510(k) Data Aggregation

    K Number
    K974356

    Validate with FDA (Live)

    Device Name
    SURGEON'S GLOVES
    Date Cleared
    1998-01-29

    (71 days)

    Product Code
    Regulation Number
    878.4460
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The surgeon's gloves are sterile gloves worn by surgeons, healthcare workers or similar personnel during surgical procedures to prevent contamination between the user and the patient.

    Device Description

    Surgeon's Gloves or Surgical Gloves. Anatomical in shape. Hand Specific (Left or Right). Available in sizes 6, 6½, 7, 7½, 8, 8½, 9. Made from Natural Rubber Latex, Sulphur, Zinc Oxide, ZDEC, ZDBC, Titanium Dioxide, Antioxidant (Styrenated Phenol), and Absorbable Corn Starch. Designed to comply with the physical measurements of ASTM D 3577.

    AI/ML Overview

    The provided document describes a 510(k) submission for Surgeon's Gloves (SUR-G GLOV®) and focuses on demonstrating substantial equivalence to predicate devices rather than proving performance against specific acceptance criteria for an AI-based device. Therefore, many of the requested elements for an AI device study (such as sample sizes for test/training sets, expert qualifications, adjudication methods, MRMC studies, standalone performance, and ground truth types) are not applicable and not present in this document.

    However, the document does contain information regarding the performance characteristics of the gloves, which can be interpreted as acceptance criteria based on the ASTM D 3577 standard.

    Here's the breakdown of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The device's performance is compared against the ASTM D 3577 (91) standard. This standard serves as the acceptance criteria for various physical characteristics of surgeon's gloves.

    CharacteristicsAcceptance Criteria (ASTM D 3577 (91))Reported Device Performance (SUR-G GLOV®)
    Width Size
    6$76 \pm 6$$76 \pm 3$
    $83 \pm 6$$83 \pm 3$
    7$89 \pm 6$$90 \pm 3$
    $95 \pm 6$$95 \pm 3$
    8$102 \pm 6$$103 \pm 3$
    $108 \pm 6$$109 \pm 3$
    9$114 \pm 6$$115 \pm 3$
    Length (min)265280
    Thickness
    Palm mm0.10$0.21 \pm 0.02$
    Finger mm0.10$0.23 \pm 0.02$
    Cuff mm0.10$0.21 \pm 0.02$
    Tensile Strength Min
    (Unaged)24 Mpa27 Mpa
    (Aged)18 Mpa24 Mpa
    Elongation @ break
    (Unaged)750% min830%
    (Aged)560% min800%
    Modulus @ 500 % elongation (max.)5.5 Mpa4.0 Mpa
    External powder levelNA (Not Available in ASTM standard)$80 \pm 25$ mg
    Internal powder levelNA (Not Available in ASTM standard)$90 \pm 25$ mg

    2. Sample sized used for the test set and the data provenance

    The document states "A test certificate of the above is in Appendix B." but Appendix B is not provided in the extracted text. Therefore, the specific sample sizes used for testing these performance characteristics are not available. The data provenance (country of origin, retrospective/prospective) is also not available from the provided text.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not Applicable. This is a physical device (surgical gloves), and its performance is measured against established engineering and material standards (ASTM D 3577), not by expert interpretation or ground truth derived from expert consensus.

    4. Adjudication method for the test set

    Not Applicable. As per point 3, the assessment is against a standard, not expert adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not Applicable. This is not an AI-based device, and therefore, no MRMC study or AI assistance is relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not Applicable. This is not an AI algorithm.

    7. The type of ground truth used

    The "ground truth" for the device's performance is adherence to the ASTM D 3577 standard for surgical gloves, which defines acceptable physical properties.

    8. The sample size for the training set

    Not Applicable. This is not an AI device that requires a training set. The gloves are manufactured and tested against specifications.

    9. How the ground truth for the training set was established

    Not Applicable. No training set is relevant for this device. The "ground truth" (i.e., the standard specifications) for manufacturing and testing are established by ASTM International.