Search Filters

Search Results

Found 7 results

510(k) Data Aggregation

    K Number
    K233601

    Validate with FDA (Live)

    Date Cleared
    2024-08-02

    (267 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Fixone Meniscal Repair is intended to suture the meniscus tear to facilitate percutaneous or endoscopic soft tissue procedures.

    Device Description

    The Fixone Meniscal Repair are all-inside meniscal repair devices. Each device includes two peek anchors, pre-tied with non-absorbable suture pre-loaded into a needle delivery system with an adjustable length depth limiter. The device is made to insert into the joint, place and fix the first anchor(T1) around the ruptured meniscus cartilage, and then place and fix the second anchor(T2) to make one Stitch to seal the torn cartilage. The device is provided sterile for single use only.

    AI/ML Overview

    This document describes the performance testing for the "Fixone Meniscal Repair" device. Primarily, it focuses on bench testing to demonstrate substantial equivalence to predicate devices, rather than a clinical study evaluating diagnostic or therapeutic efficacy with human subjects.

    Here's a breakdown based on the provided text, addressing your questions where information is available:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document lists various performance tests and their corresponding requirements. However, it does not provide specific numerical results or pass/fail outcomes for each criteria within the summary. It generally states that the tests "demonstrated that the device performs in a substantially equivalent manner to the prior related devices."

    Test ItemRequirementsReported Device Performance
    BiocompatibilityLeveraged from K203523 and K230892 for anchor/suture; ISO 10993-1 for other components.Demonstrated substantial equivalence (implied pass)
    Extraction TestVisual clarity/no foreign particles, pH difference ≤ 1.5, potassium permanganate difference ≤ 2.0 mL, residue after evaporation ≤ 1.0mg, UV spectrum (250-350nm) max absorbance ≤ 0.1, heavy metals not exceeding standard lead solution.Demonstrated substantial equivalence (implied pass)
    Extractable Color (Suture)MFDS guideline 2021-3 10.non-absorbable suture.Demonstrated substantial equivalence (implied pass)
    EO ResidualISO 10993-7, Ethylene oxide Sterilization residuals.Demonstrated substantial equivalence (implied pass)
    Sterility TestISO 10993-7, Ethylene oxide Sterilization residuals.Demonstrated substantial equivalence (implied pass)
    Comparison Test (Pull-out)Not explicitly stated, but compared to FAST-FIX Meniscal Repair System and Fixone Meniscal Repair.Demonstrated substantial equivalence (implied pass)
    Comparison Test (Fatigue)Not explicitly stated, but compared to FAST-FIX Meniscal Repair System and Fixone Meniscal Repair.Demonstrated substantial equivalence (implied pass)
    Shelf Life TestingReal-time equivalent (RTE) and Accelerated aging temperature, Taa as per AAMI TIR17 and ASTM F 1980-02; Aging temperature: 55±2°C, aging time: 225 days.Demonstrated substantial equivalence (implied pass)
    Comparison Test (Benchtop performance simulating clinical usage)Not explicitly stated.Demonstrated substantial equivalence (implied pass)
    External SurfaceNo damage to appearance in visual inspection.Demonstrated substantial equivalence (implied pass)
    Length (Suture)Not less than 95.0% of 750mm when laid out smooth without tension.Demonstrated substantial equivalence (implied pass)
    Diameter (Suture)According to MFDS Notification No.2020-20, "Nonabsorbable Surgical Suture, 4. Diameter" with Table 1.Demonstrated substantial equivalence (implied pass)
    Needle AttachmentMeet test requirements of MFDS notification No.2020-97 for "Surgical suture, non-absorbable, with/without needle".Demonstrated substantial equivalence (implied pass)
    Tensile Strength (Knot)Meet test requirements of MFDS notification No.2020-97 for "Surgical suture, non-absorbable, with/without needle".Demonstrated substantial equivalence (implied pass)
    Endotoxin TestUSP 42 <85>, Bacterial endotoxin test (LAL).Conducted for each batch; batches not passing are discarded.
    Pyrogen TestUSP 42 : 2019, <151> Pyrogen test.Demonstrated substantial equivalence (implied pass)

    2. Sample Size Used for the Test Set and Data Provenance:

    The document describes bench testing, not clinical studies with human subjects or test sets in the context of diagnostic accuracy. Therefore, information about patient sample size, country of origin, or retrospective/prospective nature of data for a "test set" is not applicable. The samples used for mechanical and material testing would be device prototypes/production units.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    This question is not applicable as the studies described are bench tests of mechanical and material properties, not studies requiring expert interpretation of clinical data to establish ground truth.

    4. Adjudication Method for the Test Set:

    Not applicable. Adjudication methods like 2+1 or 3+1 are used for clinical studies where expert consensus on interpretations is needed.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance:

    This information is not provided because an MRMC study is not mentioned. The device is a meniscal repair device, not an AI-powered diagnostic tool.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done:

    Not applicable. This device is a physical medical device, not an algorithm.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" for the bench tests would be established through objective engineering and material science standards and measurements. For example, the "ground truth" for tensile strength is a numerical value measured by a testing machine according to a defined protocol. For biocompatibility, it's adherence to ISO 10993 standards and a lack of adverse biological reactions in defined tests.

    8. The Sample Size for the Training Set:

    Not applicable. This device is not an AI algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K230892

    Validate with FDA (Live)

    Date Cleared
    2023-06-09

    (70 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Fixone Biocomposite Anchors are intended for fixation of suture (soft tissue) to bone in the shoulder, foot/ankle, knee and elbow in the following procedures:

    Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction;

    Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair;

    Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Posterior Oblique Ligament Repair, lliotibial Band Tenodesis;

    Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction.

    The following indications are for the S and SL type anchors only:

    Hand/Wrist: Scapholunate Ligament Reconstruction, Carpal Ligament Reconstruction, Repair/Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP, and MCP joints for all digits, digital tendon transfers

    Hip: Capsular repair, Acetabular Labral repair

    Device Description

    The Fixone Biocomposite Anchor is intended for reattaching soft tissue to bone with sutures. The anchor is manufactured from biodegradable materials (PLGA copolymer and β-TCP). A nonabsorbable UHMWPE suture is inserted into the anchor. The anchor is implanted using a provided driver.

    B,N,BN type(Knotted suture anchors) are 'screw-in' anchors used alone or in combination with Kc,KcN type(Knotless suture anchors) for soft-tissue repair.

    Kc,KcN type(Knotless suture anchor) are used for a soft-tissue repair in combination with the B,N,BN type.

    SL type 'push-in' anchors that are ideal for soft-tissue repair in narrow or deep areas.

    S type are 'push-in' anchors that are ideal for soft-tissue repair in the foot, ankle, hand and wrist

    The Fixone Biocomposite Anchor consists of cannulated anchors with an eyelet. They are pre-loaded on an insertion device. The Fixone Biocomposite Anchor is intended to provide secure reattachment of the soft tissue to bone.

    Devices are provided sterile. Single use only.

    AI/ML Overview

    The provided text describes a medical device, the "Fixone Biocomposite Anchor," and its 510(k) submission to the FDA. This submission aims to demonstrate substantial equivalence to previously cleared devices. However, the document does not contain information related to acceptance criteria or a study proving the device meets those criteria in the context of an AI/ML medical device.

    The document discusses:

    • Device Description: What the Fixone Biocomposite Anchor is, its components (PLGA copolymer and β-TCP anchor, UHMWPE suture), and how it's implanted.
    • Intended Use: Fixation of suture (soft tissue) to bone in various anatomical locations (shoulder, foot/ankle, knee, elbow, hand/wrist, hip).
    • Predicate Device Comparison: A detailed table comparing the subject device to several predicate devices (K171299, K192032, K192484, K213008) across various attributes like indications for use, materials, dimensions, fixation mechanism, suture characteristics, and sterilization.
    • Performance Data: Mentions "Bench test were performed. Bench testing included biocompatibility, mechanical testing including EO residues, shelf-life testing, pyrogenicity testing and endotoxin monitoring. The tests demonstrated that the device performs in a substantially equivalent manner to the prior related devices."

    Crucially, there is no mention of:

    • AI/ML components: The device is a physical biocomposite anchor, not a software algorithm.
    • Clinical study or diagnostic performance metrics: The performance data refers to bench testing for physical and material properties, not clinical outcomes, sensitivity, specificity, or AUC for a diagnostic tool.
    • Ground truth, expert readers, sample sizes for AI models, or MRMC studies. These are concepts relevant to evaluating AI/ML diagnostic or predictive devices, which the Fixone Biocomposite Anchor is not.

    Therefore, I cannot fulfill the request as the provided input does not contain the necessary information about acceptance criteria and a study proving an AI/ML device's performance. The device described is a traditional medical implant, and its evaluation focuses on physical and biological equivalence, not AI performance.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222423

    Validate with FDA (Live)

    Date Cleared
    2023-03-23

    (224 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The non-absorbable suture anchors are intended to reattach soft tissue to bone in orthopedic surgical procedures. The Fixone All Suture Anchors may be used in either arthroscopic or open surgical procedures. After the suture strands are anchored to the bone, they may be used to reattach soft tissue, such as ligaments, tendons, or joint capsules to the bone. In conjunction with appropriate postoperative immobilization throughout the healing period, the suture anchor systems stabilize the damaged soft tissue.

    Device Description

    Fixone All suture anchor consists of one "fix Suture" and one,two or three non-absorbable Sutures. The nonabsorbable suture is manufactured from UHMWPE. They are pre-loaded on a handled insertion device. This device is provided sterile, for single use only. This device is medical device. Prescription use only.

    The color additive copper-phthalocyanine blue, color additive D&C green 6, color additive D&C violet 2 and color additive D&C black 4 are according to FDA regulations and it is approved for use in medical applications (§74.3045 – FDA), (§74.3206 –FDA), (§74.1602 –FDA) and (§74.3054 – FDA) respectively.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device called "Fixone All Suture Anchor." This document primarily deals with the substantial equivalence of a new version of the suture anchor to previously cleared predicate devices. It focuses on engineering and biocompatibility testing rather than clinical study data involving AI or human readers evaluating image-based diagnostics.

    Therefore, the information required to answer the prompt regarding acceptance criteria and study that proves the device meets the acceptance criteria, specifically related to AI in a clinical setting (e.g., sample size for test set, data provenance, expert ground truth, MRMC study, standalone performance, training set details), is not present in the provided text.

    The document discusses acceptance criteria and proof of mechanical performance and biocompatibility for the suture anchor. Here's what can be extracted based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance (Mechanical & Biocompatibility)

    Please note that the acceptance criteria are generally implied by the "Requirements" column and "Pass" result, indicating the device met those requirements. Specific numerical requirements are given where applicable.

    Anchor Performance

    Test ItemRequirementsReported Device Performance
    Insertion StrengthInsert each anchor with 5mm/min rate into 10pcf/40pcf combined urethane block.Pass
    Pull-out (FAS-1652802iw)Pull out each anchor with 5mm/min rate from 10pcf/40pcf combined urethane block.Pass
    Fatigue test (FAS-1652802iw)Pull out each anchor with 20N to 60N cycle loading and 1Hz/40,000 cycles from 10pcf/40pcf combined urethane block.Pass
    Pull-out (FAM-7615802fg)Pull out each anchor with 5mm/min rate from 10pcf/40pcf combined urethane block.Pass
    Fatigue test (FAM-7615802fg)Pull out each anchor with 20N to 60N cycle loading and 1Hz/40,000 cycles from 10pcf/40pcf combined urethane block.Pass

    Suture Performance (USP 0, UHMWPE)

    Test ItemRequirementsReported Device Performance
    AppearanceThe test article shall be no damage to the appearance in visual inspection.Pass
    LengthThe length of suture was measured while the strand is laid out smooth, without tension, on a plane surface: the length of the strand is not less than 95.0 percent of 750mm.Pass
    Tensile StrengthFor USP 0 size suture, tensile strength must be larger than 2.16 kgf.Pass
    Needle AttachmentFor USP 0 size suture, needle attachment must be larger than 1.50 kgf.Pass

    Suture Performance (USP 2, UHMWPE)

    Test ItemRequirementsReported Device Performance
    LengthThe length of suture was measured while the strand is laid out smooth, without tension, on a plane surface: the length of the strand is not less than 95.0 percent of 750mm.Pass
    Tensile StrengthFor USP 2 size suture, tensile strength must be larger than 3.52 kgf.Pass

    Suture Performance (USP 0, USP 2, UHMWPE)

    Test ItemRequirementsReported Device Performance
    AppearanceThe test article shall be no damage to the appearance in visual inspection.Pass
    DiameterFor USP 0 size suture, diameter must be within 0.350.399mm. For USP 2 size suture, diameter must be within 0.500.599mm.Pass
    Tensile strengthWith 25~35cm/min, record the load-displacement curves and maximum load. (N)Pass

    Biocompatibility (UHMWPE + PET)

    Test ItemTest Method / Test CriteriaReported Result
    CytotoxicityISO 10993-5, Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicityPass
    Systemic toxicity testISO 10993-11, Biological evaluation of medical devices - Part 11: Tests for systemic toxicityPass
    Pyrogen TestISO 10993-11 Test for systemic toxicity, pyrogen testPass
    Intracutaneous reactivity testISO 10993-10, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitizationPass
    Maximization sensitizationISO 10993-10, Test for irritation and skin sensitization, Maximization test for delayed hypersensitivityPass
    Genotoxicity testISO 10993-3, Genotoxicity test OECE 471, Bacterial reverse mutation testPass
    Implantation testISO 10993-6, Tests for local effects after implantation, Annex D test methods for implantation in bonePass
    Hemolysis testISO 10993-4, Biological evaluation of medical devices - Part 4: Selection of tests for interactions with bloodPass
    Subchronic toxicityISO 10993-11 Tests for systemic toxicityPass

    Biocompatibility (Only UHMWPE)

    (Same tests and results as UHMWPE + PET table, excluding PET interaction)

    Test ItemTest Method / Test CriteriaReported Result
    CytotoxicityISO 10993-5, Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicityPass
    Systemic toxicity testISO 10993-11, Biological evaluation of medical devices - Part 11: Tests for systemic toxicityPass
    Pyrogen TestISO 10993-11 Test for systemic toxicity, pyrogen testPass
    Intracutaneous reactivity testISO 10993-10, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitizationPass
    Maximization sensitizationISO 10993-10, Test for irritation and skin sensitization, Maximization test for delayed hypersensitivityPass
    Genotoxicity testISO 10993-3, Genotoxicity test OECE 471, Bacterial reverse mutation testPass
    Implantation testISO 10993-6, Tests for local effects after implantation, Annex D test methods for implantation in bonePass
    Hemolysis testISO 10993-4, Biological evaluation of medical devices - Part 4: Selection of tests for interactions with bloodPass
    Subchronic toxicityISO 10993-11 Tests for systemic toxicityPass

    Sterilization/Endotoxin

    Test ItemTest Method / Test CriteriaReported Result
    Endotoxin testUSP 43 <85>, Bacterial endotoxin test (LAL)Pass

    2. Sample size used for the test set and the data provenance:
    This information is not provided in the context of an AI/image-based diagnostic test set. The document refers to "bench tests" performed on physical samples of the anchors and sutures. The number of samples for each mechanical test is not specified, only that "each anchor" or "the test article" was evaluated. The data provenance is implied to be from AJU Pharm Co., Ltd.'s internal testing, likely at their facility or a contracted lab. No country of origin for this specific testing data is given, but the company address is in South Korea (A-207, 697, Pangyo-ro, Seongnam-si, Gyeonggi-do, Korea). The data is retrospective of the device design.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
    Not applicable for this type of device (mechanical and biocompatibility testing). Ground truth is established by objective physical measurements and standardized biological assays.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
    Not applicable for this type of device.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
    Not applicable. This is not an AI-assisted diagnostic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
    Not applicable. This is not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
    For mechanical tests: Physical measurements against engineering specifications (e.g., tensile strength, length, diameter, pull-out force).
    For biocompatibility tests: Standardized laboratory assays with pass/fail criteria based on ISO standards.
    For sterilization/endotoxin: Standardized laboratory assays (e.g., LAL test) with pass/fail criteria.

    8. The sample size for the training set:
    Not applicable. This is not a machine learning/AI device.

    9. How the ground truth for the training set was established:
    Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K203523

    Validate with FDA (Live)

    Date Cleared
    2021-12-25

    (389 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Fixone Hybrid Knotless Anchor is intended for fixation of suture (soft tissue) to bone in the shoulder, foot/ankle, knee and elbow in the following procedures:

    Shoulder: Rotor Cuff Repair, Bankart Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction;

    Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair;

    Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Posterior Oblique Ligament Repair, lliotibial Band Tenodesis;

    Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction

    Device Description

    The Fixone Hybrid Knotless Anchor is intended for reattaching soft tissue to bone with sutures. The anchor is manufactured from biodegradable materials (PLGA copolymer and β-TCP) and nonabsorbable PEEK tip. A nonabsorbable suture manufactured from cobraided UHMWPE and PET fibers is inserted into the anchor. The anchor is implanted using a provided driver.

    This device is could used with instrument that manufactured by Aju Pharm Co.,Ltd. It is consist of 11 models. It provide non-sterile (user must sterilization before use).

    AI/ML Overview

    Here's a summary of the acceptance criteria and the study that proves the device meets them, based on the provided FDA 510(k) summary for the Fixone Hybrid Knotless Anchor.

    1. Table of Acceptance Criteria and Reported Device Performance

    Test ItemAcceptance CriteriaReported Performance
    Mechanical Testing
    External Surface (ASTM F2502, USP<28>)Not explicitly stated in table, but implied to meet standards.Pass
    MeasurementWithin 5% of the dimensions specified in specifications.Pass
    Insertion Torque (ASTM F1839)Not less than 0.15 Nm.Pass
    Fixation StrengthNot less than 100 N.Pass
    Tensile Strength (suture threads)Not less than 34.5 N.Pass
    Torsional TestMaximum torque should not be less than insertion torque. Anchor should endure movement of tissues and bones (Arthrex pull-out and cyclic pull-out testing cited as supporting this).Pass
    Extraction TestNot explicitly stated in table.Pass
    pHDifference should be 1.5 and less.Pass
    Potassium Permanganate Reducing SubstancesDifference of consumption of potassium permanganate should be 2.0 mL and less.Pass
    Residue after EvaporationWeight of the residue should be 1.0 mg and less.Pass
    Heavy MetalsAny brown color produced within 10 minutes in the tube containing the extract of the prepared sample does not exceed that in the tube containing the standard lead solution.Pass
    UV Spectrum (250nm~350nm)Maximum absorbance between 250 to 350 nm should be 0.1 and less.Pass
    Property (Visual inspection)Test solution should be clear and have no foreign particles.Pass
    Biocompatibility (Anchor)
    Cytotoxicity (ISO 10993-5)Meets ISO 10993-5 criteria.Pass
    Maximization test for delayed hypersensitivity (ISO 10993-10)Meets ISO 10993-10 criteria.Pass
    Intracutaneous reactivity (ISO 10993-10)Meets ISO 10993-10 criteria.Pass
    Acute Systemic toxicity (ISO 10993-11)Meets ISO 10993-11 criteria.Pass
    Material-mediated Pyrogen (ISO 10993-11, USP39<151>)Meets ISO 10993-11 criteria.Pass
    Sub acute Systemic toxicity (ISO 10993-11)Meets ISO 10993-11 criteria.Pass
    Bacterial reverse mutation test (ISO 10993-3, OECD 471)Meets ISO 10993-3 and OECD 471 criteria.Pass
    Mammalian erythrocyte micronucleus test (ISO 10993-3, OECD 474)Meets ISO 10993-3 and OECD 474 criteria.Pass
    Implantation test (ISO 10993-6, Annex D for bone)Meets ISO 10993-6 criteria for local effects after implantation in bone.Pass
    Biocompatibility (Suture)
    Cytotoxicity (ISO 10993-5)Meets ISO 10993-5 criteria.Pass
    Maximization test for delayed hypersensitivity (ISO 10993-10)Meets ISO 10993-10 criteria.Pass
    Intracutaneous reactivity (ISO 10993-10)Meets ISO 10993-10 criteria.Pass
    Acute Systemic toxicity (ISO 10993-11)Meets ISO 10993-11 criteria.Pass
    Material-mediated Pyrogen (ISO 10993-11)Meets ISO 10993-11 criteria.Pass
    Sub acute Systemic toxicity (ISO 10993-11)Meets ISO 10993-11 criteria.Pass
    Bacterial reverse mutation test (ISO 10993-3, OECD 471)Meets ISO 10993-3 and OECD 471 criteria.Pass
    Mammalian erythrocyte micronucleus test (ISO 10993-3, OECD 474)Meets ISO 10993-3 and OECD 474 criteria.Pass
    Implantation test (ISO 10993-6, 4 & 12 weeks)Meets ISO 10993-6 criteria for local effects after implantation.Pass
    Biological risk assessmentAssessment according to ISO 10993-1, EN ISO 14971, FDA G95-1, EU Medical Device Directive. Examines components, literature, and history of safe use.-
    Biocompatibility (Tip)
    Cytotoxicity (ISO 10993-5)Meets ISO 10993-5 criteria.Pass
    Maximization test for delayed hypersensitivity (ISO 10993-10)Meets ISO 10993-10 criteria.Pass
    Intracutaneous reactivity (ISO 10993-10)Meets ISO 10993-10 criteria.Pass
    Acute Systemic toxicity (ISO 10993-11)Meets ISO 10993-11 criteria.Pass
    Material-mediated Pyrogen (ISO 10993-11)Meets ISO 10993-11 criteria.Pass
    Sub acute Systemic toxicity (ISO 10993-11)Meets ISO 10993-11 criteria.Pass
    Bacterial reverse mutation test (ISO 10993-3, OECD 471)Meets ISO 10993-3 and OECD 471 criteria.Pass
    Mammalian erythrocyte micronucleus test (ISO 10993-3, OECD 474)Meets ISO 10993-3 and OECD 474 criteria.Pass
    Implantation test (ISO 10993-6, Annex D for bone)Meets ISO 10993-6 criteria for local effects after implantation in bone.Pass
    Biocompatibility (Colored Suture)
    Cytotoxicity (ISO 10993-5)Meets ISO 10993-5 criteria.Pass
    Maximization test for delayed hypersensitivity (ISO 10993-10)Meets ISO 10993-10 criteria.Pass
    Intracutaneous reactivity (ISO 10993-10)Meets ISO 10993-10 criteria.Pass
    Acute Systemic toxicity (ISO 10993-11)Meets ISO 10993-11 criteria.Pass
    Material-mediated Pyrogen (ISO 10993-11)Meets ISO 10993-11 criteria.Pass
    Sub chronic Systemic toxicity (ISO 10993-11)Meets ISO 10993-11 criteria.Pass
    Bacterial reverse mutation test (ISO 10993-3, OECD 471)Meets ISO 10993-3 and OECD 471 criteria.Pass
    Mammalian erythrocyte micronucleus test (ISO 10993-2, OECD 474)Meets ISO 10993-2 and OECD 474 criteria.Pass
    Implantation test (ISO 10993-6, 4 & 12 weeks)Meets ISO 10993-6 criteria for local effects after implantation.Pass
    Biocompatibility (Driver shaft - Stainless steel 304)
    Cytotoxicity (ISO 10993-5)Meets ISO 10993-5 criteria.Pass
    Intracutaneous reactivity test (ISO 10993-10)Meets ISO 10993-10 criteria.Pass
    Maximization sensitization (ISO 10993-10)Meets ISO 10993-10 criteria.Pass
    Sterilization & Shelf Life
    Pyrogen (Material-mediated)Meets ISO10993-11, USP39<151> criteria.Pass
    Shelf life testing (Accelerated aging)Real-time equivalent (RTE) and Accelerated aging temperature (Taa) as per AAMI TIR17 and ASTM F 1980-02 confirmed stability at 55±2°C for 229 days.Pass
    Shelf life testing (Real-time)Physicochemical changes and packaging tested at 0, 1, 2, 3, 4, and 5 years for validity and safety per ISO 11607-1:2006, ISO 11607-2:2006. Valid for 5 years.Pass
    EO sterilization Validation (EO residual test)ISO11135(2014) standards met.Pass
    Endotoxin (USP 43 <85>, LAL)Endotoxin level conforms.Pass

    2. Sample Size for the Test Set and Data Provenance

    The document describes bench testing for mechanical and biocompatibility properties. It doesn't specify a "test set" in the context of patient data or clinical images. The tests were performed on the device components themselves.

    • Sample Size: Not explicitly stated as a single number across all tests. Each test (e.g., tensile strength, insertion torque, various biocompatibility tests) would have its own sample size, which is not detailed but assumed to be sufficient for the relevant standards (e.g., ASTM, ISO, USP). For tensile strength, it mentions "measuring 10 suture threads."
    • Data Provenance: The studies were bench tests conducted by the manufacturer, AJU Pharm Co., Ltd., which is located in Korea. The details provided do not indicate the use of retrospective or prospective patient data, as this is a physical device rather than a diagnostic or AI-driven system.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    This information is not applicable to this type of medical device submission. The Fixone Hybrid Knotless Anchor is a physical surgical implant, not a device that generates or interprets diagnostic images or data requiring expert review for ground truth. The "ground truth" here is established through adherence to recognized international and national standards (e.g., ISO, ASTM, USP) for mechanical properties, material properties, and biocompatibility.

    4. Adjudication Method for the Test Set

    This is not applicable as there is no "test set" in the clinical imaging or diagnostic sense requiring expert adjudication. The performance is assessed against predefined quantitative and qualitative criteria in bench testing.

    5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study

    This is not applicable. The device is a surgical anchor, not an AI-assisted diagnostic tool or a system that involves human readers interpreting output. Therefore, there is no concept of improving human reader performance with or without AI assistance in this context.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    This is not applicable. The Fixone Hybrid Knotless Anchor is a physical medical device, not a software algorithm or an AI system.

    7. Type of Ground Truth Used

    The "ground truth" in this submission is established through:

    • International Standards: Adherence to widely recognized standards from organizations like ISO (e.g., ISO 10993 for biocompatibility, ISO 11135 for sterilization, ISO 11607 for packaging), ASTM (e.g., ASTM F2502, F1839, F1980-02 for mechanical and aging tests), and USP (e.g., USP<28> for material properties, USP39<151> for pyrogen, USP 43 <85> for endotoxin).
    • Predicate Device Comparison: The primary method for proving substantial equivalence is demonstrating that the test results of the Fixone Hybrid Knotless Anchor are comparable to a legally marketed predicate device (Fixone Biocomposite Anchor, K192484) and meet relevant performance benchmarks. The new device is substantially similar, with the only noted difference being the material of the tip (PEEK in the new device versus PLGA + β-TCP in the predicate).

    8. Sample Size for the Training Set

    This is not applicable. There is no "training set" as this is not a machine learning or artificial intelligence device.

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable for the reasons mentioned above.

    Ask a Question

    Ask a specific question about this device

    K Number
    K193497

    Validate with FDA (Live)

    Date Cleared
    2021-01-05

    (385 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Fixone Biocomposite Interference Screws are indicated for fixation of bone or soft tissue grafts during Anterior and/or Posterior Cruciate Ligament (ACL/PCL) reconstruction.

    Device Description

    The Fixone Biocomposite Interference Screws are intended to be used for fixation of bone-tendon-bone or soft tissue grafts during anterior or posterior cruciate ligament reconstruction surgery. This device is could used with instrument that manufactured by Aju Pharm Co.,Ltd. It is consist of 6 models. It provide non-sterile (user must sterilization before use).

    AI/ML Overview

    The provided text describes a medical device called "Fixone Biocomposite Interference Screw" and outlines its clearance process with the FDA. It includes a 510(k) summary focusing on bench testing and biocompatibility.

    Here's an analysis of the acceptance criteria and study information, based on the provided document:

    1. A table of acceptance criteria and the reported device performance

    Test ItemAcceptance CriteriaReported Device Performance
    External surfaceASTM F2502 and USP<28> (Implied: device meets these standards)Pass
    MeasurementWhen tested by vernier calipers, the tolerance should be within ±5%Pass
    Tensile strengthThe average strength of 10 stands is not less than 100N.Pass
    Extractable colorCompare with the matching solution: any color present is not more intense than of the appropriate matching solution.MD2014-00216 (Implied: Pass)
    Insertion torqueTested with ASTM F2502-Standard specification and test methods for bioabsorbable plates and screws for internal fixation implants (Implied: meets specified performance within this standard)Pass
    Fixation strength (Push-out test)Tested with ASTM F2502-Standard specification and test methods for bioabsorbable plates and screws for internal fixation implants; Tested subject and predicate device under same condition for the comparison (Implied: comparable to predicate)AJU-M20201006-01 (Implied: Pass or equivalent to predicate)
    Extraction test pHThe difference should be 1.5 and less.Pass
    Potassium permanganate reducing substancesThe difference of the consumption of potassium permanganate should be 2.0 mL and less.Pass
    Residue after evaporationRecord the weight of the residue should be 1.0mg and less.Pass
    Heavy metalsAny brown color produced within 10 minutes in the tube containing the extract of the prepared sample does not exceed that in the tube containing the standard lead solution.MD2014-00132 (Implied: Pass)
    UV spectrum (250nm~350nm)Maximum absorbance between 250 to 350 nm should be 0.1 and less.Pass
    Property (Visual inspection)When observing it with the naked eye, test solution should be clear and have no foreign particles.Pass
    Performance Testing of Fixone® Biocomposite Anchor (Pull-out test)Pull-out test by immersion time of saline solution to evaluate two bioabsorbable suture anchors. [Absorption time (soaking time) : 4 weeks, 6 weeks, 12 weeks and 26 weeks] (Implied: comparable to predicate, within acceptable limits)Ajum20170428-01 (Implied: Pass or equivalent to predicate)
    ASTM F1839-08 (Polyurethane Foam for Orthopaedic Devices)(Implied: Compliance with this standard for material properties)Pass
    Biocompatibility Tests:
    CytotoxicityISO 10993-5 Biological evaluation of medical devices - Part5: Tests for in vitro cytotoxicityPass
    Acute systemic toxicity testISO 10993-11 Biological evaluation of medical devices - Part 11: Tests for systemic toxicityPass
    Pyrogen TestISO 10993-11 Test for systemic toxicity, pyrogen testPass
    Intracutaneous (intradermal) reactivity testISO 10993-10 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitizationPass
    Maximization test for delayed hypersensitivityISO 10993-10 Test for irritation and skin sensitization, Maximization test for delayed hypersensitivityPass
    Bacterial reversion mutation testISO 10993-3 Genotoxicity test OECE 471, Bacterial reverse mutation testPass
    Mammalian erythrocyte micronucleus testISO 10993-3 Genotoxicity test OECE 471, Bacterial reverse mutation testPass
    Implantation testISO 10993-6 Tests for local effects after implantation, Annex D test methods for implantation in bonePass
    Bioabsorbable screws testASTM F2502 Standard specification and test methods for bioabsorbable plates and screws for internal fixation implantsPass
    Subchronic toxicity testISO 10993-11 Biological Evaluation of Medical Devices Part 11- Test for systemic toxicityPass

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document primarily details bench testing for mechanical, physical, and biocompatibility properties. It does not mention a clinical test set or human subject data.

    • Sample Size (for bench tests): The "Tensile strength" test mentions an "average strength of 10 stands," indicating a sample size of at least 10 for that specific test. Other tests, such as those related to "Extractable color" and "Residue after evaporation," refer to "a quantity of suture, equivalent to not less than 250mg" and standard protocols, but specific sample counts for each bench test are generally not provided.
    • Data Provenance: The bench tests were conducted by AJU Pharm Co., Ltd. in Korea (per the applicant information). These are laboratory tests, not clinical data from patients. They are inherently retrospective in the sense that they are conducted on manufactured devices in a controlled environment.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. This device approval is based on bench testing and biocompatibility assessments, not on studies requiring human experts to establish "ground truth" for diagnostic or clinical performance. The "ground truth" for the tests performed are the defined standards and specifications (e.g., ASTM, ISO standards, quantified limits for chemical properties).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. As noted above, this is not a diagnostic or clinical performance study that would involve expert adjudication of results. The results are quantitative measurements against predefined criteria.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a medical device (an interference screw for ACL/PCL reconstruction) and not an AI-powered diagnostic or assistive tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the device's performance is established by:

    • Industry Standards: ASTM F2502, ASTM F1839-08, various ISO 10993 series for biocompatibility.
    • Pharmacopoeial Standards: USP<28> (for external surface).
    • Defined Quantitative Limits: Specific values for pH difference, potassium permanganate consumption, residue after evaporation, maximum absorbance, and tensile strength (e.g., "not less than 100N").
    • Comparison to Predicate Device: The fixation strength (push-out test) directly compares the subject device to the predicate device under the same conditions. This implies the predicate device's established performance serves as a comparative ground truth.

    8. The sample size for the training set

    Not applicable. This is not a machine learning model, so there is no "training set."

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K192484

    Validate with FDA (Live)

    Date Cleared
    2020-03-11

    (183 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Fixone Biocomposite Anchors are intended for fixation of suture (soft tissue) to bone in the shoulder, foot/ankle, knee and elbow in the following procedures:

    Shoulder: Rotator Cuff Repair, Bankart Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair. Capsular Shift or Capsulolabral Reconstruction:

    Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair;

    Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair. Posterior Oblique Ligament Repair. lliotibial Band Tenodesis:

    Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction.

    Device Description

    The Fixone Biocomposite Anchor is intended for reattaching soft tissue to bone with sutures. The anchor is manufactured from biodegradable materials (PLGA copolymer and β-ΤCP). A nonresorbable suture manufactured from cobraided UHMWPE and PET fibers is inserted into the anchor is implanted using a provided driver.

    This device is could used with instrument that manufactured by Aju Pharm Co.,Ltd. It is consist of 18 models. It provide non-sterile (user must sterilization before use).

    AI/ML Overview

    The provided text describes a medical device, the "Fixone Biocomposite Anchor," and its 510(k) premarket notification. This document focuses on demonstrating substantial equivalence to a predicate device through bench testing and biocompatibility assessments, rather than presenting a study of AI/algorithm performance. Therefore, many of the requested criteria related to AI studies (e.g., sample size for test/training sets, ground truth establishment, expert adjudication, MRMC studies) are not applicable to this document.

    However, I can extract information related to the acceptance criteria and the studies performed for this medical device.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are primarily based on established international standards (ASTM, ISO, USP) and predefined limits for specific tests. The reported device performance for all listed tests is "Pass," indicating that the device met the established criteria.

    Test ItemAcceptance CriteriaReported Device Performance
    Bench Testing - Anchor Performance
    External surface / MeasurementASTM F2502 and USP<28>Pass
    Insertion torqueN/A (implied to meet predicate performance)Pass
    Fixation strengthN/A (implied to meet predicate performance)Pass
    Tensile strengthN/A (implied to meet predicate performance)Pass
    Extractable colorN/A (implied to meet predicate performance)Pass
    Extraction test - pHThe difference should be 1.5 and less.Pass
    Extraction test - Potassium permanganate reducing substancesThe difference of the consumption of potassium permanganate should be 2.0 mL and less.Pass
    Extraction test - Residue after evaporationRecord the weight of the residue should be 1.0mg and less.Pass
    Extraction test - Heavy metalsAny brown color produced within 10 minutes in the tube containing the extract of the prepared sample does not exceed that in the tube containing the standard lead solution.Pass
    Extraction test - UV spectrum (250nm~350nm)Maximum absorbance between 250 to 350 nm should be 0.1 and less.Pass
    Extraction test - PropertyWhen observing it with the naked eye, test solution should be clear and have no foreign particles.Pass
    Performance comparison test (Fixone vs. Arthrex) - Insertion forceN/A (implied to demonstrate substantial equivalence to predicate)Pass
    Performance comparison test (Fixone vs. Arthrex) - Insertion torqueASTM F2502Pass
    Performance comparison test (Fixone vs. Arthrex) - Pull-out testN/A (implied to demonstrate substantial equivalence to predicate)Pass
    Performance comparison test (Fixone vs. Arthrex) - Fatigue testThe anchor shall be inserted into the pilot hole. Apply the load 50% of Pull-out test (offset 10%, Sine shape) - 30° (worst case)Pass
    Performance Testing of Fixone® Biocomposite Anchor (Pull-out test) - Short Immersion (5min - 24hr)Pull-out test by immersion time of saline solution to evaluate two bioabsorbable suture anchors. ASTM F1839-08, Standard specification for RGID Polyurethane Foam for Use as a Standard Material for Testing Orthopaedic Devices and Instruments. (Specific quantitative criteria not explicitly listed but implied successful comparison).Pass
    Performance Testing of Fixone® Biocomposite Anchor (Pull-out test) - Long Immersion (4 weeks - 26 weeks)Pull-out test by immersion time of saline solution to evaluate two bioabsorbable suture anchors. ASTM F1839-08, Standard specification for RGID Polyurethane Foam for Use as a Standard Material for Testing Orthopaedic Devices and Instruments. (Specific quantitative criteria not explicitly listed but implied successful comparison).Pass
    Shelf-life testingReal-time equivalent (RTE) and Accelerated aging temperature, Taa as per AAMI TIR17 and ASTM F 1980-02. Aging temperature: 55±2℃, aging time: 225 days. (Specific quantitative criteria not explicitly listed but implied successful demonstration of 5-year shelf-life).Pass
    Biocompatibility - Anchor
    CytotoxicityISO 10993-5(2009) Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicityPass
    Acute systemic toxicity testISO 10993-11(2009) Biological evaluation of medical devices - Part 11: Tests for systemic toxicityPass
    Pyrogen TestISO 10993-11 Test for systemic toxicity, pyrogen testPass
    Intracutaneous (intradermal) reactivity testISO 10993-10(2013) Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitizationPass
    Maximization test for delayed hypersensitivityISO 10993-10(2013) Test for irritation and skin sensitization, Maximization test for delayed hypersensitivityPass
    Bacterial reverse mutation testISO 10993-3, Genotoxicity test OECE 471, Bacterial reverse mutation testPass
    Mammalian erythrocyte micronucleus testISO 10993-3, Genotoxicity test OECE 471, Bacterial reverse mutation testPass
    Implantation testISO 10993-6, Tests for local effects after implantation, Annex D test methods for implantation in bonePass
    Bioabsorbable screws testASTM F2502 Standard specification and test methods for bioabsorbable plates and screws for internal fixation implantsPass
    Subchronic toxicity testISO 10993-11 Biological Evaluation of Medical Devices Part 11- Test for systemic toxicityPass
    Biocompatibility - Suture
    CytotoxicityISO 10993-5, Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicityPass
    Systemic toxicity testISO 10993-11, Biological evaluation of medical devices - Part 11: Tests for systemic toxicityPass
    Pyrogen TestISO 10993-11 Test for systemic toxicity, pyrogen testPass
    Intracutaneous reactivity testISO 10993-10, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitizationPass
    Maximization sensitizationISO 10993-10, Test for irritation and skin sensitization, Maximization test for delayed hypersensitivityPass
    Genotoxicity testISO 10993-3, Genotoxicity test OECE 471, Bacterial reverse mutation testPass
    Implantation testISO 10993-6, Tests for local effects after implantation, Annex D test methods for implantation in bonePass
    Hemolysis testISO 10993-4, Biological evaluation of medical devices - Part 4: Selection of tests for interactions with bloodPass
    Biocompatibility - Final Product
    LAL test (Bacterial Endotoxin test)USP 38 NF 33 <85> Bacterial Endotoxin test "Gel clot method"Pass

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document does not specify exact sample sizes for each particular bench test. It indicates that "Bench test were performed" and "The tests demonstrated that the device performs in a substantially equivalent manner to the predicate device." These tests are typically conducted in a laboratory setting.

    • Sample Size: Not explicitly stated as a number of individual anchors or sutures. Each test would have involved a specific number of samples as per the relevant ASTM or ISO standard.
    • Data Provenance: The document states "AJU Pharm Co.,Ltd. / Korea" as the manufacturer. The testing was likely conducted in a laboratory setting, presumably in Korea or by a certified testing facility contracted by the manufacturer. This is prospective data generation through controlled laboratory experiments.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This is not an AI/algorithm study involving expert review of medical images or patient data. The "ground truth" for this device's performance is established by adhering to widely accepted engineering, biocompatibility, and material testing standards (ASTM, ISO, USP).

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable. This is not an AI/algorithm study involving expert adjudication. The "adjudication" is inherent in the pass/fail criteria defined by the standardized testing methods.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a medical device for orthopedic fixation, not an AI diagnostic or assistance tool.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a medical device, not an AI algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" is defined by the technical specifications and performance requirements established by recognized standards organizations (ASTM, ISO, USP) for medical devices of this type. For example:

    • Bench Tests: Mechanical properties (e.g., insertion torque, pull-out strength, fatigue) are measured against predefined engineering limits or compared directly to a predicate device's performance.
    • Biocompatibility Tests: Biological response (e.g., cytotoxicity, systemic toxicity, irritation, implantation response) is assessed against criteria specified in ISO 10993 series for acceptable biological safety.
    • Shelf-life Tests: Stability over time is assessed against physical, chemical, and functional criteria at accelerated and real-time aging conditions according to AAMI TIR17 and ASTM F 1980-02.

    8. The sample size for the training set

    Not applicable. There is no training set as this is not an AI/machine learning device.

    9. How the ground truth for the training set was established

    Not applicable. There is no training set as this is not an AI/machine learning device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K192032

    Validate with FDA (Live)

    Date Cleared
    2020-03-03

    (217 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Fixone Biocomposite Small Anchor are intended to be used fixation to bone in the foot, ankle, knee, hand, wrist, elbow, shoulder, and hip.

    Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clabicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction

    Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Metatarsal Ligament Repair, Hallux Valgus reconstruction, digital tendon transfers. Mid-foot reconstruction

    Knee: Medial Collateral Ligament Repair, Lateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis

    Hand Wrist: Scapholunate Ligament Reconstruction, Carpal Ligament Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PP, DIP, and MCP joints for all digits, digital tendon transfers Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction

    Hip: Capsular repair, Acetabular Labral repair

    Device Description

    The Fixone Biocomposite Small Anchor is intended for reattaching soft tissue to bone with sutures. It consists of Driver, Anchor, Non-absorbable suture and needle. A needle was attached to the end of the suture so that it could penetrate the damaged soft tissue to enable it to be used for OPEN surgery without arthroscopy. The anchor is manufactured from biodegradable materials (PLGA copolymer and beta-TCP). A non absorbable suture manufactured from co-braided UHMWPE and PET fibers is inserted into the anchor is implanted using a provided driver.

    AI/ML Overview

    The provided text is a 510(k) summary for the Fixone Biocomposite Small Anchor, a medical device intended for soft tissue to bone fixation. It details the device's characteristics, intended use, and performance data used to demonstrate substantial equivalence to a legally marketed predicate device.

    Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided document:

    Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the Fixone Biocomposite Small Anchor are primarily based on demonstrating substantial equivalence to a predicate device (Arthrex SutureTak Suture Anchor) through a series of bench tests and biocompatibility evaluations. The "Requirements" column in the table implicitly defines the acceptance criteria, and the "Results" column indicates whether these criteria were met.

    1. Table of Acceptance Criteria and Reported Device Performance:
    Test Item / CategoryAcceptance Criteria (Defined by Standard/Comparison)Reported Performance / Result
    Material Performance
    External surface MeasurementNot explicitly stated (likely visual/dimensional inspection)Pass
    Insertion torqueASTM F2502 and USP <28> (implies meeting specified torque ranges)Pass
    Fixation strengthNot explicitly stated (implied to be equivalent to predicate or safe for intended use)Pass
    Tensile strengthNot explicitly stated (implied to be equivalent to predicate or safe for intended use)Pass
    Extractable colorNot explicitly stated (likely visual inspection for absence of unapproved color)Pass
    Extraction testNot explicitly stated (general requirement for extraction testing)Pass
    pHThe difference should be 1.5 and less.Pass
    Potassium permanganate reducing substancesThe difference of the consumption of potassium permanganate should be 2.0 mL and less.Pass
    Residue after evaporationRecord the weight of the residue should be 1.0mg and less.Pass
    Heavy metalsAny brown color produced within 10 minutes in the tube containing the extract of the prepared sample does not exceed that in the tube containing the standard lead solutionPass
    UV spectrum (250nm~350nm)Maximum absorbance between 250 to 350 nm should be 0.1 and less.Pass
    PropertyWhen observing it with the naked eye, test solution should be clear and have no foreign particles.Pass
    Performance Testing of Fixone® Biocomposite Anchor (Pull-out test)ASTM F1839-08 (Standard for Rigid Polyurethane Foam, implies testing in a relevant bone model with expected retention)Pass
    Biocompatibility (Anchor)
    CytotoxicityISO 10993-5(2009) Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicityPass
    Acute systemic toxicity testISO 10993-11(2009) Biological evaluation of medical devices - Part 11: Tests for systemic toxicityPass
    Pyrogen TestISO 10993-11 Test for systemic toxicity, pyrogen testPass
    Intracutaneous (intradermal) reactivity testISO 10993-10(2013) Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitizationPass
    Maximization test for delayed hypersensitivityISO 10993-10(2013) Test for irritation and skin sensitization, Maximization test for delayed hypersensitivityPass
    Bacterial reverse mutation testISO 10993-3, Genotoxicity test OECD 471, Bacterial reverse mutation testPass
    Mammalian erythrocyte micronucleus testISO 10993-3, Genotoxicity test OECD 471, Bacterial reverse mutation testPass
    Implantation testISO 10993-6, Tests for local effects after implantation, Annex D test methods for implantation in bonePass
    Bioabsorbable screws testASTM F2502 Standard specification and test methods for bioabsorbable plates and screws for internal fixation implantsPass
    Subchronic toxicity testISO 10993-11 Biological Evaluation of Medical Devices Part 11- Test for systemic toxicityPass
    Biocompatibility (Suture)
    CytotoxicityISO 10993-5, Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicityPass
    Systemic toxicity testISO 10993-11, Biological evaluation of medical devices - Part 11: Tests for systemic toxicityPass
    Pyrogen TestISO 10993-11 Test for systemic toxicity, pyrogen testPass
    Intracutaneous reactivity testISO 10993-10, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitizationPass
    Maximization sensitizationISO 10993-10, Test for irritation and skin sensitization, Maximization test for delayed hypersensitivityPass
    Genotoxicity testISO 10993-3, Genotoxicity test OECD 471, Bacterial reverse mutation testPass
    Implantation testISO 10993-6, Tests for local effects after implantation, Annex D test methods for implantation in bonePass
    Hemolysis testISO 10993-4, Biological evaluation of medical devices - Part 4: Selection of tests for interactions with bloodPass

    Study Details:

    The provided document describes a bench testing and biocompatibility testing study conducted to demonstrate the substantial equivalence of the Fixone Biocomposite Small Anchor to its predicate device. This is a common approach for Class II medical devices seeking 510(k) clearance, where clinical trials are often not required if substantial equivalence can be demonstrated through non-clinical data.

    1. Sample Size used for the test set and the data provenance:

      • Sample Size: The document does not explicitly state the specific number of units/samples used for each bench test (e.g., number of anchors for pull-out tests, number of extracts for chemical analysis). It simply states that "Bench test were performed." and "The tests demonstrated that the device performs in a substantially equivalent manner to the predicate device." For biocompatibility, it states "Test result" for each test, implying a sufficient number of test articles were used as per the specified ISO/ASTM standards.
      • Data Provenance: The tests were performed by AJU Pharm Co.,Ltd. in Korea, which is the manufacturer. The data is retrospective, as it's presented after the tests were conducted.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not Applicable. This type of study (bench testing and biocompatibility) does not involve human experts establishing "ground truth" for a test set in the way an AI diagnostic study would where radiologists label images. The "ground truth" here is established by adherence to recognized international standards (ISO, ASTM, USP, OECD) and their inherent testing methodologies, which are designed to be objective and reproducible.
    3. Adjudication method for the test set:

      • Not Applicable. Adjudication methods (like 2+1 or 3+1 by experts) are pertinent to studies involving human interpretation or clinical outcomes, typically in AI/diagnostic device evaluations. For bench testing and biocompatibility, the results are typically quantitative or qualitative "pass/fail" outcomes based on pre-defined standard criteria.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, an MRMC comparative effectiveness study was not done. This type of study is relevant for AI-powered diagnostic devices where human readers (e.g., radiologists) use the device to assist in interpretation. The Fixone Biocomposite Small Anchor is a physical implant, not a diagnostic tool.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • Not Applicable. As a physical medical device (biocomposite anchor), the concept of "standalone algorithm only performance" does not apply. The device's performance is measured directly through physical and chemical properties and biological interactions, not through an algorithm.
    6. The type of ground truth used:

      • The "ground truth" for this device is based on established industry standards and regulatory requirements for medical device safety and performance. This includes:
        • Biocompatibility Standards: ISO 10993 series for biological evaluation of medical devices (e.g., cytotoxicity, systemic toxicity, irritation, sensitization, genotoxicity, implantation).
        • Material and Mechanical Testing Standards: ASTM (e.g., F2502 for bioabsorbable devices, F1839 for foam testing) and USP (pharmacopeial standards).
        • Comparison to Predicate Device: The performance and safety profile of the predicate device (Arthrex SutureTak Suture Anchor) acts as a de facto "ground truth" for establishing substantial equivalence in terms of intended use, principle of operation, and general safety/performance characteristics.
    7. The sample size for the training set:

      • Not Applicable. There is no "training set" in the context of this device's evaluation. This is not an AI/machine learning device that requires training data. The device's design and manufacturing processes are established based on engineering principles and regulatory guidance, not learned from data.
    8. How the ground truth for the training set was established:

      • Not Applicable. Since there is no training set, there is no ground truth established for one.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1