Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K170623

    Validate with FDA (Live)

    Date Cleared
    2018-02-27

    (363 days)

    Product Code
    Regulation Number
    864.7470
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ACE Hemoglobin A lc (HbA lc) Reagent is intended for the quantitative determination of percent hemoglobin A lc in venous whole blood collected in K2-EDTA tubes using the ACE Axcel® Clinical Chemistry Systems. This test is intended for use in clinical laboratories and physician office laboratories to monitor long term blood glucose control in individuals with diabetes mellitus. For in vitro diagnostic use only.

    Device Description

    The ACE Hemoglobin A1c (HbA1c) Reagent assay requires a pretreatment step of denaturation of the whole blood samples, which is performed off-line. The red blood cells in the sample are lysed by the Hemoglobin Denaturant and the hemoglobin chains are hydrolyzed. For determination of HbA1c, a latex agglutination inhibition assay is used. In the absence of HbA1c in the sample, the synthetic polymer containing the immunoreactive portion of HbA 1c Agglutinator Reagent will agglutinate with the antibody-coated microparticles in the HbA1c Antibody Reagent. The presence of HbA1c in the blood sample competes for the antibody binding sites and inhibits agglutination. The increase in absorbance, monochromatically at 692 mm, is inversely proportional to the HbA1c present in the sample. For the determination of total hemoglobin, all hemoglobin derivatives in the sample are converted to alkaline hematin. The reaction produces a green colored solution. which is measured bichromatically at 573 nm/692 nm. The intensity of color produced is directly proportional to the total hemoglobin concentration in the sample. The concentrations of both HbA 1 c and total hemoglobin are measured, the ratio is calculated, and the result reported as percent HbA1c.

    AI/ML Overview

    1. Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the ACE Hemoglobin A1c (HbA1c) Reagent device are implicitly established by demonstrating comparable performance to the predicate device, the DCA 2000+ System for Hemoglobin A1c (K951361), across a range of analytical performance characteristics. While explicit numerical acceptance criteria are not always stated, the study aims to show that the new device's performance aligns with acceptable standards for HbA1c measurement in clinical diagnostics.

    Here's a table summarizing the reported device performance:

    Performance CharacteristicAcceptance Criteria (Implied by Predicate & Clinical Relevance)Reported Device Performance (ACE Alera & ACE Axcel)
    Limit of Quantitation (LoQ)Clinically relevant lower limit for HbA1c measurement.ACE Alera: 2.5% HbA1c ACE Axcel: 2.5% HbA1c
    Linearity (HbA1c)Strong correlation (r² close to 1) and a regression equation with a slope near 1 and y-intercept near 0 across the measuring range, indicating accurate and proportional measurement of HbA1c.ACE Alera (Range 2.7%-13.0% HbA1c): y = 0.987x + 0.3, r² = 0.9948 ACE Axcel (Range 2.4%-13.1% HbA1c): y = 0.954x + 0.3, r² = 0.9936
    Linearity (Total Hemoglobin)Strong correlation (r² close to 1) and a regression equation with a slope near 1 and y-intercept near 0 across the measuring range of total hemoglobin.ACE Alera (Range 1.4-22.2 g/dL): y = 1.006x + 0.10, r² = 0.9978 ACE Axcel (Range 1.2-21.8 g/dL): y = 0.997x + 0.20, r² = 0.9964
    Precision (Within-Run %CV)Low %CV for different HbA1c levels, indicating consistent results within a single analytical run. Typically, <5% for routine assays.ACE Alera (In-house): Range from 1.3% to 2.8% ACE Axcel (In-house): Range from 1.6% to 2.4% ACE Alera (POLs): Range from 0.9% to 2.7% ACE Axcel (POLs): Range from 1.1% to 2.2%
    Precision (Total %CV)Low total %CV, indicating good reproducibility over time and across different runs. Typically, <5% for routine assays.ACE Alera (In-house): Range from 2.4% to 3.1% ACE Axcel (In-house): Range from 2.4% to 2.9% ACE Alera (POLs): Range from 1.6% to 3.1% ACE Axcel (POLs): Range from 1.8% to 4.3%
    Comparative Analysis (In-house vs. Predicate)Strong correlation (correlation coefficient > 0.97) and regression parameters (slope near 1, intercept near 0) indicating agreement with the predicate device. Confidence intervals for slope should include 1 and for intercept should include 0.ACE Alera (n=101, Range 3.2-12.8% HbA1c): y = 0.979x + 0.05, Correlation = 0.9839, SE = 0.32, CI slope (0.944-1.015), CI intercept (-0.21-0.31) ACE Axcel (n=102, Range 2.5-12.8% HbA1c): y = 0.983x - 0.03, Correlation = 0.9832, SE = 0.34, CI slope (0.948-1.019), CI intercept (-0.29-0.24)
    Comparative Analysis (POLs vs. Predicate)Similar strong correlation and regression parameters to in-house comparative analysis, demonstrating robust performance in typical clinical laboratory settings.ACE Alera (POLs): Correlation range 0.9892 to 0.9945. Slopes generally close to 1 (e.g., 0.967, 0.984, 0.981). Intercepts generally close to 0 (e.g., 0.34, -0.02, -0.09). ACE Axcel (POLs): Correlation range 0.9885 to 0.9960. Slopes generally close to 1 (e.g., 1.000, 0.993, 0.980). Intercepts generally close to 0 (e.g., -0.28, -0.12, 0.02).
    Analytical SpecificityNo significant interference from common endogenous substances or therapeutic compounds within specified concentrations.Interferents: Bilirubin (≤ 53 mg/dL), Triglycerides (≤ 1100 mg/dL), Ascorbic Acid (≤ 6 mg/dL), Sodium Fluoride (≤ 1200 mg/dL), Acetaldehyde (≤ 100 mg/dL) showed no significant interference.
    Cross-ReactivityNo significant interference from common hemoglobin variants or modified hemoglobins.Non-Interfering: Acetylated Hb (2000 mg/dL), Carbamylated Hb (2000 mg/dL), Labile A1c (1440 mg/dL), Non-glycated Hb (HbA0) (1725 mg/dL), HbA1a+b fraction (100 mg/dL) showed no significant interference. Known Interferences (within certain concentrations): HbD (≤ 36.3%), HbE (≤ 22.5%) showed no significant interference. High HbF (> 10.1%), High HbC (> 14.0%), and High HbS (> 17.1%) will result in inaccurate HbA1c results. These interferences are acknowledged and will be included in labeling.
    Measuring RangeConsistent with or broader than the predicate device to coverclinically relevant HbA1c values.Candidate Device: 2.7 – 13.0% HbA1c Predicate Device: 2.5 – 14.0% HbA1c (The candidate device's range is slightly narrower at the upper end but still covers the critical clinical range).

    Study Details:

    2. Sample Size and Data Provenance for Test Set:

    • Linearity: 11 samples were used for both HbA1c and Total Hemoglobin linearity studies, run in 4 replicates each, for both the ACE Alera and ACE Axcel systems.
    • Precision (In-house): 4 samples (A, B, C, D) were tested, but the number of runs/replicates to calculate SD and %CV is not explicitly stated in the table. Typically, precision studies involve multiple replicates over several days.
    • Precision (Physician Office Labs - POLs): 4 samples were tested across 3 POLs for each instrument (ACE Alera and ACE Axcel). Similar to in-house, the specific number of runs/replicates per POL for SD and %CV calculation is not detailed.
    • Comparative Analysis (In-house): 101 samples for ACE Alera and 102 samples for ACE Axcel were compared against the predicate device (DCA 2000+).
    • Comparative Analysis (POLs): 50 samples per POL for a total of 150 samples for ACE Alera, and 52 samples for one POL and 50 samples for the other two POLs (total 152 samples) for ACE Axcel were compared against the predicate device (DCA 2000+).
    • Analytical Specificity/Cross-Reactivity: The number of samples for these studies is not explicitly stated, but typically involves spiking known concentrations of interferents into samples and measuring the effect.
    • Data Provenance: The studies were conducted in-house by Alfa Wassermann Diagnostic Technologies, LLC, and in external Physician Office Laboratories (POLs). Given the nature of performance validation for a diagnostic device, these studies are prospective, as samples are analyzed using the new device and compared against a reference method or predicate. The "country of origin of the data" is implicitly the United States, where the manufacturer and the POLs are located.

    3. Number of Experts and their Qualifications for Ground Truth:

    The document does not mention the use of "experts" in the traditional sense (e.g., radiologists, pathologists) to establish ground truth for the test set. For an in vitro diagnostic device measuring a quantitative analyte like HbA1c, the ground truth is typically established by:

    • Reference Methods: Highly accurate and precise analytical methods, often traceable to international standards (e.g., NGSP, IFCC), which are considered the "gold standard" for measuring the analyte.
    • Predicate Devices: Comparison to a legally marketed device that has already established its safety and effectiveness.

    In this case, the ground truth for the comparative studies was derived from the predicate device (DCA 2000+ System for Hemoglobin A1c), which is itself NGSP Certified and traceable to IFCC reference materials.

    4. Adjudication Method for the Test Set:

    Not applicable. Adjudication methods (like 2+1, 3+1) are typically used in clinical studies involving interpretation of medical images or complex diagnostic assessments by human readers, where discrepancies between readers need to be resolved. For a quantitative in vitro diagnostic device, the ground truth is established analytically through reference methods or predicate comparison, not through expert consensus requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    Not applicable. MRMC studies are generally performed for image-based diagnostic aids or other devices where human interpretation plays a significant role in the diagnostic outcome. This document describes the analytical performance validation of an in vitro diagnostic reagent, which is a quantitative measurement, not an interpretative task for human readers in the context of an MRMC study. Therefore, no effect size of human readers improving with AI vs. without AI assistance is relevant or reported.

    6. Standalone (Algorithm Only) Performance Study:

    Yes, this entire submission focuses on the standalone performance of the ACE Hemoglobin A1c (HbA1c) Reagent when used with the ACE Alera® and ACE Axcel® Clinical Chemistry Systems. The studies presented (linearity, precision, comparative analysis, specificity, cross-reactivity) all evaluate the direct analytical performance of the device itself, without human intervention for interpretation beyond standard laboratory procedures for operating the instrument and processing samples.

    7. Type of Ground Truth Used:

    The ground truth used for the comparative analysis studies was the predicate device, the DCA 2000+ System for Hemoglobin A1c. The document explicitly states that the DCA Hemoglobin A1c test method is National Glycohemoglobin Standardization Program (NGSP) Certified and is traceable to International Federation of Clinical Chemistry (IFCC) reference materials and test methods. This indicates that the predicate device serves as a highly standardized and accepted reference for HbA1c measurement.

    8. Sample Size for the Training Set:

    The document does not explicitly mention a "training set" in the context of machine learning or AI models. This device is a diagnostic reagent kit for a clinical chemistry system, not a software algorithm that requires a separate training phase with a distinct dataset. The performance data presented are for the validation of the finalized device.

    9. How the Ground Truth for the Training Set Was Established:

    As there is no "training set" in the context of this device being a reagent for a clinical chemistry system, this question is not applicable. The device's design and formulation would have been developed through internal R&D, likely using internal validation and optimization experiments, but these are not typically referred to as a "training set" with established ground truth in the same way as for AI/ML models. The ground truth for the validation of the device's performance (as described above) was established by comparison to the NGSP/IFCC-traceable predicate device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K131351

    Validate with FDA (Live)

    Date Cleared
    2013-08-15

    (97 days)

    Product Code
    Regulation Number
    862.1050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ACE Alkaline Phosphatase Reagent is intended for the quantitative determination of alkaline phosphatase activity in serum and lithium heparin plasma using the ACE, ACE Alera, and ACE Axcel Clinical Chemistry Systems. Measurements of alkaline phosphatase are used in the diagnosis and treatment of liver, bone, parathyroid, and intestinal diseases. This test is intended for use in clinical laboratories and physician office laboratories. For in vitro diagnostic use only.

    The ACE Amylase Reagent is intended for the quantitative determination of α-amylase activity in serum and lithium heparin plasma using the ACE, ACE Alera, and ACE Axcel Clinical Chemistry Systems. Amylase measurements are used primarily for the diagnosis and treatment of pancreatitis (inflammation of the pancreas). This test is intended for use in clinical laboratories and physician office laboratories. For in vitro diagnostic use only.

    The ACE ALT Reagent is intended for the quantitative determination of alanine aminotransferase activity in serum and lithium heparin plasma using the ACE, ACE Alera, and ACE Axcel Clinical Chemistry Systems. Alanine aminotransferase measurements are used in the diagnosis and treatment of certain liver diseases (e.g., viral hepatitis and cirrhosis) and heart diseases. This test is intended for use in clinical laboratories and physician office laboratories. For in vitro diagnostic use only.

    The ACE AST Reagent is intended for the quantitative determination of aspartate aminotransferase activity in serum and lithium heparin plasma using the ACE, ACE Alera, and ACE Axcel Clinical Chemistry Systems. Measurements of aspartate aminotransferase are used in the diagnosis and treatment of certain types of liver and heart disease. This test is intended for use in clinical laboratories and physician office laboratories. For in vitro diagnostic use only.

    Device Description

    In the ACE Alkaline Phosphatase Reagent assay, alkaline phosphatase catalyzes the hydrolysis of colorless p-nitrophenyl phosphate to p-nitrophenol and inorganic phosphate. In an alkaline solution (pH 10.5), p-nitrophenol is in the phenoxide form and has a strong absorbance at 408 nm. The rate of increase in absorbance, monitored bichromatically at 408 nm/486 nm, is directly proportional to the alkaline phosphatase activity in the sample.

    In the ACE Amylase Reagent assay, α-amylase hydrolyzes the 2-chloro-p-nitrophenyl-α-D-maltotrioside substrate to release 2-chloro-p-nitrophenol and form 2-chloro-p-nitrophenyl-α-D-maltoside, maltotriose and glucose. The rate of increase in absorbance, monitored bichromatically at 408 nm/ 647 nm, is directly proportional to the α-amylase activity in the sample.

    In the ACE ALT Reagent assay, alanine aminotransferase converts the L-alanine and α-ketoglutarate substrates in the reagent to L-glutamate and pyruvate, respectively. Lactate dehydrogenase (LDH) catalyzes the oxidation of the reduced cofactor to the cofactor. The rate of conversion of the reduced cofactor to the cofactor can be determined by monitoring the decrease in absorbance bichromatically at 340 nm/647 nm. This rate of conversion from the reduced cofactor to the cofactor is a function of the activity of ALT in the sample.

    In the ACE AST Reagent assay, aspartate aminotransferase converts the L-aspartate and α-ketoglutarate in the reagent to oxaloacetate and L-glutamate, respectively. The oxaloacetate undergoes reduction, with concurrent oxidation of NADH to NAD+ in the malate dehydrogenase-catalyzed indicator reaction. NADH absorbs strongly at 340 nm, whereas NAD+ does not. Therefore, the rate of conversion of NADH to NAD+ can be determined by monitoring the decrease in absorbance bichromatically at 340 nm/647 nm. This rate of conversion from NADH to NAD+ is a function of the activity of AST in the sample. Lactate dehydrogenase is added to prevent interference from endogenous pyruvate, which is normally present in blood.

    AI/ML Overview

    Here's an analysis of the provided information regarding the acceptance criteria and study for the ACE reagents:

    Summary of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for these in vitro diagnostic reagents (ALP, Amylase, ALT, AST) appear to be primarily demonstrated through comparisons with predicate devices and comprehensive performance characteristics like precision, linearity, and interference. The documentation focuses on demonstrating that the new devices perform equivalently to the existing predicate devices and meet established performance expectations for clinical chemistry assays.

    1. Table of Acceptance Criteria and Reported Device Performance

    Since this document describes multiple reagents and doesn't explicitly state pass/fail acceptance values for each performance metric, I will summarize the demonstrated performance and what can be inferred as the "acceptance criteria" (i.e., that the results are comparable to established predicate device performance and within acceptable clinical ranges).

    Performance MetricAcceptance Criteria (Inferred)Reported Device Performance
    PrecisionLow total CV% (generally < 5-10% for clinical assays, often lower for high-precision analytes). Consistent across different instruments (ACE, Alera, Axcel) and sample types (serum, plasma).ALP (Serum): Low: 1.9-2.3% Total CV; Mid: 1.0-1.3% Total CV; High: 0.7-1.7% Total CV.ALP (Plasma): Low: 3.7-5.1% Total CV; Mid: 3.3-4.0% Total CV; High: 2.9-3.1% Total CV.Amylase (Serum): Low: 2.1-3.4% Total CV; Mid: 1.2-1.4% Total CV; High: 1.2-1.9% Total CV.Amylase (Plasma): Low: 3.2-6.0% Total CV; Mid: 1.5-1.6% Total CV; High: 1.6-2.1% Total CV.ALT (Serum): Low: 3.3-4.0% Total CV; Mid: 1.6-2.0% Total CV; High: 1.5-1.9% Total CV.ALT (Plasma): Low: 4.0-4.7% Total CV; Mid: 1.0-1.4% Total CV; High: 0.7-1.4% Total CV.AST (Serum): Low: 3.3-5.7% Total CV; Mid: 1.5-1.8% Total CV; High: 1.3-1.6% Total CV.AST (Plasma): Low: 3.7-5.4% Total CV; Mid: 1.1-1.4% Total CV; High: 1.3-1.5% Total CV.POL Precision (ACE & Alera): Total CVs generally within similar ranges, demonstrating acceptable performance in Point of Care settings. Example: ALP Low on ACE ranges from 2.5-4.1% Total CV across POL sites.
    Matrix Comparison (Serum vs. Plasma)Slopes close to 1.0, intercepts close to 0, and correlation coefficients (R) close to 1.0 (e.g., >0.98 or 0.99) with narrow confidence intervals, indicating interchangeability of sample types.ALP: Slopes 0.983-1.017, Intercepts -6.5 to -8.3, Correlations 0.9952-0.9982.Amylase: Slopes 0.977-0.994, Intercepts -1.76 to 1.7, Correlations 0.9994-0.9996.ALT: Slopes 0.985-1.003, Intercepts -3.35 to -3.6, Correlations 0.9986-0.9994.AST: Slopes 0.998-1.006, Intercepts 0.3 to 1.5, Correlations 0.9993-0.9998. All indicate a strong agreement between serum and plasma samples.
    Method Comparison (vs. In-House ACE and POL sites)Slopes close to 1.0, intercepts close to 0, and correlation coefficients (R) close to 1.0 (e.g., >0.98 or 0.99) with narrow confidence intervals, indicating consistency across different instruments and sites.In-House ACE vs. POL ACE:ALP: Slopes 0.977-0.989, Intercepts -9.5 to -2.8, Correlations 0.9987-0.9997.• AMY: Slopes 0.970-0.974, Intercepts 1.5-3.9, Correlations 0.9995-0.9998.• ALT: Slopes 0.982-1.021, Intercepts -4.7 to -2.3, Correlations 0.9978-0.9993.• AST: Slopes 0.992-1.019, Intercepts -0.6 to 2.4, Correlations 0.9989-0.9994.In-House ACE vs. POL Alera:ALP: Slopes 0.997-1.029, Intercepts -6.6 to -4.1, Correlations 0.9986-0.9992.• AMY: Slopes 0.960-1.010, Intercepts 3.0-5.8, Correlations 0.9991-0.9995.• ALT: Slopes 0.970-1.019, Intercepts -3.5 to 2.4, Correlations 0.9977-0.9986.• AST: Slopes 1.004-1.040, Intercepts 0.5-1.8, Correlations 0.9992-0.9995.All indicate strong agreement between different sites and initial in-house testing, demonstrating substantial equivalence.
    Detection Limits (LoB, LoD, LoQ)Values below the clinical reference ranges and suitable for detecting low levels of analytes.ACE Alera (Approximate):ALP: LoB 2.8, LoD 0.9, LoQ 4.8Amylase: LoB 0.2, LoD 3.3, LoQ 5.6ALT: LoB 1.6, LoD 4.8, LoQ 4.1AST: LoB 2.2, LoD 3.1, LoQ 3.3
    LinearityCorrelation coefficient (R^2) close to 1.0 (e.g., >0.99) over the specified measuring range, with slopes near 1 and intercepts near 0 for the regression equation.ACE Alera: ALP: Linear to 1400 U/L, R^2 = 0.9993Amylase: Linear to 1900 U/L, R^2 = 0.9974ALT: Linear to 480 U/L, R^2 = 0.9992AST: Linear to 450 U/L, R^2 = 0.9992
    InterferencesNo significant interference at stated concentrations of common interferents (Icterus, Hemolysis, Lipemia, Ascorbic Acid).The document lists the tested concentrations of interferents (e.g., Icterus up to 70.6 mg/dL for ALP, Hemolysis up to 500 mg/dL for ALT, Lipemia up to 1000 mg/dL for ALP/Amylase, Ascorbic Acid 6 mg/dL for all). The implication, by inclusion in the performance data without negative remarks, is that these levels did not cause unacceptable interference.

    2. Sample Size Used for the Test Set and Data Provenance

    • Precision (Serum vs. Plasma):
      • In-House: Each dataset (low, mid, high for serum and plasma) involved "n=20" (number of replicates, likely over multiple days, contributing to within-run and total precision calculations).
      • POL Precision (ACE & Alera): For each analyte (ALP, AMY, ALT, AST) and each POL site (POL 1, POL 2, POL 3), there were 2 to 3 sample levels (Low, Mid, High), with a reported "n" for each (e.g., n=24 for ALT/AST in initial in-house, but the POL tables don't explicitly state the 'n' for each specific mean/SD/CV, implying a standard number of replicates as per precision studies).
    • Matrix Comparison (Serum vs. Plasma):
      • ALP: ACE (108 pairs), ACE Alera (108 pairs), ACE Axcel (62 pairs).
      • Amylase: ACE (104 pairs), ACE Alera (101 pairs), ACE Axcel (52 pairs).
      • ALT: ACE (54 pairs), ACE Alera (52 pairs), ACE Axcel (56 pairs).
      • AST: The number of pairs for AST in the serum vs. plasma matrix comparison is not explicitly stated in the provided snippet. However, based on the pattern of other analytes, it would likely be similar (e.g., 50+ pairs).
    • Method Comparison (In-House vs. POL Sites):
      • ALP: 49-50 samples per site.
      • Amylase: 51 samples per site.
      • ALT: 44-49 samples per site.
      • AST: 50 samples per site.
    • Linearity: Not explicitly stated as an "n" for samples, but rather as "low level tested," "upper level tested," and "linear to" values, which typically involve preparing a dilution series from a high concentration sample.
    • Data Provenance: The studies are labeled "In-House" and "POL" (Point of Care). This suggests:
      • Country of Origin: Likely the USA, given the FDA 510(k) submission.
      • Retrospective or Prospective: These types of performance studies for IVDs are typically prospective, with samples analyzed specifically for the study. The method comparison data often uses a mix of native patient samples and spiked samples to cover the measuring range.

    3. Number of Experts Used to Establish Ground Truth and Their Qualifications

    This document describes the performance of IVD reagents on clinical chemistry systems. The "ground truth" here is not subjective, human interpretation (like in imaging AI), but rather the quantitative measurement of analytes.

    • Number of Experts: Not applicable in the context of IVD reagent performance. The "ground truth" is established by the analytical method itself, or by comparison to a recognized reference method or a legally marketed predicate device.
    • Qualifications of Experts: Not applicable. The "experts" would be qualified laboratory professionals operating the instruments and performing the biochemical assays according to established protocols.

    4. Adjudication Method for the Test Set

    Not applicable. As described above, the "truth" for these quantitative measurements is derived directly from the biochemical reactions and instrument readings, not subjective human judgment requiring adjudication. The predicate device's established performance serves as a comparative benchmark.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. This is a submission for in vitro diagnostic reagents, not an AI-assisted diagnostic device that involves human readers interpreting images or complex data. Therefore, an MRMC study is not relevant.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, in essence, the performance data presented is "standalone" in the context of the device's function. The ACE reagents, when used on the ACE, ACE Alera, and ACE Axcel Clinical Chemistry Systems, operate as an automated system to quantify the target analytes. The performance metrics (precision, linearity, method comparison, interferences) reflect the intrinsic analytical performance of the regent-analyzer combination without human intervention influencing the measurement itself. Human operators are involved in sample loading, quality control, and result review, but not in directly influencing the quantitative output in a way that would require a human-in-the-loop comparison for algorithm performance.

    7. The Type of Ground Truth Used

    The "ground truth" in this context is established by:

    • Comparison to Predicate Devices: The primary method is demonstrating substantial equivalence to previously cleared devices (K113253, K931786, K930104, K113436, K113382). This means the new reagents provide results that are analytically comparable to those already accepted by the FDA.
    • Expected Analytical Performance: Meeting industry-standard requirements for precision (low CV%), accuracy (linearity, inter-instrument/site agreement via regression analysis), and specificity (minimal interference).
    • Expected Values/Ranges: The devices are expected to produce results that align with established "expected values" for healthy individuals.

    8. The Sample Size for the Training Set

    Not applicable. These are chemical reagents for quantitative diagnostic tests, not machine learning algorithms that require a "training set" in the conventional sense. The "training" for such systems involves analytical validation experiments to define reagent stability, reaction kinetics, and instrument parameters.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable for the same reason as point 8. The "ground truth" for developing and validating these reagents is based on fundamental principles of analytical chemistry, biochemical reactions, and extensive internal testing to ensure the reagents perform as intended within the specified analytical parameters.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1