510(k) Data Aggregation
(78 days)
Aearo Surgical N95 Respirators, N9508C and N9511 are NIOSH approved N95 respirators intended for single use by operating room personnel and other health care workers to protect both the patients and the health care workers from transfer of microorganisms, blood and body fluids, and airborne particulate materials. This includes use as a procedure mask, isolation mask or dental face mask. This device also meets CDC Guidelines for TB Exposure Control.
These masks are molded cup masks, with a center layer of polypropylene meltblown material sandwiched by inner and outer layers of nonwoven material. The mask has 2 elastic headbands. No fiberglass media is used in this product.
Here's an analysis of the provided text regarding the Aearo Company N9508C and N9511C Surgical N95 Respirators, broken down into the requested sections:
1. Table of Acceptance Criteria and Reported Device Performance
| Performance Characteristic | Test Method | Acceptance Criteria | Reported Device Performance (N9508C and N9511C) |
|---|---|---|---|
| Fluid Resistance Performance (mmHg) | ASTM F-1862 @ 80 mmHg | 29 of 32 must pass (no fluid penetration) | 32 of 32 passed (no fluid penetration) |
| Flammability Class | 16 CFR 1610 | Flame spread must be within upper and lower control limits; meets Class I | No flame spread on 10 of 10 samples, meets Class I |
| Filter Efficiency | NIOSH, 42CFR Part 84 | >95% Efficient | PASS (>95% Efficient, due to NIOSH N95 approval) |
| Breathing Resistance (Exhalation) | NIOSH 42CFR Part 84 | $\le$ 25 mm H2O @ 85 lpm | PASS ($\le$ 25 mm H2O @ 85 lpm, due to NIOSH N95 approval) |
| Breathing Resistance (Inhalation) | NIOSH 42CFR Part 84 | $\le$ 35 mm H2O @ 85 lpm | PASS ($\le$ 35 mm H2O @ 85 lpm, due to NIOSH N95 approval) |
| Bacterial Filtration Efficiency | ASTM F2101-01 | >99.9% | >99.9% on all samples tested |
| Biocompatibility: Cytotoxicity | ISO-10993-1 (Agar Overlay) | Cytotoxicity score of 2 or less | Score of 0 (no detectable reactivity) |
| Biocompatibility: Sensitization | ISO-10993-1 (Closed Patch) | No visible change, Score of 0 | No visible change, Score of 0 (None of the patched animals had any visible change at the test site 48 hours after the challenge dose, using albino guinea pigs) |
| Biocompatibility: Primary Skin Irritation | ISO-10993-1 (Dermal Irritation, 16 CFR 1500 modified) | Negligible, Score of 0 | Negligible, Score of 0 (No irritation evident on intact or abraded skin after 24 or 72 hours, using 3 rabbits) |
2. Sample Size Used for the Test Set and Data Provenance
- Fluid Resistance (ASTM F-1862): 32 masks tested. The provenance is not explicitly stated as retrospective or prospective, but due to it being part of a 510(k) submission for a new device, it would be considered prospective testing specifically for this device. The country of origin for the data is not specified beyond "Nelson Laboratories" which is a US-based lab.
- Flammability (16 CFR 1610): 10 samples tested. (Prospective testing by Nelson Laboratories, US).
- Filter Efficiency (NIOSH 42CFR Part 84): Not explicitly stated, but would be part of the NIOSH certification process for N95 respirators. The NIOSH approval numbers (TC-84A-4268 and TC-84A-4261) indicate the device has met these requirements. (Prospective testing, likely US-based for NIOSH certification).
- Breathing Resistance (NIOSH 42CFR Part 84): Not explicitly stated, but part of NIOSH certification. (Prospective testing, likely US-based).
- Bacterial Filtration Efficiency (ASTM F2101-01): "All samples tested," exact number not specified, but typically involves multiple replicates. (Prospective testing by Nelson Laboratories, US).
- Biocompatibility (ISO-10993):
- Cytotoxicity: "None of the mask samples tested," exact number not specified. (Prospective testing by Nelson Laboratories, US).
- Primary Skin Irritation: 3 rabbits. (Prospective testing by AppTec Laboratory Services, US).
- Delayed Hypersensitivity/Sensitization: Albino guinea pigs (number not specified, but usually several for animal studies). (Prospective testing by Northview Pacific Laboratories, US).
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
For this type of device (Surgical N95 Respirators), the "ground truth" is established through standardized laboratory testing protocols (e.g., ASTM, NIOSH, ISO standards) rather than expert consensus on medical images or clinical outcomes.
- The experts involved are the qualified laboratory personnel performing these rigorous, standardized tests. Their qualifications are inherent in their ability to correctly execute these specific test methods and interpret their results according to the standards. No specific number or individual qualifications are provided for these lab personnel in the document, as the focus is on the adherence to established test methodologies.
4. Adjudication Method for the Test Set
Not applicable for this type of device testing. The results are quantitative measurements or direct observations (e.g., presence/absence of fluid penetration, flame spread, irritation) that either meet or do not meet predefined criteria. There is no subjective interpretation requiring an adjudication process.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices that rely on human interpretation, often with AI assistance, rather than for performance testing of physical protective equipment.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done
This question is not applicable. The device is a physical N95 respirator, not an algorithm or AI system. The performance tests are themselves "standalone" in the sense that they measure the physical properties and performance of the respirator without a human-in-the-loop component for interpreting the device's function.
7. The Type of Ground Truth Used
The ground truth used for these performance tests is based on:
- Standardized Test Methodologies: The acceptance criteria and performance measurements are derived directly from established international and national standards (ASTM, NIOSH, ISO, CFR) for the performance of medical devices and personal protective equipment.
- Physical Measurements and Chemical/Biological Assays: The ground truth is confirmed through objective laboratory measurements and assays (e.g., measuring pressure drop, bacterial filtration efficiency, observing fluid penetration, analyzing cellular reactions).
8. The Sample Size for the Training Set
Not applicable. This device is a physical product (respirator), not an AI or machine learning model that requires a training set.
9. How the Ground Truth for the Training Set was Established
Not applicable, as there is no training set for a physical device.
Ask a specific question about this device
(90 days)
Aearo Surgical N95 Respirators, N9504C, N9504CS and N9514 are intended for single use by operating room personnel and other health care workers to protect both the patients and the health care workers from transfer of microorganisms, blood and body fluids, and airborne particulate materials. This includes use as a procedure mask. isolation mask or dental face mask. This device also meets CDC Guidelines for TB Exposure Control.
These masks are molded cup masks, with a center layer of polypropylene meltblown material sandwiched by inner and outer layers of nonwoven material. The mask has 2 elastic headbands. No fiberglass media is used in this product.
Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance:
| Performance Characteristic | Test Method | Acceptance Criteria | Reported Device Performance (N9504C, N9504CS, N9514C) | Predicate Device Results (Gerson Isolair APR Model 2735) |
|---|---|---|---|---|
| Fluid Resistance | ASTM F-1862 @ 80 mmHg | 29 of 32 show no fluid penetration (from narrative) / 32 of 32 pass (from table) | 32 of 32 pass | 32/32 pass |
| Flammability Class | 21 CFR Part 58 | Flame spread must be within upper and lower control limits / Meets Class I | No flame spread on 10 of 10 samples, meets Class I | Meets Class I |
| Filter Efficiency | NIOSH, 42CFR Part 84 | ≥95% Efficient | Average 97.36% (N9514C), 97.66% (N9504C), and 97.83% (N9504CS) efficient of 20 samples | Average 96.86% efficient of 20 samples |
| Breathing Resistance (Inhalation) | NIOSH 42CFR Part 84 (Inhalation) | ≤35.0 mm H2O @ 85 lpm (implied as max for inhalation) | Average 6.2 mm H2O (N9514C), 7.6 mm H2O (N9504C), and 8.0 mm H2O (N9504CS) @ 85 lpm (average of 3 samples each) | Average 15.2 mm H2O on 3 samples |
| Biocompatibility - Cytotoxicity | ISO-10993-1 / Agar Overlay | Score of 2 or less | Score of 0 | N/A (not provided for predicate) |
| Biocompatibility - Sensitization | ISO-10993-1 / Closed Patch Test | No visible change, Score of 0 | No visible change, Score of 0 | N/A (not provided for predicate) |
| Biocompatibility - Primary Skin Irritation | ISO-10993-1 / 16 CFR 1500 (rabbits) | Negligible, Score of 0 | Negligible, Score of 0 | N/A (not provided for predicate) |
| Bacterial Filtration Efficiency (BFE) | ASTM F2101-01 | Not explicitly stated as acceptance criteria in the table, but narrative states "showed a percent BFE of greater than 99.9% on all samples tested" for adequacy of mask as a respiratory barrier for bacteria. | Greater than 99.9% on all samples tested (using Staphylococcus aureus biological aerosol) | N/A (not explicitly provided for predicate, likely covered by NIOSH certification for predicate as well) |
Notes on Acceptance Criteria:
- The document sometimes states the acceptance criteria in the narrative and sometimes directly in the table. The table combines this information.
- For Fluid Resistance, the narrative states "29 of 32 show no fluid penetration," while the table states "32 of 32 pass." The reported performance of "32 of 32 pass" clearly meets both.
- The breathing resistance acceptance criteria are derived from NIOSH 42CFR Part 84, which specifies maximum allowed resistance for inhalation and exhalation. The provided data are for inhalation.
2. Sample Size Used for the Test Set and Data Provenance:
- Fluid Resistance: 32 masks
- Flammability Class: 10 samples
- Filter Efficiency: 20 samples
- Breathing Resistance: 3 samples for each model (N9514C, N9504C, N9504CS), totaling 9 samples.
- Bacterial Filtration Efficiency (BFE): "all samples tested" (specific number not given, but described using a biological aerosol of Staphylococcus aureus)
- Biocompatibility - Cytotoxicity: "None of the mask samples tested" (specific number not given)
- Biocompatibility - Primary Skin Irritation: 3 rabbits
- Biocompatibility - Delayed Hypersensitivity: Albino guinea pigs (specific number not given, but based on ISO 10993-10, 2002 requirements)
Data Provenance: The studies were conducted by various laboratories, including Nelson Laboratories and AppTec Laboratory Services. The location of these labs (e.g., country of origin) is not specified. The studies appear to be prospective tests performed specifically for this 510(k) submission, confirming the device's performance against established standards.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:
This section is not applicable to this type of device (N95 respirators) and testing. The "ground truth" for these performance characteristics (fluid resistance, flammability, filtration efficiency, biocompatibility) is established by adherence to standardized, objective test methods (e.g., ASTM, NIOSH, ISO) and laboratory measurements, not by expert consensus or interpretations of complex data like medical images. The expertise lies in conducting the tests according to the specified protocols and interpreting the objective results.
4. Adjudication Method for the Test Set:
This section is not applicable. Adjudication methods like "2+1" or "3+1" are typically used in studies where individual expert assessments need to be combined or resolved, such as in clinical image interpretation or diagnostic evaluations. The tests performed for this device are objective laboratory measurements with numerical or categorical outcomes (e.g., pass/fail, percentage, score of X).
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
This section is not applicable. MRMC studies are used to evaluate the diagnostic performance of medical devices that involve human interpretation, often assisted by AI. This submission is for an N95 respirator, which is a physical barrier device, not a diagnostic tool requiring human interpretation or AI assistance.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
This section is not applicable. "Standalone performance" refers to the performance of an AI algorithm without human involvement. As stated above, this device is not an AI algorithm or a diagnostic tool.
7. The Type of Ground Truth Used:
The ground truth for these tests is based on objective, standardized laboratory measurements and physical property assessments using internationally recognized standards (ASTM, NIOSH, ISO).
- Fluid Resistance: Direct measurement of liquid penetration.
- Flammability: Observation of flame spread under controlled conditions.
- Filter Efficiency: Measurement of particle penetration through the material.
- Breathing Resistance: Measurement of pressure drop across the respirator.
- Bacterial Filtration Efficiency: Measurement of bacterial penetration.
- Biocompatibility: In vitro (cytotoxicity) and in vivo (irritation, sensitization on rabbits/guinea pigs) biological responses as per ISO standards.
8. The Sample Size for the Training Set:
This section is not applicable. The N95 respirator is a manufactured physical product, not a learning algorithm. Therefore, there is no "training set" in the context of machine learning or AI.
9. How the Ground Truth for the Training Set Was Established:
This section is not applicable for the same reason as point 8.
Ask a specific question about this device
(60 days)
The Pleats Plus ™ 1054 and 1054S N95 Respirators and Surgical Masks are intended for single use by operating room personnel and other health care workers to protect both the patients and the health care workers from transfer of microorganisms, blood and body fluids, and airborne particulate materials. This includes use as a procedure mask, isolation mask or dental face mask. This device also meets CDC Guidelines for TB Exposure Control.
These masks are pleated, 3-ply masks, with a center layer of polypropylene meltblown material sandwiched by inner and outer layers of nonwoven material. The mask has 2 braided synthetic elastic headbands and a flexible wire tie nosepiece that allows the respirator to form to the bridge of the wearers nose. No fiberglass media is used in this product.
The Abbreviated 510(K) for Aearo Company Pleats Plus™ 1054 and 1054S Surgical N95 Respirators (K063013) describes the acceptance criteria and the studies performed to demonstrate that the device meets these criteria.
1. Table of Acceptance Criteria and Reported Device Performance
| Performance Characteristics | Test Method | Acceptance Criteria / Predicate Device Results | Reported Device Performance (Pleats Plus 1054 and 1054S) |
|---|---|---|---|
| Fluid Resistance Performance | ASTM 1862-00a @ 120mmHg | 29 of 32 to pass / PASS | PASS |
| Flammability Class | 21 CFR part 58 | Flame spread must be within upper and lower control limits / No flame spread on 10 of 10 samples, meets Class I | No flame spread on 10 of 10 samples, meets Class I |
| Filter Efficiency | NIOSH, 42CFR Part 84 | ≥95% Efficient / Average 99.11% efficient of 20 samples | Average 98.45% efficient for 20 samples |
| Breathing Resistance | NIOSH 42CFR Part 84 | ≤35.0 mm H2O @ 85 lpm / average 3.5 mm H2O @ 85 lpm for 3 samples | Average 1.1 mm H2O @ 85 lpm for 3 samples |
| Biocompatibility: Cytotoxicity | ISO-10993-1, Agar overlay method | Score of 2 or less / Score of 0 | Score of 0 |
| Biocompatibility: Sensitization | ISO 10993-10:2002 (study based on) | No visible change, Score of 0 | Score of 0 |
| Biocompatibility: Primary Skin Irritation | ISO 10993-10:2002 (rabbits) | Negligible, Score of 0 | Score of 0 |
| Bacterial Filtration Efficiency (BFE) | ASTM F2101-01 (Staphylococcus aureus) | (Not specified as acceptance criteria in table, but reported as an additional test) | Greater than 99.9% on all samples tested |
2. Sample Sizes Used for Each Test and Data Provenance
| Test / Performance Characteristic | Sample Size Used | Data Provenance / Testing Body |
|---|---|---|
| Fluid Resistance Performance | 32 samples (implied from "29 of 32 to pass") | Not explicitly stated, likely Nelson Laboratories (common for such tests) |
| Flammability Class | 10 samples | Nelson Laboratories |
| Filter Efficiency | 20 samples | NIOSH (for certification), specific lab not stated for comparative |
| Breathing Resistance | 3 samples | NIOSH (for certification), specific lab not stated for comparative |
| Biocompatibility: Cytotoxicity | Not specified | Nelson Laboratories |
| Biocompatibility: Sensitization | Not specified | AppTec Laboratory Services (rabbits) |
| Biocompatibility: Primary Skin Irritation | Not specified | AppTec Laboratory Services (rabbits) |
| Bacterial Filtration Efficiency (BFE) | Not specified | Nelson Laboratories |
The data provenance is from laboratory testing (e.g., Nelson Laboratories, AppTec Laboratory Services, NIOSH certification). These are prospective tests performed specifically for device evaluation. The country of origin for the manufacturing facility is Handan Hengyong Protective & Clean Products Co., LTD. Jiankang East Street Yongnian County High-Tech Development Area Hebei province, P.R. China, however, the testing laboratories are US-based (Nelson Labs and AppTec).
3. Number of Experts and Qualifications for Establishing Ground Truth
This document describes the testing of a physical medical device (N95 respirator) against established performance standards and methods. It does not involve a "ground truth" expert consensus in the same way a diagnostic imaging AI algorithm would. Instead, the "ground truth" is defined by:
- Standardized test methods: ASTM 1862-00a, 21 CFR part 58, NIOSH 42CFR Part 84, ISO-10993-1, ASTM F2101-01.
- Regulatory bodies and accredited laboratories: NIOSH approval TC-84A-4320, Nelson Laboratories, AppTec Laboratory Services.
Therefore, the concept of "number of experts used to establish ground truth" with specific qualifications (like radiologists) is not applicable here. The "experts" are the scientists and technicians in the accredited labs performing the standardized tests according to the protocols set by regulatory bodies.
4. Adjudication Method for the Test Set
The concept of an adjudication method (like 2+1, 3+1) is typically relevant for studies where human interpretation or consensus is required (e.g., medical image reading). For the physical testing of the N95 respirator, the results are quantitative and objective measurements against predefined acceptance criteria. Therefore, no adjudication method as described is applicable or performed in this context. The results are reported directly from the laboratory tests.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is relevant for evaluating the impact of AI algorithms on human reader performance, typically in diagnostic tasks. This submission is for a physical device (N95 respirator), not an AI-powered diagnostic tool.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
No, this question is not applicable as the device is a physical N95 respirator, not an algorithm or software. Therefore, there is no "standalone algorithm" performance to report.
7. Type of Ground Truth Used
For this device, the "ground truth" is based on:
- Standardized physical and chemical test results: Objective measurements demonstrating performance against established industry standards and regulatory requirements (e.g., ASTM, NIOSH, ISO, CFR).
- Predicate device comparison: The device's performance is also compared against a legally marketed predicate device (Pleats Plus 1050), establishing substantial equivalence based on meeting similar performance characteristics.
8. Sample Size for the Training Set
This question is not applicable as the device is a physical N95 respirator, not a machine learning algorithm that requires a training set.
9. How the Ground Truth for the Training Set Was Established
This question is not applicable as the device is a physical N95 respirator and does not use a training set in the context of machine learning.