510(k) Data Aggregation

    K Number
    K250282

    Validate with FDA (Live)

    Device Name
    Persona C HR
    Date Cleared
    2025-04-30

    (89 days)

    Product Code
    Regulation Number
    892.1650
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    PERSONA C HR is a mobile X-ray device used for radiological guidance and visualization during diagnostic, interventional and surgical procedures.

    PERSONA C HR device can be used on all patients, except pediatric patients, within the limits of the device.

    Examples of clinical applications could be Orthopedic surgery, General surgery, Cardiac procedures, Thoracic surgery, Vascular procedures, Pain therapy and Urological procedures.

    Device Description

    PERSONA C HR is a C-arm mobile unit with flat panel detector.

    It allows imaging under the following modes:

    • Low Dose Fluoroscopy,
    • High Quality Fluoroscopy,
    • High Quality + Fluoroscopy
    • Digital radiography (Snapshot),
    • Fluoroscopy in Road Mapping mode (optional),
    • Fluoroscopy in DSA mode (optional).

    It is provided with a 30x30 cm Flat Panel detector and a 25 kW X-ray generator.

    The unit is composed of: Monitor unit, Stand with c-arm, X-ray commands, Printer (optional).

    The device acquires images employing X-rays emitted by an X-ray source which can produce up to 25kW.

    The unit, which is powered by a single-phase electrical supply, generates ionizing radiations (X-rays) which, as they pass through the patient, reach a flat panel detector that produces radiological images.

    An x-ray collimator, placed on the monobloc, is responsible for limiting the emission beam of ionizing radiation.

    The images collected by the detector are sent to a video processor, which, after processing the information received, allows radiological images to be displayed on the display monitor.

    AI/ML Overview

    I am sorry, but the provided text does not contain the detailed information necessary to fully address all aspects of your request regarding acceptance criteria and a study proving device performance. The document is an FDA 510(k) clearance letter and summary, which confirms substantial equivalence to a predicate device but does not typically include the granular details of performance studies you are looking for, such as specific acceptance criteria thresholds, detailed test set data, expert adjudication methods, or MRMC study results with effect sizes.

    Specifically, the document states: "Substantial equivalence was supported by engineering-based performance testing, including evaluation of image quality using test phantoms and cadaver images by an 'American Board of Radiology' radiologist." However, it does not provide the specifics of this evaluation in the way you've outlined in your request.

    Therefore, I cannot generate the table of acceptance criteria with reported device performance or the detailed study breakdown that you've requested beyond what is implicitly stated about image quality evaluation.

    If you have a different document that contains this information (e.g., a detailed clinical study report or a more extensive validation report), I would be happy to analyze it for you.

    Ask a Question

    Ask a specific question about this device

    K Number
    K211777

    Validate with FDA (Live)

    Device Name
    ARCO FP-S
    Date Cleared
    2021-08-13

    (65 days)

    Product Code
    Regulation Number
    892.1650
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ARCO FP-S is a mobile X-ray device with a Flat Panel detector used for radiological guidance and visualization during diagnostic, interventional and surgical procedures. In particular, the device is to be applied during othopaedic, neuro, abdominal, vascular, thoracic and cardiac procedures. ARCO FP-S device can be used on all patients except paediatric patients within the limit of the device.

    Device Description

    ARCO FP-S is a mobile C-arm with the solid-state image detector "Flat Panel". It will be realized in 2 models, depending on the size of the detector: Model: ARCO FP-SR21 Dual focus 0.3/0.6 mm, rotating anode X-Ray tube, generator nominal power 5 KW. Detector size 21x21 cm, 1024x1024 pixel. Models: ARCO FP-SR30 Dual focus 0.3/0.6 mm, rotating anode X-Ray tube, generator nominal power 5 KW. Detector size 30x30 cm, 1536 x 1536 pixel. The system consists of: C-arm Stand, X-RAY MONOBLOC, rotating anode, FLAT PANEL DETECTORS: 2121cm or 3030cm, MONITOR LCD 27" QUAD HD format 16:9, CONTROL PANEL, X-RAY COLLIMATOR, SYSTEM CONTROLLER, X-RAY HAND SWITCH, ARTICULATED ARM FOR MONITOR SUPPORT, ANTISCATTERING GRID. Optional components include FOOTSWITCH, two-foot pedal, LASER LOCALISER, Dose Area Meter, DSA Packages, I.R. remote controls, DICOM Classes, EXTERNAL VIDEO OUTPUT-Wireless transmission, DICOM DATA Wi-Fi transmission to Ethernet/PACS, NFC LOGIN, Wireless monitor trolley, model TROTTER W, for duplicating the image of the monitor on board.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device called ARCO FP-S, a mobile X-ray device. The submission aims to demonstrate substantial equivalence to a predicate device (ARCO FP K182086). The study is primarily a non-clinical bench test and radiologist's image evaluation to confirm that the changes made to the device (specifically, the digital x-ray receptor panels) do not negatively impact its performance or safety compared to the predicate.

    Here's the breakdown of the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are not explicitly stated in a quantitative table format with pass/fail thresholds in the provided document. Instead, the document asserts "substantial equivalence" based on various performance specifications relative to the predicate device and the evaluation of image quality by a radiologist. The implicit acceptance criterion is that the performance of the ARCO FP-S, especially concerning image quality, must be at least equivalent to, or better than, the predicate device.

    Performance CharacteristicAcceptance Criteria (Implicit: at least equivalent to predicate)Reported Device Performance (ARCO FP-S)Conclusion (Implicit)
    Detector (21x21 Model)- Active detector size: ~207x207 mm (predicate PIXIUM 2121S)- PaxScan 2121DXV (new)Met
    - Pixel size: ~154 μm (predicate PIXIUM 2121S)- Active detector size: 1024x1024 pixel, 205x205 mmMet
    - Detective Quantum Efficiency (DQE) RQA5:- Pixel size: 205 μmMet
    - @1.0 Lp/mm: 56% (predicate PIXIUM 2121S)- DQE RQA5:Met
    - @2.0 Lp/mm: 46% (predicate PIXIUM 2121S)- @1.0 Lp/mm: 65% (improved from predicate)
    - Modulation Transfer Function (MTF) RQA5:- @2.0 Lp/mm: 40% (slightly lower for 21x21 vs 46% for predicate, but still considered substantially equivalent given overall performance)
    - @1.0 Lp/mm: 59% (predicate PIXIUM 2121S)- MTF RQA5:Met
    - @2.0 Lp/mm: 29% (predicate PIXIUM 2121S)- @1.0 Lp/mm: 50% (slightly lower for 21x21 vs 59% for predicate, but still considered substantially equivalent given overall performance)
    - Nyquist: 11% (predicate PIXIUM 2121S)- @2.0 Lp/mm: 22% (slightly lower for 21x21 vs 29% for predicate, but still considered substantially equivalent given overall performance)
    Detector (30x30 Model)- Active detector size: ~301x301 mm (predicate PIXIUM 3030S)- Nyquist: 15% (improved from predicate)
    - Pixel size: ~154 μm (predicate PIXIUM 3030S)- PaxScan 3030DXV (new)Met
    - Detective Quantum Efficiency (DQE) RQA5:- Active detector size: 1516x1516 pixel, 294x294 mmMet
    - @1.0 Lp/mm: 57% (predicate PIXIUM 3030S)- Pixel size: 194 μmMet
    - @2.0 Lp/mm: 48% (predicate PIXIUM 3030S)- DQE RQA5:Met
    - Modulation Transfer Function (MTF) RQA5:- @1.0 Lp/mm: 65% (improved from predicate)
    - @1.0 Lp/mm: 59% (predicate PIXIUM 3030S)- @2.0 Lp/mm: 44% (slightly lower for 30x30 vs 48% for predicate, but still considered substantially equivalent given overall performance)
    - @2.0 Lp/mm: 29% (predicate PIXIUM 3030S)- MTF RQA5:Met
    - Nyquist: 11% (predicate PIXIUM 3030S)- @1.0 Lp/mm: 55% (slightly lower for 30x30 vs 59% for predicate, but still considered substantially equivalent given overall performance)
    Maximum frame per second25 (predicate)- @2.0 Lp/mm: 23% (slightly lower for 30x30 vs 29% for predicate, but still considered substantially equivalent given overall performance)Met
    Workstation Monitor Resolution1920 x 1080 (predicate)- Nyquist: 15% (improved from predicate)
    Diagnostic Image QualityAcceptable diagnostic quality (predicate)30 (improved from predicate)Met
    2560 x 1440 (4K) (Improved from predicate)Met
    A Board Certified Radiologist reviewed the still and moving images acquired by the newer digital receptor panels and found them to be of good and acceptable diagnostic quality.Met

    2. Sample size used for the test set and the data provenance

    • Sample Size for Test Set: Not explicitly stated in terms of a numerical count of images or cases. The document mentions a "Radiologist's Images evaluation" where the radiologist reviewed "the still and moving images acquired by the newer digital receptor panels." It does not specify how many images or what types of images were reviewed.
    • Data Provenance: The images were "acquired by the newer digital receptor panels" (PaxScan), implying they are current acquisitions using the modified device. The document does not specify the country of origin or whether the data was retrospective or prospective. Given it's a bench test, the images were likely acquired in a controlled setting for testing purposes.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Number of Experts: One ("A Board Certified Radiologist").
    • Qualifications of Experts: "Board Certified Radiologist." No specific information about years of experience is provided.

    4. Adjudication method for the test set

    • Adjudication Method: Not applicable or "None" in the typical sense of multiple readers and consensus. A single Board Certified Radiologist evaluated the images.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done

    • MRMC Study: No, an MRMC comparative effectiveness study was not explicitly stated or described. The evaluation involved a single radiologist reviewing images to confirm diagnostic quality, not a comparison of human readers with and without AI assistance. The device itself is a C-arm X-ray system, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Standalone Performance: Not applicable for this device as it is an X-ray imaging system, not an AI algorithm performing diagnostic interpretations. The "performance" being evaluated is the quality of the raw images produced by the system.

    7. The type of ground truth used

    • Type of Ground Truth: "Expert consensus" in a limited sense, as it was the subjective assessment of a single Board Certified Radiologist who determined the images were of "good and acceptable diagnostic quality." This is essentially expert opinion on image interpretability for diagnostic purposes. It is not pathology, outcomes data, or an objective gold standard.

    8. The sample size for the training set

    • Sample Size for Training Set: Not applicable. The ARCO FP-S is a hardware device (X-ray system) with upgraded software. There is no mention of a machine learning component requiring a "training set" of images for an AI algorithm. The software upgrade is described as "FUNCTIONALLY IDENTICAL" to the predicate's software.

    9. How the ground truth for the training set was established

    • Ground Truth for Training Set: Not applicable, as there is no specified training set for a machine learning algorithm.
    Ask a Question

    Ask a specific question about this device

    K Number
    K211542

    Validate with FDA (Live)

    Device Name
    PRIMO S
    Date Cleared
    2021-07-09

    (51 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intended for digital image capture use in general radiographic examinations, wherever conventional screen-film systems may be used, excluding fluoroscopy, angiography and mammography. The kit allows imaging of the skull, chest, shoulders, spine, abdomen, pelvis, and extremities.

    Device Description

    PRIMO S is an image acquisition and processing software application, in radiography mode for Flat Panel detectors. The software is specifically designed for integration with production equipment of the SEDECAL group. The PRIMO S application will be used on different types of Sedecal equipment / systems:

    • mobile units
    • fixed installations.

    The PRIMO S VP application provides the following functions:

    • User login: the device is usable only by authenticated users
    • Management of the operator interface GUIs and setup of the application itself
    • The operator interface GUI must reserve a space on the monitor for the Sedecal equipment/system GUI (choice of examination (APR), X-ray generator commands, collimator, stand, etc.)
    • Management of patient data through manual entry and reception from the DICOM WORKLIST service
    • Management of image processing algorithms for each type of examination
    • Management of the automatic advancement procedures of the operations during the study.
    • Image acquisition and processing
    • Saving in Hard Disk of the acquired images
    • Automatic and manual image stitching procedure
    • Off-line image editing and optimization using process and graphic functions
    • Documentation of images and study data using DICOM services of STORE, PRINT, CDROM, MPPS, RDSR, STORAGE COMMITMENT
    • Application configuration setup
    • Export and automatic saving of images on external support (USB key)

    The application communicates with the Sedecal equipment through software modules (DLL) for:

    • The choice of the examination, made by the operator through a GUI defined by Sedecal (APR)
    • Send the exposure parameters foreseen by the selected exam (kV, mA, mAs, ms, collimator aperture, stand position, ...)
    • Receive the system status parameters and the exposure result
    AI/ML Overview

    While the provided text describes the PRIMO S device and its 510(k) submission, it explicitly states:

    "6. Clinical testing. Not required for a determination of substantial equivalence."

    This indicates that a clinical study with detailed acceptance criteria and performance metrics, as requested in your prompt, was not performed or required for the FDA clearance of this specific device. The clearance was based on substantial equivalence to a predicate device, supported by non-clinical testing and adherence to various standards.

    Therefore, I cannot provide the specific information you requested regarding validation studies, sample sizes, expert involvement, adjudication methods, MRMC studies, standalone performance, or ground truth establishment for a clinical test set, as such a study was not conducted or reported in this document.

    The document primarily focuses on:

    • Device Description: What PRIMO S is and what it does.
    • Indications for Use: The medical conditions and body parts it's intended for.
    • Technological Characteristics Comparison: How it compares to its predicate device (Sedecal SA K130883) in terms of X-ray generator, digital detectors, panel sizes, operating system, etc.
    • Non-Clinical Testing: A list of standards (IEC 62304, EN/IEC 62366-1, ISO 14971, NEMA DICOM, ISO 15223-1) that were employed in development, and mention of software validation based on FDA guidance and cybersecurity considerations.

    Since there's no clinical trial data, I cannot populate the table or answer the specific questions about the study design that would prove the device meets acceptance criteria based on clinical performance.

    Ask a Question

    Ask a specific question about this device

    K Number
    K182086

    Validate with FDA (Live)

    Device Name
    ARCO FP
    Date Cleared
    2019-08-20

    (383 days)

    Product Code
    Regulation Number
    892.1650
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ARCO FP is a mobile x-ray device used for radiological guidance and visualization during diagnostic, interventional and surgical procedures. In particular, the device is to be applied during orthopaedic, neuro, abdominal, vascular, thoracic and cardiac procedures. Arco FP device can be used on all patients except pediatric patients within the limits of the device.

    Device Description

    ARCO FP is a mobile x-ray device.

    AI/ML Overview

    This document is a 510(k) clearance letter for a medical device (ARCO FP) and contains no information about acceptance criteria or a study proving that acceptance criteria are met. The document focuses on the regulatory clearance of the device based on substantial equivalence to a predicate device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1