Found 12 results

510(k) Data Aggregation

    K Number
    K231338

    Validate with FDA (Live)

    Date Cleared
    2023-05-10

    (2 days)

    Product Code
    Regulation Number
    872.3260
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    3MIM Clinpro™ Clear 2.1% Sodium Fluoride Treatment is indicated for the treatment of hypersensitive teeth.

    Device Description

    3M™ Clinpro™ Clear 2.1% Sodium Fluoride Treatment is an aqueous fluoride coating that is applied topically to the tooth surfaces for the treatment of hypersensitive teeth. The product is sweetened with xylitol and sucralose. Clinpro 2.1% Sodium Fluoride Treatment contains 9,500 ppm fluoride and added calcium and phosphate. 3M Clinpro Clear 2.1% Sodium Fluoride Treatment uses the L-Pop™ unit dose dispensing system designed by 3M. Each L-Pop contains 0.5 ml (0.5 grams) of the coating and is offered in mint and melon flavors and a flavorless version.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device called "3M™ Clinpro™ Clear 2.1% Sodium Fluoride Treatment." This type of submission focuses on demonstrating substantial equivalence to a predicate device, rather than conducting a full clinical trial to prove efficacy and safety through detailed acceptance criteria and studies as one might see for novel high-risk devices.

    Therefore, the information typically available for a detailed acceptance criteria and performance study (as requested in the prompt) such as prospective studies, ground truth establishment by multiple experts with adjudication, MRMC studies, or standalone algorithm performance, is not present in this type of FDA submission.

    Instead, the document details bench testing to demonstrate that the subject device performs similarly to or better than its predicate devices based on certain physical properties and release characteristics. The acceptance criteria here are mainly "Pass/Fail" based on whether the device meets specific thresholds or shows comparable performance to the predicate devices in these bench tests.

    Here's an attempt to extract and present the available information in the requested format, with the understanding that many aspects of your request (e.g., human-in-the-loop studies, multi-reader studies, detailed expert ground truth, training set information) are not applicable or not provided in a 510(k) submission for a dental varnish.


    Acceptance Criteria and Device Performance Study for 3M™ Clinpro™ Clear 2.1% Sodium Fluoride Treatment

    Note: This 510(k) submission primarily relies on demonstrating substantial equivalence through bench testing against predicate devices, rather than clinical efficacy studies with human subjects. Therefore, many of the requested details, particularly those related to human reader studies, multi-expert ground truth establishment, and training/test set specifics for AI/ML models, are not applicable or not provided in this type of regulatory document for a tooth varnish. The "acceptance criteria" here refer to the performance thresholds set for the bench tests.

    1. Table of Acceptance Criteria and Reported Device Performance

    Physical Properties/CharacteristicTest MethodAcceptance Criteria (Specification)Reported Device Performance (3M Clinpro Clear 2.1% Sodium Fluoride Treatment)Predicate Device (3M Vanish Varnish) PerformanceReference Device (Fluor Protector S) Performance
    Dentin tubule occlusion (In-vitro)SEMPass/Fail: Must visually occlude (cover) exposed dentin tubulesPassPassPass
    Dentin fluid flow reduction (In-vitro)Modified Pashley TestPass/Fail: Must demonstrate visual reduction in fluid flow when treated compared to when they were untreatedPassPassPass
    Fluoride ReleaseISO 17730-2020Pass/Fail: Fluoride release potential is ≥ 14 µg F/mm² after 60 minutesPassPassPass
    Cumulative fluoride release potential (1 hour)Internal 3M Test MethodPass/Fail: Cumulative fluoride release potential at 1 hour is >0 µg/cm²/gPassPassPass
    Cumulative fluoride release potential (24 hours)Internal 3M Test MethodPass/Fail: Cumulative fluoride release potential at 24 hours is >0 µg/cm²/gPassPassPass
    Calcium Release (1 hour)ICP-AESPass/Fail: Cumulative calcium release potential at 1 hour is >0 µg/cm²/gPassPassNot tested - not claimed
    Calcium Release (24 hours)ICP-AESPass/Fail: Cumulative fluoride release potential at 24 hours is >0 µg/cm2/gPassPassNot tested - not claimed
    Phosphorus (Phosphate) Release (1 hour)ICP-AESPass/Fail: Cumulative phosphorus (phosphate) release potential at 1 hour is >0 µg/cm2/gPassPassPass
    Phosphorus (Phosphate) Release (24 hours)ICP-AESPass/Fail: Cumulative phosphorus (phosphate) release potential at 24 hours is >0 µg/cm2/gPassPassPass
    ConsistencyInternal 3M Test MethodPass/Fail: The reading number of 1/32th inch is 32-52.PassPassNot applicable (3M requirement)
    pHInternal 3M Test MethodPass/Fail: 6.0-8.0PassNot applicablePass

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a "test set" in the context of patient data or clinical images, as the evaluation is based on in-vitro bench testing of the device material. The sample size for each bench test (e.g., number of replicates, number of specimens) is typically detailed in the internal test reports, but not provided in this summary. The data provenance is from laboratory bench testing conducted by 3M, not from specific patient populations or countries.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    Not applicable. The "ground truth" for this device's performance is established by standardized laboratory testing methods and physical measurements, not by human expert interpretation of clinical data or images.

    4. Adjudication Method for the Test Set

    Not applicable. As the "test set" is material performance data from laboratory tests, no adjudication method is required.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No. This type of study (comparing human readers with and without AI assistance) is not applicable to a dental varnish product. This device is not an imaging AI or diagnostic tool.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a physical product (dental varnish), not an algorithm or software. Its performance is evaluated through material science and physical property tests.

    7. The Type of Ground Truth Used

    The ground truth used is based on physical and chemical measurements and observations from in-vitro laboratory bench tests, conforming to established standards (e.g., ISO 17730-2020) or validated internal methods. For example, "visually occlude exposed dentin tubules" for SEM, or measurable fluoride release values.

    8. The Sample Size for the Training Set

    Not applicable. This device is a physical product, not a machine learning model, so there is no "training set" in the context of data for model development.

    9. How the Ground Truth for the Training Set was Established

    Not applicable. See point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K230270

    Validate with FDA (Live)

    Date Cleared
    2023-02-06

    (6 days)

    Product Code
    Regulation Number
    872.3250
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    3M VitCal Liner/Base is indicated for lining and basing applications under dental restorations. 3M VitCal Liner/Base is also indicated for direct pulp capping.

    Device Description

    3M VitCal liner/base is a two-part paste/paste, light-cured system. The paste/paste materials are contained in a Micro-Mix Syringe Dispensing System. This dispensing system provides simultaneous dispensing and mixing of each component for a consistent mix and direct application into the mouth. The composition is a combination of a resin-modified glass ionomer and calcium-based material. The first paste contains a radiopaque fluoro-aluminosilicate glass and calcium filler. The second paste contains a modified polyalkenoic acid. VitCal liner/base provides the major benefits of glass ionomer cements including adhesion to the tooth structure and sustained fluoride release. Additionally, VitCal linerbase offers a high pH environment and a combination of a prolonged working time with an on-demand set time achieved by light curing.

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification for a dental product, 3M™ VitCal Liner/Base. It details the device's characteristics, indications for use, and a comparison to predicate devices to establish substantial equivalence.

    It is important to note that this document is for a DENTAL LINER/BASE material, NOT an AI-powered medical device for diagnosis or image analysis. Therefore, the requested information regarding AI study design, such as MRMC studies, expert adjudication for ground truth, sample sizes for training/test sets in an AI context, and effect sizes of human readers with AI assistance, is not applicable to this submission.

    The acceptance criteria and "study" proving the device meets them, in this context, refer to bench testing of physical, chemical, and biological properties (biocompatibility) to demonstrate that the new dental material performs comparably or acceptably to existing predicate devices.

    Here's the relevant information extracted from the document, tailored to the type of device:


    1. Table of Acceptance Criteria and the Reported Device Performance (Bench Test Results):

    Physical Properties Acceptance CriteriaUnits/Additional Details3M VitCal Liner/Base Reported PerformanceTheraCal LC (Primary Predicate) Reported Performance3M Vitrebond Plus (Secondary Predicate) Reported Performance
    Compressive Strength ≥ 69 MPaMPa103 ± 10Could not be cured effectively102 ± 6
    Flexural Strength ≥ 10 MPaMPa25 ± 316 ± 227 ± 1
    Surface Hardness (Barcol) ≥ 80% (0 mm)% with curing light distance 0 mm (using ~950 mW/cm² for 20s)98 ± 8%84 ± 3%109 ± 15%
    Surface Hardness (Barcol) ≥ 80% (7 mm)% with curing light distance 7 mm (using ~950 mW/cm² for 20s)93 ± 3%60 ± 4%101 ± 6%
    Surface Hardness (Barcol) ≥ 80% (0 mm)% with curing light distance 0 mm (using ~400 mW/cm² for 30s)94 ± 4%92 ± 2%
    Surface Hardness (Barcol) ≥ 80% (7 mm)% with curing light distance 7 mm (using ~400 mW/cm² for 30s)89 ± 11%Not tested.101 ± 8%
    Radiopacity ≥ 1.0 mm Aluminum StandardmmAl1.52 ± 0.051.56 ± 0.011.33 ± 0.04
    Calcium Release > 0 µg/g (Pass/Fail)Cumulative amount for seven daysPassPassNot applicable.
    Fluoride Release > 0 µg/g (Pass/Fail)Cumulative amount for seven daysPassNot applicable.Pass
    Phosphate Release > 0 µg/g (Pass/Fail)Cumulative amount for seven daysPassNot applicable.Not applicable.
    Working Time > 1.5 minutesminutes2.44 ± 0.0N/A, one part2.5 per IFU
    Shear Bond Strength to Dentin > 4 MPaMPa, adhesion to 3M Scotchbond Universal6 ± 32 ± 19 ± 3
    MPa, adhesion to 3M Adper Single Bond Plus Adhesive6 ± 2Not tested.13 ± 1
    MPa, adhesion to 3M Scotchbond Multi-Purpose Plus Adhesive8 ± 3Not tested.13 ± 2
    MPa, adhesion to Kuraray CLEARFIL™ SE BOND Adhesive6 ± 3Not tested.8 ± 3
    Dentin Tubule Occlusion: Tag formation observed under SEMYes/NoYesYesYes

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size for Bench Tests: The specific number of samples (e.g., n=5, n=10) for each individual bench test (e.g., compressive strength, flexural strength) is not explicitly detailed in this summary. However, standard deviations are provided for most quantitative measurements (e.g., 103 ± 10), implying multiple replicates were used for testing. For "Dentin Tubule Occlusion," it states "Only 1 lot of 3M VitCal was tested."
    • Data Provenance: The data is generated from laboratory bench testing of the dental material. The document does not specify a country of origin for the testing, but the submitting company (3M ESPE Dental Products) is based in St. Paul, MN, USA. The testing is prospective in the sense that it's performed specifically to characterize the device for this submission.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    • This question is not applicable as the "test set" here refers to physical and chemical properties measured in a lab, not clinical images or patient data analyzed by experts.
    • For biocompatibility assessment, it states: "A Diplomate of the American Board of Toxicology has assessed the safety of this product and has determined that it is safe for its intended use." This is an expert opinion on the safety, but not "ground truth" derived from patient data.

    4. Adjudication Method for the Test Set:

    • This question is not applicable as there is no expert reading or interpretation of "cases" involved. Performance is based on direct physical and chemical measurements.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

    • No, an MRMC study was NOT done. This type of study is relevant for AI systems that assist human readers in tasks like image interpretation, which is not the function of this dental liner/base material.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • No, this is not an algorithm. This is a physical dental material. The "standalone performance" is represented by the bench test results of the material itself.

    7. The Type of Ground Truth Used:

    • The "ground truth" for this device is based on established industry standards and physical/chemical measurements (e.g., ISO 9917-2 standards for dental cements, ASTM/ISO standards for mechanical properties, and analytical chemistry methods for ion release). For biocompatibility, it's based on toxicological assessments against established biological safety standards (e.g., ISO 10993).

    8. The Sample Size for the Training Set:

    • Not applicable. This device is a material, not an AI model requiring a training set.

    9. How the Ground Truth for the Training Set was Established:

    • Not applicable. This device is a material, not an AI model, and therefore has no "training set."
    Ask a Question

    Ask a specific question about this device

    K Number
    K191122

    Validate with FDA (Live)

    Date Cleared
    2019-05-29

    (30 days)

    Product Code
    Regulation Number
    872.3275
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Permanent cementation of pediatric crowns:

    • Stainless Steel crowns
    • Zirconia crowns
    • 3M™ Pediatric Esthetic Crowns
    Device Description

    3MTM RelyX™ Pediatric Resin Modified Glass Ionomer Cement is a radiopaque, fluoride-releasing, resin-modified glass ionomer luting cement. RelyX Pediatric Cement is intended for the cementation of preformed pediatric crowns made of stainless steel, zirconia, and resin composite material.

    RelyX Pediatric Cement consists of a base paste and a catalyst paste packaged in the Automix delivery system for direct delivery in consistent mix ratios. The mixing tip attachment offers convenience over traditional hand mixed cement systems and delivers consistent mix ratios. The cement is available in a white shade. The Automix syringe contains 8.5 g of cement material. The product provides both selfcuring and light curing mechanisms which makes it uniquely versatile for the cementation of different restorative materials.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and the study details for the 3M™ RelyX™ Pediatric Resin Modified Glass Ionomer Cement, based on the provided text:

    Important Note: The provided document is a 510(k) summary for a dental cement. It does not describe an AI medical device, but rather a traditional medical device that undergoes physical and chemical performance testing to demonstrate substantial equivalence to predicate devices. Therefore, many of the requested fields related to AI, such as sample sizes for test/training sets, data provenance, ground truth experts, adjudication methods, and MRMC studies, are not applicable and thus not present in this type of document.


    Acceptance Criteria and Reported Device Performance

    The acceptance criteria for this device are implied by the demonstration of "equivalent or superior bonding" and other physical properties compared to predicate devices. The document does not explicitly list numerical acceptance criteria in a table format for each property, but rather states that the "results submitted in the 510(k) demonstrated equivalent or superior bonding."

    Here's a table based on the reported performance relative to predicate devices:

    Performance MetricAcceptance Criteria (Implied)Reported Device Performance
    Adhesion to DentinEquivalent to or superior to predicate devices (K111185, K001730)Demonstrated equivalent or superior bonding
    Adhesion to EnamelEquivalent to or superior to predicate devices (K111185, K001730)Demonstrated equivalent or superior bonding
    Adhesion to 3M Pediatric Esthetic CrownAdequate performance for indicated use (specific to the new indication)Demonstrated equivalent or superior bonding
    Compressive StrengthPerformance comparable to predicate devicesData included in submission to evaluate performance for new indication
    Flexural StrengthPerformance comparable to predicate devicesData included in submission to evaluate performance for new indication
    Fluoride ReleasePerformance comparable to predicate devicesData included in submission to evaluate performance for new indication
    Linear ExpansionPerformance comparable to predicate devicesData included in submission to evaluate performance for new indication
    RadiopacityPerformance comparable to predicate devicesData included in submission to evaluate performance for new indication
    Color StabilityPerformance suitable for esthetic properties (specific to the new indication)Data included in submission to evaluate performance for new indication
    BiocompatibilityMeets biocompatibility requirements (ISO 10993-1:2018)Confirmed to meet requirements; no additional testing needed due to identical formulation as predicate device
    Shelf Life24 months at room temperatureEstablished as 24 months based on predicate device testing

    Study Details

    1. Sample size used for the test set and the data provenance:

      • Sample Size: Not explicitly stated for each test. The document mentions "data from in vitro testing," which typically involves multiple samples for each test (e.g., several samples per group for adhesion tests, compressive strength, etc.), but the exact number is not provided.
      • Data Provenance: The studies were in vitro testing. The country of origin of the data is not specified, but the submitter is 3M ESPE Dental Products, located in St. Paul, MN, USA. The testing was conducted to support a 510(k) submission to the U.S. FDA. The studies are by nature prospective in the sense that the manufacturer performed them to gain clearance for the device.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This is not applicable as this is not an AI/diagnostic device. The "ground truth" for material properties is established through standardized physical and chemical testing methods (e.g., ISO standards), not expert consensus.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • Not applicable for this type of in vitro material performance testing. Standardized test methods and measurements are used, not human adjudication.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. This is not an AI medical device.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not applicable. This is not an AI medical device.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For the physical and chemical properties (adhesion, strength, fluoride release, etc.), the "ground truth" is determined by the results obtained from standardized in vitro test methods (e.g., per FDA Guidance "Dental Cements - Premarket Notification" and ISO 9917-2:2017). For biocompatibility, the ground truth is established by meeting the requirements of ISO 10993-1:2018.
    7. The sample size for the training set:

      • Not applicable. This is not an AI medical device, so there is no "training set."
    8. How the ground truth for the training set was established:

      • Not applicable. This is not an AI medical device, so there is no "training set" or corresponding ground truth establishment process in that context.
    Ask a Question

    Ask a specific question about this device

    K Number
    K183476

    Validate with FDA (Live)

    Date Cleared
    2019-01-02

    (16 days)

    Product Code
    Regulation Number
    872.3690
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    · Direct anterior and posterior restorations (including occlusal surfaces)
    · Core build-ups
    · Splinting
    · Indirect restorations including inlays, onlays, and veneers

    Device Description

    3MTM FiltekTM Universal Restorative, is a visible-light activated, restorative composite optimized to create esthetic anterior and posterior restorations. This material provides excellent strength and low wear for durability and improved esthetics with higher level of visual opacity. The shades are body like opacity enabling up to a 2mm depth-of-cure. The pink opaquer can be placed in 1mm thick increments. All shades are radiopaque. Filtek Universal Restorative is offered in the following shades: A1. A2, A3. A3.5. A4, B1. B2, D3. XW, and PO. The pink opaque shade option can be used to mask discolored or stained tooth structure.

    Filtek™ Universal Restorative is packaged in traditional syringes, for dispensing restorative on a pad outside the mouth, and single-dose capsules for dispensing restorative intraorally. The capsules are dispensed using the 3M ESPE Restorative Dispenser. Capsules may be warmed up to 70°C/158°F for one hour in a commercial warmer prior to use.

    AI/ML Overview

    This document is a 510(k) summary for a dental restorative material, not an AI/ML medical device. Therefore, it does not contain the information required to answer the prompt regarding AI model performance, acceptance criteria, or specific study details like MRMC comparative effectiveness studies, ground truth establishment for AI training/test sets, or expert qualifications for reviewing AI outputs.

    The document primarily focuses on establishing substantial equivalence to a predicate device based on material composition, physical properties, and intended use for a traditional dental restorative. It outlines typical material property tests (e.g., Compressive Strength, Depth of Cure) for dental composites.

    Therefore, I cannot provide the requested information based on the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K141081

    Validate with FDA (Live)

    Date Cleared
    2014-05-08

    (13 days)

    Product Code
    Regulation Number
    872.3690
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    • Direct anterior and posterior restorations (including occlusal surfaces)
    • Base/liner under direct restorations
    • Core build-ups
    • Splinting
    • Indirect restorations including inlays, onlays and veneers
    • Restorations of deciduous teeth
    • Extended fissure sealing in molars and premolars
    • Repair of defects in porcelain restorations, enamel, and temporaries
    Device Description

    3MTM ESPETM FiltekTM Bulk Fill Posterior Restorative material is a visible-light activated, restorative composite optimized to create posterior restorations simpler and faster. This bulk fill material provides excellent strength and low wear for durability. The shades are semi-translucent and low stress curing, enabling up to a 5 mm depth-ofcure. With excellent polish retention, Filtek™ Bulk Fill Posterior Restorative is also useful for anterior restorations that call for a semi-translucent shade. All shades are radiopaque. Filtek™ Bulk Fill Posterior Restorative is offered in A.I. A2. A3. B1. and C2 shades. Filtek™ Bulk Fill Posterior Restorative is applied to the tooth following use of a methacrylate-based dental adhesive, such as manufactured by 3M™ ESPETM. which permanently bonds the restoration to the tooth structure.

    Filtek™ Bulk Fill Posterior Restorative is packaged in traditional syringes, for dispensing restorative on a pad outside the mouth, and single-dose capsules for dispensing restorative intraorally. The capsules are dispensed using the 3M ESPE Restorative Dispenser.

    AI/ML Overview

    This document is a 510(k) summary for a dental restorative material, not a medical imaging or AI-based device, so many of the requested criteria (e.g., sample size for test set, number of experts, MRMC study, training set information) are not applicable.

    However, I can extract the acceptance criteria and performance as presented for the physical properties of the dental material.

    Acceptance Criteria and Reported Device Performance for Filtek™ Bulk Fill Posterior Restorative:

    The submission indicates that "Test results for the following physical properties were included in this submission: Compressive Strength, Diametral Tensile Strength, Flexural Strength, Flexural Modulus, Surface Hardness, Radiopacity, Water Sorption, Water Solubility, Volumetric Shrinkage, Wear, Depth of Cure, Cusp Deflection and Polish Retention."

    The overall "acceptance criterion" for this 510(k) submission is Substantial Equivalence to the predicate devices. The document explicitly states:

    "Filtek™ Bulk Fill Posterior Restorative is substantially equivalent to the predicate devices, Filtek™ Supreme Ultra Universal Restorative, SonicFill, Sonic-Activated Bulk Fill Composite and Tetric EvoCeram Bulk Fill in terms of intended use, indications for use, physical properties, and technological characteristics. FiltekTM Bulk Fill Posterior Restorative is substantially equivalent to the predicate device, Filtek™ Supreme Ultra Universal Restorative in terms of formulation."

    The study that proves the device meets the acceptance criteria (i.e., is substantially equivalent) involves a comparison of its technological characteristics and physical properties against predicate devices.

    1. Table of Acceptance Criteria and Reported Device Performance:

    Since specific numerical acceptance criteria (e.g., "Compressive Strength > X MPa") and corresponding reported performance are not detailed in this summary, the table will reflect the comparative nature of the substantial equivalence claim. The "acceptance criteria" here are essentially "similar to or better than predicate devices" for the listed properties.

    Acceptance Criterion (Technological/Physical Property)Reported Device Performance (Filtek™ Bulk Fill Posterior Restorative)Predicate Device(s)
    Photoinitiator systemX (Present)Filtek™ Supreme Ultra Universal Restorative, Tetric EvoCeram Bulk Fill
    Methacrylate-based resin matrixX (Present)All Predicates (Filtek™ Supreme Ultra U.R., SonicFill, Tetric EvoCeram B.F.)
    Compatible with methacrylate-based dental adhesivesX (Compatible)Filtek™ Supreme Ultra U.R., Tetric EvoCeram B.F.
    Inorganic fillersX (Present)All Predicates
    Oxide fillers are silane treatedX (Present)Filtek™ Supreme Ultra U.R., SonicFill
    Bulk fill (up to 4 mm depth of cure)X (Achieved, for Class I from occlusal aspect)SonicFill, Tetric EvoCeram Bulk Fill
    Bulk fill (5 mm depth of cure, Class II)X (Achieved with multi-site light-curing)SonicFill, Tetric EvoCeram Bulk Fill
    Light-induced polymerization mechanismX (Operates via this mechanism)All Predicates
    Dispensing system: single-use capsuleX (Available)All Predicates
    Dispensing system: reusable syringeX (Available)Filtek™ Supreme Ultra U.R., Tetric EvoCeram Bulk Fill
    Recommended for load-bearing occlusal surfacesX (Recommended)All Predicates
    Additional Physical Properties Tested (Data not detailed but stated to be included in submission):Performance "readily pass the ISO 6874 Depth of Cure requirement" and generally support substantial equivalence in physical properties.Predicate devices' performance as comparative benchmark.
    Compressive StrengthData included
    Diametral Tensile StrengthData included
    Flexural StrengthData included
    Flexural ModulusData included
    Surface HardnessData included
    RadiopacityData included
    Water SorptionData included
    Water SolubilityData included
    Volumetric ShrinkageData included
    WearData included
    Depth of Cure (ISO 4049 & ISO 6874)Data included, "readily pass the ISO 6874 Depth of Cure requirement."
    Cusp DeflectionData included
    Polish RetentionData included
    Biocompatibility (ISO 10993 series, ISO 7405)Assessed to be safe for intended use by a Diplomate of the American Board of Toxicology.Filtek™ Supreme Ultra U.R. (evaluated against earlier versions of same standards).
    Product Standards (ISO 4049)Tested against ISO 4049:2009.Filtek™ Supreme Ultra U.R. (evaluated against same standard version for this submission).

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not specified in the 510(k) summary. The document mentions "Test results for the following physical properties were included in this submission," implying laboratory testing, but does not provide specific sample quantities for each test.
    • Data Provenance: Not explicitly stated regarding country of origin or whether it was retrospective/prospective. As these are material property tests, they are typically conducted in a laboratory setting.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable in the context of material property testing for substantial equivalence. The "ground truth" for material properties is established by adherence to recognized international standards (e.g., ISO).
    • However, for biocompatibility, it states: "A Diplomate of the American Board of Toxicology assessed the safety of the product." This indicates at least one, highly qualified expert was involved in the safety assessment.

    4. Adjudication method for the test set:

    • Not applicable. Adjudication methods like 2+1 or 3+1 are used for expert consensus on clinical diagnoses or interpretations, which is not the nature of this submission. The tests are based on objective physical measurements per ISO standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This device is a dental restorative material, not an AI-assisted diagnostic or imaging tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is not an algorithm or AI device.

    7. The type of ground truth used:

    • The "ground truth" for the performance claims is based on International Standardization Organization (ISO) standards for dental materials. Specifically, ISO 4049 and ISO 6874 are mentioned for certain properties, and the ISO 10993 series and ISO 7405 for biocompatibility.

    8. The sample size for the training set:

    • Not applicable. This is a physical material, not a machine learning model, so there is no "training set."

    9. How the ground truth for the training set was established:

    • Not applicable. (See #8)
    Ask a Question

    Ask a specific question about this device

    K Number
    K120453

    Validate with FDA (Live)

    Date Cleared
    2012-03-19

    (34 days)

    Product Code
    Regulation Number
    872.3690
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    • Base under Class I and II direct restorations .
    • Liner under direct restorative materials .
    • . Pit and fissure sealant
    • Restoration of minimally invasive cavity preparations (including small, ◆ non stress-bearing occlusal restorations)
    • Class III and V restorations .
    • . Undercut blockout
    • Repair of small enamel defects .
    • Repair of small defects in esthetic indirect restorations .
    • Repair of resin and acrylic temporary materials .
    • As a core build-up where at least half the coronal tooth structure is . remaining to provide structural support for the crown
    Device Description

    Filtek™ Bulk Fill Flowable Restorative, is a low viscosity, visible-light activated, radiopaque, flowable. This low stress flowable material is semitranslucent enabling a 4mm depth of cure. The restorative is packaged in capsules and syringes. The shades offered with Filtek™ Bulk Fill flowable are U (Universal), A1, A2, and A3. Filtek™ Bulk Fill flowable contains bisGMA, UDMA, bisEMA(6) and Procrylat resins. The fillers are a combination of ytterbium trifluoride filler with a range of particle sizes from 0.1 to 5.0 microns and zirconia/silica with a particle size range of 0.01 to 3.5 µm. The inorganic filler loading is approximately 64.5% by weight (42.5% by volume). A dental adhesive is used to permanently bond the restoration to the tooth structure.

    Filtek™ Bulk Fill Flowable Restorative is a modification of predicate device Filtek™ Supreme Ultra Flowable Restorative. The formulation was modified to increase depth of cure, while decreasing polymerization shrinkage stress.

    AI/ML Overview

    The provided text describes a 510(k) submission for a dental restorative material, not an AI/ML powered device. As such, the requested information regarding acceptance criteria, study details (sample size, data provenance, ground truth, MRMC study, standalone performance), and training set details for a device meeting acceptance criteria and proven by a study is not present.

    The document focuses on demonstrating substantial equivalence of the Filtek™ Bulk Fill Flowable Restorative to predicate devices. This is achieved through bench testing of physical properties and an assessment of biocompatibility and risk management.

    Here's an analysis of the provided text in relation to your request, highlighting what is not available:

    1. A table of acceptance criteria and the reported device performance:

    • Not Available directly as a table for "acceptance criteria" and "reported device performance" in the AI/ML context.
    • The document states: "This 510(k) submission includes data from bench testing to evaluate the performance of Filtek™ Bulk Fill Flowable Restorative compared to predicate devices Filtek™ Supreme Ultra Flowable Restorative, Adaptable Composite Resin Restorative Material and Revolution™ Formula 2. Standards utilized include ISO 4049.2009 and ISO 6874:2005. Properties evaluated include Compressive Strength, Diametral Tensile Strength, Flexural Strength, Flexural Modulus, Surface Hardness, Radiopacity, Water Sorption, Water Solubility, Cusp Deflection, Volumetric Shrinkage, Wear and Depth of Cure."
    • The "acceptance criteria" here are implied to be meeting or being comparable to the performance of the predicate devices and relevant ISO standards for these physical properties. The reported device performance for each of these specific properties is not explicitly listed in the summary but would have been part of the full 510(k) submission.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • Not available. The document mentions "bench testing" but does not specify sample sizes for these tests, nor the "provenance" in terms of subject or data origin, as this is a material science evaluation, not a clinical study involving human data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    • Not applicable/available. This section refers to AI/ML ground truth establishment. For this dental material, "ground truth" is established by standard material science testing and adherence to ISO standards.
    • One "Diplomate of the American Board of Toxicology" assessed biocompatibility, which is a different type of expert review, not related to an AI/ML ground truth for a test set.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable/available. Adjudication methods are relevant for expert consensus in AI/ML performance evaluation and are not described here for material property testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable/available. This is a question specifically for AI/ML devices. No human reader studies are mentioned, as the device is a dental restorative material, not an AI aid for diagnosis or treatment planning.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable/available. This question is also specific to AI/ML algorithms.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Not applicable/available in the AI/ML sense. The "ground truth" for this device's performance is based on established scientific measurement methods for material properties (e.g., compressive strength, depth of cure) as defined by ISO standards 4049.2009 and 6874:2005. Biocompatibility "ground truth" is derived from toxicology assessments and ISO 10993 and 7405 standards.

    8. The sample size for the training set:

    • Not applicable/available. There is no "training set" as this is not an AI/ML device.

    9. How the ground truth for the training set was established:

    • Not applicable/available. There is no "training set."

    In summary, the provided document describes the 510(k) submission for a non-AI dental material. Therefore, most of the requested information pertaining to AI/ML device evaluation criteria and study design is not present or applicable. The document focuses on demonstrating substantial equivalence based on material properties, biocompatibility, and risk management through bench testing and adherence to established standards, rather than clinical efficacy through human evaluations or AI model performance.

    Ask a Question

    Ask a specific question about this device

    K Number
    K111185

    Validate with FDA (Live)

    Date Cleared
    2011-05-05

    (8 days)

    Product Code
    Regulation Number
    872.3275
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This device is intended for use as a dental cement. RelyX™ Luting Plus Automix (Lexus-2) is indicated for Luting:

    • Luting porcelain fused to metal crowns and bridges to tooth structure, amalgam, composite or glass ionomer core build ups;
    • Luting metal inlays, onlays or crowns;
    • Luting pre-fabricated and cast post cementation
    • Luting orthodontic appliances
    • Luting crowns made with all-alumina or all zirconia cores such as Procera® AllCeram
    Device Description

    3M™ ESPE™ RelyX™ Luting Plus Automix (Lexus-2) is a radiopaque, fluoride-releasing, resinmodified glass ionomer luting cement. It is self-curing with an option for tack light curing of excess cement. RelyX Luting Plus Automix (Lexus-2) consists of a base (Paste A) and catalyst (Paste B) packaged in an automix (dual barrel) syringe. The dual barrel syringe dispenses paste A and Paste B. The cement is available in a white shade.

    AI/ML Overview

    The provided text describes the acceptance criteria and a study demonstrating that the device, RelyX™ Luting Plus Automix, meets these criteria.

    1. Acceptance Criteria and Reported Device Performance

    The acceptance criteria are established by comparing the performance of the new device (RelyX™ Luting Plus Automix, referred to as Lexus 2) against a legally marketed predicate device (3M™ ESPE™ RelyX™ Luting Plus Cement, K022476). The reported device performance is indicated by "pass" for all tested properties, signifying that Lexus 2 met or exceeded the design specifications, which are implicitly acceptable as they are compared to a predicate device.

    PropertiesMethodLexus 2 Design SpecificationLexus2 Ave and Predicate (K022476) Performance
    Flexural strength (FS) (MPa)ISO 9917-2≥ 10 MPapass
    Radio-opacity (mm of Al)ISO 9917-2≥1.0mm of Alpass
    Adhesion to dentin (MPa)3M ESPE internal≥2 MPapass
    Adhesion to enamel (MPa)3M ESPE internal≥2 MPapass
    Adhesion to Metal (MPa)3M ESPE internal≥2 MPapass
    Adhesion to Lava™ (MPa) zirconia3M ESPE internal≥2 MPapass
    Adhesion to composites3M ESPE internal≥2 MPapass
    Adhesion to amalgam3M ESPE internal≥2MPapass
    Adhesion to Titanium (MPa)3M ESPE internal≥2 MPapass
    Adhesion to Lithium disilicate (MPa)3M ESPE internal≥2 MPapass
    Fluoride release (µg F/g) at 90 days3M ESPE internal≥500pass
    Film thickness (microns)ISO 9917-2≤ 25pass
    Work time (s)3M ESPE internal≥ 90spass
    Set time (s)3M ESPE internal<480spass

    2. Sample size used for the test set and the data provenance

    The document does not explicitly state the specific sample sizes used for each individual bench test. It mentions that the data is obtained from "Bench Test Data Comparison," and the methods used for certain tests are "ISO 9917-2" and "3M ESPE internal" methods.

    The data provenance is from internal testing conducted by 3M ESPE. There is no mention of country of origin for the data or whether it was retrospective or prospective, but it implies a controlled laboratory environment designed to compare the new device to the predicate.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This study concerns the physical and chemical properties of a dental cement, evaluated through standardized bench tests and internal laboratory methods, rather than expert interpretation of medical images or patient data. Therefore, there is no "ground truth" established by experts in the context of medical diagnosis.

    4. Adjudication method for the test set

    Not applicable. This study does not involve human interpretation or subjective assessment that would require an adjudication method. The evaluations are based on predefined quantifiable metrics and established testing protocols.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not a study involving human readers or AI assistance. It is a bench test comparison of the physical and chemical properties of a dental cement.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an algorithm or AI-based device. It is a dental cement, and its performance is assessed through laboratory bench tests.

    7. The type of ground truth used

    The "ground truth" in this context is established by:

    • International Standards: ISO 9917-2 for Flexural Strength, Radio-opacity, and Film Thickness. These standards define the methods and potentially acceptable ranges for these properties.
    • Predicate Device Performance: The primary "ground truth" for substantial equivalence is the performance of the legally marketed predicate device (RelyX™ Luting Plus Cement, K022476). The new device is considered substantially equivalent if its performance (as measured by design specifications and actual test results) is comparable to or better than the predicate's and meets the specified criteria.
    • Internal 3M ESPE Specifications: For properties like adhesion, fluoride release, work time, and set time, "3M ESPE internal" methods and "Design Specifications" are used, implying established internal criteria and methods by the manufacturer.

    8. The sample size for the training set

    Not applicable. This is a bench test study for a physical product, not a machine learning model, so there is no training set in the AI/ML sense.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K110131

    Validate with FDA (Live)

    Date Cleared
    2011-01-21

    (3 days)

    Product Code
    Regulation Number
    872.3690
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The product is indicated for inlays, onlays, veneers, and full crown restorations, including crowns on implants.

    Device Description

    The product is a strong, wear-resistant and highly esthetic mill block that provides a fast and easy-to-use alternative to porcelain blocks for milling CAD/CAM indirect restorations. The material is specially processed to enhance its properties for use in CAD/CAM milling procedures.

    AI/ML Overview

    The provided text describes a 510(k) submission for a dental restorative material, not a software or AI-powered medical device. Therefore, the information requested about acceptance criteria, study details, ground truth, and AI-specific metrics (like MRMC studies, human reader improvement, standalone performance, and training set information) is not applicable or cannot be extracted from this document.

    The document primarily focuses on demonstrating substantial equivalence to predicate dental devices through bench testing of physical properties.

    However, I can extract the general "acceptance criteria" in the context of this device being approved based on comparison to predicates and meeting certain material properties.

    1. Table of Acceptance Criteria and Reported Device Performance

    Property EvaluatedAcceptance Criteria (Implied by Predicate Equivalence)Reported Device Performance (Implied by Substantial Equivalence Determination)
    Flexural StrengthComparable to predicate devicesDeemed substantially equivalent
    Flexural ModulusComparable to predicate devicesDeemed substantially equivalent
    Compressive StrengthComparable to predicate devicesDeemed substantially equivalent
    Diametral Tensile StrengthComparable to predicate devicesDeemed substantially equivalent
    Water SorptionComparable to predicate devicesDeemed substantially equivalent
    Water SolubilityComparable to predicate devicesDeemed substantially equivalent
    Fracture ToughnessComparable to predicate devicesDeemed substantially equivalent
    BiocompatibilitySafe for intended use (based on risk assessment)Deemed safe for intended use

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not specified for each test. The document states "data from bench testing" was used.
    • Data Provenance: Bench testing, likely conducted by 3M ESPE Dental Products. No information about country of origin, retrospective or prospective is relevant as it's a material test.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

    • Not applicable. This is a material science study, not one requiring expert human interpretation for "ground truth" as you would find in an imaging or diagnostic device. The "ground truth" here is the measured physical properties of the material and its comparison to established predicate devices.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not applicable. This concept is relevant for studies involving human interpretation or subjective assessments, not for bench testing of material properties.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This is not an AI-powered device.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is not an AI-powered device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The ground truth for this type of material device is derived from standardized bench test measurements of physical and chemical properties and comparison to the established performance of legally marketed predicate devices. Biocompatibility was assessed through a "risk assessment techniques and consideration of FDA & internationally recognized guidelines."

    8. The sample size for the training set:

    • Not applicable. This is not an AI-powered device, so there is no "training set."

    9. How the ground truth for the training set was established:

    • Not applicable. This is not an AI-powered device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K092141

    Validate with FDA (Live)

    Date Cleared
    2010-03-10

    (238 days)

    Product Code
    Regulation Number
    872.3260
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    • Treatment of hypersensitive teeth
    • Use on exposed dentin and root sensitivity
    • Under temporary restoratives and cements where post-operative sensitivity is of concern
    Device Description

    Vanish Varnish is a topically applied, flavored cavity varnish containing sodium fluoride and calcium phosphate in a rosin based preparation. The varnish is an insoluble viscous liquid that forms a film on tooth surfaces.

    AI/ML Overview

    The provided text is a 510(k) summary for the Vanish™ Varnish, 5% Sodium Fluoride White Varnish. It details the device's classification, predicate devices, description, and intended use. However, it does not contain any information about acceptance criteria, study methodologies (like sample sizes, data provenance, expert ground truth, adjudication methods, MRMC studies, or standalone algorithm performance), or training set details.

    The document mainly focuses on establishing substantial equivalence to previously marketed predicate devices based on properties, intended use, and composition. It mentions that "The data provided in this 510(k) submission shows that the composition is safe based on the biocompatibility assessment conducted based on ISO10993 and ISO 7405." This indicates some testing was done for safety, but not for effectiveness or performance metrics that would typically involve a multi-reader, multi-case study or a standalone algorithm performance test.

    Therefore, I cannot provide the requested information. The document primarily confirms the device's features and its equivalence to other legally marketed devices as per the 510(k) submission process, rather than detailing specific performance studies with acceptance criteria for an AI/ML powered device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K073395

    Validate with FDA (Live)

    Date Cleared
    2008-01-17

    (45 days)

    Product Code
    Regulation Number
    872.3275
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    1. Lining and basing applications under composite, amalgam, ceramic or metal restorations.
    2. Treatment of exposed root surface sensitivity
    3. Site specific protective coating for newly erupted teeth and other tooth surfaces including non-cavitated lesions
    Device Description

    GILB-2, Glass Ionomer protective coating is classified as Dental Cement (21 C.F.R§872.3275) because it is a device intended to provide protection as a cavity liner/basing material as well as in treatment of exposed root surfaces hypersensitivity. GILB-2 is a site specific protective coating for newly erupted teeth and other tooth surfaces including non-cavitated lesions.

    GILB-2 is a two part liquid/paste system. The liquid/paste materials are contained in the Clicker™ Dispensing System manufactured by 3M ESPE. This dispensing system provides simultaneous dispensing of each component for a consistent mix.

    The chemical composition is similar to predicate glass ionomer dental cement devices. The data provided in this 510(k) submission shows that the composition is safe based on the biocompatibility assessment conducted based on ISO10993 and ISO 7405.

    The performance testing results provided in the submission confirms the performance as substantially equivalent to the predicate devices with common indications.

    AI/ML Overview

    The provided text is a 510(k) summary for a dental cement device named GILB-2. It describes the device, its intended use, and claims substantial equivalence to predicate devices based on chemical composition and performance testing. However, it does not contain specific acceptance criteria, detailed study designs, or performance metrics from any clinical or performance studies.

    Therefore, I cannot provide the requested information in the format specified because the input text does not contain that level of detail.

    Here's why and what can be inferred:

    • Acceptance Criteria and Reported Device Performance: This information is completely absent. The document states "The performance testing results provided in the submission confirms the performance as substantially equivalent to the predicate devices with common indications," but no actual criteria or results are listed.
    • Sample size, data provenance, number of experts, adjudication method, MRMC study, standalone performance, ground truth type, training set size, and training set ground truth: All of these details relate to specific performance studies, whether clinical or non-clinical, that would have been conducted to demonstrate substantial equivalence. These details are not present in the 510(k) summary provided.

    In summary, none of the requested information regarding acceptance criteria, study details, sample sizes, expert involvement, or ground truth methods is available in the provided document. The document focuses on regulatory classification, device description, and claiming substantial equivalence to predicate devices generally, rather than detailing the specific studies used to demonstrate that equivalence.