Search Filters

Search Results

Found 8 results

510(k) Data Aggregation

    K Number
    K223269

    Validate with FDA (Live)

    Date Cleared
    2023-05-18

    (206 days)

    Product Code
    Regulation Number
    882.1330
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AD-TECH Depth Electrodes (Spencer Probe Depth Electrodes) are intended for temporary (< 30 days) use with recording, monitoring and stimulation equipment for the recording, monitoring and stimulation of electrical signals at the subsurface level of the brain. The recording of electrical activity supports definition of the location of epileptogenic foci and brain mapping.

    Device Description

    The Depth Electrodes (Spencer Probe Depth Electrode) are intended for recording, monitoring and stimulation at sub-surface levels of the brain. These electrodes are provided sterile, disposable and single patient use. The Depth Electrodes (Spencer Probe Depth Electrodes) provide the patient contact device. The Depth Electrodes (Spencer Probe Depth Electrodes) connect to an electrode cable that is applied to the user's equipment. The electrodes are used by physicians.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study information for the Spencer Probe Depth Electrodes, as extracted from the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Hazard AddressedTest Method UsedAcceptance CriterionReported Device Performance (Summary of Test Results)
    Image ArtifactASTM F2119-13No Criteria, Descriptive Statement1.5T: Adoption from 3.0T tests since 3.0T is the worst-case. 3.0T: Image distortion of 19 mm from the edge of the device.
    Magnetically Induced Displacement ForceASTM F2052-15<45° deflection1.5T: Adoption from 3.0T tests since 3.0T is the worst-case. 3.0T: pass (meaning deflection was less than 45°).
    Magnetically Induced TorqueASTM F2213-17, Low friction surface method<45° rotation1.5T: Adoption from 3.0T tests since 3.0T is the worst-case. 3.0T: pass (meaning rotation was less than 45°).
    RF Induced HeatingASTM F2182-19e2<6°C1.5T & 3.0T: Under the conditions defined in the labeling, the MR induced heating will be less than <6°C.
    Accessory Stay Flange ConductivityN/A (Descriptive Test)Negligible conductivityResults showed that conductivity was negligible, and use of the optional accessory Stay Flange would not lead to any additional safety concerns in association with MRI.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size:
      • Image Artifact, Magnetically Induced Displacement Force, Magnetically Induced Torque: "The device with the maximum mass/linear length" (for image artifact) and "The device with maximum mass" (for displacement force and torque) were tested. This implies a single representative model or a limited number of "worst-case" configurations. The exact number of individual devices or instances is not specified, but it's likely a small, controlled sample to demonstrate compliance with MR safety.
      • RF Induced Heating: "devices with different lengths, different electrode designs, and different insertion depths were studied." Again, the exact number isn't quantified, but it indicates a range of device configurations were tested.
    • Data Provenance: The studies were conducted by "Ad-Tech, in conjunction with University of Houston." The data is thus likely laboratory-generated performance data specific to the device under controlled testing conditions, not derived from patient data (retrospective or prospective). The country of origin of the data would be the USA (University of Houston).

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This type of information is not applicable to the provided document. The studies described are engineering performance tests related to MR safety, not clinical studies requiring expert ground truth for diagnostic or therapeutic accuracy. The acceptance criteria are based on established engineering standards (ASTM and ISO).

    4. Adjudication Method for the Test Set

    This is not applicable as the studies are technical performance tests with objective, measurable criteria, not clinical assessments requiring adjudication of human readings or interpretations.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size of AI Improvement

    This is not applicable. The document describes the technical performance and safety testing of a medical device (depth electrodes) in the MR environment, not an AI-powered diagnostic or therapeutic tool that would typically undergo MRMC studies.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    This is not applicable. The device is a physical medical instrument, not an algorithm.

    7. Type of Ground Truth Used

    The "ground truth" for these tests are the objective measurements against established engineering and safety standards:

    • Physical properties: Measured displacement, rotation, and temperature changes in a controlled MR environment.
    • Standard compliance: The tests aim to demonstrate compliance with ASTM and ISO standards for MR safety. The "ground truth" is therefore adherence to these physical principles and regulatory standards.

    8. Sample Size for the Training Set

    This is not applicable. The device is a physical electrode, not a machine learning model, so there is no training set involved.

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable for the same reason as above.

    Ask a Question

    Ask a specific question about this device

    K Number
    K223276

    Validate with FDA (Live)

    Date Cleared
    2023-05-18

    (206 days)

    Product Code
    Regulation Number
    882.1330
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ad-Tech Anchor Bolts are optional accessories for use with Depth Electrodes. The Anchor Bolts may be applied when it is desired to minimize concerns about potential cerebrospinal fluid (CSF) leakage and infection of the subdural space while stabilizing the electrode. Anchor Bolts are secured in the skull to provide an access point for Depth Electrodes.

    Device Description

    The device under review is a family of Anchor Bolts. Anchor Bolts are optional accessories to Depth Electrodes. These Anchor Bolts provide an optional access point through the skull and stabilization support for Depth Electrodes.

    AI/ML Overview

    The furnished document is an FDA 510(k) Premarket Notification stating substantial equivalence for the "Anchor Bolts as Accessories to Depth Electrodes" device. It outlines the device description, intended use, comparison with a predicate device, and performance data related to its safety in the Magnetic Resonance (MR) environment.

    However, the information provided does not describe an AI medical device or a study involving human readers or AI assistance. The performance data is solely for the physical device's safety in an MR environment, not for an AI algorithm's diagnostic or assistive performance.

    Therefore, I cannot fulfill your request as it pertains to acceptance criteria and studies for an AI medical device, multi-reader multi-case studies, or AI algorithm performance.

    The document discusses the following acceptance criteria and performance for the physical device:

    1. Table of Acceptance Criteria and Reported Device Performance

    Hazard AddressedTest Method UsedAcceptance CriterionMedical Device Configuration TestedSummary of Test Results and pass/fail (if Appropriate)
    Image Artifact (1.5T)ASTM F2119-13No Criteria, Descriptive StatementDevice with the maximum mass/linear lengthAdoption from 3.0T tests since 3.0T is the worst-case.
    Magnetically Induced Displacement Force (1.5T)ASTM F2052-15<45° deflectionThe device with maximum massAdoption from 3.0T tests since 3.0T is the worst-case.
    Magnetically Induced Torque (1.5T)ASTM F2213-17, Low friction surface method<45° rotationThe device with maximum massAdoption from 3.0T tests since 3.0T is the worst-case.
    RF Induced Heating (1.5T)ASTM F2182-19e2<6°CBased on ISO10974 methodology; devices with different lengths, electrode designs, and insertion depths.Under the condition defined in the labeling, the MR induced heating will be less than <6°C
    Image artifact (3.0T)ASTM F2119-13No Criteria, Descriptive StatementDevice with the maximum mass/linear lengthImage distortion of 19 mm from the edge of the device
    Magnetically Induced Displacement Force (3.0T)ASTM F2052-15<45° deflectionThe device with maximum masspass
    Magnetically Induced Torque (3.0T)ASTM F2213-17, Low friction surface method<45° rotationThe device with maximum masspass
    RF Induced Heating (3.0T)ASTM F2182-19e2<6°CBased on ISO10974 methodology; devices with different lengths, electrode designs, and insertion depths.Under the condition defined in the labeling, the MR induced heating will be less than <6°C

    The following information, as requested in your prompt, is not applicable or not available in the provided text because the document is not about an AI/software device:

    1. Sample size used for the test set and the data provenance: Not applicable. This is not a study on an algorithm's performance on a dataset of patient data. It is testing on physical device configurations.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for an AI algorithm is not established here.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    6. The type of ground truth used (expert concensus, pathology, outcomes data, etc): Not applicable.
    7. The sample size for the training set: Not applicable.
    8. How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ad-Tech Subdural Electrodes (Strip/Intraoperative Strip, Grid/Intraoperative Grid, Dual-Sided Interhemissheric, Multi-Strip and Split Grid, Intraoperative Mapping Grid) are intended for temporary (< 30 days) use with recording, monitoring, and stimulation equipment for the recording, monitoring, and stimulation of electrical signals on the surface of the brain. The recording of electrical activity supports definition of epileptogenic foci and brain mapping.

    Device Description

    The device under review is a family of Subdural Electrodes. These electrodes provide surface brain contact to support recording, monitoring and stimulation from user supplied equipment.

    The family of Subdural Electrodes under review are used under the direction of neurosurgeons and other skilled physicians to support their clinical needs for subdural electrodes on the brain. Based upon the needs of their clinical practice and particular patients, various 2-dimensional geometric shapes (length and width) resulting in variations of Subdural Electrode body shapes and orientation configurations are necessary.

    All variations of Subdural Electrodes under review consist of the same materials and fundamental design as the predicate Subdural Electrodes. Either round discs or rectangular electrode contact material are sandwiched between two layers of silicone substrate electrode body material. The brain contacting side of the silicone substrate body has material removed to expose an amount of electrode contact surface area. The subdural electrode wires between the electrode contact and connector contacts at the most distal end of the subdural electrode tail pass through a tube for interface with the user's equipment.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Ad-Tech Subdural Electrodes based on the provided document:

    This document is a 510(k) summary for a medical device, which typically aims to demonstrate substantial equivalence to a predicate device rather than providing extensive clinical study data that would be found in a PMA (Premarket Approval) application. Therefore, you will find information related to design and performance verification rather than large-scale clinical trials.

    1. Table of Acceptance Criteria and Reported Device Performance

    Device: Ad-Tech Subdural Electrodes (Strip/Intraoperative Strip, Grid/Intraoperative Grid, Dual-Sided Interhemispheric, Multi-Strip and Split Grid, Intraoperative Mapping Grid)

    Acceptance Criteria / Performance RequirementReported Device Performance
    Electrical Resistance: Measure electrical resistance from the electrode contact to connector. Verification of electrical resistance from electrode contact to connector meet specification, no electrical contact or connector cross connects to another electrode contact or connector.Pass: Verification of electrical resistance from electrode contact to connector meet specification, no electrical contact or connector cross connects to another electrode contact or connector. (The specific resistance values/specifications are not provided in this summary but are implied to be met).
    Dimensional Characteristics: Dimensional requirements are met by manufacturing. Verification of the dimensional characteristics, electrode contact placement is confirmed using the same manufacturing methods and processes as the predicate electrodes.Pass: Verification of the dimensional characteristics, electrode contact placement is confirmed using the same manufacturing methods and processes as the predicate electrodes. (The specific dimensional requirements are not provided, but the document mentions modifications like range of contact spacing (1-20mm) and range of round electrode contact diameters (1.96-6.35mm), and exposed surface area (1.08-19.6 mm²), all confirmed through manufacturing methods.)
    Stimulation Effect: Stimulation at 30 µC/cm² does not affect the electrodes. Verification that under the anticipated duration of electrical stimulation at the maximum limit does not affect the electrode resistance.Pass: Verification that under the anticipated duration of electrical stimulation at the maximum limit does not affect the electrode resistance. (Confirmed that charge density limits of < 30 µC/cm² for the electrode contact exposed diameter surface area are discussed in the Directions for Use, and that no change in Subdural Electrode resistance was verified for durations of maximum stimulation charge density.)
    Shelf Life: Confirmation of subdural electrode functional performance and sterile barrier pouch integrity (seal peel per ASTM F88 and bubble leak per ASTM F2096) after 2-year aging and simulated distribution per ASTM 4169-16 and bubble leak (ASTM F2096) after accelerated aging (ASTM F1980) and real-following time aging of two years and after simulated distribution by ASTM 4169-16.Pass: Verified by application of referenced standards and conformance with specifications. (Specific test results are not detailed, but adherence to ASTM F88, F2096, F4169-16, and F1980 is claimed, demonstrating 2-year shelf life and barrier integrity.)
    Sterility: EO sterilization of the subject device was validated according to ISO 11135:2014 to demonstrate that it is provided with a minimum sterility assumed level (SAL) of 10-6.Pass: EO sterilization of the subject device was validated according to ISO 11135:2014 to demonstrate that it is provided with a minimum sterility assumed level (SAL) of 10-6.
    Biocompatibility: (Not changed from predicate device, therefore no new testing required for this submission.)N/A: Biocompatibility was not provided to support a determination of substantial equivalence for this submission because material and manufacturing processes have not been changed from those were cleared with the predicate device.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a "test set" in the context of clinical data for algorithmic performance. Instead, it details performance tests conducted on the device hardware.

    • Sample Size for Performance Tests: The document does not explicitly state the number of devices or samples used for each performance test (Electrical Resistance, Dimensional Characteristics, Stimulation Effect, Shelf Life, Sterility). It implies that sufficient samples were tested to demonstrate compliance.
    • Data Provenance: The data provenance is laboratory testing and manufacturing verification of the device's physical and electrical characteristics, adhering to industry standards (e.g., ASTM, ISO). This is not retrospective or prospective clinical data from human subjects.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts

    This information is not applicable as the submission is for a hardware device (electrodes) and focuses on engineering/performance verification rather than a diagnostic algorithm requiring ground truth established by medical experts. The "ground truth" for these tests would be the established engineering specifications and international standards.

    4. Adjudication Method for the Test Set

    The concept of an "adjudication method" (like 2+1 or 3+1) is relevant for clinical studies where multiple human readers assess cases and their disagreements are resolved. This is not applicable to the performance verification tests described for these physical electrodes. The "adjudication" in this context is the comparison of test results against predefined specifications and industry standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. The submission explicitly states: "Clinical data was not provided to support a determination of substantial equivalence." This type of study would be for evaluating the performance of a diagnostic or assistive AI algorithm, which is not the subject of this 510(k).

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Study was done

    No, a standalone algorithm study was not done. This submission is for physical medical electrodes, not an algorithm.

    7. The Type of Ground Truth Used

    For the performance tests described, the "ground truth" is based on:

    • Established engineering specifications for electrical resistance and dimensional characteristics.
    • Recognized international standards (e.g., ISO 11135:2014 for sterilization, ASTM F88, F2096, F4169-16, F1980 for shelf life and packaging).
    • Clinically established physiological limits (e.g., charge density limits of < 30 µC/cm² for brain tissue).

    8. The Sample Size for the Training Set

    This information is not applicable as this 510(k) submission is for a physical medical device (electrodes) and does not involve an AI algorithm with a training set.

    9. How the Ground Truth for the Training Set was Established

    This information is not applicable as this 510(k) submission is for a physical medical device (electrodes) and does not involve an AI algorithm with a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K181544

    Validate with FDA (Live)

    Date Cleared
    2018-08-09

    (58 days)

    Product Code
    Regulation Number
    882.1330
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ad-Tech Anchor Bolts are optional accessories for use with Depth Electrodes. The Anchor Bolts may be applied when it is desired to minimize concerns about potential cerebrospinal fluid (CSF) leakage and infection of the subdural space while stabilizing the electrode.

    Anchor Bolts are secured in the skull to provide an access point for Depth Electrodes.

    Device Description

    The device under review is a family of Anchor Bolts. Anchor Bolts are optional accessories to Depth Electrodes. These Anchor Bolts provide an optional access point through the skull and stabilization support for Depth Electrodes.

    AI/ML Overview

    The provided text describes the 510(k) premarket notification for "Anchor Bolts as Accessories to Depth Electrodes." This document primarily focuses on demonstrating substantial equivalence to a predicate device through non-clinical performance testing. It does not describe a software-as-a-medical-device (SaMD) or AI/ML-based device that would require the kind of complex clinical study with human readers and ground truth adjudication methods typically seen for such applications.

    Therefore, many of the requested criteria, such as expert opinions, MRMC studies, standalone algorithm performance, and training set information, are not applicable or mentioned in this type of submission. The information provided is for a physical medical device.

    However, I can extract the relevant "acceptance criteria" (performance requirements for the device) and the "study" (non-clinical performance tests) that prove the device meets these criteria.


    Device Name: Anchor Bolts as Accessories to Depth Electrodes

    Device Type: Physical Medical Device (Accessory to Depth Electrodes)

    Study Focus: Non-clinical performance tests to demonstrate substantial equivalence to a predicate device.


    Here's the information as it relates to the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Test (Acceptance Criteria)Reported Device Performance (Result)
    Torque required to insert to pre-drilled skull hole:Pass: All results ≤ 1 in-lbs. (Acceptance Criteria: ≤ 8 in-lbs.)
    Axial tension to pull Anchor Bolt out of pre-drilled skull hole:Pass: All results > 48 lbf. (Acceptance Criteria: ≥ 40 lbf.)
    Axial tension to pull Depth Electrode from the Anchor Bolt:Pass: All results > 140 grams. (Acceptance Criteria: ≥ 100 grams.)
    Anchor Bolt gasket resist CSF leakage:Pass: No leaks observed after 24 hours. (Acceptance Criteria: No leakage at the Depth Electrode, Anchor Bolt after 24 hours, under 15 mmHg hydrostatic pressure.)
    Sterility:Pass: Terminally ethylene oxide sterilized devices by a validated EtO sterilization cycle, per ANSI/AAMI/ISO 11135-1:2007 half-cycle method. The sterility assurance level (SAL) is ≤ SAL of 10-6.
    Sterile barrier integrity:Pass: Evaluated by simulated distribution, seal peel, and bubble emission testing for the double Tyvek pouch after terminal sterilization.
    Anchor Bolt silicone gasket shelf life:Pass: The same acceptance criteria were met. (Ability of real-time aged Anchor Bolts with Silicone Gaskets to meet Depth Electrode retention requirements and minimize cerebrospinal fluid (CSF) leakage were evaluated to the methods used for Axial tension to pull Depth Electrode from the Anchor Bolt and Anchor Bolt gasket resist CSF leakage.)
    Biocompatibility:Pass: Following test methods applied to final finished form Anchor Bolts (patient contacting materials), with passing results: Cytotoxicity, Sensitization, Intracutaneous Reactivity, Acute Systemic Toxicity, Material-Mediated Pyrogenicity, Genotoxicity (Salmonella Typhimurium and Escherichia Coli Reverse Mutation Assay, Mouse Lymphoma Mutagenesis Assay), Hemocompatibility, Neurotoxicity, Subacute/Subchronic Toxicity, Implantation (Neurological (Brain) Implantation Test).

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: For the mechanical performance tests (Torque, Axial tension, CSF leakage), a sample size of N=10 was generally used. For the CSF leakage test, N=22 samples were used. Other tests like sterility, sterile barrier integrity, shelf life, and biocompatibility applied broader methods/principles to the devices.
    • Data Provenance: The data is from non-clinical bench testing conducted by the manufacturer, Ad-Tech Medical Instrument Corporation. The document does not specify the country of origin for the data beyond the company's address in Oak Creek, WI, USA. This is retrospective data from device testing.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    • Not Applicable. This submission is for a physical medical device and relies on engineering and material science testing methods (e.g., tensile testers, torque wrenches, established biocompatibility protocols) rather than human expert interpretation of medical images or clinical outcomes. Therefore, no "medical experts" in the sense of radiologists or similar were used to establish ground truth for these non-clinical performance tests.

    4. Adjudication Method for the Test Set:

    • None. As the tests are non-clinical, objective measurements, there is no human interpretation or "adjudication" required in the way it would be for a diagnostic AI. The results are quantitative and pass/fail based on pre-defined thresholds.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

    • No. This is not a diagnostic imaging device or an AI/ML-based device. Thus, MRMC studies are not relevant to demonstrating the safety and effectiveness of an anchor bolt for depth electrodes.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not Applicable. This is a physical medical device, not a software algorithm.

    7. The Type of Ground Truth Used:

    • The "ground truth" for the performance tests is based on objective, quantitative measurements against defined engineering specifications and industry standards for mechanical performance, sterility, and biocompatibility. For instance, the "ground truth" for axial tension is the measured force in lbf, compared to a ≥ 40 lbf requirement. The "skull bone" simulation was polyurethane foam (40 pcf).

    8. The Sample Size for the Training Set:

    • Not Applicable. There is no "training set" as this is not an AI/ML device.

    9. How the Ground Truth for the Training Set was Established:

    • Not Applicable. There is no "training set."
    Ask a Question

    Ask a specific question about this device

    K Number
    K170442

    Validate with FDA (Live)

    Date Cleared
    2017-09-07

    (205 days)

    Product Code
    Regulation Number
    882.4310
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Cranial Drill Bits and accessories are intended to be used for drilling holes in the skull for neurological procedures, such as brain biopsy, brain contacting electrode and electrode accessory device placement.

    Device Description

    The device under review is a family of cranial drill bits and accessories. These cranial drill bits and accessories are applied to create a hole through the skull in support of neurological procedures. Cranial drill bits are twist drill bits used for drilling holes in the skull, supporting access to the brain for brain biopsy, brain contacting electrode placement, electrode accessory placement such as Anchor Bolts or other needs as determined by the user. The cranial drill bit variations are 16 cm to 30 cm length and 2.4 mm to 3.2 mm outer diameter. Accessories to the cranial drill bits include: Drill Stop, Drill Stop Wrench, Drill Sleeve Guide, Guide Block.

    AI/ML Overview

    The provided document is a 510(k) summary for Ad-Tech Medical Instrument Corporation's Cranial Drill Bits and Accessories. It outlines the device's indications for use, technological characteristics, and performance tests for demonstrating substantial equivalence to predicate devices. This document does not pertain to an AI/ML powered medical device, an autonomous AI device, or other medical device software. Therefore, the questions regarding acceptance criteria and studies for AI-powered devices are not applicable.

    Here's an analysis of the provided information, addressing the closest relevant aspects:

    Acceptance Criteria and Device Performance for (Non-AI) Cranial Drill Bits and Accessories

    The acceptance criteria are not explicitly titled as "acceptance criteria" but are embedded within the "Summary of Requirement" for the performance tests conducted. The device performance is reported as "Pass" for all listed tests.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Summary of Requirement)Reported Device Performance
    Drill through simulated skull material in < 1 min.Pass
    Comparison of drilled hole diameter to the desired hole size (with Guide Block, Drill Sleeve Guide and Drill Stop).Pass
    Comparison of drilled hole diameter to the desired hole size (without Guide Block, Drill Sleeve Guide and Drill Stop).Pass
    Guide Block compatible with stereotactic frame.Pass

    2. Sample size and data provenance for the test set

    The document does not specify the sample size used for the tests or the provenance of any data (e.g., country of origin, retrospective/prospective). It only mentions "simulated skull material" for the drilling efficiency test.

    3. Number of experts and qualifications for ground truth establishment

    This information is not available in the provided document. As this device is mechanical and not an AI/ML product, expert review in this context would likely be related to design validation and usability, not establishing ground truth for algorithmic performance.

    4. Adjudication method for the test set

    This information is not available in the provided document. Given the nature of the device (mechanical drill bits), typical adjudication methods like 2+1 or 3+1 used for diagnostic AI are not applicable.

    5. Multi-reader multi-case (MRMC) comparative effectiveness study results

    A multi-reader multi-case (MRMC) comparative effectiveness study was not done as this is a mechanical device, not an AI-powered device intended for diagnostic or assistive interpretation by human readers. Therefore, there is no effect size of human readers improving with AI assistance to report.

    6. Standalone (algorithm only without human-in-the-loop performance) study results

    A standalone performance study was not done because this is a mechanical device and does not involve an algorithm.

    7. Type of ground truth used

    The ground truth used for these performance tests appear to be based on engineering specifications and measurable physical properties rather than clinical outcomes, expert consensus, or pathological diagnoses. For example, "drilling through simulated skull material in < 1 min" uses a timing measurement against a specified material, and "comparison of drilled hole diameter to the desired hole size" relies on a direct physical measurement against a设计 specification.

    8. Sample size for the training set

    This information is not applicable as this is a mechanical device and does not involve a training set for an AI/ML algorithm.

    9. How the ground truth for the training set was established

    This information is not applicable as this is a mechanical device and does not involve a training set for an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K163355

    Validate with FDA (Live)

    Date Cleared
    2017-04-12

    (133 days)

    Product Code
    Regulation Number
    882.1330
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ad-Tech Depth Electrodes (Depth Electrodes, Foramen Ovale Depth Electrodes, Macro Micro Depth Electrodes, Spencer Probe Depth Electrodes, Wyler Sphenoidal Depth Electrodes) are intended for temporary (< 30 days) use with recording, monitoring and stimulation equipment for the recording, monitoring and stimulation of electrical signals at the subsurface level of the brain. The recording of electrical activity supports definition of epileptogenic foci and brain mapping.

    Device Description

    The device under review is a family of Depth Electrodes. These electrodes provide sub-surface brain contact to support recording, monitoring and stimulation from user supplied equipment.

    AI/ML Overview

    This document describes the Ad-Tech Medical Instrument Corporation's Depth Electrodes (K163355) and their substantial equivalence to predicate devices, rather than a study proving the device meets specific acceptance criteria in the context of an AI/ML medical device.

    Therefore, most of the requested information regarding acceptance criteria, study design, expert involvement, and ground truth for an AI/ML device is not applicable to this 510(k) summary for a traditional medical device.

    However, I can extract the acceptance criteria (defined as "Summary of Requirement" in the document) and the reported performance ("Result") for the physical device testing.

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criterion (Summary of Requirement)Reported Device Performance (Result)
    Measure electrical resistance from the electrode contact to connector.Pass
    Dimensional requirements are met by manufacturing.Pass
    Stimulation at 30 $\mu$ C/cm2 does not affect the electrodes.Pass
    The optional accessory Stay Flange supports adequate Depth Electrode retention.Pass

    The following information is not applicable to this document as it pertains to an AI/ML software device, and this 510(k) is for a set of physical medical electrodes:

    1. Sample size used for the test set and the data provenance.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts.
    3. Adjudication method for the test set.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance.
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc).
    7. The sample size for the training set.
    8. How the ground truth for the training set was established.

    Additional relevant information from the document:

    • Intended Use: The Ad-Tech Depth Electrodes are intended for temporary (< 30 days) use with recording, monitoring, and stimulation equipment for the recording, monitoring, and stimulation of electrical signals at the subsurface level of the brain. The recording of electrical activity supports the definition of epileptogenic foci and brain mapping.
    • Predicate Devices: The device was deemed substantially equivalent to K053358, K964644, K990788, K041604, K802151, and K151790 (all previous Depth Electrodes from Ad-Tech Medical Instrument Corporation or PMT Corporation).
    • Study Type: The submission focused on demonstrating substantial equivalence based on technical characteristics and performance testing of the physical device, rather than clinical trials or AI performance evaluations.
    Ask a Question

    Ask a specific question about this device

    K Number
    K152769

    Validate with FDA (Live)

    Device Name
    FREMAP ELECTRODE
    Date Cleared
    2016-05-06

    (224 days)

    Product Code
    Regulation Number
    874.1820
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The FREMAP Cranial Nerve Electrodes are designed for intraoperative (=< 10 hours) monitoring/recording from and stimulation to the facial nerve or facial nerve root at the CP angle during skull based surgical procedures when comnected to the MEE-1000A or MEE-2000 Neural Function Measuring System.

    Device Description

    The device under review is a Facial nerve Root Evoked Muscle Action Potential (FREMAP) electrode. The electrode is used to elicit the muscle action potentials by electric stimulation directly to the facial nerve. It is commonly placed on the facial nerve or facial nerve root at the CP-angle and is used for monitoring (elicit) the muscle action potential response to preserve the facial nerve function especially during acoustic neuroma surgery or tumor removal (such as meningioma) at the CP-angle. The FREMAP electrodes are non-active and are intended to be connected to Nihon Kohden Corporation's Neural Function Measuring System Neuromaster MEE-1000A or MEE-2000

    AI/ML Overview

    The document provided is a 510(k) premarket notification for a medical device called the "FREMAP Cranial Nerve Electrode." It compares the device to several predicate devices to demonstrate substantial equivalence, rather than presenting a study where the device's performance is measured against acceptance criteria in a clinical or simulated clinical setting.

    Therefore, the requested information regarding acceptance criteria, reported device performance, sample sizes, expert ground truth, adjudication methods, MRMC studies, standalone performance, training set details, and ground truth establishment cannot be fully provided from the given document.

    However, I can extract the performance tests performed to demonstrate substantial equivalence, which serve as a form of acceptance criteria in the context of this 510(k) submission.

    Here's a breakdown of what can be extracted and what cannot:

    1. A table of acceptance criteria and the reported device performance:

    The document lists performance tests conducted to establish substantial equivalence. It states that "Test requirements were met or a process monitor is in place to confirm compliance." Specific quantitative results or typical acceptance ranges for these tests are not provided in the document.

    TestSummary of RequirementReported Device Performance
    Electrical ResistanceMeasure electrical resistance from the electrode contact to connector.Met (or process monitor in place to confirm compliance)
    FREMAP electrode Connector compatibility with monitor equipmentVerify the FREMAP electrode Connector is able to connect to neural function measuring systems (MEE-1000A or MEE-2000).Met (or process monitor in place to confirm compliance)
    Electrical continuity and resistance after tensile testingVerify electrical continuity and resistance properties are met after tensile load was applied to the FREMAP electrode.Met (or process monitor in place to confirm compliance)
    Electrode separation under tensile testEvaluate risk of electrode head separation with a load under tensile test.Met (or process monitor in place to confirm compliance)
    Electrical continuity and resistance after simulated use.Apply a simulated stimulation potential for the 10 hour anticipated duration of intraoperative use.Met (or process monitor in place to confirm compliance)
    Overall lengthConfirm dimensional requirements are met.Met (or process monitor in place to confirm compliance)
    Biocompatibility (Cytotoxicity, Sensitization, Intracutaneous Test and Acute System Toxicity)Evaluations for these specific biological effects of manufactured and sterilized FREMAP Electrodes.Performed and presumably met, though specific results not listed.

    2. Sample size used for the test set and the data provenance:

    • Sample Size: The document does not specify the sample sizes for any of the performance tests listed. This is not a clinical study but rather a series of engineering and biological safety tests.
    • Data Provenance: The document does not specify the country of origin of the data or whether the data is retrospective or prospective. These are lab tests performed by the manufacturer.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This information is not applicable as the document describes engineering and biocompatibility tests, not a study requiring expert-established ground truth for device performance in a diagnostic or clinical decision-making context.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • This information is not applicable for the type of tests described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC comparative effectiveness study was not done. The FREMAP electrode is a passive cranial nerve electrode for monitoring and stimulation, not an AI-powered diagnostic or assistive device for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • This information is not applicable. The device is a passive electrode, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For the performance tests mentioned (electrical resistance, continuity, tensile strength, dimensions, biocompatibility), the "ground truth" would be the engineering specifications and established ISO/ASTM standards for medical device materials and components, and biological safety standards (e.g., ISO 10993 for biocompatibility). The document does not detail these specific standards or criteria, only that the tests were performed and requirements were met.

    8. The sample size for the training set:

    • This is not applicable as the FREMAP electrode is not an AI/ML device that requires a training set.

    9. How the ground truth for the training set was established:

    • This is not applicable as the FREMAP electrode is not an AI/ML device that requires a training set.

    In summary: The provided document is a 510(k) submission focused on demonstrating substantial equivalence through a comparison of technological characteristics and safety/performance testing of the device itself (an electrode), rather than an efficacy study of a diagnostic or AI-powered system involving clinical data and human interpretation. Therefore, many of the requested elements are not relevant or present in this type of regulatory submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K152547

    Validate with FDA (Live)

    Device Name
    DNAP Electrode
    Date Cleared
    2016-03-04

    (178 days)

    Product Code
    Regulation Number
    882.1330
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DNAP Cranial Nerve Electrodes are designed for intraoperative monitoring/recording for less than or equal to 10 hours from the cochlear cranial nerves during skull based surgical procedures when connected to the MEE-1000A or MEE-2000 Neural Function Measuring System.

    Device Description

    The DNAP Cranial Nerve Electrode (DNAP electrode) is intended for intraoperative nerve monitoring/recording during acoustic neuroma surgery or tumor removal (such as meningioma) at the CP-angle during surgical procedures. The DNAP electrodes are non-active and are intended to be connected to Nihon Kohden Corporation's Neural Function Measuring System MEE-1000A or MEE-2000 Neural Function Measuring System.

    The DNAP electrode is a single patient use, disposable cranial nerve electrode used for recording cranial nerve action potentials. The DNAP electrode may be used for recording the action potentials of cochlear nucleus elicited by auditory stimulation. The user places the electrode on the dorsal cochlear nuclei. Access to the dorsal cochlear nuclei may be through the foremen of Luschka. This potential is called DNAP (dorsal cochlear nucleus action potential), and used to preserve cochlear nerve function especially during acoustic neuroma surgery or tumor removal (such as meningioma) at the CP-angle.

    AI/ML Overview

    The provided document is a 510(k) summary for the Ad-Tech Medical Instrument Corporation's DNAP Cranial Nerve Electrode. It outlines the device's indications for use and compares its technological characteristics and performance to a predicate device (Cueva Cranial Nerve Electrode K044061) to demonstrate substantial equivalence.

    Here's an analysis of the acceptance criteria and study information:

    Acceptance Criteria and Reported Device Performance

    Acceptance Criteria / RequirementReported Device Performance and Results
    Electrical Resistance (from electrode contact to connector)Pass
    DNAP electrode Connector compatibility (with monitor equipment)Pass
    Electrical continuity and resistance after tensile testingPass
    Electrode separation under tensile testPass (No electrode head separation)
    Electrical continuity and resistance after simulated use (for ≤ 10 hours intraoperative use)Pass
    Overall length (dimensional requirements)Pass
    Biocompatibility (Cytotoxicity, Sensitization, Intracutaneous, Acute Systemic Toxicity)Performed and deemed acceptable

    Study Information

    1. Sample size used for the test set and the data provenance:

      • The document does not specify the exact sample size for the test sets used in the performance benches. It refers to "evaluations" and "tests" without detailing the number of devices or trials.
      • Data provenance is not explicitly stated (e.g., country of origin, retrospective/prospective). Given these are bench tests for a medical device seeking premarket notification in the US, it's highly likely they were conducted in a controlled lab environment.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This question is not applicable to the type of studies presented. The performance tests (electrical resistance, tensile strength, compatibility, biocompatibility) are objective bench tests, not requiring expert human interpretation or "ground truth" establishment in the clinical sense.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • This question is not applicable as the studies described are objective bench tests, not involving subjective human assessment or consensus methods for ground truth adjudication.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, an MRMC comparative effectiveness study was not done. This device is an electrode for intraoperative monitoring, not an AI-powered diagnostic tool that assists human readers/interpreters. The focus is on the physical and electrical performance of the electrode itself.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • No, a standalone algorithm performance study was not done. This device is a physical medical instrument (an electrode), not an algorithm or software. Its function is to record signals, which are then interpreted by human clinicians using an external Neural Function Measuring System.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The "ground truth" for the performance tests consists of objective, predefined engineering and biocompatibility standards and specifications. For instance, electrical resistance must fall within a specified range, and tensile strength must meet a certain threshold without failure.
    7. The sample size for the training set:

      • This is not applicable. The device is an electrode, not a machine learning model. Therefore, there is no "training set."
    8. How the ground truth for the training set was established:

      • This is not applicable as there is no training set for this type of device.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1