Found 2 results

510(k) Data Aggregation

    K Number
    K973697

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1998-04-08

    (194 days)

    Product Code
    Regulation Number
    886.5916
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CONTEX OK™ (siflufocon A) Rigid Gas Permeable Contact Lens for Orthokeratology is indicated for use in the reduction of myopic refractive error in non-diseased eyes. The lens is indicated for daily wear in an orthokeratology fitting program for the temporary reduction of myopia of up to 3.00 diopters. The lens may be disinfected using a chemical disinfection system only. Note: To maintain the orthokeratology effect of myopia reduction lens wear must be continued on a prescribed wearing schedule.

    Device Description

    The CONTEX OK™ (orthokeratology) contact lens is a rigid gas permeable contact lens in a reverse geometry design. The lens material siflufocon A, is a fluoro silicone acrylate polymer which contains D & C Green #6 as a color additive. The CONTEX OK™ (orthokeratology) contact lens has the following dimensions: Chord Diameter................Approximately 6.5 to 13.0 mm Center Thickness for Low Minus Lens..................................0.10 to 0.30 mm for Plus Lens:........................................0.20 to 0.70 mm Base Curve........................................6.50 to 11.00 mm Secondary Curves........................ 0.1 to 2 mm. Flatter or Steeper than Base Curve Peripheral Curves........................ 0.1 to 2 mm. Flatter or Steeper than Base Curve Powers........................................-10.00 to +5.00 Diopters Aspheric Lens Eccentricity ................................................................................................................................................... (Oblate, Prolate or Tangent Conic) The physical properties of the lens are: Refractive Index...... 1.43 (Nd at 25°) Light Transmittance... > 92.5% (370-760 nm) Wetting Angle ... 24.0 (Contact Receding Angle) Specific Gravity.... 1.25 Hardness.... 82 Water Content... <1% Oxygen Permeability.....(cm²/sec) (ml O₂ x Hg) 81 x10⁻¹¹Dk at 35°C CONTEX OK™ (orthokeratology) contact lenses produce a temporary reduction of myopia by changing the shape (flattening) of the cornea, which is elastic in nature. Flattening the cornea reduces the focusing power of the eye, and if the amount of corneal flattening is properly controlled, it is possible to bring the eye into correct focus and compensate for myopia. Contact lenses rest directly on the corneal tear layer and can influence the corneal shape. After the contact lens is removed, the cornea retains its altered shape for part or all of the remainder of the day. A retainer lens must be used each day to maintain the corneal flattening, or the myopia will revert back to the pre-treatment level.

    AI/ML Overview

    This document describes an orthokeratology contact lens (CONTEX OK™) intended for the temporary reduction of myopia. Below is an analysis of its acceptance criteria and the study performed.

    1. Table of Acceptance Criteria (Implied) and Reported Device Performance

    The acceptance criteria are implied through the intended use and the clinical study results, as no explicit numerical acceptance targets are stated. The device is shown to be effective if it can achieve a temporary reduction in myopic refractive error.

    Acceptance Criteria (Implied from Intended Use)Reported Device Performance
    Safety: Minimal adverse reactions and slit-lamp findings.Adverse Reactions: No eyes experienced adverse reactions, slit-lamp findings requiring treatment, or symptoms/problems/complaints requiring treatment. Slit-Lamp Findings: 188 reports of Grade 1 findings (6 edema, 108 staining, 24 injection, 50 tarsal abnormality). No Grade 2 or higher findings. Findings reduced in incidence over time. Symptoms/Complaints: 425 reports, largest being blur following lens wear (20.2%) and variable vision (17.2%) due to corneal changes; other complaints consistent with RGP lenses.
    Effectiveness: Temporary reduction of myopic refractive error up to 3.00 diopters.Refractive Power Reduction: - Average reduction: 1.69 diopters (range 0.25 to 4.25 D). - 32% of eyes had 0.25-1.00 D reduction. - 32% of eyes had 1.25-2.00 D reduction. - 23% of eyes had 2.25-3.00 D reduction. - 9% of eyes had 3.25-4.00 D reduction. - 1% of eyes had 4.25 D reduction. - 106 out of 110 eyes showed some reduction. - Limit for full reduction to emmetropia: -3.50 diopters. - For initial myopia >3.75D, average reduction was 2.75D.
    Visual Acuity Improvement:Uncorrected Visual Acuity (Final Visit): - 39% of eyes achieved 20/20 or better. - 71% of eyes achieved 20/40 or better. Visual Acuity with contact lenses (Final Visit): - 99 eyes (90%) 20/20 or better - 109 eyes (99%) 20/40 or better - 1 eye 20/70. - Some eyes experienced 1-3 line drops in visual acuity, attributed to residual astigmatism.
    Corneal Flattening (Keratometry):Keratometry: - 89% of eye meridians flattened. - Average flattening: 1.17 diopters. - Greatest flattening: 4.13 diopters. - Corneal astigmatism change: 8% no change, 32% decrease <1D, 41% increase <1D, 16% increase >1D.

    2. Sample Size Used for the Test Set and the Data Provenance

    • Sample Size (Test Set): 138 eyes were initially enrolled, with 136 eyes dispensed contact lenses. 110 eyes completed a minimum of three months of contact lens wear. The performance data is generally reported on these 110 eyes that completed the study.
    • Data Provenance: The document does not explicitly state the country of origin. It describes a "Preclinical Studies: Described in the original 510 (k) submission K941883" and "Clinical Studies" for the K9736097 submission, suggesting the data is part of the regulatory submission for the United States. The study appears to be prospective clinical trial, observing patients over a three-month period.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    The document does not provide information on the number or qualifications of experts used to establish a ground truth for the test set. Clinical studies concerning contact lenses typically rely on measurements taken by trained optometrists or ophthalmologists, but this information is not detailed here.

    4. Adjudication Method for the Test Set

    The document does not describe any specific adjudication method (e.g., 2+1, 3+1, none) for the test set data. Clinical measurements were likely taken by the participating clinicians.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This study is a clinical trial evaluating the safety and effectiveness of a medical device (contact lens) in patients, rather than an AI-assisted diagnostic tool.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    No, a standalone (algorithm only) performance study was not done. This product is a physical medical device (contact lens), not an algorithm or AI system.

    7. The Type of Ground Truth Used

    The "ground truth" for evaluating the contact lens's performance was based on direct clinical measurements and observations of the patients using the device, including:

    • Refractive Power: Measured reduction in myopic refractive error (diopters).
    • Visual Acuity: Measured visual acuity (e.g., 20/20, 20/40).
    • Keratometry: Measured changes in corneal flattening (diopters).
    • Slit-Lamp Findings: Direct observation of ocular health.
    • Adverse Events/Symptoms/Complaints: Patient reported and clinically observed adverse effects.

    8. The Sample Size for the Training Set

    This product is a physical medical device (contact lens), not an AI algorithm. Therefore, there is no "training set" in the context of machine learning. The clinical study enrolled 138 eyes, with 110 eyes completing three months of wear. This would be considered the sample for evaluating the device's performance.

    9. How the Ground Truth for the Training Set Was Established

    As this is not an AI algorithm, there is no "training set" or "ground truth" for a training set in the machine learning sense. The clinical data collected from the 110 eyes that completed the study served as the evidence for safety and effectiveness.

    Ask a Question

    Ask a specific question about this device

    K Number
    K962934

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1996-08-27

    (29 days)

    Product Code
    Regulation Number
    886.5916
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AIRPERM (siflufocon A) Rigid Gas Permeable Spherical, Aspherical, Toric and Multifocal Contact Lens is indicated for daily wear for the correction of refractive ametropia (myopia, hyperopia, astigmatism and presbyopia) in non-aphakic persons with non-diseased eyes. The lens may be disinfected using a chemical disinfection system only.

    Device Description

    THE AIRPERM (siflufocon A) RIGID GAS PERMEABLE MULTIFOCAL CONTACT LENSES FOR DAILY WEAR is an aspheric type multifocal which contains an aspherical (conicoid) curve on either the front or back surface. The lens is manufactured by the same method, i.e., lathing, as that used for the AIRPERM (siflufocon A) RIGID GAS PERMEABLE ASPHERIC CONTACT LENSES FOR DAILY WEAR except that the eccentricity of the lens periphery is made higher in order to impart a multifocal effect into the contact lens. The eccentricity for the aspheric single vision lenses varies from 0 to 0.8 and for the multifocal lenses varies from 0.9 to 1.5. All parameters of the AIRPERM (siflufocon A) RIGID GAS PERMEABLE MULTIFOCAL CONTACT LENSES FOR DAILY WEAR are described and measured in exactly the same way as those for the AIRPERM (siflufocon A) RIGID GAS PERMEABLE ASPHERIC CONTACT LENSES FOR DAILY WEAR. The lenses have the following parameters: Chord Diameter Approximately 6.5 to 11.5 mm, Center Thickness for Low Minus Lens: 0.05 to 0.30 mm, for Plus Lens: 0.10 to 0.70 mm, Base Curve 6.50 to 9.00 mm, Powers -20.00 to +12.00 Diopters, Add Powers +1.00 to + 3.00, Eccentricity 0.1 to 1.5. The physical properties of the lens are: Specific Gravity 1.25, Refractive Index 1.43 (589 nm at 25°), Light Transmittance Clear > 92% (380-780 nm), Tinted >88.0% (380-780 nm), Wetting Angle (CLMA Method) 24.0 ± 2, Water Content <1%, Oxygen Permeability 81 x 10-11Dk at 35°C (cm²/sec) (ml O2 x Hg) Polarographic method of I Fatt, Hardness (Shore) 82. The Airperm (siflufocon A) Multifocal Rigid Gas Permeable Contact Lens is available in the lens material siflufocon A a fluorosilicone acrylate polymer which contains D & C Green #6 as a color additive. The aspheric multifocal functions on the basis of a variation in optical power in the form of increasing plus power from the lens center to the periphery. This serves as a bifocal or multifocal add. The aspheric surface is a member of the conicoid family and is defined by the apical radius and eccentricity, e. The eccentricity value describes the rate of flattening of the posterior aspheric surface or the rate of steepening of the anterior surface. The amount of reading addition is the result of the relationship between the apical radius and the eccentricity value.

    AI/ML Overview

    I am sorry, but the provided text does not contain information about the acceptance criteria for a device, nor does it describe a study proving such criteria.

    The document describes the:

    • AIRPERM (siflufocon A) RIGID GAS PERMEABLE MULTIFOCAL CONTACT LENSES FOR DAILY WEAR including its physical properties and parameters.
    • INDICATIONS (USES) and CONTRAINDICATIONS (REASONS NOT TO USE) for the contact lenses.
    • PRECAUTIONS related to the lenses.
    • SELECTION OF PATIENTS for whom these lenses are suitable.
    • FITTING PROCEDURE OUTLINE detailing steps for pre-fitting examination, initial lens power and diameter selection, and initial base curve selection.
    • Adjunct Solutions used with the lenses.

    Therefore, I cannot extract the requested information regarding acceptance criteria and a corresponding study from this text.

    Ask a Question

    Ask a specific question about this device