurement.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the AIV Clinical Electronic Thermometer:

    Acceptance Criteria and Device Performance Study for AIV Clinical Electronic Thermometer

    The provided document describes a 510(k) Premarket Notification for a replacement clinical electronic thermometer. The core of a 510(k) submission is to demonstrate substantial equivalence to a legally marketed predicate device, not necessarily to provide extensive clinical performance data against specific acceptance criteria.

    Based on the document, the "acceptance criteria" are primarily implicit in the demonstration of substantial equivalence to existing predicate devices. The study proving this substantial equivalence is limited to bench testing.

    1. Table of Acceptance Criteria and Reported Device Performance

    Given the nature of a 510(k) for a replacement device, the "acceptance criteria" are largely derived from the characteristics and performance of the predicate devices. The document explicitly states that the AIV device uses "the same type of construction and have the same technological characteristics" as the predicate devices.

    Acceptance Criteria (Implicit from Predicate Devices)Reported Device Performance (AIV Clinical Electronic Thermometer)
    Intended Use: Quantitative detection of body temperature (oral, axial, rectal)Same as Predicate: "for the quantitative detection of body temperature, which can be obtained orally, under the arm (axially) or rectally"
    Patient Usage: ReusableSame as Predicate: "Reusable."
    Technical Characteristics: Thermistor technologySame as Predicate: "Thermistor technology"
    Design/Appearance: Probe with shaft and body assembly using a protective sheathSame as Predicate: "Probe with shaft and body assembly that uses a protective sheath" and "Thermistor in metal tipped plastic shaft, flexible, shielded, multi conductor electrical cable and monitor connector."
    Type of Construction: Thermistor in metal-tipped plastic shaft, flexible, shielded cableSame as Predicate: "Thermistor in metal tipped plastic shaft, flexible, shielded, multi conductor electrical cable and monitor connector."
    Connector Design: Selected to fit appropriate monitorsSame as Predicate: "Connectors are selected to fit the appropriate monitors" (A list of compatible monitors is provided, implying functional compatibility).
    Cable Length: Various specified lengthsSame as Predicate: "Various specified lengths."
    Wire Material: Braided shield, tin/copper with elastomer jacketSame as Predicate: "Braided shield, tin/copper with elastomer jacket."
    Sterility: Used non-sterileSame as Predicate: "Used non-sterile."
    Performance: Function as intended (implicitly, accuracy and reliability comparable to predicates)"Bench Testing demonstrates that the AIV devices perform as intended." (No specific numerical performance metrics like accuracy ranges, response times, or drift are provided in this summary. This statement relies on the assumption that "as intended" means consistent with the performance of substantially equivalent predicate devices.) "These devices do not raise new issues of safety and effectiveness, nor do they alter the fundamental technology of the predicate devices."

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a sample size for a test set in the traditional sense of a clinical trial or a structured validation study with a defined dataset. The "study" referenced is bench testing.

    • Data Provenance: The data provenance is not specified beyond "bench testing." It is not clinical data (patient data), so concepts like "country of origin" are not applicable in this context. It's an in-house prospective test by the manufacturer.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    Not applicable. This submission is for a medical device (thermometer) and relies on bench testing and comparison to predicate devices, not on expert clinical interpretation or ground truth establishment in the way an AI diagnostic algorithm would.

    4. Adjudication Method for the Test Set

    Not applicable. There is no "adjudication method" described as the testing is not based on expert review of ambiguous cases.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, a MRMC comparative effectiveness study was not done. This type of study is relevant for evaluating the impact of an AI system on human reader performance, typically in diagnostic imaging. The AIV Clinical Electronic Thermometer is a direct measurement device, not an AI diagnostic tool that assists human readers.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done

    Yes, in spirit, a standalone performance was done, in that the device's performance was evaluated independently through bench testing. However, it's not an "algorithm only" performance as the device is a physical electronic thermometer. The "standalone" evaluation here refers to the device itself performing its intended function (temperature measurement) without a human's interpretative cognitive task being assisted by AI. The human is involved in placing and operating the device, but the temperature reading itself is generated by the device.

    7. The Type of Ground Truth Used

    The "ground truth" for the bench testing would be established by reference temperature standards and calibrated measurement equipment. While not explicitly detailed, standard engineering and metrology practices dictate that the thermometer's readings would be compared against highly accurate and traceable temperature sources to verify its performance specifications (e.g., accuracy, precision). This is essentially physics/engineering-based ground truth, not clinical or pathology-based.

    8. The Sample Size for the Training Set

    Not applicable. This device is a physical electronic thermometer; it does not involve machine learning or AI algorithms that require a "training set" of data.

    9. How the Ground Truth for the Training Set was Established

    Not applicable. As there is no training set for an AI algorithm, there is no ground truth established for one.

    Ask a Question

    Ask a specific question about this device

    K Number
    K052544

    Validate with FDA (Live)

    Date Cleared
    2006-09-29

    (379 days)

    Product Code
    Regulation Number
    870.2700
    Age Range
    All
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    These devices are intended to be used as replacement accessories to Original Equipment Manufacturer monitors used in the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate. These systems are used to monitor patients who are either well or poorly perfused, in hospitals, hospital type facilities, intra-hospital transport and home environments.

    Device Description

    AIV's Pulse Oximeter Adapter Cables are direct replacements for similar cables manufactured by the Original Equipment Manufacturers (OEM) for their respective monitors. They use the same type of construction and have the same technological characteristics as the predicate devices (OEM). The adapter cable connects the OEM pulse oximeter sensor with the OEM monitor.

    AI/ML Overview

    This document describes the 510(k) summary for Pulse Oximeter Adapter Cables from American IV Products, Inc. The submission aims to demonstrate substantial equivalence to existing legally marketed predicate devices.

    1. Table of Acceptance Criteria and Reported Device Performance

    For this device, the "acceptance criteria" are based on demonstrating substantial equivalence to predicate devices, focusing on key characteristics like intended use, construction, and performance. The "reported device performance" reflects that these criteria have been met through testing and comparison.

    Acceptance Criteria (Comparison to Predicate Device)Reported Device Performance
    Intended Use: Connects OEM pulse oximeter sensor with OEM monitor for continuous, non-invasive SpO2 and pulse rate monitoring.Same as predicate devices. The AIV adapter cable connects the OEM pulse oximeter sensor with the OEM monitor for continuous, non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate.
    Type of Construction: Flexible, shielded, multi-conductor electrical cable.Same as predicate devices. AIV cables use the same type of construction.
    Connector Design: Adapter cable connectors are keyed to fit the appropriate monitors.Same as predicate devices. Adapter cable connectors are keyed to fit the appropriate monitors.
    Installation into System: Installed between monitor and sensor using keyed connectors.Same as predicate devices. Installed in system between monitor and sensor using keyed connectors.
    Target Patient Population: Patients that are well or poorly perfused.Same as predicate devices.
    Patient Use/Reuse: Reusable.Same as predicate devices.
    Sterility: Non-sterile.Same as predicate devices.
    IEC 60601-1 Testing: Passed Safety Testing.Passed Safety Testing. The company reports conformity to consensus standards relating to Electrical/EMC/Mechanical/Safety.
    IEC 60601-1-2 Testing: Passed or equivalent to OEM adapter cable if OEM system not passing with AIV cable.Passed or shown to be equivalent. When the OEM system did not pass with the AIV cable, the OEM system was tested with the OEM adapter cable to show AIV equivalent to the OEM adapter cable. The company reports conformity to consensus standards relating to Electrical/EMC/Mechanical/Safety.
    Overall Performance: Device performs as intended and does not raise new safety/effectiveness issues or alter fundamental technology.Bench Testing demonstrates that the AIV devices perform as intended. The devices do not raise new issues of safety and effectiveness, nor do they alter the fundamental technology of the predicate devices. This is a conclusion based on the comparison and bench testing.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for the Test Set: Not explicitly stated as a separate "test set" in the context of an AI/algorithm-driven device. For these types of accessories, the testing primarily involves bench testing and comparison to the physical and electrical characteristics of multiple predicate devices. The number of specific AIV part numbers tested is 13, each corresponding to a different OEM part number for various monitor systems.
    • Data Provenance: The testing is primarily retrospective, comparing the AIV cables to existing OEM devices and standards. The data is generated from bench testing within the manufacturer's environment to ensure conformity and equivalence. The country of origin of the data is not specified but is inferred to be the manufacturer's location in the USA (Hanover, MD).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of device (adapter cables) does not typically involve "experts" establishing a "ground truth" for a test set in the way an AI diagnostic algorithm would. Instead, engineering and quality assurance personnel perform instrumental measurements and comparisons. The submission mentions "Director of Regulatory Affairs and Quality Assurance" (Gregory Falk), suggesting that qualified personnel within the company are responsible for overseeing the testing and ensuring compliance. Specific numbers or detailed qualifications of technical experts involved in the bench testing are not provided in this summary.

    4. Adjudication Method for the Test Set

    Not applicable in the context of this device. Adjudication methods (e.g., 2+1) are typically used for interpreting ambiguous cases in diagnostic imaging or similar scenarios where human interpretation is a variable. For adapter cables, compliance is determined by meeting physical, electrical, and performance specifications through direct measurement and comparison, not by consensus among interpreters.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is an adapter cable, not an AI or diagnostic algorithm, so no MRMC study or AI assistance comparison would be performed.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is an adapter cable, not an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" in this context is established by:

    • Predicate Device Specifications: The technical and performance specifications of the legally marketed OEM pulse oximeter adapter cables.
    • Consensus Standards: International (e.g., IEC 60601-1, IEC 60601-1-2) and national consensus standards for medical electrical equipment safety and electromagnetic compatibility.
    • Functional Intent: The confirmed ability of the AIV cables to perform the intended function of connecting the sensor to the monitor to allow continuous SpO2 and pulse rate monitoring, without degrading the performance of the overall system.

    8. The Sample Size for the Training Set

    Not applicable. There is no "training set" as this is a hardware accessory, not an AI or machine learning algorithm.

    9. How the Ground Truth for the Training Set was Established

    Not applicable. There is no training set or ground truth in the AI/ML sense for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K040392

    Validate with FDA (Live)

    Date Cleared
    2004-09-14

    (210 days)

    Product Code
    Regulation Number
    884.2720
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    These devices are intended to be used as replacement accessories to Corometrics and Hewlett Packard Monitors, to measure fetal heart rate and uterine contractions in the gravid patient.

    AIV Transducer Model Numbers: FM10835 and FM10836. These devices are a direct replacement for the EPIC EFU200-20, Ultrasound Transducer for measuring fetal heart rate.

    AIV Transducer Model Numbers: FM10839 and FM10840. These devices are a direct replacement for the Corometric's 2264LAX. TOCO Transducer for measuring uterine contractions.

    AIV Transducer Model Number: FM10841. This device is a direct replacement for the EPIC EFT200-20, TOCO Transducer for measuring uterine contractions.

    Device Description

    AIV's ultrasound (US) and tocodynamometer (TOCO) transducers are replacements for similar transducers manufactured by Corometrics and Hewlett Packard for their respective monitors. The AIV transducers are also a replacement for similar transducers manufactured by Epic for use on Corometrics and Hewlett Packard monitors.

    The US transducers are used to detect the fetal heart rate using Doppler shift technology. The TOCO transducers detect uterine activity using a strain gauge for evaluating contractions. These transducers are intended to be a direct replacement for the Corometrics, Hewlett Packard and Epic transducers.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the fetal monitoring transducers, formatted as requested:

    Acceptance Criteria and Device Performance

    Acceptance CriteriaReported Device Performance
    Acoustic Output < 20mW/cm² averageAcoustic output testing shows power is less than 20mW/cm² average.
    Assembly design equivalent to predicate deviceAIV assembly design is equivalent to predicate device assembly design.
    Performance as intended and equivalent to predicate device assembliesBench Testing demonstrates that the AIV devices perform as intended and are equivalent to predicate device assemblies.
    BiocompatibilityAIV plastics have conformed to consensus standards relating to Biocompatibility.
    Does not raise new issues of safety and effectiveness"These devices do not raise new issues of safety and effectiveness, nor do they alter the fundamental technology of the predicate devices."
    Does not alter fundamental technology of predicate devices"These devices do not raise new issues of safety and effectiveness, nor do they alter the fundamental technology of the predicate devices."
    Intended Use: Measure fetal heart rate and uterine contractions in the gravid patient (same as predicate)The AIV devices are intended to be used as replacement transducer accessories for Corometrics and Hewlett Packard monitors, for use in measuring fetal heart rates and uterine contractions, which is "Same" as the predicate devices.
    Anatomical Sites (same as predicate)The ultrasound transducer is placed on the maternal abdomen aimed at the fetal heart; the TOCO transducer is placed on the maternal abdomen over the fundal area of the uterus (Same as predicate).
    Target Patient Population (same as predicate)Gravid patients, especially during labor (Same as predicate).
    FHR Range (same as predicate: dependent upon monitor specifications)Dependent upon monitor specifications (Same as predicate).
    Uterine Activity Range (same as predicate: dependent upon monitor specifications)Dependent upon monitor specifications (Same as predicate).
    Patient Use/Reuse: Reusable (same as predicate)Reusable (Same as predicate).
    Sterility: Non-sterile (same as predicate)Non-sterile (Same as predicate).
    Description of Patient Attachment (same as predicate)These devices attach to the gravid patient with elastic straps around the waist (Same as predicate).
    Cable Length: 10 feet (predicate varied)10 feet (predicate varied, e.g., Epic had 8 feet). This is a point of minor difference but not flagged as non-equivalent.
    Accessories (same as predicate)Transducer belts and ultrasonic gel (Same as predicate).
    Connector Design (same as predicate)Transducer cable connectors are color-coded and keyed to fit the appropriate fetal monitors (Same as predicate).
    Operational Characteristics (Pulsed Doppler)AIV FM10835 = Pulsed Doppler; AIV FM10836 = Pulsed Doppler (Same as predicate).
    Ultrasound Center Frequency (approx. 1.0 MHz)AIV FM10835 = 1.0 MHz; AIV FM10836 = 1.0 MHz (predicate 5700LAX = 1.151 MHz, EFU200-20 = 1.0 MHz). The AIV devices match one predicate directly and are close to the other.

    Study Information

    1. Sample size used for the test set and the data provenance:

      • The document does not specify a sample size for a test set. This was a 510(k) premarket notification for a replacement accessory, focusing on equivalence to predicate devices rather than a new clinical study with a test set of patients.
      • Data Provenance: Not applicable in the context of a clinical test set. The data presented is from bench testing and conformity to standards.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable. There is no mention of a test set with ground truth established by experts. The determination of substantial equivalence relies on comparison to existing, legally marketed devices and specified technical performance.
    3. Adjudication method for the test set:

      • Not applicable. No test set requiring adjudication is described.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. This is a medical device (transducer) for measuring physiological signals, not an AI-assisted diagnostic tool for human readers. Therefore, an MRMC study is not relevant or mentioned.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • No. This device is an accessory transducer, not a standalone algorithm. The "performance" assessment is based on physical and technical characteristics and its ability to function as a direct replacement for predicate transducers.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The "ground truth" in this context is the performance and specifications of the predicate devices (Corometrics, Epic, Hewlett Packard transducers) and established consensus standards (e.g., for biocompatibility, acoustic output limits). The AIV transducers aim to meet or be equivalent to these.
    7. The sample size for the training set:

      • Not applicable. This device is not an AI algorithm that requires a training set.
    8. How the ground truth for the training set was established:

      • Not applicable. As above, this device is not an AI algorithm.
    Ask a Question

    Ask a specific question about this device

    K Number
    K040819

    Validate with FDA (Live)

    Date Cleared
    2004-08-16

    (139 days)

    Product Code
    Regulation Number
    884.2740
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    These devices are intended to be used as replacement accessories to Corometrics Monitors, to measure fetal heart rate in the gravid patient.

    Device Description

    AIV's ultrasound (US) and tocodynamometer (TOCO) transducers are replacements for similar transducers manufactured by Corometrics and Hewlett Packard for their respective monitors. The AIV transducers are also a replacement for similar transducers manufactured by Epic for use on Corometrics and Hewlett Packard monitors. The US transducers are used to detect the fetal heart rate using Doppier shift technology. These transducers are intended to be a direct replacement for the Corometrics transducers.

    AI/ML Overview

    The provided document is a 510(k) Summary for a medical device (Transducers for Ultrasound Fetal Monitoring) and does not contain detailed information about a clinical study or specific acceptance criteria for performance metrics beyond general functional equivalence.

    However, I can extract the relevant information that is present and indicate where specific details you requested are not available in this document.

    Here's an attempt to answer your request based only on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Inferred from comparison to predicate)Reported Device Performance (from "Performance Data and Conclusions")
    Acoustic Output ≤ 20mW/cm² averageAcoustic output testing shows power is less than 20mW/cm² average.
    Assembly design equivalent to predicate deviceAIV assembly design is equivalent to predicate device assembly design.
    Biocompatibility (Conforms to consensus standards)AIV plastics have conformed to consensus standards relating to Biocompatibility.
    Perform as intended (measure fetal heart rate)Alv assembly design is equivalent to producted by devices perform as intended and are equivalent to predicate device assemblies.
    Does not raise new safety or effectiveness issues, nor alter fundamental technology of predicate devicesThese devices do not raise new issues of safety and effectiveness, nor do they alter the fundamental technology of the predicate devices.

    2. Sample size used for the test set and the data provenance

    • Sample Size for Test Set: Not specified. The document primarily focuses on demonstrating equivalence through design and testing of fundamental characteristics rather than a clinical trial with a defined test set of patients.
    • Data Provenance: Not specified. Given the nature of the tests (acoustic output, biocompatibility, assembly design comparison), it's likely laboratory-based. No patient data or clinical study data is mentioned.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. The "ground truth" for these types of engineering and material tests would be established through accredited laboratory methods and technical specifications, not expert clinical consensus.

    4. Adjudication method for the test set

    Not applicable. There's no indication of an adjudication method used. The "performance data" describes physical and material properties rather than clinical outcomes requiring expert review.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. This device is a passive transducer for fetal monitoring, not an AI-powered diagnostic tool. Therefore, an MRMC study or AI-assistance evaluation is not relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    No. This device is hardware (transducer) and does not involve an algorithm for standalone performance evaluation in the context of an AI device.

    7. The type of ground truth used

    • Acoustic Output: Likely established through standardized acoustic power measurement techniques and equipment.
    • Assembly Design: Established by expert engineering comparison against the predicate device's design and specifications.
    • Biocompatibility: Established by adherence to recognized consensus standards (e.g., ISO 10993 series) through laboratory testing.
    • Performance as Intended: Inferred equivalence based on the above tests and similarity to the predicate.

    8. The sample size for the training set

    Not applicable. This is not an AI/machine learning device, so there is no "training set."

    9. How the ground truth for the training set was established

    Not applicable (no training set).

    Ask a Question

    Ask a specific question about this device

    K Number
    K032789

    Validate with FDA (Live)

    Date Cleared
    2003-10-09

    (31 days)

    Product Code
    Regulation Number
    880.5725
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    These devices are intended to be used as replacement door assemblies for Baxter Flo-Gard 6201 and Baxter Flo-Gard® 6301 volumetric infusion pumps, for use to deliver drug solutions, enteral feedings and blood.

    Device Description

    AIV's pump door assemblies are replacements for similar doors manufactured by Baxter Healthcare Corporation for their Flo-Gard® 6201 and Flo-Gard® 6301 Volumetric Infusion Pumps. The door is part of a linear peristaltic pumping mechanism. The door assemblies are intended to be direct replacement for the predicate device door assembly.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and supporting study for the AIV Replacement Pump Head Door Assembly:

    Acceptance Criteria and Device Performance

    The core acceptance criterion for the AIV Replacement Pump Head Door Assembly is its performance in terms of flow volume accuracy.

    Acceptance CriteriaReported Device Performance
    Flow volume within 7% of programmed volume for a programmed flow rate"Bench Testing demonstrates that the AIV devices perform as intended and are equivalent to predicate device assemblies." (Implicitly, this means they meet the 7% tolerance, as the predicate device has the same performance criteria.)

    Study Information

    The provided document describes a bench testing study, but lacks detailed numerical results beyond the general statement of equivalence.

    1. Sample size for the test set and data provenance:

      • Sample Size: Not explicitly stated. The document refers to "Bench Testing" and "AIV devices," but does not provide the number of units tested.
      • Data Provenance: Not explicitly stated, but can be inferred as retrospective as it's a comparison to established predicate devices. The study was conducted by the manufacturer, American IV Products, Inc. (AIV). The country of origin of the data is likely USA, as the submitting company is based in Maryland.
    2. Number of experts used to establish the ground truth for the test set and their qualifications:

      • Not applicable. This was a bench test comparing the functional performance of the AIV device to its predicate. Ground truth in this context would likely be established by validated test equipment and calibration standards, not expert human assessment.
    3. Adjudication method for the test set:

      • Not applicable for this type of bench testing. Performance would be objectively measured using calibrated instruments against a defined specification.
    4. Multi-Reader Multi-Case (MRMC) comparative effectiveness study:

      • No, an MRMC comparative effectiveness study was not conducted. This device is a component of an infusion pump, and its performance is assessed via engineering bench tests, not human interpretation of medical cases.
    5. Standalone (algorithm-only) performance:

      • Yes, a standalone performance evaluation was done in the sense that the AIV device was tested on its own (or in place of the predicate device component) to verify its functional performance. This is an engineering test, not an algorithm-only test in the context of AI/ML.
    6. Type of ground truth used:

      • The ground truth for this study is the established performance specifications of the predicate device (Baxter Flo-Gard 6201 and 6301) and general engineering standards for flow rate accuracy. This is based on objective, quantifiable measurements.
    7. Sample size for the training set:

      • Not applicable. This is not an AI/ML device that requires a training set. The "design" and "materials" were presumably developed through engineering processes, prototyping, and testing, but not in the sense of a dataset for machine learning.
    8. How the ground truth for the training set was established:

      • Not applicable, as there is no training set in the AI/ML sense. The "ground truth" for the design of the AIV device was the existing, legally marketed predicate devices and their known performance characteristics.