510(k) Data Aggregation

    K Number
    K251320

    Validate with FDA (Live)

    Device Name
    XTRA4
    Manufacturer
    Date Cleared
    2025-09-11

    (135 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    XTRA4 is a corneal storage solution for storage of human cornea suitable for keratoplasty for up to 14 days under refrigeration (2-8°C).

    Device Description

    XTRA4 (ref. EXT4 001-01) is a sterile hypothermic corneal storage medium intended for human corneal storage between 2 and 8°C for up to 14 days. It is designed as a single use 20 mL solution in glass vials. XTRA4 is intended to be used only by physicians or highly skilled personnel such as eye bank operators. The preservation of donor cornea is traditionally performed by storing them at temperatures between 2°C and 8°C in a storage medium containing deswelling agent(s).

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    DEN200063

    Validate with FDA (Live)

    Device Name
    Kerasave
    Manufacturer
    Date Cleared
    2022-05-02

    (578 days)

    Product Code
    Regulation Number
    886.4320
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Kerasave is indicated for storage of human corneas at 2-8°C for up to 14 days. It is intended for prescription (Rx) use by physicians or highly skilled personnel, such as Eye Bank operators.

    Device Description

    Kerasave is made of a buffered corneal storage medium, which provides basic nutrients for cell maintenance during storage of donor corneas at 2-8°C for up to 14 days, at physiological pH. The device also includes antimicrobial agents. The antibiotics are dissolved in the solution and an antifungal agent is formulated as a tablet for stability reason and constitutes integral part of the device; it shall be dissolved in the medium prior to use.

    AI/ML Overview

    Here's a summary of the acceptance criteria and the study proving the device meets those criteria, based on the provided text:

    Acceptance Criteria and Device Performance for Kerasave

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document defines "Special Controls" which serve as the acceptance criteria for the Kerasave device. These are primarily evaluated through non-clinical performance testing.

    Acceptance CriterionReported Device Performance
    Non-clinical performance testing must demonstrate the device performs as intended under anticipated conditions of use.Corneal Endothelial Cell Layer Preservation: Kerasave was non-inferior to Optisol-GS (FDA-cleared comparator) for Endothelial Cell Density (ECD) and Hexagonality (HEX) at Day 14. An exploratory analysis also showed non-inferiority for Coefficient of Variation (%CV) after removing an outlier. (Tables 4, 5, 7)
    (i) Following performance characteristics of the cornea following storage must be demonstrated:
    (A) Endothelial cell densityNon-inferior to Optisol-GS at Day 14 (Table 4).
    (B) Endothelial cell morphology (pleomorphism: Hexagonality (HEX) and polymegathism: %CV)HEX: Non-inferior to Optisol-GS at Day 14 (Table 5). %CV: Failed to statistically clear non-inferiority initially, but exploratory analysis (excluding outlier) showed non-inferiority at Day 14 (Table 6, 7).
    (C) Corneal transparencyNo statistical difference in clarity scores between Kerasave and Optisol-GS at Day 14. However, Kerasave showed increased opacity (score 1 to 2) in 7/27 corneas, compared to 2/27 for Optisol-GS. Increased edema and folds/striae were observed in these corneas. (Table 9, and related text)
    (D) Central corneal thicknessCCT changes were similar between Kerasave and Optisol-GS arms, with no statistical difference (Table 8).
    (ii) Antimicrobial activity of the device must be demonstrated at the initial and maximum labeled storage time.Demonstrated effective inhibition of bacterial growth (streptomycin and gentamicin) and fungal growth (Amphotericin B). Amphotericin B reduced Candida sp. population by 5-log and Fusarium sp. population by ~1-log. (Antimicrobial Evaluation section)
    (iii) Characterization of all preservatives, including antifungals, must include:
    (A) Characterization of impurities, heavy metal analysis, concentration, and dissolutionAmphotericin B in Kerasave was identified as equivalent to the reference substance. Impurity profile met USP/EP limits. Heavy metals met acceptance (<10 ppm). Uniform distribution of Amphotericin B demonstrated after dissolution. (Chemical Characterization section)
    (B) Chemical activity of all preservatives over the labeled use life of the device.Amphotericin B concentration remained above 11% after 72 hours of light exposure, indicating stability. Osmolality and pH remained within acceptance limits for 14 days after Amphotericin B dissolution. Shelf-life studies support stability and sterility for 24 months (device) and 6 months (Amphotericin B tablets). (Chemical Characterization, Shelf-Life/Sterility sections)
    (2) Performance data must demonstrate the sterility of the device.Achieved through filter sterilization, dry heat, and gamma irradiation (SAL 10-6). Bacterial endotoxin levels were below recommended limits. Supported by real-time and accelerated aging studies. (Shelf-Life/Sterility section)
    (3) The device must be demonstrated to be biocompatible and non-pyrogenic.Biocompatibility evidenced by non-cytotoxic, non-sensitizer, non-irritant, non-toxic (acute and subacute/subchronic systemic), and non-pyrogenic results according to ISO 10993 and FDA Guidance. (Table 3)
    (4) Performance data must support the claimed shelf life by demonstrating continued sterility, controlled bioburden, package integrity, and device functionality over the intended shelf life.Shelf-life of 24 months (Kerasave) and 6 months (Amphotericin B tablets) established, supported by sterility & stability in real-time & accelerated aging studies. (Shelf-Life/Sterility section)
    (5) The device and each of its components (e.g., antifungal, antibiotic, medium) must be demonstrated to be compatible with their respective commercial container closure system/packaging.Not explicitly detailed in the provided text beyond the general statement of "supported by real-time and accelerated aging studies, as well as after short term storage at high temperatures" under Shelf-Life/Sterility. A specific study demonstrating "compatibility with container closure system/packaging" isn't individually called out with results.
    (6) An analysis must be provided that identifies and evaluates any contribution to the development and spread of antimicrobial resistance.FDA determined low likelihood of antimicrobial resistance emergence/spread due to concentrations of antibiotics/antifungal in Kerasave. (Antimicrobial Evaluation section)

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Twenty-seven (27) paired donor corneas were used for the performance testing related to endothelial cell layer preservation, corneal clarity, and central corneal thickness.
    • Data Provenance: The text does not explicitly state the country of origin. It indicates it was an ex-vivo (bench) study using donor corneas, suggesting it's not patient data and is prospective for the purpose of the study.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The study involved comparisons of corneal storage conditions and measurements performed using a Konan specular microscope, optical coherence tomography (OCT), and slit lamp examination. It is implied these measurements were performed by trained personnel, but the text does not specify the number or qualifications of experts used to establish "ground truth" as typically defined for AI/ML device evaluations (e.g., expert consensus on image interpretation). The "ground truth" here is objective measurements and observations of the corneal health parameters.

    4. Adjudication Method for the Test Set

    Not applicable. This was a direct measurement study comparing a new device against a predicate device on biophysical parameters, not an expert-based subjective assessment requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, an MRMC comparative effectiveness study was not conducted. This study's focus was on the performance of the corneal storage medium itself, not on human readers interpreting medical images or data with or without AI assistance.

    6. Standalone (Algorithm Only) Performance Study

    Not applicable. Kerasave is a physical corneal storage medium, not an AI algorithm. Therefore, a standalone algorithm-only performance study was not performed.

    7. Type of Ground Truth Used

    The ground truth for the performance parameters (Endothelial Cell Density, Hexagonality, Coefficient of Variation, Central Corneal Thickness, Corneal Clarity) was established through direct objective measurements using established laboratory techniques such as Konan specular microscopy, optical coherence tomography, and slit lamp examination.

    8. Sample Size for the Training Set

    Not applicable. Kerasave is a physical medical device, not an AI/ML algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As Kerasave is a physical medical device, there is no training set or associated ground truth establishment for an AI/ML model.

    Ask a Question

    Ask a specific question about this device

    K Number
    K162013

    Validate with FDA (Live)

    Device Name
    Eusol-C
    Date Cleared
    2016-09-15

    (56 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    0 - 21
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Eusol-C is indicated when it is desired to store explanted corneas for up to 14 days. It is intended for use by physicians or highly skilled personnel, such as Eye Bank operators.

    Device Description

    Not Found

    AI/ML Overview

    The provided text is related to a 510(k) premarket notification for a medical device called "Eusol-C" which is intended for storing explanted corneas. However, the document does not contain any information regarding acceptance criteria, device performance, study details (sample size, provenance, expert details, adjudication methods), or training set information.

    The document is primarily an FDA clearance letter and an "Indications for Use" statement. It confirms that the device is substantially equivalent to legally marketed predicate devices and outlines general regulatory compliance requirements.

    Therefore, I cannot provide the requested information based on the given text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K063617

    Validate with FDA (Live)

    Device Name
    EUSOL-C
    Manufacturer
    Date Cleared
    2007-02-08

    (65 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Eusol-C is sterile medium for hypothermic corneal storage intended for corneal storage at 4℃ for up to 14 days, to be used by physicians or highly skilled personnel.
    Eusol-C is indicated when it is desired to store explanted corneas for up to 14 days. It is intended for use by physicians or highly skilled personnel, such as Eye Bank operators.

    Device Description

    Eusol-C is sterile medium for hypothermic corneal storage intended for corneal storage at 4°C for up to 14 days, to be used by physicians or highly skilled personnel, such as Eye Bank operators. EUSOL-C can be stored at 2-25°C until ready for use. Based on stability studies, peak temperatures up to 42°C do not affect EUSOL-C parameters and performances.

    AI/ML Overview

    This document is a 510(k) summary for Eusol-C, a sterile medium for hypothermic corneal storage. It describes the device, its intended use, and indicates that it is substantially equivalent to a legally marketed predicate device (Eusol-C Corneal Storage Media, K032422).

    Here's an analysis of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided text does not explicitly state specific acceptance criteria in terms of numerical targets or thresholds for device performance. Instead, it states that "The following tests have been done on Eusol-C:" and lists general categories of tests. The overall conclusion is that "Eusol-C is equivalent in safety and efficacy to the legally-marketed predicate device."

    Acceptance Criteria CategoryReported Device Performance
    Bench TestsDone
    - Physical TestsDone
    - Microbiological TestsDone
    - Biological Tests and AssaysDone
    - Performance testsDone
    Clinical TestsNot required
    Overall ConclusionEquivalent in safety and efficacy to legally-marketed predicate device.
    StabilityCan be stored at 2-25°C. Peak temperatures up to 42°C do not affect EUSOL-C parameters and performances.

    2. Sample Size Used for the Test Set and Data Provenance

    The document mentions "Bench tests" and "Performance tests" but does not specify the sample size for these tests. It also does not provide details on data provenance (e.g., country of origin, retrospective or prospective) for any of the non-clinical tests. Clinical tests were "Not required."

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the document. Given that clinical tests were not required, it's unlikely that "experts" in the traditional sense (e.g., radiologists) were used to establish ground truth for a clinical test set. The non-clinical tests would rely on established laboratory testing methodologies and standards.

    4. Adjudication Method for the Test Set

    This information is not provided and is not applicable since clinical tests requiring adjudication were not conducted according to the submission.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and its effect size

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The document states "Clinical tests: Not required."

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) study was done

    This question is not applicable as the device (Eusol-C) is a sterile medium for corneal storage, not an algorithm or AI-based diagnostic tool. Therefore, an "algorithm only" or "standalone" performance study in the context of AI is not relevant.

    7. The Type of Ground Truth Used

    For the non-clinical tests, the "ground truth" would have been established through a combination of:

    • Standard laboratory assays and measurements: For physical, microbiological, and biological tests.
    • Established performance specifications: For "Performance tests," likely comparing against a predicate device or pre-defined benchmarks for corneal viability/integrity after storage in Eusol-C.
    • The document implies comparison to a predicate device for equivalence, suggesting the predicate's performance served as a de-facto ground truth or benchmark.

    8. The Sample Size for the Training Set

    This question is not applicable as the device is not an AI/ML algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    This question is not applicable as the device is not an AI/ML algorithm that requires a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K032422

    Validate with FDA (Live)

    Device Name
    EUSOL-C
    Manufacturer
    Date Cleared
    2003-12-15

    (132 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Eusol-C is indicated for corneal storage for up to 14 days.
    Eusol-C is indicated for the storage of donor corneas for up to 14 days.

    Device Description

    Eusol-C is sterile medium for hypothermic corneal storage intended for corneal storage at 4℃ for up to 14 days.

    AI/ML Overview

    This document describes a 510(k) submission for the Eusol-C corneal storage medium. The information provided is primarily focused on regulatory submission and does not detail a clinical study with an AI device or a direct comparison of device performance against established acceptance criteria in the manner requested for an AI/device performance study.

    Based on the provided text, here's a breakdown of the information, with explicit notes where the requested details are not present:

    The device in question, Eusol-C, is a sterile medium for hypothermic corneal storage, not an AI-powered device. Therefore, the questions related to AI performance, such as MRMC studies, standalone algorithm performance, and training/test set details for AI, are not applicable.

    1. Table of acceptance criteria and the reported device performance

    Acceptance CriteriaReported Device Performance
    Not explicitly defined in the document. The submission aims to demonstrate substantial equivalence to a predicate device.Not explicitly defined in the document as a "performance metric" against a pre-set acceptance criteria. The submission states "Eusol-C is equivalent in safety and efficacy to the legally-marketed predicate device" (Optisol-GS Corneal Storage Media, K924165).

    Note: For medical devices, "acceptance criteria" are usually specific metrics (e.g., cell viability percentage, storage duration without adverse events) that the new device must meet to be considered safe and effective. This document focuses on demonstrating equivalence to a predicate device, rather than meeting novel performance criteria.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: Not applicable. No clinical "test set" in the context of an AI study or a patient-based clinical trial is described. The submission refers to "non-clinical tests" and explicitly states "Clinical tests: Not required."
    • Data Provenance: Not applicable for a clinical test set. The non-clinical tests (long-term stability, accelerated aging, container closure, performance characteristics, sterilization validation, comparison with predicate device) would have been conducted by the manufacturer, AL.CHI.MI.A. S.r.l. in Italy.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not applicable. This is not a study requiring expert ground truth for image interpretation or diagnosis.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. This is not a study requiring adjudication of expert opinions.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This is not an AI device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is not an AI device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Not applicable in the context of expert-derived ground truth for diagnoses. For the non-clinical tests, the "ground truth" would be established by standard laboratory methods and comparisons to the predicate device's characteristics. For example, for "long-term stability," the ground truth would be chemical and physical properties measured over time.

    8. The sample size for the training set

    • Not applicable. This is not an AI device and thus has no "training set."

    9. How the ground truth for the training set was established

    • Not applicable. This is not an AI device.