Found 3 results

510(k) Data Aggregation

    K Number
    K063769

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2007-03-30

    (100 days)

    Product Code
    Regulation Number
    872.4880
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Alpha-Bio Bone Fixation Screw System® is used to stabilize and fixate bone grafts, bone filling materials, and / or barrier membranes used for regeneration of bone in the oral cavity. The system includes titanium screws 1.2mmd and 1.6mmd diameter, and drill instrument for fixating the screw to the bone in the oral cavity at regeneration procedure.

    Device Description

    The Alpha-Bio Bone Fixation Screw System consists of titanium alloy Gr-5 screws with 1.2mmd and 1.6mmd diameter, and with 4mm-12mm lengths. This system includes drill instrument for fixating the screw to the bone in the oral cavity at regeneration procedure

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device, the Alpha-Bio Tec® Bone Fixation Screw System. This document focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study to prove the device meets specific acceptance criteria in terms of performance metrics like sensitivity, specificity, or accuracy.

    Therefore, the requested information regarding acceptance criteria, device performance, sample sizes for test/training sets, ground truth establishment, expert qualifications, adjudication methods, MRMC studies, or standalone performance studies is largely not applicable to this type of submission.

    Here's an explanation based on the provided text:

    1. Table of acceptance criteria and reported device performance: Not applicable. For a 510(k) submission, especially for a device like a bone fixation screw system, "acceptance criteria" are generally related to demonstrating substantial equivalence in terms of materials, intended use, and mechanical performance, rather than clinical performance metrics. The document states "No performance standards have been established for such devices under Section 514 of the Federal Food, Drug, and Cosmetic Act." and instead refers to compliance with recognized standards for biocompatibility, materials, and resistance (see point 1 below for details).

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective): Not applicable. This submission does not describe a clinical study with a "test set" in the context of AI/diagnostic device evaluation. The "performance and safety testing" mentioned (Section J) refers to bench testing to demonstrate physical properties and biocompatibility, not clinical outcomes from patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience): Not applicable. No clinical "ground truth" or expert review of clinical data is described in this submission.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable. No clinical "test set" requiring adjudication is described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a bone fixation screw system, not an AI-powered diagnostic tool. MRMC studies are not relevant here.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This device is a physical surgical implant, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable. The "ground truth" for a device like this relates to its physical and biological properties.

    8. The sample size for the training set: Not applicable. There is no AI component or "training set" for this device.

    9. How the ground truth for the training set was established: Not applicable.


    However, the document does contain information relevant to the device's "acceptance" based on recognized standards and equivalence to a predicate device:

    Here's a breakdown of the relevant information from the provided text, re-framed to address the spirit of the request where possible:

    1. Table of "Acceptance Criteria" (Standards Compliance) and Reported Device Performance (as implied by equivalence):

    "Acceptance Criteria" (Standards Complied With)Reported Device Performance (Implied by Equivalence and Testing)
    ISO 7405:1997: Dentistry Preclinical Evaluation of Biocompatibility of Medical Devices Used in Dentistry - Test Methods for DentalBiocompatibility: The device is considered biocompatible, as demonstrated by compliance with ISO 7405:1997. Specific test results are not detailed in this summary but are implied to be satisfactory.
    ASTM F136-02a: Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications (UNS R56401)Material Composition (Screws): The titanium alloy screws conform to this standard for surgical implant applications, ensuring appropriate material properties for strength, corrosion resistance, and biocompatibility.
    ASTM F1350-02: Standard Specification for Wrought 18 Chromium-14 Nickel-2.5 Molybdenum Stainless Steel Surgical Fixation Wire (UNS S31673)Material Composition (Implied for other components/instruments or previous versions): Although the screws are Titanium Gr-5, compliance with this standard suggests other components (or potentially the drill instrument, or for general material quality understanding for the submission) meet standards for surgical stainless steel, ensuring appropriate material properties. (Note: The document explicitly states screws are Titanium Gr-5, so this standard might apply to other system components, though not explicitly stated for the screws themselves).
    ISO 13402:1995: Surgical and dental hand instruments -- Determination of resistance against autoclaving, corrosion and thermal exposureInstrument Durability and Sterilization Resistance: The surgical and dental hand instruments (e.g., drill instrument) in the system meet standards for resistance to autoclaving, corrosion, and thermal exposure, ensuring their reusability and integrity during sterilization processes.
    UL 544 (1998): Standard for Medical and Dental Equipment - Ed. 4.0Electrical Safety (if applicable): While a bone screw system often doesn't have electrical components, if the drill instrument or associated equipment is electrical, compliance with UL 544 ensures electrical safety standards are met.
    Substantial Equivalence to Predicate Devices (Straumann K050515; K011698)Overall Safety and Effectiveness: The device is demonstrated to be substantially equivalent to previously cleared predicate devices in terms of intended use, indications for use, technological characteristics, performance, and user interface. This implies a comparable safety and effectiveness profile to already-marketed devices. Specific "safety and performance testing" (Section J) was conducted, presumably bench tests to confirm physical and mechanical equivalency.

    Summary of the Study (or Basis for Acceptance):

    The "study" presented here is a 510(k) Premarket Notification designed to demonstrate substantial equivalence to legally marketed predicate devices, not to prove meeting specific clinical performance acceptance criteria in the manner of an AI diagnostic device.

    • Basis for Acceptance: The device is accepted because it has been shown to be substantially equivalent to predicate devices (Straumann K050515; K011698) in terms of intended use, indications for use, technological characteristics (Titanium alloy Gr-5, specific diameters and lengths), and performance. This equivalence is supported by compliance with recognized international and ASTM standards for biocompatibility, material composition, and instrument durability, as well as general "safety and performance testing" (likely bench testing) that confirmed comparable physical and mechanical properties.
    • Key points from the submission:
      • Device Description: Consists of titanium alloy Gr-5 screws (1.2mmd and 1.6mmd diameter, 4mm-12mm lengths) and a drill instrument for fixation in oral cavity bone regeneration procedures.
      • Intended Use/Indications for Use: Stabilize and fixate bone grafts, bone filling materials, and/or barrier membranes for bone regeneration in the oral cavity.
      • Predicate Devices: Straumann K050515; K011698.
      • Equivalence Justification: No unique applications, indications, materials, or specifications compared to the predicate. Intended use and indications were previously cleared for the predicate. Technical characteristics are similar. Safety and performance testing (bench-level) was conducted.
      • Conclusion: The Alpha-Bio Tec® is substantially equivalent to its predicate devices and "raises no new safety and/or effectiveness issues."
    Ask a Question

    Ask a specific question about this device

    K Number
    K063770

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2007-03-26

    (96 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Alpha Bio Orthodontic Fixation Screw System® are intended to provide a fixed anchorage point for attachment of orthodontic appliances to facilitate the orthodontic movement of teeth. It is used temporarily and is removed after orthodontic treatment has been completed. Screws are intended for single use only.

    Device Description

    Alpha Bio Orthodontic Fixation Screw System® OFS, OFSR consists of titanium alloy Gr-5 mini thread self tapping implants, are available with 1.6 and 2mm diameter and 8, 10, 12 length.

    AI/ML Overview

    This submission (K063770) is for a dental device, specifically the Alpha-Bio Tec® Orthodontic Fixation Screw System. As such, the information provided focuses on demonstrating substantial equivalence to a predicate device through adherence to recognized performance standards and material specifications, rather than clinical performance studies with acceptance criteria, ground truth, or human reader comparisons typically associated with AI/software devices.

    Therefore, most of the requested information regarding acceptance criteria, study design elements (sample size, data provenance, expert adjudication, MRMC, standalone performance), and ground truth establishment is not applicable to this type of device submission.

    Here's what can be extracted and what cannot:

    1. A table of acceptance criteria and the reported device performance:

    • Acceptance Criteria: No specific numerical "acceptance criteria" for clinical performance are stated in this document. Instead, acceptance is based on compliance with recognized standards and demonstration of substantial equivalence to a predicate device. The performance is assessed against these standards and the predicate.
    • Reported Device Performance: The document states that "Safety and performance testing" was conducted and that "The Alpha-Bio Tec® System is substantially equivalent to its predicate devices as cited above and raises no new safety and/or effectiveness issues." However, the specific results of these performance tests are not detailed in this summary. Instead, compliance with the following standards is asserted:
    Acceptance Criteria (Compliance with Standards/Predicate)Reported Device Performance (Claimed Compliance)
    ISO 7405:1997 (Biocompatibility)Complies
    F136-02a: 2004 (Titanium Alloy Spec)Complies
    ASTM F1350-02 (Stainless Steel Wire Spec)Complies
    ISO 13402:1995 (Resistance against autoclaving, corrosion etc.)Complies
    UL 544 (1998) (Medical and Dental Equipment)Complies
    Substantial Equivalence to Osteomed Orthodontic Screw System L.P. (K031936)Alpha-Bio Tec® is substantially equivalent in intended use, indications, technological characteristics, performance, and user interface.

    2. Sample size used for the test set and the data provenance:

    • Not Applicable. This submission does not describe a "test set" in the context of a clinical performance study with patient data. The evaluation relies on laboratory testing against standards and comparison to a predicate device's established safety and effectiveness.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not Applicable. This type of "ground truth" is not established for this device, as it's not an AI/diagnostic software.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not Applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not Applicable. This is not an AI-assisted device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not Applicable. This is not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • Not Applicable. The "ground truth" for this device's acceptance is its physical and chemical properties meeting established engineering and biocompatibility standards, and its functional equivalence to a legally marketed predicate device.

    8. The sample size for the training set:

    • Not Applicable. This is not a machine learning or AI device that requires a training set.

    9. How the ground truth for the training set was established:

    • Not Applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K063364

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2007-03-14

    (127 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Alpha-Bio Dental Implant System® is indicated for use in surgical and restorative applications for placement in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's chewing function. The Alpha-Bio Dental Implant System® is indicated also for immediate loading when good primary stability is achieved and with appropriate occlusal loading.

    Device Description

    The Alpha-Bio Dental Implant System® consists of one and two stage endosseous form dental implants, internal and external hexagonal; internal octagonal hexagonal; one piece implants system; cover screws and healing caps; abutment systems and superstructures; surgical instruments.

    AI/ML Overview

    The provided text focuses on regulatory submissions and approvals for the Alpha-Bio Tec® Dental Implant System and does not contain information about acceptance criteria or a study proving that the device meets specific performance criteria.

    Therefore, I cannot provide the requested table and details as the information is not present in the given document. The document primarily identifies the device, its intended use, predicate devices, and regulatory compliance with established standards (like ISO and ASTM) rather than specific performance studies with acceptance criteria.